K123350 · MedShape, Inc. · MBI · Mar 7, 2013 · Orthopedic
Device Facts
Record ID
K123350
Device Name
ECLIPSE SOFT TISSUE FASTENER
Applicant
MedShape, Inc.
Product Code
MBI · Orthopedic
Decision Date
Mar 7, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MedShape, Inc., Eclipse™ Soft Tissue Fastener is intended for fixation of soft tissue to bone in the shoulder, foot/ankle, knee, hand/wrist and elbowing procedures: Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Anchiles Tendon Reconstruction and Tendon Transfers Knee: Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal Joint Arthroplasty (basal thumb joint arthroplasty), Carpal Liqament Reconstructions and Repairs and Tendon Transfers Elbow: Biceps Tendon Reattachment and Ulnar or Radial collateral Ligament Reconstruction
Device Story
Eclipse™ Soft Tissue Fastener is a sterile, single-use orthopedic implant; used for fixation of soft tissue (ligament/tendon) to bone. Device consists of PEEK sleeve and PEEK insert; utilizes interference fit principle. Surgeon inserts device into pre-drilled bone hole; outer sleeve expands to compress soft tissue against bone wall, securing it in place. Used in clinical/surgical settings by orthopedic surgeons. Benefits include stable tissue-to-bone fixation during reconstructive procedures.
Clinical Evidence
Bench testing only. Testing included monotonic soft tissue fixation strength (pull-to-failure) in Sawbone® bone analogue, dimensional verification, and material safety testing (biocompatibility and MRI compatibility).
Technological Characteristics
Implant-grade PEEK sleeve and insert; interference fit mechanism; sterile, single-use; dimensions/form factor designed for bone hole insertion; no software or electronic components.
Indications for Use
Indicated for fixation of soft tissue to bone in shoulder, foot/ankle, knee, hand/wrist, and elbow orthopedic procedures requiring ligament or tendon reattachment.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K123350 (1/2)
Image /page/0/Picture/1 description: The image shows the logo for MedShape. The logo consists of a stylized letter "M" with a square on top and the word "MEDSHAPE" to the right of the letter. The text is in a blocky, sans-serif font.
MAR 7 2013
Traditional 510(k) Page 10 of 124
## 510(k) Summary
January 23td, 2013 Date Submitted:
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92(e).
- A. Submitter: MedShape, Inc. 1575 Northside Drive, Suite 440 Atlanta, Georgia 30318
- B. Company Contact: Jack Griffis Vice President. Research & Development (678) 235-3311 (direct) (404) 249-9158 (fax) jack.griffis@medshape.com
| C. | Device Information:<br>Trade Name(s): | Eclipse™ Soft Tissue Fastener with (optional) pre-<br>loaded ForceFiber® Sutures |
|----|---------------------------------------|---------------------------------------------------------------------------------------------|
| | Common Name(s): | Fastener, Fixation, bone, non-degradable<br>Suture, Non-absorbable, Synthetic, Polyethylene |
| D. | Classification Name: | Fastener, Fixation, Non-degradable, Soft Tissue<br>HWC/MBI 21 CFR 888.3040 |
| | | Suture, Non-absorbable, Synthetic, Polyethylene<br>GAT 21 CFR 878.500 |
- E. Predicate Device(s): Arthrex® Tenodesis Screws, K051726 MedShape, Inc., ExoShape® Soft Tissue Anchor, K101808
#### ட் Physical Description:
The proposed Eclipse™ Soft Tissue Fastener is a sterile, single use, orthopedic implant intended to be used for fixation of tissue including ligament or tendon to bone and bone tendon bone. The Eclipse™ Soft Tissue Fastener is designed to use the principles of interference fit to reattach soft tissue intended for insertion into a hole created in bone.
The Eclipse™ Soft Tissue Fastener body is comprised of an implant grade PEEK sleeve and implant grade PEEK insert. The outer sleeve is expanded into the bone hole, compressing the soft tissue aqainst the bone wall; fastening it into place.
{1}------------------------------------------------
- G. Indications for Use:
The Eclipse™ Soft Tissue Fastener is indicated for fixation of soft tissue to bone in the shoulder, foot/ankle, knee, hand/wrist and elbow in the following procedures:
- Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Tenodesis, Acromio-Clavicular Separation Biceps Repair. Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
- Lateral Stabilization, Medial Stabilization, Achilles Tendon Foot/Ankle: Repair. Hallux Valgus Reconstruction. Mid-foot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Anchiles Tendon Reconstruction and Tendon Transfers
- Knee: Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
- Hand/Wrist: Scapholunate Ligament Reconstruction, Ulmar Collateral Collateral Ligament Reconstruction, Radial Ligament Reconstruction, Carpometacarpal Joint Arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and Repairs and Tendon Transfers
- Biceps Tendon Reattachment and Ulnar or Radial collateral Elbow: Ligament Reconstruction
- H. Comparison of Technological Characteristics:
The Eclipse™ Soft Tissue Fastener is substantially equivalent in function and intended use to the following predicate devices:
### Arthrex® Tenodesis Screws, K051726 MedShape, Inc., ExoShape® Soft Tissue Anchor, K101808
All fasteners are comprised of implant grade PEEK. All fasteners have the same indications for use. In addition, functional performance testing has been conducted in Sawbone® bone analoque. This testing included monotonic soft tissue fixation strength (pull-to-failure) and other dimensional verification and material safety testing (both bio and MRI compatibility). Analysis of the results supports the conclusion that the proposed device is substantially equivalent to the predicate devices.
1/23/13
Jáck Griffis Vice President, Research & Development
Image /page/1/Picture/14 description: The image shows the logo for Medshape. The logo consists of a stylized letter "a" with a square on top, followed by the word "MEDSHAPE" in block letters. The logo is black and white and appears to be a company logo.
1575 Northside Drive, Suite 440 Atlanta, GA 30318 877-343-7016 phone 404-249-9158 fax
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three faces in profile, representing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 7, 2013
MedShape, Incorporated % Mr. Jack Griffis Vice President, Research and Development 1575 Northside Drive Northwest, Suite 440 Atlanta, Georgia 30318
Re: K123350
Trade/Device Name: Eclipse™ Soft Tissue Fastener Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, GAT Dated: January 24, 2013 Received: February 4, 2013
Dear Mr. Griffis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
#### Page 2 - Mr. Jack Griffis
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Ering Deith
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number: _ 123350
Device Name: Eclipse™ Soft Tissue Fastener
Indications for Use:
The MedShape, Inc., Eclipse™ Soft Tissue Fastener is intended for fixation of soft tissue to bone in the shoulder, foot/ankle, knee, hand/wrist and elbowing procedures:
- Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
- Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Anchiles Tendon Reconstruction and Tendon Transfers
- Knee: Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
- Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal Joint Arthroplasty (basal thumb joint arthroplasty), Carpal Liqament Reconstructions and Repairs and Tendon Transfers
- Elbow: Biceps Tendon Reattachment and Ulnar or Radial collateral Ligament Reconstruction
Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)
R Over-The-Counter
(21 CFR 801.9)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Casey L. Hanley, Ph.D.
Division of Orthopaedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.