K123327 · Ge Healthcare Coils (Usa Instruments, Inc.) · MOS · Jan 11, 2013 · Radiology
Device Facts
Record ID
K123327
Device Name
1.5T GEM RT OPEN ARRAY
Applicant
Ge Healthcare Coils (Usa Instruments, Inc.)
Product Code
MOS · Radiology
Decision Date
Jan 11, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The GEM RT Open Array Coil that is a part of the Oncology Suite is a receive-only RF coil designed for use with 1.5T MRI systems manufactured by GE Healthcare. The indications for use include the head, neck, and brachial plexus anatomies and vasculature imaging. The nucleus excited is hydrogen.
Device Story
1.5T GEM RT Open Array is a receive-only, 8-channel, 8-element posterior phased array RF coil; designed as an addition to GE Healthcare's GEM suite of coils. Device inserts into GEM table cradle at head or foot end; operates as an accessory to Optima MR450w MRI systems. Input: hydrogen nucleus signals during MRI scanning. Output: raw RF signal data processed by MRI system to generate diagnostic images. Used in clinical radiology settings by trained MRI technologists/radiologists. Provides high-resolution imaging of head, neck, and brachial plexus; aids clinicians in diagnostic assessment of these regions.
Clinical Evidence
No clinical data required. Substantial equivalence supported by non-clinical bench testing and sample clinical images.
Technological Characteristics
Receive-only, 8-channel, 8-element phased array RF coil; integrated preamplifiers. Compliant with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-2-33 (electrical safety), and NEMA MS-9 (SNR/uniformity). Designed for 1.5T MRI systems.
Indications for Use
Indicated for imaging of head, neck, and brachial plexus anatomies and vasculature in patients undergoing 1.5T MRI procedures.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text 'K123327' on the first line. The second line of the image says 'Page 1 of 3'. This indicates that the image is likely a page from a document, and it is the first page of a three-page document. The text is in a simple, sans-serif font.
### GE Healthcare 510(k) Premarket Notification Submission
#### 510(k) Summary
# JAN 1 1 2013
In accordance with 21 CFR 807.92 the following summary of information is provided:
December 27, 2012 Date:
Submitter: GE Healthcare Coils, (USA Instruments, Inc) Establishment Registration Number: 1529041 l 515 Danner Dr. Aurora, OH 44202-9273 USA
Primary Contact Person:
Andrew Menden Regulatory Affairs Manager GE Healthcare (GE Medical Systems, LLC) Establishment Registration Number: 2183553 3200 N Grandview Blvd., Mail Code - W-827 Waukesha, WI-53188 USA Phone: 262-521-6223 Fax: 414-908-9585
Secondary Contact Person:
Glen Sabin Regulatory Affairs Director GE Healthcare (GE Medical Systems LLC.) Establishment Registration Number: 2183553 3200 N Grandview Blvd., Mail Code - W-827 Waukesha, WI-53188 USA Phone: 262-521-6848 Fax: 262-364-2785
Device: Trade Name: Common/Usual Name: Classification Names:
1.5T GEM RT Open Array Coil, Magnetic Resonance, Specialty 21CFR 892.1000 - Magnetic resonance diagnostic device
| Product Code: | 90MOS |
|----------------------|------------------------------------------------------------------------------------------------------------|
| Predicate Device(s): | K103335, GEM Option for 1.5T MRI Systems |
| Device Description: | The 1.5T GEM RT Open Array is a receive only 8-<br>Channel 8-element Posterior Phased Array, for use as an |
ર - S
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The logo is black and white.
K123327 Page 2 of 3
GE Healthcare 510(k) Premarket Notification Submission
option to the Optima MR450w MR Systems with GEM suites. It is a posterior array which can be inserted into the GEM Table cradle (K103335) at either the head or foot end.
Intended Use:
The GEM RT Open Array Coil that is a part of the Oncology Suite is a receive-only RF coil designed for use with 1.5T MRI systems manufactured by GE Healthcare. The indications for use include the head, neck, and brachial plexus anatomies and vasculature imaging. The nucleus excited is hydrogen.
Technology:
Comparison with GEM Option for 1.5T MRI Systems
The GE 1.5T GEM RT Open Array is a multi-element phased array RF Receive only coil with integrated preamplifiers. The 1.5T GEM RT Open Array coil operates on the same principles and is an addition to the GEM suite of coils (K103335). The GEM RT Open Array is designed to fit into the GEM table at the head or foot end adjacent to where the existing integrated posterior array in the GEM table resides.
Determination of Substantial Equivalence:
Summary of Non-Clinical Tests:
The GE 1.5T GEM RT Open Array has used the same non-clinical voluntary standards to demonstrate substantial equivalence of safety and performance:
IEC 60601-1: Electrical Safety - compliant with all applicable sections
IEC 60601-1-2: Electromagnetic Compatibility compliant with all applicable sections (i.e. electrostatic discharge)
IEC 60601-2-33: Electrical Safety - compliant with all applicable sections
NEMA MS-9: SNR and Uniformity of Phased Array Coils - compliant with all applicable sections
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the alphanumeric string "K123327". The text is rendered in a simple, sans-serif font. The characters are uniformly sized and spaced, making the string easily readable. The background is plain white.
Page 3 of 3
## GE Healthcare 510(k) Premarket Notification Submission
The following quality assurance measures were applied to the development of the system:
- . Risk Analysis
- . Requirements Reviews
- . Design Reviews
- Testing on unit level (Module verification) .
- . Integration testing (System verification)
- Performance testing (Verification) .
- Safety testing (Verification) .
- Simulated use testing (Validation) .
Summary of Clinical Tests:
The subject of this premarket submission, 1.5T GEM RT Open Array, did not require clinical studies to support substantial equivalence. Sample clinical images have been included in this submission.
Conclusion: GE Healthcare considers the 1.5T GEM RT Open Array to be as safe, as effective, and performance is substantially equivalent to the predicate device.
Image /page/2/Picture/15 description: The image is a blank white canvas, devoid of any discernible objects, shapes, or colors. The uniformity of the white background creates a sense of emptiness and simplicity. There are no textures or patterns visible, resulting in a smooth and featureless surface. The image lacks any focal point or visual interest, presenting a minimalist and abstract composition.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a circular seal for the Department of Health & Human Services (HHS) in the USA. The seal features the department's name in a circular arrangement around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
r
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
January 11, 2013
GE Healthcare Coils (USA Instruments, Inc) % Mr. Andrew Menden 1515 Danner Dr. Aurora, OH 44202-9273
Re: K123327
Trade/Device Name: 1.5T GEM RT Open Array Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: October 25, 2012 Received: October 26, 2012
Dear Mr. Menden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for association to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 - Mr. Menden
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11: You desire spearle and the Office of In Vitro Diagnostics and Radiological Health at (301) 796-005), please note the regulation entitled, "Misbranding by reference to premarket 5 150. And 1100, productions regarding the reporting the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
under the MDR regulation (27 Of Reacony/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the I ou may obtain oner generar mennersational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Michael D. O'Hara
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined in a stylized, cursive font. The intertwined letters are enclosed within a circular border. The logo is presented in black and white.
GE Healthcare 510(k) Premarket Notification Submission
### Indications for Use
510(k) Number (if known): K123327
1.5T GEM RT Open Array Device Name:
Indications for Use:
The GEM RT Open Array Coil that is a part of the Oncology Suite is a receive-only RF coil designed for use with 1.5T MRI systems manufactured by GE Healthcare. The indications for use include the head, neck, and brachial plexus anatomies and vasculature imaging. The nucleus excited is hydrogen.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael J. D. O'Hara
(Division Sign Off)
Division of Radiolog
Office of In Vitro
5100a <1.23327
Page 1 of 1 .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.