DIMENSION CHEMISTRY III CALIBRATORY

K123321 · Siemens Healthcare Diagnostics · JIX · Feb 15, 2013 · Clinical Chemistry

Device Facts

Record IDK123321
Device NameDIMENSION CHEMISTRY III CALIBRATORY
ApplicantSiemens Healthcare Diagnostics
Product CodeJIX · Clinical Chemistry
Decision DateFeb 15, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Indications for Use

The CHEM III CAL is an in vitro diagnostic product for the calibration of Ammonia (AMM), Carbon Dioxide (ECO2) and Ethyl Alcohol (ETOH) assays on the Dimension® clinical chemistry system.

Device Story

Dimension Chemistry III Calibrator (CHEM III CAL) is a three-level, liquid, aqueous calibrator kit containing ammonium bicarbonate, sodium carbonate, and ethyl alcohol. Used in clinical chemistry laboratories to calibrate Ammonia, Carbon Dioxide, and Ethyl Alcohol assays on the Dimension clinical chemistry system. Provides standardized reference values to ensure accuracy of patient sample measurements. Product is ready-to-use; stored at 2–8°C. Healthcare providers use the calibrated system to obtain quantitative diagnostic results for patient management.

Clinical Evidence

No clinical data. Bench testing only. Stability testing performed per CLSI EP25-A guideline; established 12-month shelf life and open-vial stability (7 days on-system, 30 days refrigerated).

Technological Characteristics

Aqueous multi-analyte calibrator; contains ammonium bicarbonate, ethyl alcohol, sodium carbonate. Traceability: Ammonia (ACS Grade ammonium sulfate), CO2 (NIST SRM 351), Ethyl Alcohol (USP grade). Storage: 2–8°C. Form factor: 6-vial kit (2.5 mL/vial).

Indications for Use

Indicated for the calibration of Ammonia, Carbon Dioxide, and Ethyl Alcohol assays on the Siemens Dimension RxL Max clinical chemistry system.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k123321 B. Purpose for Submission: New calibrator material C. Measurand: Ammonia, Carbon Dioxide (CO₂), Ethyl Alcohol D. Type of Test: Not applicable. E. Applicant: Siemens Healthcare Diagnostics F. Proprietary and Established Names: Dimension Chemistry III Calibrator (CHEM III CAL) G. Regulatory Information: 1. Regulation section: 21 CFR 862.1150 2. Classification: Class II 3. Product code: JIX 4. Panel: 75 (Clinical Chemistry) {1} H. Intended Use: 1. Intended use(s): Refer to Indications for Use. 2. Indication(s) for use: The CHEM III CAL is an in vitro diagnostic product for the calibration of Ammonia (AMM), Carbon Dioxide (EC02) and Ethyl Alcohol (ETOH) assays on the Dimension® clinical chemistry system. 3. Special conditions for use statement(s): None 4. Special instrument requirements: Siemens Dimension RxL Max I. Device Description: The Dimension® Chemistry III Calibrator (CHEM III CAL) is a multi-analyte, aqueous product containing ammonium bicarbonate, ethyl alcohol and sodium carbonate. It is packaged as a kit of six vials with two vials each of Levels 1, 2 and 3, containing 2.5 mL per vial. Level 1 is made by adding preservatives to purified water and is also the base for Levels 2 and 3. Levels 2 and 3 are prepared by adding the targeted quantities of ammonia, carbon dioxide, and ethyl alcohol to levels 2 and 3. J. Substantial Equivalence Information: 1. Predicate device name(s): Dimension Vista® Chemistry 3 Calibrator (CHEM 3 CAL) 2. Predicate 510(k) number(s): k062334 3. Comparison with predicate: {2} | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Same | Calibration of ammonia, carbon dioxide, and ethyl alcohol assays | | Preparation | Same | Liquid, ready to use | | Storage | Same | 2 – 8°C | | Matrix | Same | Aqueous, containing ethyl alcohol, ammonium bicarbonate, and sodium bicarbonate | | Traceability | Same | Ammonia - ASC Grade ammonium sulfate Carbon Dioxide – NIST SRM 351 Ethyl alcohol – USP grade ethyl alcohol | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Number of calibrator levels | 3 | 2 | | Levels for specific analytes | Ammonia level 1: 0 μg/dL level 2: 707 μg/dL level 3: 1405 μg/dL | Ammonia Level 1: 17 μg/dL Level 2: 1804 μg/dL | | | CO2 level 1: 0 mmol/L level 2: 25 mmol/L level 3: 50 mmol/L | CO2 level 1: 0 mmol/L level 2: 47 mmol/L | | | Ethyl Alcohol level 1: 0 mg/dL level 2: 100 mg/dL level 3: 315 mg/dL | Ethyl Alcohol level 1: 0 mg/dL level 2: 308 mg/dL | K. Standard/Guidance Document Referenced (if applicable): CLSI EP25-A L. Test Principle: Not applicable {3} M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Not applicable b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: The traceability of the three analytes in the Dimension Chemistry III Calibrator (CHEM III CAL) is as follows: Ammonia – traceable to ACS Grade Ammonium Sulfate Carbon Dioxide (CO₂) – traceable to NIST SRM 351 Ethyl Alcohol – traceable to USP Grade Ethyl Alcohol Value assignment: Ammonia – the assigned values for ammonia are based on results from 3 instruments, 3 reagent lots, and 9 test runs. Each run generates 10 replicates per calibrator level resulting in 90 total replicates per level. Carbon Dioxide – the assigned values for CO₂ are based on results from 3 instruments, 3 reagent lots, and 9 test runs. Each run generates 5 replicates per calibrator level resulting in 45 total replicates per level. Ethyl Alcohol – the assigned values for alcohol are based on results from 3 instruments, 3 reagent lots, and 9 test runs. Each run generates 5 replicates per calibrator level resulting in 45 total replicates per level. The labeling states that obtained values should fall within the specified range provided in the labeling and that laboratories should establish appropriate acceptance criteria when using this product for its intended use. Stability: Calibrator shelf life is determined in real-time. Protocols and acceptance {4} criteria were reviewed and determined to be acceptable. The testing supports the sponsor’s claimed shelf life stability of 12 months when stored at 2-8°C. Calibrator opened vial stability is determined in real-time. Protocols and acceptance criteria were reviewed and determined to be acceptable. The testing supports the sponsor’s claimed opened vial stability of 30 days when stored at 2-8°C. Storage recommendations are provided in the labeling. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Not applicable b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable {5} 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Not applicable N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...