The Access & Support Catheter is intended for use during coronary and peripheral interventional procedures to guide and support guidewires, traverse discrete portions of the vasculature, allow for guidewire exchanges and provide a conduit for infusion of saline solution, diagnostic contrast agents and therapeutic agents.
Device Story
Access & Support Catheter is an over-the-wire (OTW) single-lumen catheter used in coronary and peripheral interventional procedures. Device facilitates guidewire support, navigation through vasculature, guidewire exchange, and infusion of saline, contrast, or therapeutic agents (max 300 psi). Operated by clinicians in clinical settings. Features atraumatic tapered tip, HDPE shaft, and distal platinum/iridium markerbands for radiopacity. Distal 40 cm includes hydrophilic coating to reduce friction. Proximal end features PEBA strain relief and HDPE manifold with female luer connection. Compatible with 5 Fr introducer sheaths and 6 Fr guiding catheters. Available in various lengths (65-150 cm) and guidewire compatibilities (0.014", 0.018", 0.035"). Single-use, sterile device.
Single-lumen catheter; HDPE shaft; PEBA strain relief; HDPE manifold; Pt/Ir markerbands; hydrophilic coating on distal 40 cm. Sterilized via EO. Dimensions: 65-150 cm length; 0.014", 0.018", 0.035" guidewire compatibility. Max infusion pressure 300 psi. Standalone mechanical device.
Indications for Use
Indicated for patients undergoing coronary and peripheral interventional procedures requiring guidewire support, navigation of vasculature, guidewire exchange, or infusion of saline, contrast, or therapeutic agents.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Minnie Support Catheter - Vascular Solutions, Inc. (K082337)
Quick-Cross Support Catheter - Spectranetics, Inc. (K033678)
Submission Summary (Full Text)
{0}------------------------------------------------
K 123311
JAN 1 4 2013
## 510(k) Summary
Date Prepared: October 5, 2012
510(k) Submitter QXMédical, LLC
1843 Buerkle Road White Bear Lake, MN 55110
Contact Person Fernando Di Caprio Phone: (651) 338-5774 Email: fernando.dicaprio@qxmedical.com
| General Information | |
|---------------------|------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Access & Support Catheter |
| Common Name | Support catheter |
| Classification Info | Percutaneous catheter per 21 CFR 870.1250 (Class II); Product Code: DQY |
| Predicate Device | Minnie Support Catheter - Vascular Solutions, Inc. (K082337)<br>Quick-Cross Support Catheter - Spectranetics, Inc. (K033678) |
## Device Description
The Access & Support Catheter is an over-the-wire (OTW) single-lumen catheter with an atraumatic tapered tip. The catheter is offered in twelve (12) size models as outlined in the following table.
| Model # | Guidewire<br>Compatibility | Shaft Outer Diameter<br>(proximal / distal) | Working<br>(Effective)<br>Length | Markerband<br>Quantity | Markerband<br>Spacing |
|----------|----------------------------|---------------------------------------------|----------------------------------|------------------------|-----------------------|
| 9002-001 | 0.014" | 0.039"/0.0265" | 65cm | 3 | 15 mm |
| 9002-002 | 0.014" | 0.039"/0.0265" | 90cm | 3 | 15 mm |
| 9002-003 | 0.014" | 0.039"/0.0265" | 135cm | 3 | 15 mm |
| 9002-004 | 0.014" | 0.039"/0.0265" | 150cm | 3 | 15 mm |
| 9002-005 | 0.018" | 0.044"/0.0305" | 65cm | 3 | 15 mm |
| 9002-006 | 0.018" | 0.044"/0.0305" | 90cm | 3 | 15 mm |
| 9002-007 | 0.018" | 0.044"/0.0305" | 135cm | 3 | 15 mm |
| 9002-008 | 0.018" | 0.044"/0.0305" | 150cm | 3 | 15 mm |
| 9002-009 | 0.035" | 0.063"/0.0505" | 65cm | 3 | 50 mm |
| 9002-010 | 0.035" | 0.063"/0.0505" | 90cm | 3 | 50 mm |
| 9002-011 | 0.035" | 0.063"/0.0505" | 135cm | 3 | 50 mm |
| 9002-012 | 0.035" | 0.063"/0.0505" | 150cm | 3 | 50 mm |
The catheter shaft consists of high density polyethylene (HDPE) with distal platinum/iridium markerbands. The distal 40 cm outer portion of all catheters have a hydrophilic coating. The proximal portion includes a PEBA strain relief with a HDPE manifold which incorporates a female luer connection which communicates with the catheter lumen. The catheter through lumen is used to pass the catheter over the appropriate guidewire or for infusion. When used for
{1}------------------------------------------------
infusion, the maximum infusion pressure should not exceed 300 psi. The catheter guidewire compatibility size and length are printed on the strain relief. All models are compatible with 5 Fr introducer sheaths and 6 Fr guiding catheters. All devices are provided sterile and intended for single use only.
#### Intended Use / Indications
The Access & Support Catheter is intended for use during coronary and peripheral interventional procedures to guide and support guidewires, traverse discrete portions of the vasculature, allow for guidewire exchanges and provide a conduit for infusion of saline solution, diagnostic contrast agents and therapeutic agents.
#### Comparison to Predicates
The indication for use of the Access & Support Catheter is comparable to the predicate devices with the differences being mainly semantics. Both devices are intended to support access to the vasculature, allow for guidewire exchanges, and provide a conduit for fluid deliveries. The Access & Support Catheter uses the terms "discreet portions of the vasculature" as a more descriptive term compared to the Quick-Cross general "access of the vasculature." For comparison, the Minnie Support Catheter also uses the "discreet regions" terms as further description of the type of vasculature the catheter can be used within.
The Access & Support Catheter is technologically similar to the predicate devices in that they all are single lumen catheters utilized to exchange guidewires and/or infuse intravascular procedural agents as necessary. The Minnie Catheter is constructed of similar materials (polyethylene and hydrophilic coating) with similar working lengths ranging in size from 65cm to 150cm, and the same dimensional guidewire compatibility (0.014", and 0.035"). The Quick-Cross Support Catheter is constructed of similar materials (HDPE, Pt/Ir, and hydrophilic coating) and ranges in length from 90cm to 135cm with the same guidewire compatibilities. The following table outlines a comparison of the subject device to the predicate devices.
{2}------------------------------------------------
| Device Component /<br>Performance Parameter | QXMédical Access &<br>Support Catheter | Quick-Cross Catheter<br>(K033678) | Minnie Catheter<br>(K082337) |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Manifold/Proximal Hub | HDPE | Same | PE |
| Strain Relief | PEBA | Molded from hub (HDPE) | unknown |
| Catheter Shaft | Single lumen, HDPE | Same | Single lumen, PE |
| Coating | Hydrophilic | Same | Same |
| Markers - Material (Quantity) | Pt/Ir (3) | Same | Material unknown (3) |
| Infusion Pressure | 300 psi | Same | Same |
| Outer Diameter | Varies by model<br>Proximal:<br>0.039", 0.044", 0.063"<br>Distal:<br>0.0265", 0.0305", 0.0505"<br>Tip:<br>0.0195", 0.0225", 0.0405" | Varies by model<br>Proximal:<br>0.039", 0.044", 0.063"<br>Distal:<br>0.0260, 0.030, 0.050"<br>Tip:<br>0.020, 0.023, 0.041" | Varies by model<br>Proximal:<br>0.039, 0.044, 0.063"<br>Distal:<br>0.026, 0.030, 0.050"<br>Tip:<br>0.020. 0.023, 0.041" |
| Effective (Working) Length | All three diameters<br>65, 90, 135, 150 cm | Varies by model<br>90, 135, 150 cm | Varies by model<br>65, 90, 135, 150 cm |
| Guidewire Diameter (min.) | 0.014", 0.018", 0.035" | Same | Same |
| Guide Catheter (max.) | 5Fr | Same | 5Fr or 6Fr |
| Access Sheath (max.) | 5Fr | Same | Same |
| Single Use, Sterile | Yes | Same | Same |
| Sterilization Method | EO | Unknown | Same |
.
{3}------------------------------------------------
## Summary of Non-Clinical Performance Data
The Access & Support Catheter has been evaluated using the following in vitro bench testing to confirm the performance characteristics of the device.
- Visual inspections .
- Dimensional evaluations
- Luer syringe compatibility .
- . Guidewire compatibility
- . Sheath compatibility
- . Guidewire support
- . Catheter stiffness & flexibility, guidewire retraction & re-insertion
- . Catheter fatigue
- . Catheter kink resistance
- Markerband retention
- Leakage/burst (Pressure & Vacuum)
- . Torque testing
- . Tensile testing
- . Corrosion testing
- Simulated Use .
- . Coating Adherence
- . Particulate Evaluations
- . Shelf life
- Package Integrity ●
Biocompatibility tests were completed to ensure all materials utilized to construct the Access & Support Catheter were biocompatible. Biocompatibility tests included:
- Cytotoxicity ●
- . Sensitization
- Irritation / Intracutaneous Reactivity ●
- Systemic Toxicity (Acute) ●
- . Genotoxicity
- Hemolysis .
- Immunology (Compliment Activation) .
- In Vivo Thromboresistance .
- Pyrogenicity (material mediated) ●
- Physicochemical Tests .
All tests demonstrated the materials, manufacturing processes, and design of the Access & Support Catheter met the established performance criteria and will perform as intended. The above non-clinical tests demonstrate that the device is safe and effective and performs as safely and effectively as the legally marketed predicate devices.
{4}------------------------------------------------
## Substantial Equivalence Comparison Summary
Based upon the intended use and descriptive information provided in this pre-market notification, the QXMédical Access & Support Catheter has been shown to be substantially equivalent to the predicate devices. The QXMédical Access & Support Catheter raises no new questions of safety or effectiveness when compared to the predicate devices and is, therefore, substantially equivalent.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human services.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
QXMedical, LLC c/o Mark Job . Regulatory Technology Services, LLC 1394 24th Street NW Buffalo, MN 55313
Re: K123311
Trade/Device Name: VELO Access and Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: November 20, 2012 Received: November 21, 2012
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
JAN 1 4 2013
{6}------------------------------------------------
Page 2 - Mr. Mark Job
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Matthew G. Hillebrenner
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known): K123311
Device Name: QXMédical Access & Support Catheter
Indications for Use:
The Access & Support Catheter is intended for use during coronary and peripheral interventional procedures to guide and support guidewires, traverse discrete portions of the vasculature, allow for guidewire exchanges and provide a conduit for infusion of saline solution, diagnostic contrast agents and therapeutic agents.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M.A. Lillehem
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.