K123217 · Cardea Associates, Inc. · DPS · Feb 5, 2013 · Cardiovascular
Device Facts
Record ID
K123217
Device Name
CARDEASCREEN
Applicant
Cardea Associates, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Feb 5, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
CardeaScreen records and measures a resting ECG from the adult and pediatric (age ≥ 14) body surface. It provides automatic ECG interpretations which are identified as "Unconfirmed" by the product until they have been over-read and confirmed by a clinician. CardeaScreen is intended for use on apparently healthy individuals and on symptomatically stable patients with known or potential cardiac conditions. This device is intended for use under the direct supervision of a licensed health care clinician.
Device Story
CardeaScreen is a portable, battery-powered or AC-mains-operated electrocardiograph (ECG) system. It consists of an ECG transmitter (data acquisition unit) and PC-based software. The transmitter connects to a 10-lead patient cable and standard ECG electrodes applied to the patient's chest. The device collects resting ECG signals and transmits them wirelessly via Bluetooth to a clinician's PC. The software performs automated ECG analysis and arrhythmia detection. Outputs are displayed on the PC for clinician review. All automated interpretations are labeled 'Unconfirmed' and require clinician over-read and confirmation. The device is used in hospital or medical settings by healthcare clinicians to assist in cardiac assessment of healthy individuals or stable patients with cardiac conditions.
Clinical Evidence
Bench testing only. Verification and validation included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), patient cable testing (ANSI/AAMI EC53), 21 CFR 898 compliance, biocompatibility (ISO 10993-1), and firmware/software verification including the ECG analysis algorithm.
Technological Characteristics
Portable 12-lead ECG system; transmitter (~6" x 4" x 1.5", ~11 oz). Connectivity: Bluetooth wireless data transfer to PC. Power: Battery or AC mains. Applied part: Type CF. Features: Pacemaker detection, leads-off detection, defibrillation protection. Software: PC-based for analysis/arrhythmia detection. Reusable, non-sterile.
Indications for Use
Indicated for recording and measuring resting ECG in adults and pediatric patients (age ≥ 14). Intended for apparently healthy individuals and symptomatically stable patients with known or potential cardiac conditions. Use under direct supervision of a licensed health care clinician.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
#### 510(k) Summary of Safety and Effectiveness ll.
# FEB 0 5 2013
# CardeaScreen™
#### General Information
| Criteria | Information |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | CardeaScreen™ |
| Model Number | CS-2020 |
| Common Name | Electrocardiograph (ECG) |
| Classification | 21 CFR 870.2340 - Electrocardiograph<br>Class II; product code: DPS |
| 510(k) Submitter | Cardea Associates, Inc.<br>13720 220th Place NE<br>Woodinville, WA 98077 USA |
| Contact Person | David M. Hadley, Ph.D.<br>President<br>Cardea Associates, Inc.<br>dave@cardeascreen.com<br>(855) 800-0760 extension 801 (phone)<br>(855) 800-0760 (fax) |
#### Device Description
The CardeaScreen™, developed by Cardea Associates, Inc. (Cardea), is small, lightweight, portable electrocardiograph, designed for collection of resting (nonambulatory) electrocardiographic (ECG) data.
The CardeaScreen device system is composed of 2 main components:
- . ECG Transmitter - this type of device is also referred to in the medical community as a "ECG recorder", or "ECG data acquisition unit". The CardeaScreen ECG transmitter is about the size of a small paperback book (~6" x 4" x 1.5") and ~11 ounces. The transmitter is battery powered or it can also be powered via AC mains power by using the medical grade power supply supplied as part of the CardeaScreen system.
A patient cable with 10 lead wires is attached to the transmitter. The lead wires terminate in connectors that are attached to commercially available ECG electrodes selected by the clinician for use. The electrodes (not part of the CardeaScreen system) adhere to the
{1}------------------------------------------------
patient's skin in the chest region and are used for detecting a patient's ECG.
- . CardeaScreen Software - the software is installed by the clinician on the clinician's personal computer (PC). The software controls the ECG transmitter's use. The transmitter cannot function in any manner to collect ECG, without the software.
Collected ECG data is transferred from the ECG transmitter to the PC using wireless (Bluetooth) transmission. The software can be used for subsequent analysis of the patient's ECG data.
### Indications for Use
CardeaScreen records and measures a resting ECG from the adult and pediatric (age ≥ 14) body surface. It provides automatic ECG interpretations which are identified as "Unconfirmed" by the product until they have been overread and confirmed by a clinician.
CardeaScreen is intended for use on apparently healthy individuals and on symptomatically stable patients with known or potential cardiac conditions.
This device is intended for use under the direct supervision of a licensed health care clinician.
| Manufacturer | Substantially<br>equivalent devices | 510(k) |
|-----------------------------------------------------------------------------------------------|-------------------------------------|---------|
| Cardiac Science Corporation<br>Bothell, WA | CareCenter MD | K093211 |
| Brentwood Medical Technology<br>Corporation,<br>Midmark Diagnostics Group<br>Versailles, Ohio | IQecg | K103640 |
#### Substantially Equivalent Devices
{2}------------------------------------------------
# Substantial Equivalence Comparison
The CardeaScreen has in many instances identical or nearly identical technological characteristics to the substantially equivalent (predicate) ECG devices. The following many similarities are noted:
- . All devices are electrocardiographs.
- Devices are all classified by FDA as Class II medical devices under . regulation 21 CFR 870.2340; code DPS.
- Devices have a common system configuration (e.g., an ECG transmitter . (data acquisition unit) and software installed on a clinician's personal computer (PC) for ECG processing/analysis.
- Devices have an identical intended use all are electrocardiographs used . in the collection and analysis of ECG data.
- The devices have very similar wording in terms of indications for use. .
- All devices are used for resting ECG data collection. .
- All are prescription use devices; no over-the-counter use. .
- Devices are used by the same personnel (e.q., healthcare clinicians). .
- All devices are used in the same use environment in hospital/medical settings. .
- All devices have the same intended patient population use in adults and/or ● pediatrics.
- All devices are not for ambulatory use. .
- The CardeaScreen and predicates are all have a common device . configuration and the same use mode in that the patient cables (leads) are integrated into the ECG transmitter and the cables terminate in clip/snap connectors that attach to ECG electrodes (applied to the patient's chest by a clinician).
- The CardeaScreen and CareCenter MD predicate ECG have wireless ECG . data transfer (via Bluetooth) of the collected ECG data to the clinician's PC. The IQecq is a wired transfer.
- All devices transfer collected ECG data continuously. .
- All devices require/are compatible with Microsoft Windows Operating . Systems and require a Windows-compatible PC.
- . Devices are all capable of interfacing with other equipment (such as servers, electronic medical records, etc.).
- All devices are similarly sized small and lightweight. .
- . Devices are all portable.
- Devices are all supplied nonsterile and are reusable. ●
- Devices are all a standard 12 lead configuration. .
- . Devices have identical or nearly identical ECG data acquisition characteristics (e.q., frequency response, filters, sensitivity, input impedance, common mode rejection, sample rate for ECG interpretation and analysis, ECG amplifier presence, simultaneous data acquisition from all leads and displayed data of resting ECGs).
{3}------------------------------------------------
- All devices have software that is installed on the user's PC and performs . ECG analysis for arrhythmia detection. Such analysis is available to the clinician for use, however the labeling of all the devices notes that device interpretations/analyses are to be confirmed by a clinician.
- All devices have comparable environmental ranges (e.g., temperature . and humidity).
- Lastly, the devices all have common safety attributes (e.g., pacemaker . and leads off detection features, technical standard conformance, input protection against defibrillation shocks, type CF applied part rating and patient isolation).
Therefore, due to the identical or nearly identical nature of the CardeaScreen, as noted in the numerous similarities itemized above, Cardea believes the CardeaScreen is substantially equivalent to the predicate ECG devices.
# Testing
The CardeaScreen underwent extensive testing and related assessments, as listed below. All testing demonstrated acceptable results.
- Design Verification (Safety) testing, including:
- IEC 60601-1 testing (electrical safety)
- IEC 60601-1-2 testing (EMC)
- ANSI/AAMI EC53 testing (patient cable)
- 21 CFR 898 Compliance assessment
- Additional design verification testing
- Firmware/Software (including ECG analysis algorithm) verification/ validation testing
- ISO 10993-1 (biocompatibility and equivalent usage in predicate devices)
# Summary of Substantial Equivalence
Based on the information provided in this 510(k) and testing completed, the CardeaScreen does not raise any new safety or effectiveness issues and is substantially equivalent to legally marketed electrocardiographs that are in commercial distribution, and have been determined to be substantially equivalent to devices in commercial distribution, prior to May 28, 1976.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
FEB 0 5 2013
Cardea Associates, Inc. c/o Mr. Jeff D. Rongero Senior Project Engineer UL LLC 12 Laboratory Drive, Research Triangle Park, NC 27709-3995
Re: K123217
Trade/Device Name: CardeaScreen™ Regulatory Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (Two) Product Code: DPS Dated: January 14, 2013 Received: January 15, 2013
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Jeff D. Rongero
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm] 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Owen P. Faris -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K123217 510(k) Number:
Device Name:
CardeaScreen™
Indications for Use:
CardeaScreen records and measures a resting ECG from the adult and pediatric (age ≥ 14) body surface. It provides automatic ECG interpretations which are identified as "Unconfirmed" by the product until they have been over-read and confirmed by a clinician.
CardeaScreen is intended for use on apparently healthy individuals and on symptomatically stable patients with known or potential cardiac conditions.
This device is intended for use under the direct supervision of a licensed health care clinician.
| Prescription Use | C |
|------------------|-----------------------------|
| | (Part 21 CFR 801 Subpart D) |
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
page 1 of 1
(Division Sign-Off) Division of Cardiovascular Devices
| 510(k) Notification - CardeaScree | 510(k) Number: K2321 |
|-----------------------------------|----------------------|
|-----------------------------------|----------------------|
Cardea Associates, Inc.
page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.