The Ethicon Endo Surgery ENSEAL® G2 Cordless Curved Jaw Tissuc Sealers are intended for use during open or laparoscopic, general and gynccological surgery to cut and scal vessels, cut, grasp and dissect tissue during surgery. Indications for use include open and laparoscopic general, gynecological surgical procedures (including urologic, thoracic, plastic and reconstructive, bowel resections, hysterectomies, cholecystectomies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectomies, etc.), or any procedure where vessel ligation (cutting and scaling), tissue grasping and dissection are performed. The devices can be used on vessels up to (and including) 7mm and bundles as large as will fit in the jaws of the instruments. The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this system for these procedures.
Device Story
Sterile, single-use, cordless RF electrosurgical instrument; powered by integrated Lithium Ion Polymer (LiPo) battery; eliminates need for external generator. Device features 3mm curved jaw with I-blade knife; operated by surgeon during open or laparoscopic procedures. User squeezes closing handle to grasp/clamp tissue; pressing energy activation button delivers bipolar RF energy to seal vessels; full handle closure advances I-blade to transect tissue. PTC technology regulates energy delivery. Benefits include cordless mobility, integrated power, and efficient vessel ligation/dissection. Device is self-contained; no user-rechargeable components.
Clinical Evidence
No human clinical studies. Evidence based on bench testing, electrical safety, biocompatibility, vessel burst pressure testing, and animal acute/survival studies. Testing demonstrated performance equivalent to predicate devices.
Technological Characteristics
Bipolar electrosurgical instrument; 3mm curved jaw; 14cm, 35cm, 45cm shaft lengths; 5mm shaft diameter. Integrated LiPo battery power source; PTC (Positive Temperature Coefficient) technology for energy regulation. Mechanical I-blade knife for transection. Single-patient use, sterile.
Indications for Use
Indicated for open or laparoscopic general, gynecological, urologic, thoracic, plastic, and reconstructive surgeries; includes bowel resections, hysterectomies, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomies; used for vessel ligation, tissue grasping, and dissection on vessels up to 7mm. Contraindicated for tubal sterilization or tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
ENSEAL Vessel Sealing and Hemostasis System (K072177)
ENSEAL Vessel Sealing and Hemostasis System (K072493)
Submission Summary (Full Text)
{0}------------------------------------------------
K123212
Page 1 of 3
Ethicon Endo-Surgery, LLC 510(k) Premarket Notification (Traditional)
ENSEAL® G2 Cordless Tissue Sealer
# 510(k) Summary
Ethicon Endo-Surgery, LLC Company 475 Calle C Guaynabo, PR 00969
Contact Liping Wu, MD, PhD, RAC Regulatory Affairs Associate II Ethicon Endo-Surgery, Inc. Telephone: (513) 337-3135 Fax: (513) 337-2930 Email: LWu@its.jnj.com
Date Prepared March 5, 2013
#### Device Name
Trade Name: Ethicon Endo- Surgery ENSEAL G2 Cordless Curved Jaw Tissue Sealer Common Name: Electrosurgical Cutting and Coagulating Instrument
# Classification Names
- · Electrosurgical, Cutting & Coagulation & Accessories (21 CFR 878.4400, Product Code PDG)
- · Electrocautery, Gynecologic and Accessories (21 CFR 884.4120, Product Code HGI)
# Predicate Devices
ENSEAL Vessel Sealing and Hemostasis System cleared under K072177 on August 29, 2007 and K072493 cleared on September 5, 2007. The ENSEAL Vessel Sealing and Hemostasis System include ENSEAL Bipolar Electrosurgical Sealing Devices and an Electrosurgical Generator.
#### Device Description:
The Ethicon Endo-Surgery ENSEAL G2 Cordless Curved Jaw Tissue Sealer is a sterile, single patient use, self-contained, disposable RF electrosurgical instrument. The cordless hand-held device is powered by a Lithium Ion Polymer (LiPo) battery with an RF (radiofrequency) generator integrated into the device handle and does not require connection to an external generator or power source. The integrated generator provides the required RF energy to seal vessels during a surgical procedure. The battery is a DC power source that is charged during manufacturing; the battery cannot be removed from the device, or be recharged by the user. The device is configured with 3mm curved jaw and is intended to cut, grasp and dissect soft tissue, and coagulate and transect vessels up to and including 7mm in diameter.
MAR 8 2013
{1}------------------------------------------------
ENSEAL® G2 Cordless Tissue Sealer
### Indications for Use:
The EES ENSEAL G2 Cordless Curved Jaw Tissue Sealers are intended for use during open or laparoscopic, general and gynecological surgery to cut and seal vessels, cut, grasp and dissect tissue during surgery.
Indications for use include open and laparoscopic general, gynecological surgical procedures (including urologic, thoracic, plastic and reconstructive, bowel resections, hysterectomies, cholecystectomies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectomies, etc.), or any procedure where vessel ligation (cutting and sealing), tissue grasping and dissection are performed. The devices can be used on vessels up to (and including) 7mm and bundles as large as will fit in the jaws of the instruments.
The devices have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this system for these procedures.
## Technological Characteristics:
The Ethicon Endo-Surgery ENSEAL G2 Cordless Curved Jaw Tissue Sealers are configured with curved jaws and are intended to coagulate and transect vessels. The instrument consists of a grip housing assembly, a rotating shaft, and a moveable jaw with an I-blade™ knife in the tip. The devices will be available in 14cm, 35cm and 45cm shaft length options, and the instruments can be rotated 360° to facilitate visualization and enable easy access to targeted tissue. The jaw is curved in a normally opened position and can be partially or fully closed by squeezing the Closing Handle. The jaw has teeth for grasping and holding targeted tissue when clamped. Bipolar energy is delivered to the clamped tissue when the Energy Activation Button is pressed. Full activation of the Closing Handle advances the I-blade™ the length of the jaws, and the targeted tissue is compressed, coagulated, and transected. The I-blade™ knife cannot be advanced to transect tissue without fully pressing the Energy Activation Button.
| Technological<br>Characteristic | Predicate Devices<br>(K072177 and K072493) | Subject Device ENSEAL<br>G2 Cordless Curved<br>Tissue Sealer |
|---------------------------------|--------------------------------------------------------|--------------------------------------------------------------|
| | Energy type | Bipolar |
| Jaw shape | Curved 3 mm | Curved 3mm |
| Shaft lengths | 14 cm, 35 cm, 45 cm | 14 cm, 35 cm, 45 cm |
| Shaft diameter | 5 mm | 5 mm |
| I-blade | Yes | Yes |
| PTC technology | Yes | Yes |
| Energy Source | Generator | Battery |
| Function | Cut and seal vessels, cut,<br>grasp and dissect tissue | Cut and seal vessels, cut,<br>grasp and dissect tissue |
Comparison of Technological Characteristics
{2}------------------------------------------------
ENSEAL® G2 Cordless Tissue Sealer
## Performance Data:
Bench testing and laboratory evaluations in an animal model were conducted to demonstrate that the Ethicon Endo-Surgery ENSEAL G2 Cordless Curved Jaw Tissue Sealers perform as intended.
Biocompatibility studies, electrical safety studies, software validation, vessel burst pressure testing and animal acute and survival studies were performed comparing the predicate device to the subject device to support the determination of substantial equivalence. Clearance was not based on human clinical studies.
## Conclusions
Based on performance testing and functional similarities to the predicate devices, the Ethicon Endo-Surgery ENSEAL G2 Cordless Curved Jaw Tissue Sealer devices are substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2013
Ethicon Endo-Surgery, LLC % Ethicon Endo-Surgery, Incorporated Liping Wu, MD, PhD, RAC Regulatory Affairs Associate II 4545 Creek Road Cincinnati, Ohio 45242
Re: K123212
Trade/Device Name: Ethicon Endo Surgery ENSEAL G2 Cordless Curved Jaw Tissue Sealers Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: PDG, HGI Dated: February 11, 2013 Received: February 13, 2013
Dear Liping Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Liping Wu, MD, PhD, RAC
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, For
# Peter D库Ruimm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1.
Ethicon Endo-Surgery, LLC 510(k) Premarket Notification (Traditional)
ENSEAL® G2 Cordless Tissuc Scaler
Indications for Use
510(k) Number (if known): J. 123212
Device Name: Ethicon Endo Surgery ENSEAL G2 Cordless Curved Jaw Tissue Sealers
Indications for Use:
The Ethicon Endo Surgery ENSEAL® G2 Cordless Curved Jaw Tissuc Sealers are intended for use during open or laparoscopic, general and gynccological surgery to cut and scal vessels, cut, grasp and dissect tissue during surgery.
Indications for use include open and laparoscopic general, gynecological surgical procedures (including urologic, thoracic, plastic and reconstructive, bowel resections, hysterectomies, cholecystectomies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectomies, etc.), or any procedure where vessel ligation (cutting and scaling), tissue grasping and dissection are performed. The devices can be used on vessels up to (and including) 7mm and bundles as large as will fit in the jaws of the instruments.
The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this system for these procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Long H. Chen -A #
for MXM
(Division Sign-Off) Division of Surgical Devices 510(k) Number K123212
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.