K123177 · Laserstim, Inc. · ILY · Jul 3, 2013 · Physical Medicine
Device Facts
Record ID
K123177
Device Name
STIMLASE 2000 LOW LEVEL LASER THERAPY SYSTEM
Applicant
Laserstim, Inc.
Product Code
ILY · Physical Medicine
Decision Date
Jul 3, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Heat Lamp Therapy Feature: The StimLase 2000 System is intended to be used for topical heating of tissue for the purpose of temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasms. The unit may be used for temporary increase in local blood circulation and/or the temporary relaxation of muscles. Electrotherapy Feature: The StimLase 2000 System is intended to be used for relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local bloodcirculation, muscle re-education, maintaining or increasing range of motion, and immediate post-surgical stimulation of calf muscles to prevent venous thrombosis.
Device Story
StimLase 2000 System provides therapeutic heating and electrotherapy; consists of base console with touch-screen, footpedal, handheld laser wand, and two disposable carbon electrodes. Laser wand utilizes gallium arsenide semiconductor diode for infrared lamp treatment; console provides power adjustment (0-550 mW) for simultaneous or separate therapy. Operated by clinician in clinical setting; output used to manage pain, muscle spasms, and circulation. Device provides physical relief through thermal and electrical stimulation; clinical decision-making guided by patient symptoms and therapeutic goals.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and performance testing to safety standards.
Indicated for patients requiring topical tissue heating for relief of minor muscle/joint pain, stiffness, arthritis, or spasms, and for patients requiring electrotherapy for muscle spasm relaxation, prevention of disuse atrophy, increased local blood circulation, muscle re-education, range of motion maintenance, or post-surgical calf muscle stimulation to prevent venous thrombosis.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
Mettler Sys*Stim 240, ME 240 Combination Neuromuscular Stimulator and Low Level Light Therapy device (K113017)
Submission Summary (Full Text)
{0}------------------------------------------------
K123177
### 510(k) Summary Pursuant to 21 CFR 807.92
3 2013 JUL
- Submitted By: LaserStim, Inc. 1. C/O FDC Services, LLC 8708 Capehart Cove Austin, Texas 78733
- 2. Contact: David C. Furr FDC Services, LLC 8708 Capehart Cove Austin, Texas 78733 512-906-9654
- 3. Product: StimLase 2000 System Regulation: 890.5500 Infrared Lamp Regulation: 882.5890 Transcutaneous electrical nerve stimulator for pain relief Class II Product Code: ILY, GZJ
- 4. Common/Trade Name:
StimLase 2000 System Heat Lamp and TENS Device
- June 30, 2013 5. Date:
#### Description:
The StimLase 2000 System is a therapeutic heating device with the capability to provide simultaneous or separate electrotherapy. The system consists of a console with selector settings, a footpedal control, and a laser wand. The unit can also accommodate two disposable OEM carbon electrodes for electrical stimulation therapy.
{1}------------------------------------------------
#### Intended Use:
Heat Lamp Therapy Feature:
The StimLase 2000 System is intended to be used for topical heating of tissue for the purpose of temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasms. The unit may be used for temporary increase in local blood circulation and/or the temporary relaxation of muscles.
#### Electrotherapy Feature:
The StimLase 2000 System is intended to be used for relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local bloodcirculation, muscle re-education, maintaining or increasing range of motion, and immediate post-surgical stimulation of calf muscles to prevent venous thrombosis.
#### Technological Characteristics:
The StimLase 2000 System is a therapeutic heating device with the capability to provide simultaneous or separate electrotherapy. The system consists of two major components, a base power control console and a hand held laser wand. The unit can also accommodate two disposable OEM carbon electrodes for electrical stimulation therapy.
The base power control console contains a key lock, a calibration port, a power adjustment dial, indicator lamps, and a touch-screen display. Power output can be adjusted from 0 to 550 mW. A foot switch is employed to activate the console. The gallium arsenide semiconductor laser diode is contained in the 7.5 inch hand held laser probe, which is used to apply infrared lamp treatment. The electrical stimulation therapy can be used separately or in conjunction with the therapeutic heating.
The StimLase 2000 System meets all requirements in 21 CFR§1040.10, 21 CFR§1040.11, and 21 CFR§1010 requirements for safety, performance. design, and labeling. In addition, the product has been tested to comply with IEC 60601 and IEC 60825.
{2}------------------------------------------------
#### Substantial Equivalence:
The StimLase 2000 System therapeutic heating feature is substantially equivalent to the Luminex Infrared Lamp System (K082969) and the Mettler Sys*Stim 240, ME 240 Combination Neuromuscular Stimulator and Low Level Light Therapy device (K113017).
Each of the predicate devices have similar technical features, indications for use, and the safety and effectiveness of the devices is equivalent.
### Conclusions:
The predicate devices and the StimLase 2000 System share similar indications, technology, and application. The StimLase 2000 product is equivalent to the predicate device products in key areas of performance that affect safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles an eagle or bird-like figure.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 3, 2013
LaserStim, Incorporated % FDC Services. LLC Mr. David C. Furr Application Correspondent 8708 Capehart Cove Austin, Texas, 78733
Re: K123177
Trade/Device Name: StimLase 2000 System Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: ILY, GZJ Dated: June 03, 2013 Received: June 05, 2013
Dear Mr. Furr:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 – Mr. David C. Furr
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
K123177 510(k) Number: StimLase 2000 System Device Name:
Indications for Use:
Heat Lamp Therapy Feature:
The StimLase 2000 System is intended to be used for topical heating of tissue for the purpose of temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasms. The unit may be used for temporary increase in local blood circulation and/or the temporary relaxation of muscles.
Electrotherapy Feature:
The StimLase 2000 System is intended to be used for relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local blood circulation, muscle re-education, maintaining or increasing range of motion, and immediate post-surgical stimulation of calf muscles to prevent venous thrombosis.
Prescription Use X (per CFR 801.109)
Over-the-counter use
Concurrence of CDRH
Or
Neil R Ogden 2013.07.03 15:12:04 -04'00'
(Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.