K123175 · Conmed Corporation · KCT · Feb 21, 2013 · General Hospital
Device Facts
Record ID
K123175
Device Name
DETACHATIP III PLUS INSTRUMENT TRAY MODEL 3-4343
Applicant
Conmed Corporation
Product Code
KCT · General Hospital
Decision Date
Feb 21, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
DetachaTip® III Plus Multi-Use Instrument Tray (REF 3-4343) is a reusable sterilization tray intended for pre-vacuum steam sterilization, organization, transportation, and storage of enclosed, reusable medical devices. DetachaTip® III Plus Multi-Use Instrument Tray is not intended to maintain sterility; it is intended to be used in conjunction with a validated FDA cleared sterilization wrap in order to maintain sterility of the enclosed medical instruments. DetachaTip® III Plus Multi-Use Instrument Tray is intended for use with DetachaTip® III instruments and ConMed reusable medical devices meeting challenge conditions defined in Table 1. The MAXIMUM recommended load is: 25 lb [11.3 kg].
Device Story
Reusable sterilization tray for organizing, transporting, and storing medical instruments during steam sterilization. Comprised of perforated polyphenylsulfone (Radel®) base and lid, silicone mat, and instrument rack. Used by healthcare providers in clinical settings. Instruments placed in tray; tray wrapped in FDA-cleared sterilization wrap; subjected to pre-vacuum steam sterilization (270°F/132°C). Tray facilitates steam penetration and protects instruments. Does not maintain sterility independently; requires external wrap. Benefits include organized processing and protection of reusable medical devices.
Clinical Evidence
Bench testing only. Device evaluated against ANSI/AAMI ST77:2006/(R)2010 standards for containment devices for reusable medical device sterilization. No clinical data provided.
Indicated for pre-vacuum steam sterilization, organization, transportation, and storage of enclosed, reusable medical devices, specifically DetachaTip® III instruments and ConMed reusable medical devices meeting defined challenge conditions (max load 25 lb).
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Conmed Corporation. The logo features a stylized "C" shape on the left, followed by the word "CONMED" in bold, sans-serif font. Below the word "CONMED" is a horizontal line, and below that is the word "CORPORATION" in a smaller, sans-serif font.
FEB 2 1 2013
# DetachaTip® III Plus Multi-Use Instrument Tray
# 510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92 for 510(k) number K123175 as of January 8, 2013.
#### A. Submitter
ConMed Corporation 525 French Road Utica, NY 13502 Establishment Registration Number: 1320894
## B. Company Contact
Anna D'Lima, RAC Regulatory Affairs Specialist ConMed Corporation 525 French Road Utica, NY 13502 Phone: 315-624-3439 Fax: 315-624-3225 e-mail: annadlima@conmed.com
#### C. Device Name
Trade Name: DetachaTip® III Plus Multi-Use Instrument Tray Common Name: Instrument Tray/Sterilization Tray, Classification Name: Sterilization Wrap Containers, Trays, Cassettes, and other Accessories Regulation Number: 21 CFR 880.6850 Class II КСТ Panel: General Hospital and Personal Use Devices
#### D. Predicate Device Name
ConMed DetachaTip® Instrument Tray ConMed Corporation 510(k) K092414
{1}------------------------------------------------
#### E. Device Description
The ConMed DetachaTip® III Plus Multi-Use Instrument Tray (REF 3-4343) is used to retain reusable instruments during steam sterilization. The tray consists of a perforated tray base and lid made from polypheny sulfone (Radel®). Latches securely hold the tray base, and handles are provided for transportation. A perforated silicone mat protects instruments placed in the bottom of the tray. A Radel® instrument rack separates and secures the DetachaTip® III instruments throughout sterilization processing and/or storage. The tray is not intended to maintain sterility. The instrument tray is to be used with an FDA cleared sterilization wrap.
DetachaTip® III Plus Multi-Use Instrument Tray is designed to fit a standard autoclave, comprised of materials resistant to corrosion, and able to withstand repeat steam sterilization cycles. The tray is designed to withstand repeated use. It is not designed to be serviced or repaired. The materials comprising the tray are biocompatible and will not contaminate reusable medical devices.
Radel® is a registered trademark of Solvay Advanced Polymers
#### F. Intended Use/ Indications for Use
DetachaTip® III Plus Multi-Use Instrument Tray (REF 3-4343) is a reusable sterilization tray intended for pre-vacuum steam sterilization, organization, transportation, and storage of enclosed, reusable medical devices. DetachaTip® III Plus Multi-Use Instrument Tray is not intended to maintain sterility; it is intended to be used in conjunction with a validated FDA cleared sterilization wrap in order to maintain sterility of the enclosed medical instruments.
DetachaTip® III Plus Multi-Use Instrument Tray is intended for use with DetachaTip® III instruments and ConMed reusable medical devices meeting challenge conditions defined in Table 1. The MAXIMUM recommended load is: 25 lb [11.3 kg].
| Table 1 – Reusable Medical Device Challenge Conditions | |
|--------------------------------------------------------|-------------------|
| Minimum<br>Inner Diameter | Maximum<br>Length |
| 5 mm | 45 cm |
| 3 mm | 32 cm |
Sterilize the DetachaTip® III Plus Multi-Use Instrument Tray using the following parameters:
| Method | Cycle | Temperature | Exposure Time | Dry Cycle Time |
|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--------------|---------------|----------------|
| Steam (wrapped) | Pre-vacuum | 270°F(132°C) | 4 minutes | 30 minutes |
| The devices should be sterilized using an FDA cleared wrap indicated for these sterilization<br>cycles. Allow for a 30 minute cool down prior to handling. | | | | |
{2}------------------------------------------------
### H. Performance Data
DetachaTip® III Plus Multi-Use Instrument Tray meets requirements of ANSI/AAMI ST77:2006/(R)2010 Containment devices for reusable medical device sterilization. Based on the nonclinical performance testing, the subject device is as safe and effective as the legally marketed predicate device.
### I. Substantial Equivalence
The DetachaTip® III Plus Multi-Use Instrument Tray and predicate device are substantially equivalent in intended use, device design, materials, performance, effectiveness, and safety. The proposed device and the predicate device are containment devices for sterilization of reusable medical devices using steam sterilization by healthcare providers. The proposed device presents no new issues of safety and efficacy. Note: Prior designs of DetachaTip® instruments shall not be used in the DetachaTip® III Plus Multi-Use Instrument Trays.
| | DetachaTip® III Plus<br>Multi-Use Instrument Tray | ConMed DetachaTip® Instrument Tray |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | | |
| | K123175 | K092414 |
| Model #/ REF | 3-4343 | 1-4327 |
| Description | Containment devices for sterilization of<br>DetachaTip® III Instruments and ConMed<br>reusable medical devices meeting challenge<br>conditions defined in Table 1 using steam<br>sterilization by healthcare providers. The device<br>should be used only in conjunction with FDA cleared<br>wrap indicated for the sterilization cycles. | Containment devices for sterilization of<br>DetachaTip® II Instruments using steam<br>sterilization by healthcare providers. The device<br>should be used only in conjunction with FDA cleared<br>wrap indicated for the sterilization cycles. |
| Sterilization<br>Method | Pre-vacuum steam sterilization | Same |
| Materials | Polyphenylsulfone (Radel), Silicone, Stainless Steel | Same |
| Device Design | Four part instrument tray<br>1) Tray lid<br>2) Inner tray includes 10 wells for DetachaTip® III<br>Instruments and 4 wells for DetachaTip® III<br>Handles<br>3) Silicone mat<br>4) Tray base for ConMed reusable medical devices<br>meeting challenge conditions defined in Table 1 | Four part instrument tray<br>1) Tray lid<br>2) Inner tray includes 10 wells for DetachaTip® II<br>Instruments<br>3) Silicone mat<br>4) Tray base for DetachaTip® II Handles |
| Performance<br>Testing | ANSI/AAMI ST77:2006/(R)2010 Containment<br>devices for reusable medical device sterilization | Same |
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 21, 2013
Ms. Anna D' Lima, RAC Regulatory Affairs Specialist ConMed Corporation 525 French Road UTICA NY 13502
Re: K123175
Trade/Device Name: DetachaTip® III Plus Multi-Use Instrument Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: January 8, 2013 Received: January 11, 2013
Dear Ms. D' Lima:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Image /page/4/Picture/7 description: The image shows a stylized graphic with overlaid text. The text appears to read "for" at the top left, followed by a signature or stylized name in the center, and the word "with" above the word "uroutts." The graphic includes geometric shapes, with a maze-like pattern on the left and a triangular shape on the right, both filled with horizontal lines.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
#### 510(k) Number (if known): K123175
Device Name: DetachaTip® III Plus Multi-Use Instrument Tray
DetachaTip® III Plus Multi-Use Instrument Tray (REF 3-4343) is a reusable sterilization tray intended for pre-vacuum steam sterilization, organization, transportation, and storage of enclosed, reusable medical devices. DetachaTip® III Plus Multi-Use Instrument Tray is not intended to maintain sterility; it is intended to be used in conjunction with a validated FDA cleared sterilization wrap in order to maintain sterility of the enclosed medical instruments.
DetachaTip® III Plus Multi-Use Instrument Tray is intended for use with DetachaTip® III instruments and ConMed reusable medical devices meeting challenge conditions defined in Table 1. The MAXIMUM recommended load is: 25 lb [11.3 kg].
| Table 1 - Reusable Medical Device Challenge Conditions | | |
|--------------------------------------------------------|-------------------|--|
| Minimum<br>Inner Diameter | Maximum<br>Length | |
| 5 mm | 45 cm | |
| 3 mm | 32 cm | |
Sterilize the DetachaTip® III Plus Multi-Use Instrument Tray using the following parameters:
| Method | Cycle | Temperature | Exposure Time | Dry Cycle Time |
|----------------------------------------------------------------------------------------------|------------|--------------|---------------|----------------|
| Steam (wrapped) | Pre-vacuum | 270°F(132°C) | 4 minutes | 30 minutes |
| The devices should be sterilized using an FDA cleared wrap indicated for these sterilization | | | | |
cycles. Allow for a 30 minute cool down prior to handling.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)[iZabet]
Page _ of
Elizabeth H. Clavelle
2013.02.20 15:19:01 -05'00'
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number;
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.