← Product Code [LLZ](/productcode/LLZ) · K123174

# CENTRICITY PACS-IW WITH UNIVERSAL VIEWER (K123174)

_GE Healthcare · LLZ · Nov 16, 2012 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K123174

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Nov 16, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Megapixel resolution and meet other technical specifications reviewed and accepted by the FDA. Typical users of this system are trained professionals, including but not limited to radiologists, physicians, nurses, medical technicians, and assistants.

## Device Story

Centricity PACS-IW with Universal Viewer is a software-only PACS component for medical image management. It receives DICOM-compliant images from modalities (CT, MR, US, NM, CR, MG, DX, PET/PT, XA, IO, RF, SC, VL). Deployed over TCP/IP networks on off-the-shelf hardware, it provides a unified workspace for image review, manipulation, and interpretation. Features include 'smart' hanging protocols based on DICOM tags, bookmarking of display states, and measurement tools (Cardio Thoracic Ratio, Ellipse). An optional Zero Footprint (ZFP) viewer allows web-based access for clinicians without local installation. The system does not generate images but facilitates clinical decision-making by providing diagnostic-quality visualization. It supports integration with GE's IDI Mammo and Advantage Workstation (AW) Server. The ZFP viewer is not intended for primary diagnosis.

## Clinical Evidence

No clinical studies were required to support substantial equivalence. Evidence consists of bench testing, including risk analysis, requirements/design reviews, usability analysis, and verification testing (unit, integration, performance, regression, system) and validation (simulated use).

## Technological Characteristics

Software-only device running on off-the-shelf computer hardware. Utilizes TCP/IP networking. Supports DICOM imaging standards. Features include ZFP web-based viewer, hanging protocols, and measurement tools. Software developed under moderate level of concern. No specific materials or energy sources as it is a software-only PACS application.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- GE Healthcare Centricity PACS-IW ([K121387](/device/K121387.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a date written in a bold, sans-serif font. The date is 'NOV 16 2012', indicating November 16th, 2012. The letters and numbers are clearly legible against a white background. The text appears to be extracted from a document or label.

25

![GE logo]

GE Healthcare 510(k) Premarket Notification Submission Centricity PACS-IW with Universal Viewer Section 5: 510(k) Summary

#### 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date: October 8, 2012

Submitter: GE Healthcare 540 West Northwest Highway Barrington, IL 60010

Primary Contact Person:

Cheryl Bork Regulatory Affairs Manager GE Healthcare Phone: 847-277-6038 Fax: 847-277-4506

Secondary Contact Person:

Jeme Wallace Regulatory Affairs Director GE Healthcare Phone: 847-277-4468 Fax: 847-939-1446

Device Trade Name: Centricity PACS-IW with Universal Viewer

Common/Usual Name: Picture Archiving and Communication System

Classification Names: 21 CFR 892.2050, System, Image Processing, Radiological

Product Code: LLZ

Predicate Device: K121387 - GE Healthcare Centricity PACS-IW

Device Description:

Centricity PACS-IW with Universal Viewer is an Internet based medical image display and interpretation software product that is part of a picture archiving and communications system. It provides users with capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images (including digital mammograms).

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The letters and the circle are black, creating a strong contrast.

#### 510(k) Premarket Notification Submission Centricity PACS-IW with Universal Viewer Section 5: 510(k) Summary .

Device Description: (cont.) Centricity PACS-IW with Universal Viewer includes features to access and manage medical imaging studies from Computed Tomography (CT), Magnetic Resonance (MR), Ultrasound (US), Nuclear medicine (NM). Computerized radiography (CR). Diaital mammography (MG),Digital x-ray (DX), Positron Emission Tomography (PET/PT), X-Ray Angiography (XA), Digital Intra-oral X-Ray (IO), Radiofluoroscopic x-ray (RF), Secondary Capture Images (SC), Visible Light (VL) Endoscopic, Microscopic and Photographic Image Storage, Slide Coordinates Microscopic Image Storage, Presentation States (PS), Key Image Notes (KIN) and other DICOM imaging modalities.

> Centricity PACS-IW with Universal Viewer is designed to be deployed over conventional TCP/IP networking infrastructure available in in most healthcare organizations and utilizes commercially available computer platforms and operating systems.

> The system does not produce any origing| medical images. All images located on the Centricity PACS, PACS-IW, and Enterprise Archive systems have been received from DICOM compliant modalities and/or image acquisition systems.

> The Universal Viewer Zero Foot Print (ZFP) clinical viewer is an optional viewer which allows trained professionals to display and manipulate images stored in Centricity Enterprise Archive or other DICOM archive devices. These trained professionals include but are not limited to physicians, radiologists, nurses, medical technicians, and assistants.

> The Universal Viewer Zero Foot Print option is not intended for primary diagnosis.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular border. There are three small decorative elements resembling stylized leaves or flourishes placed around the circle's perimeter.

510(k) Premarket Notification Submission Centricity PACS-IW with Universal Viewer Section 5: 510(k) Summary

Intended Use: Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations.

> Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Megapixel resolution and meet other technical specifications reviewed and accepted by the FDA.

Typical users of this system are trained professionals, including but not limited to radiologists, physicians, nurses, medical technicians, and assistants.

Technology: The Centricity PACS-IW with Universal Viewer device employs the same fundamental scientific technology as its predicate device, Centricity PACS-IW cleared under K121387, with the following modifications:

- . Modified viewer to present a common unified workspace for radiologists and clinicians to perform the review, manipulation and diagnostic interpretation of images and other information generated by acquisition.
- . Provides integration to Centricity PACS backend. The PACS-IW with Universal Viewer can read image data and other patient information from Centricity PACS backend.
- Implementation of a "Bookmark" feature for . image display, that enables the user to capture the Display State of an exam context as it exists at any particular time, and then restores this same state at a later time.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The logo is black and white.

- Enhanced hanging protocols ("smart" hanging . protocols) for image display, which will assess information provided in the DICOM tags to evaluate the type of study that is being displayed to determine the most appropriate method for the study to be hung.
- Enhanced image annotations and . measurements to include the ability to calculate Cardio Thoracic Ratio (CTR) (web client only) and Ellipse (ZFP only).
- Supports interface with GE Healthcare's IDI . Mammo product, in place of Cedara Mammo.
- . Supports interface with GE Healthcare's Advantage Workstation (AW) Server advanced visualization tools.

### Technology (cont.)

- Hardware minimum specifications were . modified as a result of technology advancements and obsolescence issues.
- The optional ZFP (Zero Footprint) Clinician Viewer . is a non-diagnostic DICOM image and results viewer. Zero footprint:
	- image enables an EMR or other similar o applications.
	- displays image data from GE O Healthcare's Centricity Enterprise Archive.
	- can be accessed on a client device o without requiring the user to override security settings or initiate an installation.

Centricity PACS-IW with Universal Viewer receives medical images and other information from various data sources. The information can be stored, communicated, processed and displayed within the

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined in a stylized script, enclosed within a circular frame. The letters and the frame are solid black, contrasting with the white background.

510(k) Premarket Notification Submission Centricity PACS-IW with Universal Viewer Section 5: 510(k) Summary

system or across computer networks at distributed locations, the same as its predicate devices. Centricity PACS-IW with Universal Viewer is a software-only device that runs on commercially available off-the-shelf computer hardware platforms.

The Centricity PACS-IW with Universal Viewer device will continue to have an intended use and functionality fitting within the definition of 21 CFR 892.2050. Picture Archiving and Communication Systems, Product Code LLZ.

#### Determination of Substantial Equivalence:

#### Summary of Non-Clinical Tests

The software documentation was provided at a moderate level of concern following the FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". Centricity PACS-IW with Universal Viewer complies with voluntary standards as detailed in this premarket notification submission. The following quality assurance measures were applied to the development of the system:

- . Risk Analysis
- Requirements Reviews .
- Design Reviews .
- Usability Analysis .
- . Testing on unit level (Verification)
- Integration testing (Verification) .
- . Performance testing (Verification)
- . Regression testing (Verification)
- System testing (Verification) ♥
- Simulated use testing (Validation) .

#### Summary of Clinical Tests:

The subject of this premarket notification submission, Centricity PACS-IW with Universal Viewer, did not require clinical studies to support substantial equivalence.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined within a circular border. The letters are stylized and connected, creating a unique and recognizable design. The logo is in black and white.

510(k) Premarket Notification Submission Centricity PACS-IW with Universal Viewer Section 5: 510(k) Summary

Comparison of the Intended Uses, the technological Conclusion: characteristics, and performance specifications demonstrate the functional equivalence of the subject device to the predicate device. Verification and Validation testing results demonstrate that no adverse effects have been introduced by these differences.

> Information provided in this premarket notification submission supports the Centricity PACS-IW with Universal Viewer medical device to be as safe, as effective and substantially equivalent to its predicate device.

{6}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its head facing left and its wings spread. The eagle is composed of three thick, curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

November 16, 2012

Ms. Cheryl Bork Regulatory Affairs Manager GE Healthcare-HCIT 540 W. Northwest Highway BARRINGTON IL 60010

Re: K123174

Trade/Device Name: Centricity PACS-IW with Universal Viewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 8, 2012 Received: October 25, 2012

Dear Ms. Bork:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{7}------------------------------------------------

Page 2 - Ms. Bork

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quarty of would provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Michael D. O'Hara

Janine M. Morris Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

#### Indications for Use

510(k) Number (if known): K123174

Device Name: Centricity PACS-IW with Universal Viewer

Indications for Use:

Lentricity PACS-IW with Universal viewer is a software only product that is part of a Picture Archiving and Communication System (PACS) medical device.

Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Megapixel resolution and meet other technical specifications reviewed and accepted by the FDA.

Typical users of this system are trained professionals, including but not limited to radiologists, physicians, nurses, medical technicians, and assistants.

Prescription Use Yes (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Michael D. O'Hara

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) K123174

Page 1 of

---

**Source:** [https://fda-staging.innolitics.com/device/K123174](https://fda-staging.innolitics.com/device/K123174)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).