VARI-PASS VARIABLE LENGTH ACCESS SHEATH

K123170 · Olympus Surgical Technologies America · KNY · Feb 4, 2013 · Gastroenterology, Urology

Device Facts

Record IDK123170
Device NameVARI-PASS VARIABLE LENGTH ACCESS SHEATH
ApplicantOlympus Surgical Technologies America
Product CodeKNY · Gastroenterology, Urology
Decision DateFeb 4, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.5130
Device ClassClass 2

Indications for Use

The Gyrus ACMI Variable Length Access Sheath is intended to be used to permit direct passage of catheters or other devices for the purpose of performing diagnostic and surgical procedures (e.g.,nephrostomy, cystoscopy, ureteroscopy, etc.) in the urinary tract.

Device Story

Single-use, sterile, adjustable-length access sheath and tapered dilator; used in urological procedures (nephrostomy, cystoscopy, ureteroscopy) to facilitate passage of catheters/instruments into urinary tract. Dilator attaches to sheath via locking luer; luer port allows irrigation or contrast fluid injection. Radiopaque for visualization. Operated by physicians in clinical settings. Benefits include adjustable length to match procedure requirements and simplified instrument access.

Clinical Evidence

Bench testing only. Performance data demonstrated no significant difference compared to predicate devices. Biocompatibility testing performed in compliance with ISO-10993.

Technological Characteristics

Sheath: PTFE with BiOs. Dilator: HDPE with BaS02. Features: adjustable length, tapered dilator, locking luer connector with irrigation/injection port. Radiopaque. Single-use, sterile. Biocompatibility per ISO-10993.

Indications for Use

Indicated for patients requiring diagnostic or surgical procedures in the urinary tract, including nephrostomy, cystoscopy, and ureteroscopy, where direct passage of catheters or other devices is necessary.

Regulatory Classification

Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 123170 Variable Length Access Sheath Olympus Surgical Technologies of America Gyrus ACMI, Inc. 136 Turnpike Road Southborough, MA 01772 #### Special 510(k) Notification FEB 0 4 2013 ## 510(K) SUMMARY #### General Information Manufacturer Contact Person: Date Prepared: Classification Name: Regulation / CFR Citation Number Product Code Class Review Panel Trade Name: Gyrus ACMI Inc. 136 Turnpike Rd. Southborough, MA 01772-2104 ERN: 3003790304 Deana Boushell (508) 804-2752 deana.boushell(@)Olympus-OSTA.com October 28, 2012 Urological catheter and accessories 876.5130 KNY II Gastroenterology/Urology Urological catheter and accessories Gyrus ACMI Vari-PassTM Adjustable Length Access Sheath Generic/Common Name: Predicate Device The predicate devices include: 1. Gyrus ACMI Snap-N-Peel™ Introducer 2. Gyrus ACMI Uropass® K981611 K021593 #### Intended Use The Gyrus ACMI Variable Length Access Sheath is intended to be used to permit direct passage of catheters or other devices for the purpose of performing diagnostic and surgical procedures (e.g.,nephrostomy, cystoscopy, ureteroscopy, etc.) in the urinary tract. #### Product Description Gyrus ACMI Vari-Pass™ Adjustable Length Access Sheath is a single use, sterile disposable product. The device consists of a sheath (manufactured from PTFE with BiOs) and dilator (manufactured from HDPE with BaS02) to facilitate the passage of catheters and other urological devices in the urinary tract. The sheath length is adjustable to the size necessary for the procedure. The dilator is attached to the sheath via a locking leur. The leur connector also allows for injection or irrigation or contrast fluid. For ease of visualization, the dilator and sheath are radiopaque. {1}------------------------------------------------ ## Technological features and Substanial Equivalence The proposed Gyrus ACMI Vari-Pass™ Adjustable Length Access Sheath incorporates the same technological features (peel away sheath, tapered dilator and locking mechanism with port) as previously cleared devices. The proposed Gyrus ACMI Vari-Pass™ Adjustable Length Access Sheath has the same Indications for use as previously cleared devices. The Gyrus ACMI Vari-Pass™ Adjustable Length Access Sheath uses similar materials compared to the predicate. The sheath material was changed to PTFE from FTE and though the dilator matieral is the same (HDPE) it is being sourced from a different supplier. These materials have an established history of successful clinical application in Urology. Biocompatibility testing has been performed in compliance to the relevant requirements of ISO-10993. ## Nonclinical Testing and Performance Data Nonclinical testing has been conducted and performance data demonstrates no significant difference in the performance of Gyrus ACMI Vari-Pass™ Adjustable Length Access Sheath as compared to predicate devices. #### Summary The proposed Gyrus ACMI Vari-Pass™ Adjustable Length Access Sheath, as described in this submission, is substantially equivalent to the predicates in intended use, materials, principles of operation and fundamental scientific technology . Non-clinical testing confirms that no new issues of safety and effectiveness have been raised by the proposed device modifications. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community and support. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 4, 2013 Olympus Surgical Technologies America Gyrus ACMI, Inc. % Ms. Deana Boushell Senior Regulatory Specialist 136 Turnpike Road SOUTHBOROUGH MA 01772 Re: K123170 Trade/Device Name: Gyrus ACMI Vari-Pass™ Adjustable Length Access Sheath Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: KNY Dated: December 3, 2012 Received: December 4, 2012 Dear Ms. Boushell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {3}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Herbert ﻡ Lerner -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement 510(k) Number: 23170 Device Name: Gyrus ACMI Vari-Pass™ Adjustable Length Access Sheath Indications for Use: Gyrus ACMI Vari-Pass™ Adjustable Length Access Sheath is intended to be used to permit direct passage of catheters or other devices for the purpose of performing diagnostic and surgical procedures (e.g .; nephrostomy, cystoscopy, ureteroscopy, etc.) in the urinary tract. Prescription Use: __ X OR Over-the-Counter Use: (Per 21 CFR 801.109) (PI.EASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Herbert P.Lerner -S (Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices 510(k) Number K123170
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...