K123168 · Quanta System Spa · GEX · Dec 20, 2012 · General, Plastic Surgery
Device Facts
Record ID
K123168
Device Name
Q-PLUS T+ IPL LASER SYSTEM
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 20, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Q-Plus T + IPL laser system is a laser surgical instrument for use in general surgery and dermatology intended for the incision, ablation, and vaporization of soft tissue and for the treatment of vascular and pigmented lesions, tattoo removal, and inflammatory acne (acne vulgaris).
Device Story
Q-Plus T + IPL Laser System is a multi-source surgical laser for dermatology. Inputs: electrical power; user-selected laser/IPL parameters. Operation: utilizes Q-Switched Nd:YAG (1064nm/532nm), Q-Switched Ruby (694nm) lasers, and IPL (400-1200nm range) to deliver energy to soft tissue via articulated arm or IPL handpiece. Output: laser/light energy for tissue incision, ablation, vaporization, or photothermolysis. Used in clinical settings by physicians. Safety features include remote interlock, key switch, and emergency stop. Output allows clinicians to perform dermatological procedures; benefits include removal of tattoos, pigmented/vascular lesions, and acne treatment.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence based on identical technological characteristics and performance specifications to predicate devices.
Technological Characteristics
System includes Q-Switched Nd:YAG (1064nm/532nm) and Ruby (694nm) lasers, and IPL handpiece. Optical delivery via articulated arm with 2, 3, 4, 6 mm spot sizes. Includes power supply, cooling system, and microprocessor-based controller. Safety features: remote interlock, key switch, emergency stop. Connectivity: standalone. Sterilization: not specified.
Indications for Use
Indicated for patients requiring tattoo removal (various ink colors), benign pigmented lesion removal (e.g., Café au lait, solar lentigo, Becker Nevus, Ota/Ito Nevus, Mongolian spot), hair removal, photocoagulation of dermatological vascular lesions (e.g., telangiectasia, facial/leg veins), and treatment of inflammatory acne (acne vulgaris).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K123168
Quanta System SPA - Traditional 510(k) - Q-Plus T + IPL Laser System Section 7: 510(k) Summary
## Section 7: 510(k) Summary
## C25 2 0 2012
#### Introduction:
This document contains the 510(k) Summary for the Q-Plus T +IPL laser system. The content of this summary is based on the requirements of 21 CFR 807.92(c).
Applicant / Manufacturer Name and Address:
510(k) Contact Person:
Quanta System SPA Via IV Novembre, 116 Solbiate Olona (VA) Italy, 21058
October 5, 2012
21 CFR 878.4810
Class II
GEX
Maurizio Bianchi Requlatory Affairs Manager Quanta System SPA
Q-Plus T + IPL Laser System
Email: Maurizio.bianchi@quantasystem.com Phone: +39-0331-376797 Fax: +39-0331-367815
Laser surgical instrument for use in general .
and plastic surgery and in dermatology
Date Prepared:
Device Name:
Classification:
Classification Name:
Regulation Number
Product Code:
Predicate Devices:
The Q-Plus T + IPL Laser System is claimed to be substantially equivalent to the following legally marketed predicate devices:
- K073549 Q-Plus T manufactured by Quanta System SPA
- K083207 Ultrawave III EX 1320 manufactured by Quanta System SPA
## Performance Standards:
There are no mandatory performance standards for this device.
## General Device Description:
The Q-Plus T + IPL Laser System is intended for the incision, ablation and vaporization of soft tissue for general dermatology. The device includes a Q-Switched Nd:YAG laser source with a max energy of 1000mJ at 1064nm and a max energy of 500mJ at 532nm, a Q-Switched Ruby laser source with a max energy of 1000mJ at 694nm, and an intense pulse light (IPL) hand The Q-Switched lasers are intended for treatment of pigmented lesions and tattoo piece. removal. The optical delivery system for the three laser wavelengths is an articulated arm with replaceable hand pieces with 2, 3, 4, and 6 mm spot sizes. The IPL hand piece, depending on the particular wavelength filter used, is intended for hair removal, treatment of dermatological vascular lesions, benign pigmented lesions and inflammatory acne (acne vulgaris). The IPL hand pieces are provided in a fixed or a replaceable configuration.
{1}------------------------------------------------
## Quanta System SPA – Traditional 510(k) – Q-Plus T + IPL Laser System Section 7: 510(k) Summary
The Q-Plus T + IPL laser system incorporates several safety features, including a remote interlock, a key switch and an emergency stop button.
## Indications for Use:
| Source selection<br>and wavelength | Indications for Use |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nd:YAG laser<br>(1064nm)<br>Nd:YAG laser<br>(532nm) | These wavelengths are absorbed by natural chromophores, like melanin, hemoglobin and water and by exogen pigments like tattoo ink. They are indicated for:<br>Tattoo removal:<br>1064nm: suggested for dark blue and black ink<br>532nm: suggested for red, orange, yellow, and purple ink<br>Pigmented lesion removal (benign):<br>Café au lait spot<br>Ephalides, solar lentigo (lentigines)<br>Becker Nevus<br>Ota Nevus<br>Nevus spilus |
| Ruby laser<br>(694nm) | This wavelength is absorbed by natural chromophores, like melanin, hemoglobin and water and by exogen pigments like tattoo ink. It is indicated for:<br>Tattoo removal:<br>Suggested for blue, sky blue, black, green and violet ink<br>Pigmented lesion removal (benign):<br>Café au lait spot<br>Ephalides, solar lentigo (lentigines)<br>Becker Nevus<br>Ota and Ito Nevus<br>Nevus spilus<br>Mongolian spot |
| IPL 650-1200nm<br>IPL 625-1200nm | Indicated for hair removal |
| IPL 590-1200m<br>IPL 550-1200nm<br>IPL 570-1200nm | Indicated for photocoagulation of dermatological vascular lesion (i.e. face telangiectasia), photothermolysis of blood vessels (treatment of facial and leg veins), and treatment of benign pigmented lesions |
| IPL 550-1200nm | Indicated for face vascular lesions and solar lentigines (benign pigmented lesions) |
| IPL 400-1200nm | Indicated for inflammatory acne (acne vulgaris) |
## Comparison of Technological Characteristics:
The Q-Plus T + IPL Laser System has substantially equivalent technological characteristics to the The ten subsystems in the Q-Plus T + IPL laser system contain all nine predicate devices. subsystems of the Q-Plus T (K073549) with the addition of the IPL hand piece subsystem that is cleared with the Ultrawave III EX 1320 (K083207).
The Q-Switched Nd:YAG laser source with a max energy of 1000mJ at 1064nm and a max energy of 500mJ at 532nm is identical to the Q-Switched Nd:YAG laser source utilized in the Q-Plus T device. Also, the Q-Switched Ruby laser source with a max energy of 1000mJ at 694nm is identical the Q-Switched Ruby laser source utilized in the Q-Plus T device. The optical delivery system for the three laser wavelengths is an articulated arm that is also used in the Q-Plus T device. The IPL hand piece portion of the Q-Plus T + IPL Laser System is the same as the IPL hand piece portion of the Ultrawave III EX 1320. In addition, the Q-Plus T + IPL laser system includes a power supply, a cooling system, an optical delivery system, a microprocessor based
{2}------------------------------------------------
Quanta System SPA - Traditional 510(k) - Q-Plus T + IPL Laser System Section 7: 510(k) Summary
controller, and safety features that are all either similar or identical to those in the Ultrawave III EX 1320 and Q-Plus T devices.
The laser specifications for the Q-Plus T + IPL laser system are substantially equivalent to the laser specifications for its identified predicate device with respect to the laser source, wavelengths, maximum energy, spot size, fluence, pulse width, repetition rate, beam delivery, power monitor, actuator, and aiming beam.
## Comparison of Intended Use:
The Q-Plus T + IPL laser system is a laser surgical instrument for use in general surgery and dermatology intended for the incision, ablation, and vaporization of soft tissue and for the treatment of vascular and pigmented lesions, tattoo removal, and inflammatory acne (acne vulgaris).
The intended use of the laser portion of the Q-Plus T + IPL laser system is the same as the intended use of the previously cleared Q-Plus T as no changes have been made with regard to performance or technology to the Q-Plus T. Furthermore, no changes have been made to the IPL hand piece portion of the subject device since its initial clearance in K051113 and its clearance for use with the Ultrawave III EX 1320 in K083207. Therefore, the intended use of the IPL hand piece used with the Q-Plus T + IPL laser system is substantially equivalent to the intended use of the IPL hand piece used with the Ultrawave III EX 1320.
## Substantial Equivalence:
The Quanta System Q-Plus T + IPL laser system is as safe and effective as the predicate devices. The Q-Plus T + IPL laser system has the same intended use and similar technological characteristics and principles of operation as its predicate devices. The minor technological differences between the Q-Plus T + IPL laser system and its predicate devices raise no new issues of substantial equivalence or safety and effectiveness. Thus, the Q-Plus T + IPL laser system is substantially equivalent to its identified predicate devices.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is depicted in a simple, black and white design.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002
Letter dated: December 20, 2012
Quanta System Spa % Mr. Maurizio Bianchi Regulatory Affairs Manager Via IV Novembre, 116 Solbiate Olona, VA Italy 21058
Re: K123168
Trade/Device Name: Q-Plus T + IPL Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II Product Code: GEX Dated: October 05, 2012 Received: October 09, 2012
Dear Mr. Bianchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Maurizio Bianchi
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1.1.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
K 123 68 510(k) Number (if known):
Device Name: Q-Plus T + IPL Laser System
Indications for Use:
| Source selection<br>and wavelength | Indications for Use |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nd:YAG laser<br>(1064nm)<br>Nd:YAG laser<br>(532nm) | These wavelengths are absorbed by natural chromophores, like melanin, hemoglobin and water and by exogen pigments like tattoo ink. They are indicated for:<br>Tattoo removal:<br>1064nm: suggested for dark blue and black ink<br>532nm: suggested for red, orange, yellow, and purple ink<br>Pigmented lesion removal (benign):<br>Café au lait spot<br>Ephalides, solar lentigo (lentigines)<br>Becker Nevus<br>Ota Nevus<br>Nevus spilus |
| Ruby laser<br>(694nm) | This wavelength is absorbed by natural chromophores, like melanin, hemoglobin and water and by exogen pigments like tattoo ink. It is indicated for:<br>Tattoo removal:<br>Suggested for blue, sky blue, black, green and violet ink<br>Pigmented lesion removal (benign):<br>Café au lait spot<br>Ephalides, solar lentigo (lentigines)<br>Becker Nevus<br>Ota and Ito Nevus<br>Nevus spilus<br>Mongolian spot |
| IPL 650-1200nm<br>IPL 625-1200nm | Indicated for hair removal |
| IPL 590-1200m<br>IPL 550-1200nm<br>IPL 570-1200nm | Indicated for photocoagulation of dermatological vascular lesion (i.e. face telangiectasia), photothermolysis of blood vessels (treatment of facial and leg veins), and treatment of benign pigmented lesions |
| IPL 550-1200nm | Indicated for face vascular lesions and solar lentigines (benign pigmented lesions) |
| IPL 400-1200nm | Indicated for inflammatory acne (acne vulgaris) |
Prescription Use_ X AND/OR Over-the Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden 2012.12.19 11:23:24 -05'00'
ITIES OF DEFENSE EXAMINATION (ODE)
(Division Sign-Off) Division of Surgical D 510(k) Number_
Page 1 of __ 1________________________________________________________________________________________________________________________________________________________________
Page 23 of 407
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.