ACME SURGICAL FACE MASK WITH EAR-LOOP MODEL YN-501 AB/W/G/P
K123115 · Acme Filter Mask, Inc. · FXX · May 20, 2013 · General, Plastic Surgery
Device Facts
Record ID
K123115
Device Name
ACME SURGICAL FACE MASK WITH EAR-LOOP MODEL YN-501 AB/W/G/P
Applicant
Acme Filter Mask, Inc.
Product Code
FXX · General, Plastic Surgery
Decision Date
May 20, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
Surgical Face Mask is Device that is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operation room personnel from transfer of microorganisms, body fluids and particulate material.
Device Story
Flat-pleated 3-ply surgical face mask; inner/outer layers spunbonded polypropylene; middle layer melt-blown polypropylene filter; malleable aluminum nosepiece; elastic ear-loops. Used by operating room personnel during surgical procedures. Provides barrier protection against microorganisms, body fluids, and particulates. Protects both patient and wearer. Disposable, non-sterile device.
Clinical Evidence
No clinical data. Bench testing only: Fluid Resistance (ASTM F1862-05), Bacterial Filtration Efficiency (ASTM F2101-01), Particulate Filtration Efficiency (ASTM F1215), Differential Pressure (MIL M 36954 C), Flammability (16CFR 1610), and Biocompatibility (ISO 10993).
Technological Characteristics
3-ply construction: spunbonded polypropylene layers with melt-blown polypropylene filter. Malleable aluminum nosepiece. Elastic ear-loops. Tested per ASTM F1862-05 (fluid resistance), ASTM F2101-01 (BFE), ASTM F1215 (PFE), MIL M 36954 C (Delta-P), 16CFR 1610 (flammability), and ISO 10993 (biocompatibility).
Indications for Use
Indicated for operating room personnel to protect patients and staff from transfer of bacteria, body fluids, and particulate material during surgical procedures. Device is non-sterile and disposable.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Modern Healthcare Corp. Surgical Face Mask (K063043)
Submission Summary (Full Text)
{0}------------------------------------------------
## Acme Filter Mas
No.57, Hu Shan Road, Yingge District, New Taipei City, 23941, Taiwan, R.O.C.
## " 510(k) SUMMARY "
MAY 2 0 2013
September 23, 2012
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92
The assigned 510(k) number is : __ K123115
### Submitter's Name: Acme Filter Mask Inc.
No.57, Hu Shan Road, Yingge District, New Taipei City, 23941,Taiwan, R.O.C.
### Date summary prepared:
Device Name:
- Mask, Surgical Classification name: .
- Classification number: FXX, Class II ●
- Regulation Number: 878.4040 .
- Surgical Face Mask with Ear-Loop ● Proprietary name:
YN-501AB, YN-501AW, YN-501AG for Blue, White, Green Product model: One size for different colors (Blue, White, Green)
- Common name of device: Surgical Face Mask, Disposable ●
- Surgical Face Mask, Type: Tie-on, Ear-loop, K063043 . Predicate Device:
- Official Correspondent: Dr. Jen. Ke-Min ●
E-mail: ceirs.jen@msa.hint.net (Tel) +886-3-5208829; (Fax) +886-3-5209783
#### Description of the device:
Acme Filter Mask Inc. Surgical Face Mask with Ear-Loop is flat pleated 3-ply (at least) masks with an inner and outer layer (spunbonded polypropylene) that sandwich a melt blown polypropylene filter material, also with elastic loops. The nosepiece for all Acme Filter Mask Inc. Surgical Face Masks with Ear-Loop are malleable aluminum wire. All the materials used in the construction of the Acme Filter Mask Inc. Surgical Face Masks with Ear-Loop are being used in currently marked devices.
### Labels/Labeling:
This device will be marked to medical device suppliers, Dentist and Doctor Officers, clinics, Emergency Response Professionals, Hospitals and other healthcare professional for the Intended Use purpose below:
{1}------------------------------------------------
## Acme Filter Mask Inc.
No.57, Hu Shan Road, Yingge District, New Taipei City, 23941, Taiwan, R.O.C.
### Intended Use:
Surgical Face Mask is Device that is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operation room personnel from transfer of microorganisms, body fluids and particulate material.
### Comparison to Predicate Devices:
Acme Filter Mask Inc. Surgical Face Mask with Ear-Loop is substantially equivalent for the safety and effectiveness to the Modern Healthcare Corp. Surgical Face Mask, type: Tie-on and Ear-loop:
| | Modern Healthcare Corp.<br>Surgical Face Mask<br>(K063043) | ACME Filter Mask Inc.<br>Surgical Face Mask<br>(K123115) |
|------------------------------------------------------------|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| <i>Similarity:</i> | | |
| Fluid Resistance | Fluid Resistance | Fluid Resistance |
| Flammability<br>Class | Class I<br>(No Flame Spread) | Class I<br>(No Flame Spread) |
| Regulatory Class | Class II<br>(ASTM2100-04 Low Barrier) | Class II |
| BFE(%) | Higher than 99% | Higher than 99.9% |
| <i>Difference:</i> | | |
| Type | Tie-on and Ear-loop<br>(Green, White, Blue, Pink) | Ear-Loop<br>(Green, White, Blue) |
| Delta-P | Average 2.6 | Average 3.33 ( $mmH_2O/cm^2$ )<br>for Air Exchange Pressure |
| Particulate<br>Filtration<br>Efficiency<br>Performance (%) | Average 96.8% at 0.1 micron | Average 94.79% for Solid Aerosol<br>Filtration Efficiency<br>More than 99.5% for Viral<br>Filtration Efficiency |
### Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
- Fluid Resistance (ASTM F1862-05): Synthetic Blood Penetration Resistance Test 】.
- Filtration Efficiency: Bacterial Filtration Efficiency (BFE) Test (ASTM F2101-01) II. and Particulate Filtration Efficiency (latex Particle challenge) (ASTM F1215)
- Differential Pressure (Delta-P) Test (MIL M 36954 C) III.
- Flammability Test (16CFR 1610) IV.
- Biocompatibility per ISO 10993-5 /-10 V.
It was our conclusion that performance testing met all relevant requirements of the aforementioned test standard.
{2}------------------------------------------------
# Acme Filter Mask Inc.
No.57, Hu Shan Road, Yingge District, New Taipei City, 23941, Taiwan, R.O.C.
### Discussion of Clinical Tests Performed:
Not Applicable
### Conclusions:
Acme Filter Mask Inc. Surgical Face Mask with Ear-Loop has the same intended use and technological characteristics as the predicated devices Modern Healthcare Corp., Surgical Face Mask, type: Tie-on and Ear-loop (K063043). Especially, the predicate device's types are Tie-on and Ear-loop; and the subject device just for Ear-loop. Besides, the bench testing contained in this submission supplied demonstrates that the technological characteristics do not raise any new question of safety or effectiveness.
Thus the new device is substantially equivalent to the predicate devices.
{3}------------------------------------------------
# Acme Filter Mask Inc
No.57, Hu Shan Road, Yingge District, New Taipei City, 23941, Taiwan, R.O.C.
#### COMPARISON INFORMATION 7.
We place the 510K information for the predicate device thereafter this section.
### Comparison to Predicate Devices:
Acme Filter Mask Inc. Surgical Face Mask with Ear-Loop is substantially equivalent for the safety and effectiveness to the Modern Healthcare Corp. Surgical Face Mask, type: Tie-on and Ear-loop:
| Item | Modern Healthcare Corp.<br>Surgical Face Mask<br>(K063043) | ACME Filter Mask Inc.<br>Surgical Face Mask<br>(K123115) |
|------------------------------------------------------------|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Similarity: | | |
| Fluid Resistance | Fluid Resistance | Fluid Resistance |
| Flammability<br>Class | Class I<br>(No Flame Spread) | Class I<br>(No Flame Spread) |
| Regulatory Class | Class II<br>(ASTM2100-04 Low Barrier) | Class II |
| BFE(%) | Higher than 99% | Higher than 99.9% |
| Difference: | | |
| Type | Tie-on and Ear-loop<br>(Green, White, Blue, Pink) | Ear-Loop<br>(Green, White, Blue) |
| | | |
| Delta-P | Average 2.6 | Average 3.33 (mmH2O/cm²)<br>for Air Exchange Pressure |
| Particulate<br>Filtration<br>Efficiency<br>Performance (%) | Average 96.8% at 0.1 micron | Average 94.79% for Solid Aerosol<br>Filtration Efficiency<br>More than 99.5% for Viral<br>Filtration Efficiency |
### Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
- I. Fluid Resistance (ASTM F1862-05): Synthetic Blood Penetration Resistance Test
- Filtration Efficiency: Bacterial Filtration Efficiency (BFE) Test (ASTM F2101-01) X. and Particulate Filtration Efficiency (latex Particle challenge) (ASTM F1215)
- differential Pressure (Delta-P) Test (MIL M 36954 C) XI.
- Flammability Test (16CFR 1610) XII.
- XIII. Biocompatibility per ISO 10993
It was our conclusion that performance testing met all relevant requirements of the aforementioned test standard.
{4}------------------------------------------------
## Acme Filter Mask Inc.
No.57, Hu Shan Road, Yingge District, New Taipei City, 23941,Taiwan, R.O.C.
### Discussion of Clinical Tests Performed:
Not Applicable
### Conclusions:
Acme Filter Mask Inc. Surgical Face Mask with Ear-Loop has the same intended use and technological characteristics as the predicated devices Modern Healthcare Corp., Surgical Face Mask, type: Tie-on and Ear-loop (K063043). Especially, the predicate device's types are Tie-on and Ear-loop; and the subject device just for Ear-loop. The performance tests for the Delta-P and Particulate Filtration Efficiency Performance also have the similar effectiveness. Besides, the bench testing contained in this submission supplied demonstrates that the technological characteristics do not raise any new question of safety or effectiveness.
Thus the new device is substantially equivalent to the predicate devices.
{5}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized emblem with three abstract human figures intertwined, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged around the emblem in a circular fashion.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
### May 20, 2013
Dr. Ke-Min Jen ACME Filter Mask No. 57 Hu Shan Road Yingge District, New Taipei City, Taiwan R.O.C. 23941
Re: K123115
> Trade/Device Name: Surgical Face Mask with Ear-Loop Model: YN-501AB, YN-501AW, YN-501AG for Blue, White, Green Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: April 12, 2013 Received: April 17, 2013
Dear Dr. Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{6}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Image /page/6/Picture/6 description: The image shows a signature block. The text "Sincerely yours," appears at the top. Below that is a signature and the name "Tejashri Purohit-Sheth, M.D." followed by the title "Clinical Deputy Director" and the acronym "DAGRID". Finally, the word "FOR" is at the bottom right.
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{7}------------------------------------------------
## Indications for Use
510 (K) Number ( If Known ): _________________________________________________________________________________________________________________________________________________ K123115
Device Name: Surgical Face Mask with Ear-Loop Model: YN-501AB, YN-501AW, YN-501AG for Blue, White, Green
Indications for Use:
The Surgical Face Mask of different colors (Blue, White, and Green) is a device intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operation room personnel from transfer of bacteria, body fluid and particulate material.
This Surgical Face Mask is non-sterilized and disposable.
Prescription Use
AND/OR
Over-The-Counter Use_
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ri S. Purohitsheth -S 17 13:52:16
Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental
510(k) Number:
P 5
Page_1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.