FREESTYLE AUTO-ASSIST SOFTWARE VERSON 2.0 WITH MODEL PRT24238-001
Applicant
Abbott Laboratories
Product Code
NBW · Clinical Chemistry
Decision Date
Apr 5, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Freestyle Auto-Assist Software Version 2.0 is intended for use by people with diabetes to aid in the review, analysis and evaluation of information such as blood glucose results, blood ketone test results and other data uploaded from a compatible meter, such as insulin doses and exercise data from the Freestyle Insulinx meter, in support of an effective diabetes health management program. Freestyle Auto-Assist software is not intended for the diagnosis of or screening for diabetes mellitus.
Device Story
Standalone software application for home use; retrieves blood glucose, ketone, insulin, and exercise data from compatible Abbott meters via serial/USB cables. Transforms raw meter data into graphical reports, logbooks, and statistical summaries (averages, target range adherence). Used by patients to track diabetes management; healthcare providers review generated reports to inform clinical decisions. Facilitates date/time synchronization and meter settings backup. Benefits include improved disease tracking and data visualization.
Clinical Evidence
Human factors study with 35 participants (ages 18-66, 13 Type 1, 22 Type 2 diabetes) evaluated software installation, data upload, and report generation. 32/35 subjects successfully created and viewed reports; data accuracy was 100% compared to stored meter data. Bench testing confirmed functional requirements and design specifications.
Technological Characteristics
Standalone software application; compatible with Windows XP/Vista/7 and Mac OS X. Connectivity via serial, USB, or Micro-USB cables. Implements data retrieval, graphical reporting, and statistical analysis. Complies with IEC 62366 for usability engineering. Software lifecycle processes followed for development.
Indications for Use
Indicated for people with diabetes to aid in review, analysis, and evaluation of blood glucose, ketone, insulin, and exercise data uploaded from compatible meters to support diabetes management. Not for diagnosis or screening of diabetes.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE
A. 510(k) Number: k123089
B. Purpose for Submission: New 510(k) for Diabetes Data Manager for use with the Abbott Freestyle Insulinx meter cleared under k111874
C. Manufacturer and Instrument Name: Abbott Laboratories Freestyle auto-assist software version 2.0 with model
D. Type of Test or Tests performed: Diabetes data management system
E. System Descriptions:
1. Device Description:
Freestyle Auto-Assist Software Version 2.0 is a software application that allows people with diabetes to upload data from Freestyle Insulinx, Precision Xtra™, Freestyle Freedom, Freestyle 5, Freestyle Flash, Freestyle Lite and Freestyle Freedom Lite Blood Glucose Monitoring Systems into FreeStyle Auto-Assist Software Version 2.0, where users may review blood glucose and other data in a number of different graphical presentations.
It is a stand-alone software program available for download to a home user's computer via download from an Abbott Diabetes Care (ADC) website or if a previous version of Auto-Assist software is installed on the computer, through a notification from the ADC Informatics Upgrade Server. The software will also be available to users via a CD-ROM.
2. Principles of Operation:
- Data Upload
FreeStyle Auto-Assist Software Version 2.0 retrieves each blood glucose test result or ketone result, and the date and time of the measurement, from the memory of recognized, compatible meters. It retrieves the data through either a Serial or Serial/USB cable, a Strip Port /USB cable, or a Micro-USB/USB cable. These data are viewable through and used by FreeStyle Auto-Assist Software Version 2.0 to generate reports. Additionally these data may be exported to a text file (.txt) and saved.
{1}
Page 2 of 8
- Data Entry
Through the FreeStyle Auto-Assist Software Version 2.0 user interface additional personal notes, such as insulin, meals, and exercise data may be added to the logged data collected from the meter.
- Data Access
Data from the compatible meter can be displayed through the software or saved in text file format (.txt).
- Meter Settings Backup
For compatible meters, the meter settings are retrieved from the meter. The meter settings can be saved to the computer by selecting the backup meter file option.
- Report Generation/Printout, Logbook, Glucose Average, Daily Statistics, Meal Event Averages
- Date/Time Synchronization with PC
3. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer webserver, or mobile device? Yes ☑ or No ☐.
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission: Yes ☐ or No ☑.
4. Specimen Identification:
Specimen identification is based on time and date of testing.
5. Specimen Sampling and Handling:
Data transmission from compatible meters using capillary whole blood samples
6. Calibration:
Compatible meter specific. See statement below under section J.
7. Quality Control:
Compatible meter specific. See statement below under section J.
8. Software:
FDA has reviewed the applicant’s Hazard Analysis and software Documentation: Yes ☑ or No ☐
{2}
Page 3 of 8
F. Regulatory Information:
| Device Name | Product Code | Classification | Regulation | Panel |
| --- | --- | --- | --- | --- |
| Glucose Test System | NBW: Blood Glucose Test System, Over-the-Counter | Class II | 21 CFR § 862.1345 | Clinical Chemistry (75) |
| Calculator/Data Processing Module for Clinical Use | JQP: Calculator/Data Processing Module for Clinical Use | Class I | 21 CFR § 862.2100 | Clinical Chemistry (75) |
G. Intended Use:
1. Indication(s) for Use:
The Freestyle Auto-Assist Software Version 2.0 is intended for use by people with diabetes to aid in the review, analysis and evaluation of information such as blood glucose results, blood ketone test results and other data uploaded from a compatible meter, such as insulin doses and exercise data from the Freestyle Insulinx meter, in support of an effective diabetes health management program.
Freestyle Auto-Assist software is not intended for the diagnosis of or screening for diabetes mellitus.
2. Special Condition for use Statement(s):
Over the Counter Use
H. Substantial Equivalence Information:
1. Predicate device name(s) and 510(k) numbers:
ABBOTT CoPilot™ Health Management System cleared under k062770.
2. Comparison with Predicate Device:
Similarities
| Feature | Predicate CoPilot Diabetes Data Management System | Proposed Auto-Assist™ Software Version 2.0 |
| --- | --- | --- |
| Editing of Patient Data | Does not allow user to edit data retrieved from a meter | Same |
| Intended Use | Intended for use by people with diabetes to aid in the review, analysis and evaluation of information such as blood glucose test results, blood ketone | Same |
{3}
| Feature | Predicate CoPilot Diabetes Data Management System | Proposed Auto-AssistTM Software Version 2.0 |
| --- | --- | --- |
| | test results and other data uploaded from a meter, such as insulin doses and exercise data, in support of an effective diabetes health management program. | |
| Statistical Functionality Summary | Glucose Statistics: Average blood glucose concentration Number of readings Number of days Target blood glucose range Percent of readings above, below, and within target range Standard deviation of blood glucose readings Insulin Statistics: Avg/Day Avg/Day Total | Same |
| Hardware Requirements | CD-ROM drive (not required if application is loaded from internet) Mouse Printer (if user wants to print) Internet connectivity (for host or e-mail Communication) Class I fax-enabled modem (required for fax capabilities only) Data Cable USB port | Same |
| Data Transfer | Serial or Serial/USB adapter Communication cables. | Same |
| Compatible Blood Glucose Meters | Precision XtraTM Freestyle FreedomTM FreeStyle Lite FreeStyle Freedom Lite | Same |
| Languages | English | Same |
| Data Access | Free | Same |
| Data Access to Internet | Free | Same |
| Data Access to Internet | Free | Same |
| Data Access to Internet | Free | Same |
| Data Access to Internet | Free | Same |
{4}
Page 5 of 8
Differences
| Feature | Predicate CoPilot Diabetes Data Management System | Proposed Auto-Assist™ Software Version 2.0 |
| --- | --- | --- |
| Import Capability | Event Importing:
Tab delimited files
XML Files
Database Importing:
Precision Link (V.2.5 or higher)
FreeStyle Connect
CoZMonitor | No import capability |
| Compatible Blood Glucose Meters | Precision® Xceed
Freestyle®
Freestyle Tracker®
Freestyle Navigator®
Continuous Glucose Monitoring System | Freestyle Insulinx
Freestyle 5
Freestyle Flash |
| Operating System | Microsoft Windows 98SE 2000 and XP | Microsoft Windows 7 (32 or 64 bit), Vista, or Windows XP Mac OS X Snow Leopard or Lion |
| Statistical Functionality Summary | Pump Statistics:
Avg/Day Meal Bolus
Avg/Day Correction Bolus
Avg/Day Total Bolus
Avg/Day Basal
Avg/Day Total Insulin | None |
| Hardware Requirements | Co-Pilot for FreeStyle Navigator:
Pentium 4 or above
512Mb
Kensington USB Bluetooth Adapter and compatible
Widcomm driver (if downloading data from FreeStyle Navigator) | Not compatible with FreeStyle Navigator |
| Languages | No Spanish translation | Spanish |
{5}
| Feature | Predicate CoPilot Diabetes Data Management System | Proposed Auto-Assist™ Software Version 2.0 |
| --- | --- | --- |
| Graphic Data Displays | Glucose data readings are displayed graphically along with the user-defined target range. Glucose data readings above target are displayed as purple and below target range as yellow. Triangle = Discrete data Circle = Continuous data Square = Manually entered glucose Above target = purple Within target = green Below target = yellow | Glucose data readings are displayed graphically along with the user-defined target range Glucose data readings above target are displayed as green, on target as blue, and below target as gray. Open Circle = Average blood glucose Closed Circle = Discrete blood glucose data Up-Arrow = blood glucose value greater than 350mg/dl |
| Data Transfer | Co-Pilot System for FreeStyle Navigator functionality communicates with FreeStyle Navigator via Kensington Bluetooth adapter/driver. | Micro-USB/USB adapter |
| Storage Capacity | CoPilot storage capacity is limited by the amount of available hard disk storage space | None |
I. Standard/Guidance Document Referenced (if applicable): IEC62366 Medical Devices - application of usability engineering to medical devices
J. Performance Characteristics: The performance characteristics listed below as applicable, are presented in the specific glucose clearance under k111874
1. Analytical Performance: The performance characteristics listed below as applicable, were presented in the specific glucose meter clearance under k111874
a. Accuracy: See above statement under section J(1).
b. Precision/Reproducibility: See above statement under section J(1).
c. Linearity: See above statement under section J(1).
{6}
Page 7 of 8
d. Carryover:
See above statement under section J(1).
e. Interfering Substances:
See above statement under section J(1).
2. Other Supportive Instrument Performance Data Not Covered Above:
a. Human Factors Study. A total of 32 of 35 participants were fully evaluated. Subjects were of typical computer experience and ages (age range 18 to 66 years old) that would be expected to use diabetes data management software were enrolled in the Human Factors study conducted for FreeStyle Auto-Assist Software Version 2.0. All of the study participants had diabetes: 13 were Type 1 and 22 were Type 2.
As part of the study, users installed the software on their test computer desktop and launched the software. Task 1 required the subjects to install the software, Task 2 required subjects to upload meter data and create and print reports, and Task 3 required subjects to save previously generated reports.
Users were assessed on their ability to upload meter data, create and print reports, and save reports to their desktop. The Freestyle Freedom Lite meter and the Precision Xtra meter were chosen as meters best representing the line of compatible meter including the Freestyle Insulinx meter and were preloaded with data such as blood glucose or ketone results, based on the compatible meters, and users created reports and graphs using this stored data. Stored data was automatically uploaded from a compatible meter when it was connected to a computer running FreeStyle Auto-Assist Software Version 2.0.
Based on the study, subjects were able to successfully create and view reports, and the data in these reports matched the stored data in 100% of cases.
b. Ease of Use Evaluation. Ease-of-use ratings for the Auto-Assist 2.0 software were recorded upon task completion in the Human Factors study. Participants were asked to rate the ease of use for each task, on a scale from 1 to 5, where 1 was very easy and 5 was very difficult. At the end of the session, they were also asked to provide an overall rating of the software using the same scale.
The overall success of the majority of participants led to a positive experience with the software. On a 5-point scale where 1 represented very difficult and 5 represented very easy, participants rated all tasks as
{7}
Page 8 of 8
relatively easy (Task 1: 4.3, Task 2: 4.1, and Task 3: 4.3). Even when participants struggled with a task or failures occurred, participants still thought that the software would benefit them by helping them better track their diabetes. The participants who struggled assumed they would not have difficulty if they were to use the software again. The benefits of tracking their disease and the assumption that the software was easy to learn gave participants an overall positive impression of the software, which is reflected in the overall ease of use rating of 4.2.
c. Bench Testing. Bench testing was conducted during design verification activities. The existing FreeStyle and Precision USB Data Cables were used to upload data from compatible meters to the Auto-Assist 2.0 application running on a Windows or Mac Computer. Auto-Assist 2.0 verification was performed on the following supported Operating Systems:
- Windows XP SP3 (32-bit)
- Windows Vista (32-bit and 64-bit)
- Windows 7 (32-bit and 64-bit)
- Mac OS X Snow Leopard
- Mac OS X Lion
Meter memory capacity and data rollover is a function of meter design and unchanged from those submitted in each meter pre-market application, and thus was not tested as part of FreeStyle Auto-Assist Software Version 2.0 Auto-Assist 2.0 software has been verified to be implemented according to its meter specific data upload requirements to meet 100% accuracy and data management requirements.
d. Software documentation was reviewed and demonstrated that the device was developed under appropriate software lifecycle processes.
e. The Auto-Assist Software Version 2.0 User's Guide Readability Assessment was at a 6.4 grade level using the Flesch-Kincaid readability assessment.
## K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
## L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.