BUTTOCK MUSCLE STIMULATOR MODEL WL-2413B; BUTTOCK MUSCLE STIMULATOR MODE WL-2413E
Applicant
Well-Life Healthcare Limited
Product Code
NGX · Physical Medicine
Decision Date
Jul 11, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles
Device Story
Buttock Muscle Stimulator (model WL-2413E) is a two-channel, battery-operated EMS device. System includes electronic stimulator module, shorts with integrated silicon pad electrodes, conductive gel, and lead wires. Device applies transcutaneous electrical muscle stimulation to buttocks muscles to improve muscle tone. Operated by healthy individuals in home setting. User wears shorts with electrodes in contact with skin; stimulator generates biphasic, rectangular stimulation signals. Device features four output modes, timer (20-40 minutes), and automatic shut-off. Output parameters (voltage, current, pulse duration) are controlled by internal microprocessor. Device provides non-invasive muscle stimulation; intended for OTC use.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10 standards. Software verification performed per FDA guidance.
Technological Characteristics
Powered muscle stimulator; 3x AAA battery source. ABS housing. Silicon pad electrodes (7.0 cm diameter). Biphasic, rectangular waveform. 4 output modes. Type BF patient isolation. Microprocessor-controlled. Connectivity: None. Sterilization: Not applicable (non-sterile). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10.
Indications for Use
Indicated for healthy persons seeking improvement of buttock muscle tone via transcutaneous electrical muscle stimulation (EMS).
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
Predicate Devices
SPORT-ELEC models Body Control System "BS" (K092142)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(K) Summary
The assigned 510(k) number is: K123075
#### Submitter's Identifications: 1.
Well Life Healthcare Limited 1F., No. 16, Lane 454, Jungjeng Road, Yunghe City, Taipei County 234, Taiwan, ROC
Contact: Jenny Hsieh Telephone: + 886 2 2928 2112
Date of Summary Preparation: May 25, 2013
#### 2. Name of the Device:
Trade/Device Name: Buttock Muscle Stimulator, model: WL-2413E. Regulation Number: 21 CFR 890.5850 Requlation Name: Power muscle stimulator. Regulatory Class: II Product Code: NGX
#### 3. Information of the 510(k) Cleared Device (Predicate Device):
SPORT-ELEC models Body Control System "BS"( K092142)
#### Device Description: 4.
The Buttock Muscle Stimulator, model WL-2413E, is a two channel battery operated muscle stimulation system specifically designed to exercise the buttocks' muscle. The device comprises mainly the electronic stimulator module which generates the required stimulation signals.
The Buttock Muscle Stimulator, model WL-2413E, comprises a short with silicon pad electrodes which connect the signals from the stimulator to the skin. The electrodes are placed on the inner surface of the accessories. The product is supplied with a 510(K) cleared conductive gel, a user's manual, 3AAA batteries, 2 lead wires and a storage bag.
The required power of the device is derived from 3 batteries located in a compartment protected by a removable battery cover. The electrodes are in the inner surface of the shorts which is worn as shown on the picture of user manual. There is no current passed from side to side. The user cannot access the wiring or connectors within the shorts.
The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles
{1}------------------------------------------------
#### 5. Intended Use:
The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles
#### Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are റ. as follows:
Compliance to applicable voluntary standards includes IEC 60601-2-10, as well as IEC 60601-1, and IEC 60601-1-2 requirement.
In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.
Discussion: The Compliance to applicable voluntary standards as above mentioned indicates that the new device in this submission used the same standards as that of predicate device in addition to the software validation standard. For our new device, the software validation was prepared according to FDA guidance, however the predicate device used IEC 60601-1-4 to prepare their software validation report. For this difference we believe that we used more adequate reference tool to prepare software validation report than the predicate device. Therefore, we consider that the compliance of standards included in our submission is adequate for the determination of substantial equivalence.
## 7. Comparison of Significant device features
| Comparison<br>Features | Models to be Compared | |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| Model | SPORT-ELEC Body Control System<br>"BS" | WL-2413E |
| 510(K) No. | K092142 | K123075 |
| Prescription or OTC | OTC | Same |
| Indication for use | intended for use by healthy persons to<br>apply transcutaneous electrical muscle<br>stimulation(EMS) through skin contact<br>electrode for the purpose of improvement<br>of muscle tone of the buttocks muscles | Same |
| FDA product code | NGX | Same |
| Electrode Used | Self Adhesive Electrode( 5 X 5 cm) | K082065, Silicon Pad Electrode<br>( 7.0 cm diameter) |
| Manufacturer | SPORT-ELECT | Well-Life |
{2}------------------------------------------------
## 8. Comparison of Unit Characteristics & Output Specification
| Mode or Program Name | Predicate Device | New Device | | | | |
|--------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|----------------------------|--------------------|----------------------------------------|----------------------------------------|
| Waveform (e.g., pulsed monophasic, biphasic) | Biphasic | Biphasic | | | | |
| Shape (e.g., rectangular, spike, rectified<br>sinusoidal) | Rectangular | Rectangular | | | | |
| Maximum Output Voltage (volts)-<br>(+/- 20%) | 31.2V @500Ω<br>66.0V @2KΩ<br>66.0V @10KΩ | 40.8V @500Ω<br>70.0V @2KΩ<br>106.0V @10KΩ | | | | |
| Maximum Output Current (mA)-<br>(+/- 20%) | 62.4A @500Ω<br>33.0A @2KΩ<br>6.6A @10KΩ | 81.6mA @500Ω<br>35.0mA @2KΩ<br>10.6mA @10KΩ | | | | |
| Duration of primary phase (usec) | 200 Max | 300 Max | | | | |
| Pulse Duration (usec) | 540 Max. | 720 Max. | | | | |
| Frequency (Hz) [or Rate (pps)] | 70 Max | 70 Max | | | | |
| For<br>multiphasic<br>waveforms<br>only: | Symmetrical phases?<br>Phase Duration (include units),<br>(Stage range, if applicable),<br>(both phases, if asymmetrical) | Yes<br>Not applicable | Yes<br>Not applicable | | | |
| 510(K) Number | K092142 | K123075 | | | | |
| Device Name and Model | BS | WL-2413E | | | | |
| Manufacturer | SPORT-ELECT | Well-Life | | | | |
| Power Source(s) | 1.5Vx3<br>(AAA Size) | 1.5Vx3<br>(AAA Size) | | | | |
| - Method of Line current Isolation<br>- Patient Leakage Current | Type BF<br>--- | Type BF<br>--- | | | | |
| - Normal condition (uA) | Under 0.1 | Under 0.1 | | | | |
| - single Fault condition (uA) | Under 0.5 | Under 0.5 | | | | |
| Average DC current through electrodes when<br>device is on but no pulses are being applied ( uA) | Not applicable | Not applicable | | | | |
| Number of Output Modes | 4 | 4 | | | | |
| Number of Output<br>Channels: | Synchronous or<br>Alternating?<br>Method of Channel<br>Isolation | Synchronous<br>Output Coil | Synchronous<br>Output Coil | | | |
| Regulated Current or Regulated Voltage? | Voltage | Voltage | | | | |
| Software/Firmware/Microprocessor control? | Yes | Yes | | | | |
| Automatic Overload Trip? | No | No | | | | |
| Automatic No-Load Trip? | Yes | Yes | | | | |
| Automatic Shut Off? | Yes | Yes | | | | |
| User Override control? | No | No | | | | |
| Indicator Display: | On/Off Status?<br>Low Battery?<br>Voltage/Current<br>Level? | Yes<br>Yes<br>Yes | Yes<br>Yes<br>Yes | | | |
| Timer Range (Minutes) | 20-40 | 20-40 | | | | |
| Compliance with Voluntary Standards? | IEC 60601-2-10 | IEC 60601-2-10 | | | | |
| Compliance with 21 CFR 898? | Yes | Yes | | | | |
| Weight (g) including battery | 171.6 | 80 | | | | |
| Dimensions (mm.) [W x H x D] | 87×43×69 | 64×90×20 | | | | |
| Housing Materials and construction | ABS | ABS | | | | |
| Pulse per burst | Same for each program | Same for each program | | | | |
| Burst per second | Same for each program | Same for each program | | | | |
| Bust duration | Same for each program | Same for each program | | | | |
| Duty Cycle | Same for each program | Same for each program | | | | |
| Method of achieving zero net charge for net | Method of achieving zero net charge/pulse | Biphasic symmetric | | Biphasic symmetric | Biphasic symmetric wave for each pulse | Biphasic symmetric wave for each pulse |
·
{3}------------------------------------------------
| Comparison feature | Model | | Remarks |
|-------------------------|--------------------|--------------------|---------|
| | SPORT-ELECT BS | WL-2413E | |
| Net charge | 0 | 0 | Note 1 |
| Max. phase charge | 14 uc | 24 uc | |
| Max. current Density | 0.0784 mA/cm² | 0.0997 mA/cm² | |
| Max. Average<br>current | 38.43mA | 38.644mA | |
| 500 Ω | | | |
| 2K Ω | 14.892mA | 16.907 mA | |
| (RMSA) | | | |
| 10 Ω | 4.996mA | 5.120 mA | |
| Max. Power Density | 0.00274 Watts/ cm² | 0.00399 Watts/ cm² | |
| Burst Mode | Yes | Yes | |
## 9. Significant output characteristics comparison table:
Note 1: Net Charge is Zero for the symmetrical stimulation biphasic wave
### 10. Summary for the technology comparison.
Basically the WL-2413E has the similar technological characteristics with the predicate device in the product design, material, energy source type, main program mode and the main output waveform... etc. There exist some difference in the detailed output parameters( mainly in the pulse duration and electrode sizes). Through the detailed calculation comparison of stimulation output energy for each operation mode(in particular the output current density), we found the output level in each operation mode for our WL-2413E and predicate device are very close and within the acceptable range as specified in FDA guidance. So we believe the difference in detailed output parameters does not affect the determination of substantial equivalence.
## 11. Conclusions
The Buttock Muscle Stimulator, model: WL-2413E have the same intended use and the similar technological characteristics as the cleared device- SPORT-ELEC models Body Control System "BS"( K092142). Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device.
In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles a person embracing another person, or a bird in flight, composed of three curved lines.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### July 11, 2013
Well-Life Healthcare Ltd C/O Ms. Jennv Hsieh 1F. No. 16, Lane 454, Jungjeng Road Yunghe City, Taipei County, Taiwan R.O.C.
Rc: K123075
Trade/Device Name: Well-Life Buttock Muscle Stimulator. model W1.-2413E Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: NGX Dated: June 2. 2013 Received: June 5, 2013
Dear Ms. Hsieh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (sec above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Ms. Jenny Hsieh
CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
Victor Krauthamer, Ph.D. for Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K123075
Device Name: Well-Life Buttock Muscle Stimulator, model: WL 2413E
Indications For Use:
The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles. .
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Joyce M. Whang -S
(Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)
510(k) Number _ K123075_
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.