K123070 · Sirona Dental Systems GmbH · OAS · Mar 15, 2013 · Radiology
Device Facts
Record ID
K123070
Device Name
GALILEOS FAMILY
Applicant
Sirona Dental Systems GmbH
Product Code
OAS · Radiology
Decision Date
Mar 15, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Devices of the GALILEOS family consist of an x-ray system that uses a cone beam with a rotational sequence, providing two dimensional images and three dimensional volume reconstructions of the head area, which includes ENT and dentomaxillofacial areas, for use in planning and diagnostic support in adult and pediatric care. Devices of the GALILEOS family comprise a package of PC software modules to expand SIDEXIS capabilities to handling 3D data. This includes 3D reconstruction, storage, retrieval, viewing and processing of 3D-image data.
Device Story
GALILEOS family is an extraoral dental X-ray system using cone-beam technology to generate 3D volume reconstructions of teeth, jaw, and head area. System includes support stand, height-adjustable sled, motor-driven rotatable ring, X-ray source, and image receptor. Patient positioning uses Class I laser light localizers, bite blocks, and forehead/temple supports. Operator uses control panel or remote to select modes and initiate exposure. PC software modules (SIDEXIS) perform 3D reconstruction, storage, retrieval, viewing, and processing of image data. Output includes 3D volumes, panoramic, and cephalometric images. Used in clinical settings by dental professionals for diagnostic support and treatment planning. Benefits include high-resolution 3D visualization of dentomaxillofacial structures to aid clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only. System functionality verified against requirements, risk analysis, and user perspective. Exposure programs validated using test phantoms to confirm equality with predicate device. Additional testing performed per FDA guidance for solid-state X-ray imaging devices.
Technological Characteristics
Cone-beam X-ray system; 98kV/6mA generator; image intensifier; mechanical diaphragm; Class I laser localizers. Patient-contact materials biocompatible per ISO 10993-1:2003. PC-based software for 3D reconstruction and SIDEXIS database integration. Standalone system with height-adjustable sled and rotatable ring.
Indications for Use
Indicated for adult and pediatric patients requiring diagnostic imaging and treatment planning for ENT and dentomaxillofacial areas using cone-beam computed tomography.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K123070
Page 1 of 5
## MAR 1 5 2013
510(k) Summary for Sirona Dental Systems GALILEOS family
### 1. Sponsor
Sirona Dental Systems GmbH Fabrikstrasse 31 D-64625 Bensheim Germany Contact Person: Fritz Kolle +49 6251 16 3294 Telephone: Date Prepared: September 28, 2012
### 2. Device Name
| Proprietary Name: | GALILEOS family |
|----------------------|-------------------------------------|
| Common/Usual Name: | X-ray, Tomography, Computed, Dental |
| Classification Name: | Computed tomography x-ray system |
## 3. Predicate Devices
Sirona GALILEOS (K060892), Genoray VOLUX 21C (K120263)
#### 4. Intended Use
Devices of the GALILEOS family consist of an x-ray system that uses a cone beam with a rotational sequence, providing two dimensional images and three dimensional volume reconstructions of the head area, which includes ENT and dentomaxillofacial areas, for use in planning and diagnostic support in adult and pediatric care.
Devices of the GALILEOS family comprise a package of PC software modules to expand SIDEXIS capabilities to handling 3D data. This includes 3D reconstruction, storage, retrieval, viewing and processing of 3D-image data.
Sirona Dental Systems 510(k) GALILEOS family
Page 1
{1}------------------------------------------------
K123070
Page 2 of 5
## 5. Device Description and Function
The GALILEOS family is an extraoral source dental X-ray system intended to produce Xrays for obtaining three dimensional volume reconstructions of the teeth, jaw, and the head area, which includes ENT and dentomaxillofacial areas, for use in planning and diagnostic support in adult and pediatric care.
The GALILEOS family device generates a conical x-ray beam that rotates around the patient's head within a certain angle.
The device comprises an image receptor for 3D volume exposure with an adjustable diaphragm. Three volume regions are defined through this. Class I laser beam light localizers serve to position the patient's head that may be fixed through bite block and adjustable forehead and temple supports.
From the obtained exposures a three dimensional image is reconstructed and can be viewed as well as panoramic/cephalometric images. The constructed 3D volume and simulated projection exposures as well as panoramic/cephalometric data are conveyed to SIDEXIS and stored in the SIDEXIS data base.
An operator control panel allows height adjustment, selection of mode and program, and indicates machine states.
A separate handhold push-button serves for exposure release An optional remote control is available.
## 6. Scientific Concept
The underlying scientific concept is cone-beam x-ray technology. All volumetric exposures are obtained by cone-beam technology.
Sirona Dental Systems 510(k) GALILEOS family
{2}------------------------------------------------
K123070
Page 3 of 5
## 7. Physical and Performance Characteristics
#### 7.1. Design
The GALILEOS family comprises of a support stand to which a height adjustable sled is attached. The sted carries the patient fixation, the operator control panel (easy pad) and the motor driven rotatable ring. The X-ray source and image receptor are fixed to the ring. During a scan the device generates a conical x-ray beam that rotates around the patient's head at varying angles.
Class I laser beam light localizers serve for positioning the patient's head that may be fixed through bite block and adjustable forehead and temple supports.
The exposure area is defined by the geometry of the GALILEOS family devices.
A control panel allows the user to select the exposure modes and the exposure factors, view the machine status information, control the height adjustment and turn on the laser indicator.
An optional remote control is available,
The PC software reconstructs the three-dimensional image as well as process panoramic / cephalometric images. The constructed 3D volume and simulated projection exposures as well as panoramic/ cephalometric data are conveyed to SIDEXIS and stored in the SIDEXIS database.
#### 7.2. Material Used
Materials that come into patient contact intentionally are biocompatible and evaluated according to ISO 10993-1: 2003, "Biological evaluation of medical devices ~ Part 1: Evaluation and testing within a risk management process".
#### 7.3. Physical Properties
Not applicable.
Sirona Dental Systems 510(k) GALILEOS family
Page 3
{3}------------------------------------------------
K123070
Page 4 of 5
## 8. Summary of the technological characteristics
The GALILEOS family with the new member GALILEOS Comfort"LUS is a further development of the GALILEOS (K060892).
All members of the GALILEOS family provide adapted programs of the GALILEOS and the same 3D operating principles as the predicate device GALILEOS.
The new member of the GALILEOS family the GALILEOS Comfort".US has a different source to skin distance than the GALILEOS and the image intensifier has a new mode to support the increased amount of images during an exposure.
The devices of the GALILEOS family have two type of X-ray generators. The predicate GALILEOS was capable of maximum 85kV with 7mA and the new GALILEOS Comfort Lus is capable of maximum 98kV with 6 mA.
All GALILEOS family devices now include a mechanical diaphragm.
The PC software performs identical functions and algorithms as the GALILEOS. The software has been adapted to the higher number of images.
The GALILEOS family offers a calculated panoramic view in combination with slices orthogonal to the panoramic curve ('transversal slices') as the predicate GALILEOS. The GALILEOS family offers nearly the same functionality in viewing slices, projections and volume views.
Page 4
Sirona Dental Systems 510(k) GALILEOS family
{4}------------------------------------------------
K123070
Page 5 of 5
## 9. Nonclinical Testing
The GALILEOS family system functions have been tested in a system test (covers the requirements from the function specification, the risk/hazard analysis and the functionality of the equipment from the user's perspective).
The exposure programs have been tested utilizing test phantoms. The tests evaluate the equality of exposures of proposed GALILEOS family and predicate device GALILEOS.
Additional tests have been performed taking into account FDA Guidance "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, Document issued on: August 6, 1999" adapted for image intensifier products.
#### Clinical Testing 10.
Clinical tests have not been performed.
#### Conclusion 11.
Based on a comparison of intended use, indications, construction materials, principle of operation, features, and technical data, the Sirona Dental GALILEOS family and the new member GALILEOS Comfort" is safe and effective to perform its intended use and is substantially equivalent to the predicate devices.
Sirona Dental Systems 510(k) GALILEOS family
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
March 15. 2013 -
Mr. Fritz Kolle Sirona Dental Systems GmbH Fabrikstrasse 31 BENSHEIM. GM D-64625
Re: K123070
Trade/Device Name: GALILEOS Comfort PLUS Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: February 06, 2013 . Received: February 13, 2013
Dear Mr. Kolle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Mr. Kolle
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
K123070 510{k} Number (if known):
Device Name:
GALILEOS Comfort PLUS
Indications for Use:
Devices of the GALILEOS family consist of an x-ray system that uses a cone beam with a rotational sequence, providing two dimensional images and three dimensional volume reconstructions of the head area, which includes ENT and dentomaxillofacial areas, for use in planning and diagnostic support in adult and pediatric care.
Devices of the GALILEOS family comprise a package of PC software modules to expand SIDEXIS capabilities to handling 3D data. This includes 3D reconstruction. storage, retrieval, viewing and processing of 3D-image data.
Prescription Use v (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Smh.7)
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
K123070 510(k)
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.