K123038 · E.M.S Electro Medical Systems S.A · FFK · Jan 16, 2013 · Gastroenterology, Urology
Device Facts
Record ID
K123038
Device Name
SWISS LITHOBREAKER
Applicant
E.M.S Electro Medical Systems S.A
Product Code
FFK · Gastroenterology, Urology
Decision Date
Jan 16, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4480
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SWISS LITHOBREAKER is intended to be used for the fragmentation of urinary tract calculi in the kidney, ureter, and bladder.
Device Story
Handheld, battery-operated intracorporeal lithotripter; uses mechanical impact drive to fragment urinary tract calculi. Input: manual activation via button; device transforms battery power into mechanical impact waves transmitted through probes. Output: mechanical impacts at ~3Hz (continuous or single-pulse mode). Used in clinical settings by physicians; probes inserted through endoscope working channels. Optional LithoVac lv3 suction system removes fragments. Benefits: enables minimally invasive stone fragmentation; stable energy output throughout battery life.
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing, including probe tip displacement, probe tip velocity, stone breakage/clearance, stone displacement (retropulsion), and energy output, alongside compliance with electrical safety (IEC 60601-1) and sterilization standards.
Indicated for fragmentation of urinary tract calculi in the kidney, ureter, and bladder in patients requiring lithotripsy.
Regulatory Classification
Identification
An electrohydraulic lithotriptor is an AC-powered device used to fragment urinary bladder stones. It consists of a high voltage source connected by a cable to a bipolar electrode that is introduced into the urinary bladder through a cystoscope. The electrode is held against the stone in a water-filled bladder and repeated electrical discharges between the two poles of the electrode cause electrohydraulic shock waves which disintegrate the stone.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notifications for Intracorporeal Lithotripters.”
{0}------------------------------------------------
### 510(k) Summary for the E.M.S. Electro Medical Systems SA SWISS LITHOBREAKER
### JAN 1 6 2013
#### 1. SUBMITTER/510(K) HOLDER
E.M.S. Electro Medical Systems SA Ch. de la Vuarpillière 31 CH - 1260 Nyon Switzerland
Sonia Callegaro Contact Person: 41 22 994 26 11 Telephone:
September 27, 2012 Date Prepared:
#### 2. DEVICE NAME
| Proprietary Name: | SWISS LITHOBREAKER |
|----------------------|---------------------------------------|
| Common/Usual Name: | Endoscopic intracorporeal lithotripsy |
| Classification Name: | Electrohydraulic lithotripsy |
#### 3. PREDICATE DEVICES
The proposed SWISS LITHOBREAKER is substantially equivalent to the following legally marketed medical devices:
- E.M.S. Electro Medical Systems S.A., Swiss LithoClast Master (aka Swiss ● LithoClast Ultra) (K012445)
- . The Laryngial Mask Company Limited, LMA StoneBreaker (K062475)
The probes used by the SWISS LITHOBREAKER were previously cleared by FDA in K012445, and K973788. The LithoVac lv3 suction system was previously cleared in K012445. The sterile drape was previously cleared in K110724.
#### 4. DEVICE DESCRIPTION
The EMS SWISS LITHOBREAKER is a handheld, standalone, battery operated intracorporeal lithotripter which uses a mechanical impact drive to fragment urinary tract calculi in the kidney, ureter, and bladder. The device generates around 3 impacts in continuous mode per second which are transmitted through the probes on
{1}------------------------------------------------
the stones or a lower frequency by "single-pulse" mode, depending on the activation time of the button. The probes are introduced into the working channels of suitable endoscopes. The energy is sustained and stable throughout the life of the battery and the frequency drops only at the end of the battery life. An optional suction handpiece allows extraction of fragmented stones.
The proposed device consists of the following:
- Handpiece including powerpack .
- LithoClast probes, non-sterile reusable and sterile single use .
- LithoVac lv3 suction system with suction tubes .
The SWISS LITHOBREAKER is also supplied with accessories for cleaning (endcap), periodic maintenance (O-rings), probe accessories (interface and probe guide), AAA batteries and sterile drapes.
#### 5. INTENDED USE
.
The SWISS LITHOBREAKER is intended to be used for the fragmentation of urinary tract calculi in the kidney, ureter, and bladder.
#### PRINCIPLES OF OPERATION 6.
The proposed SWISS LITHOBREAKER can be used as a lithotripter only or in combination with a suction device LithoVac lv3. The probes mounted on the handpiece transmit impact generated waves of sufficient strength to fragment urinary tract calculi of any composition. LithoVac lv3 suction system can be combined with the lithotripsy device to provide simultaneous suction to remove calculi fragments.
The SWISS LITHOBREAKER handpiece is a battery operated lithotripter which uses a mechanical impact drive to fragment urinary tract calculi in the kidney, ureter, and bladder. The device works at an impact frequency of approximately 3Hz in either a continuous mode or "single-pulse" mode, depending on the time the activation button is depressed and probes are used through the working channels of suitable endoscopes.
Probes are mounted on the handpiece and the probes are inserted into the working channel of an endoscope. The LithoVac lv3 suction system may be used in conjunction with the SWISS LITHOBREAKER handpiece and probe to allow for suctioning during procedures. The probe is brought into contact with the calculi, and the LithoBreaker is activated.
{2}------------------------------------------------
K123038 Page 3 of 4
#### 7. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICES
The proposed SWISS LITHOBREAKER is substantially equivalent to the previously cleared Swiss LithoClast Master (K012445) and the handheld LMA StoneBreaker (K062475).
Differences between the proposed SWISS LITHOBREAKER and the predicate Swiss LithoClast Master (K012445) and the handheld LMA StoneBreaker (K062475) are limited to the power sources used and the mechanism used to translate the power into impacts which are transmitted through the probes, resulting in non-significant variations in performance. The design and function of the lithotripsy probes of the proposed SWISS LITHOBREAKER are identical to the EMS LithoClast probes.
Testing demonstrated that the SWISS LITHOBREAKER fulfills the prospectively defined performance specifications. The similarities in intended use, operational characteristics, and technological characteristics between the proposed SWISS LITHOBREAKER and the predicate Swiss LithoClast Master (K012445) and LMA StoneBreaker (K062475) lead to a conclusion of substantial equivalence between the proposed and predicate devices. A side-by-side comparison of the predicate devices and the proposed device is provided in Table 5-1 at the end of this section.
#### 8. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
Testing was performed to verify compliance of the SWISS LITHOBREAKER with the following standards:
- · ANSI/AAMI ST79:2006, "Comprehensive guide to steam sterilization and sterility assurance in health care facilities"
- · ISO 17664 (2004), "Sterilization of Medical Devices Information to be provided by the Manufacturer for the Processing of Resterilizable Medical Devices"
- · ANSI/AAMI/ISO 17665-1:2006 "Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices"
- · IEC 60601-1:1988 +A1: 1991 + A2: 1995, "Medical electrical equipment, Partl : General requirements for safety"
- IEC 60601-1-2, "Medical electrical equipment. Part 1 General requirements for . safety - Part 2 Collateral standard: Electromagnetic compatibility: Requirements and Tests" (2007).
{3}------------------------------------------------
- · ISTA 2A (2008), "ISTA Preshipment Testing Procedures- Combination Tests for Packaged-Products 150 lb (68 kg) or Less"
The results of this testing confirm that the SWISS LITHOBREAKER is safe and effective for the intended use described in Section 5.
In addition, the SWISS LITHOBREAKER was tested as compared to the predicates with respect to the following:
- · Probe Tip Displacement
- · Probe Tip Velocity
- · Stone Breakage/Clearance
- · Stone Displacement (Retropulsion)
- · Energy Output
The results of this testing confirm that the SWISS LITHOBREAKER is substantially equivalent to the predicates.
#### 9. SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
No clinical testing was conducted for this submission.
#### 10. SUMMARY OF OTHER INFORMATION
No other information is available.
#### 11. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS
Based on the information and supporting documentation provided in the premarket notification, the SWISS LITHOBREAKER is substantially equivalent to the cited predicate devices. Testing demonstrates that the SWISS LITHOBREAKER fulfills prospectively defined design and performance specifications.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### January 16, 2013
EMS Electro Medical Systems SA % Ms. Suzanne Goodman, BSE, RAC Regulatory Project Manager, Medical Devices Aptiv Solutions 225 Turnpike Road SOUTHBOROUGH MA 01772
Re: K123038
> Trade/Device Name: SWISS LITHOBREAKER Regulation Number: 21 CFR § 876.4480 Regulation Name: Electrohydraulic lithotriptor Regulatory Class: Class II Product Code: FFK Dated: November 30, 2012 Received: December 6, 2012
Dear Ms. Goodman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Herbert R. Lerner
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
K123038 510(k) Number (if known):
Device Name: SWISS LITHOBREAKER
Indications for Use:
The SWISS LITHOBREAKER is intended to be used for the fragmentation of urinary tract calculi in the kidney, ureter, and bladder.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Herbert R. Lerner**
| (Division Sign-Off) | |
|---------------------------------------------|---------|
| Division of Reproductive, Gastro-Renal, and | |
| Urological Devices | |
| 510(k) Number | K123038 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.