K123010 · ACON Laboratories, Inc. · NBW · Dec 14, 2012 · Clinical Chemistry
Device Facts
Record ID
K123010
Device Name
ON CALL VIVID BLOOD GLUCOSE MONITORING SYSTEM
Applicant
ACON Laboratories, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Dec 14, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Indications for Use
The On Call® Vivid Pro Blood Glucose Monitoring System is an electrochemical enzymatic assay. It is used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertip. The On Call® Vivid Pro Blood Glucose Monitoring System is intended for multiple patient use by health care professionals in health care facilities as an aid to monitoring the effectiveness of diabetes control programs. The system should only be used with single-use, auto-disabling lancing devices. The On Call® Vivid Pro Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus, nor intended for use on neonates. The On Call® Vivid Pro Blood Glucose Test Strips are used with the On Call Vivid Pro Blood Glucose Meter in the quantitative measurement of glucose in capillary blood from the finger. The On Call® Vivid Pro Blood Glucose Control Solutions are for use with the On Call® Vivid Pro Blood Glucose Meter and Strips as a quality control check to verify the accuracy of blood glucose test results.
Device Story
System uses electrochemical enzymatic assay to measure glucose in capillary whole blood. Input: 0.8 µL blood sample applied to test strip containing glucose oxidase and mediator. Meter measures electrical current generated by reaction; calculates glucose concentration; displays result calibrated to plasma reference. Used in healthcare facilities by professionals for diabetes management. Output: quantitative blood glucose reading. Benefits: aids monitoring of diabetes control programs. Features auto-coding upon strip insertion. Quality control performed via aqueous glucose control solution.
Clinical Evidence
Clinical study conducted with lay persons and trained technicians comparing system results to YSI Model 2300 STAT PLUS reference. Results demonstrated that nonprofessional, inexperienced lay users obtained blood glucose readings comparable to trained technicians. Users reported satisfaction with ease of operation and performance. Bench testing included repeatability, intermediate precision, linearity, and environmental stress testing (temperature, humidity, altitude).
Technological Characteristics
Electrochemical enzymatic biosensor; glucose oxidase reagent. Measurement range 20-600 mg/dL. Sample: 0.8 µL capillary blood. Power: 2x CR 2032 3.0V batteries. Connectivity: Serial data port (9600 baud). Dimensions: 3.58" x 2.28" x 0.83". Operating range: 5-45°C, 10-90% RH. Hematocrit range: 20-70%. Auto-coding via strip recognition.
Indications for Use
Indicated for quantitative measurement of glucose in fresh capillary whole blood from the fingertip in patients requiring diabetes control monitoring. Intended for multiple patient use by healthcare professionals in clinical settings. Contraindicated for neonates and not intended for diabetes diagnosis or screening.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
One Touch® Ultra® Blood Glucose Monitoring System (K002134)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k123010
B. Purpose for Submission:
New device
C. Measurand:
Capillary whole blood glucose
D. Type of Test:
Quantitative amperometric assay, glucose oxidase
E. Applicant:
ACON Laboratories, Inc.
F. Proprietary and Established Names:
On Call® Vivid Pro Blood Glucose Monitoring System
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| NBW-system, test, blood glucose, prescription | Class II | 21 CFR § 862.1345 | 75-Chemistry |
| CGA-glucose oxidase, glucose | Class II | 21 CFR § 862.1345 | 75-Chemistry |
| JJX-single analyte control | Class I | 21 CFR § 862.1660 | 75-Chemistry |
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
{1}
The On Call® Vivid Pro Blood Glucose Monitoring System is an electrochemical enzymatic assay. It is used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertip. The On Call® Vivid Pro Blood Glucose Monitoring System is intended for multiple patient use by health care professionals in health care facilities as an aid to monitoring the effectiveness of diabetes control programs. The system should only be used with single-use, auto-disabling lancing devices.
The On Call® Vivid Pro Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus, nor intended for use on neonates.
The On Call® Vivid Pro Blood Glucose Test Strips are used with the On Call Vivid Pro Blood Glucose Meter in the quantitative measurement of glucose in capillary blood from the finger.
The On Call® Vivid Pro Blood Glucose Control Solutions are for use with the On Call® Vivid Pro Blood Glucose Meter and Strips as a quality control check to verify the accuracy of blood glucose test results.
3. Special conditions for use statement(s):
- For in vitro diagnostic use only
- For Professional Use
- Not intended for use on neonates
- Not for diagnosis of or screening for diabetes mellitus
- Not to be used for patients who are dehydrated, hypotensive, in shock, critically ill or in a hyperosmolar state
- Meter should be cleaned and disinfected after use on each patient
- System should only be used with single-use, auto-disabling lancing devices.
4. Special instrument requirements:
The On Call® Vivid Pro Blood Glucose Meter
Only single-use auto-disabling lancing devices should be used with the On Call® Vivid Pro Blood Glucose Monitoring System.
I. Device Description:
The On Call® Vivid Pro Blood Glucose Monitoring System is a quantitative assay for the detection of glucose in capillary whole blood sampled from the fingertip. The glucose measurement is achieved by using the amperometric detection method.
The test strip has a reagent system including glucose oxidase and a mediator that reacts with glucose in the whole blood sample to produce an electrical current. This current is measured by the meter, and after calculation by the meter, the blood glucose concentration is shown on the meter display.
{2}
The On Call® Vivid Pro Blood Glucose Monitoring System consists of the On Call Vivid Pro Blood Glucose Meter, Carrying Case, User's Manual, Warranty Card and Logbook. Materials needed but not provided are the On Call Vivid Pro Blood Glucose Test strips, The On Call Vivid Pro Blood Glucose control solutions (Level I and Level II) and disposable, single use lancing devices.
J. Substantial Equivalence Information:
1. Predicate device name(s):
One Touch Ultra Blood Glucose Monitoring System
2. Predicate 510(k) number(s):
k002134
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device On Call® Vivid Pro Blood Glucose Monitoring System | Predicate One Touch Ultra Blood Glucose Monitoring System (k002134) |
| Intended Use/Indications for Use | Intended to be used for quantitative measurement of glucose in fresh capillary whole blood from the fingertip as an aid to monitor the effectiveness of diabetes control in people with diabetes. | Same |
| Methodology | Glucose oxidase biosensor | Same |
| Calibration | Plasma-equivalent | Same |
| Test time | 5 seconds | Same |
| Sample Type | Fresh capillary whole blood from the fingertip | Same |
| Data Port | One serial data port-Data transmission was not evaluated in this submission. | Same |
| Measurement range | 20 to 600 mg/dL | Same |
Differences
{3}
| Item | Device
On Call® Vivid Pro Blood Glucose Monitoring System | Predicate
One Touch Ultra Blood Glucose Monitoring System (k002134) |
| --- | --- | --- |
| Minimum Sample size | 0.8 μL | 1.0 μL |
| Hematocrit Range | 20-70% | 30-55% |
| Operating Temperature | 5-45°C (41-113°F) | 6-44°C (43-111°F) |
| Coding | Auto Coding by meter.
Automatic recognition of the intended coding after strip insertion | Manual Coding by manually selecting code by pressing button. |
| Meter memory | Up to 500 records with time and date | 150 blood glucose and control solution tests |
| Power Source | Two (2) CR 2032 3.0V coin cell batteries | One (1) CR 2032 3.0V coin cell battery |
| Meter Size | Approximately 60g (with battery installed) | 1.5 ounces with battery (Approximately 42g) |
K. Standard/Guidance Document Referenced (if applicable):
1. EN ISO 15197: 2003, In vitro diagnostic test systems - Requirements for In Vitro Whole Blood Glucose.
2. EN 11137-1: 2006 Sterilization of healthcare products. Radiation. Requirements for development, validation and routine control of a sterilization process for medical devices.
3. EN 11137-2: 2007 Sterilization of healthcare products. Radiation. Part 2: Establishing the sterilization dose.
4. EN 11137-3: 2006 Sterilization of healthcare products. Radiation. Guidance on dosimetric aspects.
5. EN 61326-1: 2006 Class B Electrical Equipment for Measurement, Control and Laboratory Use. EMC Requirements. General Requirements.
6. EN 61326-2-6: 2006 Electrical Equipment for Measurement, Control and Laboratory Use - EMC Requirements. Particular requirements. In vitro diagnostics (IVD) medical equipment.
7. EN 61010-1:2001 Safety requirements for electrical equipment for measurement, control and laboratory use. Part 1. General requirements.
8. EN 556-1:2001 Sterilization of medical devices. Requirements for medical devices to be designated "Sterile". Requirements for terminally sterilized devices.
9. EN11607-1:2006 Packaging for terminally sterilized medical devices. Requirements for materials, sterile barrier systems and packaging systems.
10. EN 10993-5:2009 Biological evaluation of medical devices. Tests for in vitro cytotoxicity
11. EN 10993-10:2009 Biological evaluation of medical devices. Tests for irritation and delayed-type hypersensitivity.
12. IEC/EN 61010-2-101:2002 Safety requirements for electrical equipment for measurement, control and laboratory use. Particular requirements for in vitro diagnostic (IVD) medical equipment.
13. ISO 13640:2002 Stability testing of in vitro diagnostic reagents
14. EN 62366:2008 Medical Devices. Application of usability engineering to medical
{4}
devices.
15. EN 62304: 2006 Medical Device Software. Software life-cycle processes
16. ISO 14971: 2009 Medical devices – Application of Risk management to medical devices
17. CLSI EP-6 Evaluation of Linearity of Quantitative Measurement Procedures: A Statistical Approach
## L. Test Principle:
The On Call® Vivid Pro Blood Glucose Monitoring System is a quantitative assay for the detection of glucose in fresh capillary whole blood from the fingertip. Amperometric technology is used for the detection of glucose from testing the strip (with whole blood sample) on the meter.
Reagent consisting of glucose oxidase and mediator is deposited onto the reaction cell section of the test strip with printed electrodes. When a drop of whole blood sample is applied to the reaction cell on the test strip, glucose in the blood sample reacts in the presence of glucose oxidase and mediator, and the reaction yields electrons. Thus, a current signal is produced from the reaction and detected by the meter. The detected current signal is then calculated by the meter, and glucose concentration reading is then displayed on the meter display.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The sponsor performed within-run precision studies using heparinized venous blood samples at five different glucose concentrations. Each sample was tested on 3 lots of test strips on 10 meters. Ten replicates were tested per meter, test strip lot, and glucose concentration, (N=100 per test strip) for a total 300 measurements for each glucose level. Results are summarized below:
| Glucose Level | 30-50 (mg/dL) | | | 50-110 (mg/dL) | | | 110-150 (mg/dL) | | | 150-250 (mg/dL) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Test Strip Lot | 1 | 2 | 3 | 1 | 2 | 3 | 1 | 2 | 3 | 1 | 2 | 3 |
| Mean (mg/dL) | 41.7 | 42.7 | 40.1 | 78.0 | 79.8 | 77.7 | 132.3 | 131.0 | 129.3 | 200.8 | 191.7 | 196.8 |
| SD | 1.6 | 1.6 | 1.3 | 3.1 | 2.5 | 2.7 | 4.0 | 4.0 | 3.7 | 4.4 | 4.4 | 5.8 |
| CV% | 3.9 | 3.7 | 3.2 | 3.9 | 3.1 | 3.5 | 3.0 | 3.1 | 2.8 | 2.2 | 2.3 | 3.0 |
| n | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
{5}
| Glucose Level | 250-400 (mg/dL) | | |
| --- | --- | --- | --- |
| Test Strip Lot | 1 | 2 | 3 |
| Mean (mg/dL) | 278.0 | 272.7 | 273 |
| SD | 6.0 | 7.5 | 6.5 |
| CV% | 2.2 | 2.7 | 2.4 |
| N | 100 | 100 | 100 |
Intermediate precision was evaluated using three levels of glucose control solutions, Level 1, Level 2, and Level 3. Each control solution was tested using 10 test strips on 10 meters (1 test strip per meter). 3 test strip lots were tested for 10 days for a total 300 measurements for each glucose level. Results are summarized below:
| Glucose Level | Level 1 | | | Level 2 | | | Level 3 | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Test Strip Lot | 1 | 2 | 3 | 1 | 2 | 3 | 1 | 2 | 3 |
| Mean (mg/dL) | 36.0 | 36.6 | 37.2 | 118.9 | 121.2 | 122.4 | 359.3 | 355.2 | 359.7 |
| SD | 1.6 | 1.5 | 1.6 | 3.4 | 3.4 | 4.2 | 8.5 | 11.6 | 11.0 |
| CV% | 4.4 | 4.1 | 4.2 | 2.8 | 2.8 | 3.5 | 2.4 | 3.3 | 3.0 |
| n | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
## b. Linearity/assay reportable range:
Linearity was evaluated using 11 venous blood samples ranging in glucose concentrations from 11.9 to 638.9 mg/dL (11.9, 28.0, 48.0, 78.7, 110.7, 168.5, 220.2, 326.4, 449.8, 549.4 and 638.9 mg/dL) as measured by the reference method. Each sample was tested in replicates of 4 on each of 3 test strip lots. The values obtained from the On Call® Vivid Pro Blood Glucose meter were compared to those obtained from the reference method. The results from regression analysis are summarized below:
Lot #1: y=0.9998x+4.2445; R²=0.999
Lot#2: y=1.0202x+2.8045; R²=0.998
Lot#3: y=1.0318x+2.3263; R²=0.999
The results of the study support the claimed glucose measuring range of 20 to 600 mg/dL.
## c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
{6}
The On Call® Vivid Pro Blood Glucose Control solutions are traceable to the NIST SRM 917b reference material.
## Control Value Assignment:
Value assignment is based on testing the control materials on multiple meters and test strip lots. Several replicates are performed on each level of control and a mean value is obtained. The mean value is then compared to a target value which has been established for each lot of control solution. Control solution ranges are provided on the test strip vial label. The protocols were reviewed and found to be adequate.
## Stability:
**Control Solutions:** Accelerated stability studies were conducted to assess the shelf-life and open vial stability of the control solutions and test strips. Real-time stability studies are still on-going. Unopened control solutions have a 24 month shelf life and are stable for 6 months after first opening when stored at 2-30°C (36-86°F).
**Test Strips:** The sponsor claims that the unopened test strips have a 24 month shelf life and are stable for 6 months after opening when stored at 25-30°C (77-86°F).
This information is provided in the labeling of the test strips and control solution materials.
## d. Detection limit:
The reportable range is 20 to 600 mg/dL based on the linearity/reportable range studies above. The low and high detection limits for this device have been set at 20 and 600 mg/dL. Readings below 20 mg/dL and above 600 mg/dL will indicate a "Lo" and "Hi" on the meter display, respectively. See linearity study in Section M1b above.
## e. Analytical specificity:
To assess potential interferences with the OnCall Vivid Pro system the sponsor used whole blood samples that were spiked to nominal glucose concentrations of 40-60 mg/dL, 80-120 mg/dL and 300-400 mg/dL. The interfering substances were prepared at 2 concentrations to include normal (therapeutic) and high (toxic). Each sample was analyzed in quadruplicate using three different lots of test strips. The sponsor defines no significant interference as ≤10% difference. Results are presented in the table below:
{7}
| Potential Interfering Substance | Concentration at which no interference is observed | | Potential Interfering Substance | Concentration at which no interference is observed |
| --- | --- | --- | --- | --- |
| Acetaminophen | 20 mg/dL | | Maltose | 100 mg/dL |
| Ascorbic Acid | 3 mg/dL | | Mannitol | 600 mg/dL |
| Cholesterol | 500 mg/dL | | Methyl Dopa | 1.5 mg/dL |
| Conjugated Bilirubin | 50 mg/dL | | Salicylic Acid | 60 mg/dL |
| Creatinine | 5 mg/dL | | Sorbitol | 70 mg/dL |
| Dopamine | 0.09 mg/dL | | Tetracycline | 1.5 mg/dL |
| Ethanol | 400 mg/dL | | Tolazamide | 10 mg/dL |
| Fructose | 100 mg/dL | | Tolbutamide | 64 mg/dL |
| Galactose | 100 mg/dL | | Triglycerides | 3000 mg/dL |
| Hemoglobin | 500 mg/dL | | Unconjugated Bilirubin | 40 mg/dL |
| Ibuprofen | 50 mg/dL | | Uric Acid | 23.5 mg/dL |
| Lactose | 25 mg/dL | | Xylose | 200 mg/dL |
| L-Dopa (Levo-Dopa) | 3 mg/dL | | | |
The sponsor has the following limitations in their labeling:
Acetaminophen, uric acid and ascorbic acid (vitamin C) (when occurring in blood at normal or at high therapeutic concentration) do not significantly affect results. However, abnormally high concentration in blood may cause inaccurately high results.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
To assess system accuracy, results from the On Call® Vivid Pro Blood Glucose Monitoring system were compared to results from YSI Model 2300 Stat Plus Glucose Analyzer (reference). Capillary samples from 105 participants with glucose concentrations ranging from 40.0- 562.5 mg/dL (capillary) by YSI method were analyzed. To obtain blood glucose concentrations <50mg/dL and >400 mg/dL, 8 samples were allowed to glycolize or were spiked to achieve desired glucose concentrations. Three lots of test strips were used in the testing. The results relative to the reference method are summarized in the tables below:
Fingertip Sample Site On Call Vivid Result (vs. Plasma YSI with Fingertip sample):
{8}
System accuracy for glucose concentrations $< 75\mathrm{mg / dL}$
| Strip Lot | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| --- | --- | --- | --- |
| 1 | 13/17 (76.5%) | 17/17 (100.0%) | 17/17 (100.0%) |
| 2 | 9/17 (52.9%) | 17/17 (100.0%) | 17/17 (100.0%) |
| 3 | 12/17 (70.6%) | 17/17 (100.0%) | 17/17 (100.0%) |
| Combined | 34/51 (66.66%) | 51/51 (100.0%) | 17/17 (100.0%) |
System accuracy results for glucose concentrations $\geq 75\mathrm{mg / dL}$
| Strip Lot | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL | Within ± 20 mg/dL |
| --- | --- | --- | --- | --- |
| 1 | 63/96 (65.6%) | 88/96 (91.7%) | 96/96 (100.0%) | 96/96 (100.0%) |
| 2 | 62/96 (64.6%) | 85/96 (88.5%) | 96/96 (100.0%) | 96/96 (100.0%) |
| 3 | 57/96 (59.4%) | 86/96 (89.6%) | 96/96 (100.0%) | 96/96 (100.0%) |
| Combined | 182/288 (63.2%) | 259/288 (89.9%) | 288/288 (100.0%) | 288/288 (100.0%) |
Linear Regression results On Call® Vivid Pro Blood Glucose fingerstick results vs. YSI Model 2300 Stat Plus Glucose Analyzer
| Sample Site | Strip Lot | Slope | Intercept | R | R2 | N |
| --- | --- | --- | --- | --- | --- | --- |
| Fingertip | 1 | 1.0265 | -1.8674 | 0.9947 | 0.9895 | 113 |
| Fingertip | 2 | 1.0018 | 1.2672 | 0.9942 | 0.9885 | 113 |
| Fingertip | 3 | 1.0083 | -1.9030 | 0.9948 | 0.9895 | 113 |
b. Matrix comparison:
Not applicable. Only capillary whole blood samples are used.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
# User Performance Study:
A user performance evaluation was performed to compare the lay user self-test results from capillary fingertip samples and the health technician results (fingertip samples from professional user) with the reference method (YSI). Studies were performed
{9}
with 105 lay user participants and 3 lots of test strips. The samples ranged from 48.0 to 444.0 mg/dL glucose as measured by the YSI (reference) method. A summary of the results received from one lot are summarized in the tables below:
## Lay user vs. YSI:
Results for glucose concentration < 75 mg/dL
| Within ±5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| --- | --- | --- |
| 7/12 (58.3%) | 12/12 (100.0%) | 12/12 (100.0%) |
Results for glucose concentration ≥ 75 mg/dL
| Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% |
| --- | --- | --- | --- |
| 51/93 (58.3%) | 83/93 (89.2%) | 91/93 (97.8%) | 93/93 (100.0%) |
## Health Technician vs. YSI
Results for glucose concentration < 75 mg/dL
| Within ±5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| --- | --- | --- |
| 9/12 (75.0%) | 12/12 (100.0%) | 12/12 (100%) |
Results for glucose concentration ≥ 75 mg/dL
| Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% |
| --- | --- | --- | --- |
| 62/93 (75.0%) | 85/93 (91.3%) | 93/93 (100.0%) | 93/93 (100%) |
| Strip Lot | Tester | Linear Regression | R² | N |
| --- | --- | --- | --- | --- |
| MI1110428A | Layperson | y=1.0279x -5.7082 | 0.9957 | 105 |
| MI1110428A | Technician | y=1.0421x -6.5499 | 0.9949 | 105 |
## 4. Clinical cut-off:
Not applicable
## 5. Expected values/Reference range:
For people without diabetes¹:
Fasting and before meals: 70-100 mg/dL
2 hours after meal: < 140 mg/dL
¹ADA Clinical Practice Recommendations, 2011 Diabetes Care, 2011, Vol.34, supplement 1, S62-S69
{10}
11
N. Instrument Name:
On Call® Vivid Pro Blood Glucose Meter
O. System Descriptions:
1. Modes of Operation:
Each test strip is single use only and requires a sample volume of 0.8μL.
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?:
Yes ______ or No ☐ X
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?:
Yes ☐ or No ☐ X
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
The device is intended to be used with capillary whole blood from the fingertip. The whole blood sample is applied directly to the test strip by capillary action.
5. Calibration:
There is no calibration required for the On Call® Vivid Pro Blood Glucose meter. The meter is plasma-calibrated and is an auto coding meter. The coding information is on each test strip so the meter can recognize the coding information when each strip is inserted into the meter.
{11}
6. Quality Control:
Glucose control solutions (On Call® Vivid Pro Blood Glucose Control Solutions) at two different concentrations (control solution 1 and control solution 2) can be run with this device. The control solution 1 and control solution 2 are not provided with the meter. The ranges for the control solution are test strip dependant. An example of a lot specific range would be: control solution 1 (93-139mg/dL) control solution 2 (285-427 mg/dL). The meter has an algorithm to automatically recognize the control solutions to prevent control results from being stored in the internal memory as a patient result. Recommendations on when to test the control materials are provided in the labeling. The control solution readings are not included in the average of the patient results. An acceptable range for each control level is printed on the test strip vial label. The user is cautioned not to use the meter if the control result falls outside these ranges.
12
{12}
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. Altitude Study:
This study was conducted at a single site to evaluate the effect of altitude on the On Call Vivid Pro BGMS. The tested glucose range (obtained using YSI reference method) was tested at 8 different glucose concentrations from 41.7 to 567 mg/dL at sea level and 35.7 to 525 mg/dL at 8,516 ft above sea level. The glucose results obtained obtained using the On Call Vivid Pro blood glucose meter had acceptable biases to the reference to support the claim in the labeling that altitudes up to 8,516 feet have not significant affect on blood glucose measurements from the On Call Vivid Pro System.
2. Hematocrit Study:
The effect of different hematocrit levels on the performance of the On Call Vivid Pro BGMS was evaluated using venous whole blood samples with 8 different hematocrit levels (20, 25, 30, 40, 50, 60, 65, and 70%) across the glucose measuring range (target glucose concentrations 50, 100, 275, and 500 mg/dL). Three test strip lots were evaluated and results from the On Call Vivid Pro BGMS compared to the YSI method. The results support the sponsor's claimed hematocrit range of 20 – 70%.
3. Temperature and Relative Humidity Study:
Operating Conditions Study:
The sponsor performed temperature and humidity studies using venous whole blood samples to evaluate temperatures ranging from 5°C to 45°C and relative humidity from 10% to 45%. In this study, three test strip lots were tested at 7 combined temperature and relative humidity conditions (5°C/10% RH, 21°C/10% RH, 45°C/10%, 5°C/90% RH, 21°C/90% RH, 45°C/90% RH, and 21°C/45% RH). Three blood samples were adjusted to the following glucose concentration ranges: (40-60 mg/dL, 80-120 mg/dL, 500-550 mg/dL). Each sample was also compared to the YSI reference method. The results demonstrate that accurate readings can be obtained at temperatures ranging from 5 - 45°C and relative humidity conditions ranging from 10 - 90%.
4. EMC Electromagnetic Compatibility and Electrical Safety verification testing of the On Call Vivid Pro Blood Glucose Monitoring System was performed and found to be adequate.
5. Sample volume study:
The sponsor performed a sample volume study to demonstrate that 0.8μL of whole blood is sufficient volume for the On Call Vivid Pro blood glucose meter system. For this study, sample volumes ranging from 0.4μL – 1.0 μL were evaluated using three different lots of test strips. The blood glucose sample results collected using the On Call Vivid Pro blood glucose meter were compared with glucose values obtained with the YSI reference method and supported the minimum sample volume claim of 0.8 μL.
13
{13}
6. Infection control:
The device system is intended for multiple-patient use. Disinfection efficacy studies were performed on the materials comprising the meter by an outside commercial testing demonstrating complete inactivation of hepatitis B virus (HBV) with the chosen disinfectant, (DisCide Wipes) disinfectant wipe (EPA Reg. No. 10492-4). Robustness studies were also performed by the sponsor demonstrating that there was not change in performance or external materials of the meter after 10,950 times of cleaning and disinfection cycles, using DisCide disinfectant wipes, to simulate 3 years of use by professional users. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
7. Readability Assessment:
A readability assessment was conducted on the user guides, test strip package insert and control solution package insert. It was determined that the readability of the labeling, using a Flesch-Kincaid analysis, were found to be written at the 7th grade level.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.