← Product Code [LZA](/productcode/LZA) · K122998

# BLUE POWDERED NITRILE EXAMINATION GLOVE (K122998)

_Northstar Healthcare Holdings · LZA · Nov 6, 2012 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K122998

## Device Facts

- **Applicant:** Northstar Healthcare Holdings
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Nov 6, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Nitrile Powdered Examination Glove is a Disposable device intended for medical and dental purposes to be worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Textured, blue, powdered nitrile examination glove; disposable; worn on examiner's hand or finger. Used in medical and dental settings to provide a barrier against cross-contamination between patient and examiner. Device is non-sterile and single-use. Performance verified through bench testing for dimensions, physical properties, pinhole resistance, and residual powder levels per ASTM standards. Biocompatibility confirmed via skin irritation and dermal sensitization testing.

## Clinical Evidence

Bench testing only. No clinical data. Performance evaluated against ASTM D6319-10 (dimensions, physical properties), ASTM D5151-06 (pinholes), ASTM D6124-06 (residual powder), and ISO 10993-10:2010 (biocompatibility). Results confirmed device meets all specified standards; residual powder values < 10 mg/dm²; non-irritating and non-sensitizing.

## Technological Characteristics

Material: Nitrile. Form: Powdered, textured, blue, non-sterile, single-use. Dimensions: Length 230mm min; thickness 0.06-0.08mm. Standards: ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, ISO 10993-10:2010. Physical properties: 14MPa tensile strength, 500% elongation (before/after aging).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Nitrile Powdered Examination Gloves ([K974907](/device/K974907.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Abbreviated 510(k) Premarket Notification Nitrile Powdered Examination Gloves

K122998 Section 5, 510(k) Summary Page 1 of 3

5. 510(k) SUMMARY

| DATE:  | February 17, 2012                                                                 |
|--------|-----------------------------------------------------------------------------------|
| OWNER: | Northstar Healthcare Holdings<br>70 Sir John Rogerson's Quay<br>Dublin 2, Ireland |

NOV 6 2012

| OFFICIAL CORRESPONDENT: | Michael Riordan<br>Operations Manager<br>Telephone: 00353-21-4548255<br>Fax: 00353-21-4548294<br>Email: michael.riordan@mckesson.ie |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| DEVICE NAME:            | Trade Name: Textured, Blue Powdered Nitrile Examination Glove                                                                       |
|                         | Common Name: Patient Examination Gloves                                                                                             |
|                         | Classification: Patient Examination Gloves                                                                                          |
|                         | Class: Class I                                                                                                                      |
|                         | Product Code: LZA                                                                                                                   |

## PREDICATE DEVICE(S):

| Predicate<br>510(k) | Device Name                            | Indication                                                                                                                                                                                   | Clearance<br>Date | Company                           |
|---------------------|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|-----------------------------------|
| K974907             | Nitrile Powdered<br>Examination Gloves | The examination glove is a disposable<br>device intended for medical and dental<br>purposes that is worn on the examiner's<br>hand to prevent contamination between<br>patient and examiner. | 13 Mar 1998       | Smart Glove<br>Corp., SDN,<br>BHD |

DEVICE DESCRIPTION: Nitrile Powdered patient examination glove

{1}------------------------------------------------

K122998

Section 5, 510(k) Summary Page 2 of 3

## STATEMENT OF INTENDED USE:

The Nitrile Powdered Examination Glove is a Disposable device intended for medical and dental purposes to be worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## TECHNOLOGICAL CHARACTERISTICS:

The Nitrile Powdered Examination Glove is substantially equivalent to the predicate device with regard to physical characteristics, design, product features, and intended use. Both gloves are made with nitrile using similar manufacturing processes.

| Feature                          | Nitrile Powdered Examination Gloves<br>K974907<br>Predicate                                                                                                                               | Textured, Blue Nitrile Powdered Examination Gloves<br>Proposed                                                                                                                           |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                     | Intended for medical purposes that is worn on the<br>examiner's hand to prevent contamination between patient<br>and examiner.                                                            | Same                                                                                                                                                                                     |
| Indications for<br>Use Statement | The examination glove is a disposable device intended for<br>medical and dental purposes that is worn on the<br>examiner's hand to prevent contamination between patient<br>and examiner. | The examination glove is a disposable device intended for<br>medical purposes that is worn on the examiner's hand or finger<br>to prevent contamination between patient and examiner.    |
| Description                      | The powdered examination gloves made of nitrile. The<br>gloves are provided in sizes extra small, small, medium,<br>large and extra large.                                                | The powdered examination gloves made of nitrile. The gloves<br>are provided in sizes small, medium, large, and extra large.                                                              |
| Presentation                     | Gloves are provided in dispenser boxes.                                                                                                                                                   | Same                                                                                                                                                                                     |
| Material                         | Nitrile                                                                                                                                                                                   | Same                                                                                                                                                                                     |
| Sterilization                    | Non-sterile                                                                                                                                                                               | Same                                                                                                                                                                                     |
| Single Use                       | Yes                                                                                                                                                                                       | Same                                                                                                                                                                                     |
| Dimensions                       | Meets ASTM D3578-95                                                                                                                                                                       | Length Small 230 mm min.<br>Medium 230 mm min.<br>Large 230 mm min.<br>Extra Large 230 mm min.<br>Width Small 80-90 mm<br>Medium 90-100 mm<br>Large 100-110 mm<br>Extra Large 110-120 mm |
|                                  |                                                                                                                                                                                           | Thickness Finger 0.08 mm min.<br>Palm 0.06mm min                                                                                                                                         |
| Physical<br>Properties           | Meets ASTM D3578-95                                                                                                                                                                       | Before aging/after aging<br>Elongation 500% 500%<br>Tensile Strength 14MPa 14MPa                                                                                                         |
| Freedom from<br>Pinholes         | Meets ASTM D5151-92                                                                                                                                                                       | Meets ASTM D5151-06                                                                                                                                                                      |
| Residual Powder                  | Meets ASTM D6124-97                                                                                                                                                                       | Meets ASTM D6124-06                                                                                                                                                                      |

{2}------------------------------------------------

K122998

Page 3 of 3

Section 5, 510(k) Summary

Abbreviated 510(k) Premarket Notification Nitrile Powdered Examination Gloves

# ASSESSMENT OF NONCLINICAL DATA:

| Characteristic           | Standard                                                          | Device Performance                                                |
|--------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------|
| Dimension                | ASTM Standard D6319-10                                            | Meets                                                             |
| Physical Properties      | ASTM Standard D6319-10                                            | Meets                                                             |
| Freedom from<br>Pinholes | 21 CFR 800.20; ASTM D5151-<br>06                                  | Meets                                                             |
| Powder Residual          | ASTM Standard D6124-06                                            | Meets<br>Results generated<br>values below 10<br>mg/dm² of powder |
| Biocompatibility         | Primary Skin Irritation in<br>rabbits<br>(ISO 10993-10:2010)      | Gloves are non-irritating                                         |
|                          | Dermal Sensitization in the<br>guinea pig (ISO 10993-<br>10:2010) | Gloves do not<br>display any<br>potential for<br>sensitization    |

## CONCLUSIONS:

The Nitrile Powdered Examination Gloves meet the requirements of established standards ASTM D6319-10, ASTM D5151-06, ASTM D6124-06 and ISO 10993-10:2010.

Based on the comparison of intended use, design, materials and performance, the Nitrile Powdered Examination Gloves are substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The text is in all caps and is oriented to follow the curve of the circle.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

## November 6, 2012

Northstar Healthcare Holdings C/O Underwriters Laboratories, Incorporated Mr. Ned Devine 333 Pfingsten Road Northbrook, Illinois 60062

Re: K122998

Trade/Device Name: Textured Blue Powdered Nitrile Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: October 19, 2012 Received: October 22, 2012

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

## Page 2 - Mr. Devine

 Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Digitally signed by Anthony D. Watson DN: c=US, o=U.S. Government, ou=HHHS, ou=FDA, ou=People, cn=Anthony D. Watson, 0.9.2342.19200300.100.1.1=1300092402 Date: 2012.11.06 14:26:52 -05/00'

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory, Infection Control and

Dental Devices

Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## INDICATIONS FOR USE

510(k) Number (if known): K122998

#### Device Name: Textured Blue Powdered Nitrile Examination Glove

Indications for Use: The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## AND/OR

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

for ECW
(Division Sign-Off)

(Division Sign-On)
Division of Anesthesiology, General Hospital Division Control, Dental Devices

510(k) Number: K122998

Page 1 of 1

Northstar Healthcare Holdings

---

**Source:** [https://fda-staging.innolitics.com/device/K122998](https://fda-staging.innolitics.com/device/K122998)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).