K122955 · Wright Health Group , Ltd. · ELM · Mar 1, 2013 · Dental
Device Facts
Record ID
K122955
Device Name
DENTAVIT, MONARCH, ACROTONE, SENATOR, DENTORIUM
Applicant
Wright Health Group , Ltd.
Product Code
ELM · Dental
Decision Date
Mar 1, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3590
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use for these plastic teeth is to make partial or full dentures. These dentures are made to fit the patient's mouth as a removable device, not implanted. These plastic tech may also be attached to dental implants. Federal law restricts this device to sale by or on the order of a physician. It is intended to be used only by appropriately trained and qualified healthcare professionals and servicing staff in clinical environments.
Device Story
Synthetic plastic teeth (models: Dentavit, Monarch, Acrotone, Senator, Dentorium) used for denture fabrication. Produced via polymerization of polymethylmethacrylate (PMMA) with methylmethacrylate monomer, cross-linking agents, and pigments under heat and pressure. Finished teeth are checked for flash and polished. Used by trained dental professionals in clinical environments to construct removable dentures or implant-supported prosthetics. Provides patients with functional and aesthetic tooth replacement. No software or automated processing involved.
Clinical Evidence
Bench testing only. Compliance with ADA specification No. 15:2008 (Artificial Teeth), BS EN ISO 22112:2006, and biocompatibility standards ISO 10993-1 and ISO 7405:2008.
Technological Characteristics
Materials: Polymethylmethacrylate (PMMA), methylmethacrylate monomer, cross-linking agents, non-toxic pigments. Standards: ADA specification No. 15:2008, BS EN ISO 22112:2006, ISO 10993-1, ISO 7405:2008. Manufacturing: Heat and pressure polymerization. Form factor: Preformed plastic teeth for denture integration.
Indications for Use
Indicated for use as components in the fabrication of removable partial or full dentures, or for attachment to dental implants, for patients requiring prosthetic tooth replacement.
Regulatory Classification
Identification
A preformed plastic denture tooth is a prefabricated device, composed of materials such as methyl methacrylate, that is intended for use as a tooth in a denture.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
510(K) SUMMARY August 10, 2012
### SUBMITTER INFORMATION: (3003103749)
Wright Health Group Ltd. Dunsinane Industrial Estate Kingsway West Dundee, Scotland DD2 3QD UK
## FDA AGENT/CORRESPONDING OFFICIAL INFORMATION:
North American Technical Services (NATS) Corp. 30 Northport Rd Sound Beach, NY 11789 Tel: 631-744-0059 Fax: 631-744-0192 Email: natscorp@aol.com Contact: Stephen T. Micoch
### DEVICE NAME:
Name: Denture Preformed Plastic Teeth Proprietary Models: Dentavit, Monarch, Acrotone, Senator, Dentorium Classification: II Product Code: ELM Regulation: 21 CFR 872.3590
### PREDICATE DEVICES:
Union Dental S.A./ Unidesa Odi K070591 - model Replica Dental Vipi Ltd K022300 - model Vipi Dent Plus Vident K914476 - model Vitapan Huge Dental Material Co., Ltd. K101029 -- model Kaili
### DESCRIPTION:
The synthetic plastic teeth are used in the production of dentures. They are produced from a I he synuletic process. Polymethylmethacrylate or PMMA is mixed with methylmethacrylate monomer, a cross linking agent and non-toxic pigments under heat and pressure. Once the monomer, a cross iniking agoit and non teen are checked for flash and polished according to the relevant model characteristics.
# SUBSTANCIAL EQUIVALENCE AND TECHNOLOGY:
The plastic teeth have comparable chemical composition as the predicate devices. The models I he piastic tecur have comparable chesage to the predicate devices. Also, preamendment are similal in size, shape, color, and asage to the productive. The added models use the same materials and controls established for over 20 years.
MAR 0 1 2013
{1}------------------------------------------------
The intended use for these plastic teeth is to make partial or full dentures. These dentures are made to fit the patient's mouth as a removable device, not implanted. These plastic tech may also be attached to dental implants.
Federal law restricts this device to sale by or on the order of a physician. It is intended to be used only by appropriately trained and qualified healthcare professionals and servicing staff in clinical environments.
# SUMMARY OF NONCLINICAL TESTS AND DESIGN CONTROL ACTIVITIES:
The performance and safety testing activities were conducted on acceptable dental materials to establish the reliability characteristics of these teeth devices. Validation and verification of compliance with the following mandatory and voluntary standards has been made
- CE Medical Directive 93/42/EEC 2007/47/EC t
- ISO9001/GMP/ISO13485: Quality System Certification . Factory control procedures are established for production QA and QC with use of ADA specification No. 15:2008 Artificial Teeth, and BS EN ISO22112:2006, with Biocompatibility Standards ISO10993-1, ISO7405:2008
This control activity shows that there are no new questions of safety and effectiveness for the plastic teeth models made by Wright.
### CONCLUSION:
For and and the transity
The plastic teeth are substantially equivalent to the predicate teeth and have the same intended use. They are suitable for equivalent use on dentures according to the specifications stated based on the same materials, production process and technology.
Wright has been a manufacturer for about 100 years and has been in the market for over 40 years and has met preamendment rules. There are no significant health incidents in that time. These models have the same intended use and similar indications, technical characteristics and function as the predicate devices. Suitable dental industry materials have been used by Wight and predicate companies that solve biocompatibility, durability and esthetic concerns. The minor differences raise no new issues of safety or effectiveness. These models are substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines representing the staff and two snakes intertwined around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 1, 2013
Wright Health Group, Limited C/O Mr. Stephen Mlcoch Chief Executive Officer North American Technical Services (NATS) Corporation 30 Northport Road Sound Beach NY 11764
Re: K122955
Trade/Device Name: Plastic Teeth Models Dentavit, Monarch, Acrotone, Senator, and Dentorium Regulation Number: 21 CFR 872.3590 Regulation Name: Prosthetic Devices Regulatory Class: II Product Code: ELM Dated: January 15, 2013 Received: February 13, 2013
Dear Mr. Mlcoch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Mlcoch
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Kwame O. Ulmer for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K122955
#### INDICATIONS FOR USE
510(K) Number (if known):
Device Name:
Plastic teeth models Dentavit, Monarch, Acrotone, Senator, and Dentorium.
Indications for Use:
These dental plastic teeth are parts used for denture fabrication. PMMA is mixed with monomer in a polymenzation process. The plastic teeth are part of dentures made to fit a patient's mouth which are removable or may be attached to dental implant devices.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary S. Runner -S 2013.02.27 15:20:12 -05'00'
(Division Sign-Off) . Division of Anesthesiology, General Hospital ntection Control, Dental Devices
510(k) Number: K1208955
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.