WONDFO MULTI-DRUG URINE TEST CUP MODEL W2002-CU; W2003-CU; W2004-CU; W2005-CU; W2006-CU; W2007-CU; W2008-CU; W2009-CU;
Device Facts
| Record ID | K122904 |
|---|---|
| Device Name | WONDFO MULTI-DRUG URINE TEST CUP MODEL W2002-CU; W2003-CU; W2004-CU; W2005-CU; W2006-CU; W2007-CU; W2008-CU; W2009-CU; |
| Applicant | Guangzhou Wondfo Biotech Co., Ltd. |
| Product Code | DKZ · Clinical Toxicology |
| Decision Date | Nov 15, 2012 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.3100 |
| Device Class | Class 2 |
Indications for Use
Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Methamphetamine, Secobarbital, Oxazepam, Cocaine, Cannabinoids, Amphetamine, Phencyclidine, Notriptyline and Methadone, Methylenedioxymethamphetamine, Morphine, Oxycodone in human urine at the cutoff concentrations of: [Table of cut-off levels]. Configuration of the Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel can consist of any combination of the above listed drug analytes. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The test will yield preliminary positive results when prescription drugs Notriptyline, Oxazepam, Oxycodone and Secobarbital are ingested, even at or above therapeutic doses. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
Device Story
Wondfo Multi-Drug Urine Test Cup and Panel are lateral flow immunochromatographic assays; detect drugs of abuse in human urine. Principle: competitive binding; antigen-antibody interaction. Input: human urine sample. Output: visual colored lines on test strips; absence of line in test region indicates positive result (drug concentration above cutoff); presence of line indicates negative result. Control line always appears. Used in clinical or home settings; operated by clinicians or patients. Provides preliminary screening results; requires confirmatory testing via GC/MS or LC/MS for clinical decision-making. Benefits: rapid, simultaneous screening for multiple substances.
Clinical Evidence
Bench testing only. Verification and validation activities performed per design control procedures; results met predetermined acceptance criteria.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding principle. Form factors: test cup and dip card. Analyte detection via antigen-antibody interaction. No electronic components, energy source, or software algorithms.
Indications for Use
Indicated for qualitative, simultaneous detection of Amphetamine, Secobarbital, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, MDMA, Morphine, Methadone, Phencyclidine, Notriptyline, and Oxycodone in human urine. For prescription and OTC use. Preliminary results only; requires confirmatory testing (GC/MS or LC/MS).
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
- Wondfo single DOA Tests (K112071)