K122899 · C.R. Bard, Inc. · LJT · Nov 15, 2012 · General Hospital
Device Facts
Record ID
K122899
Device Name
POWERPORT CLEARVUE SLIM IMPLANTABLE PORT
Applicant
C.R. Bard, Inc.
Product Code
LJT · General Hospital
Decision Date
Nov 15, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PowerPort® ClearVUE® Slim Implantable Port with 8F Polyurethane Catheter is intended to be an implanted vascular access device designed to provide long-term, repeated access to the vascular system.
Device Story
Implantable vascular access port; consists of hard plastic port body and 8F polyurethane catheter; attached via catheter lock compression fitting. Triangular body shape; three palpation bumps on septum. Used for long-term, repeated vascular access; infusion of medications, fluids, nutrition, blood products; blood withdrawal. Operated by clinicians in clinical settings. For power injection of contrast media, must be accessed with Bard PowerLoc® Safety Infusion Set (SIS). Output is delivery of fluids/medications or withdrawal of blood samples. Benefits include reliable, repeated vascular access for patients requiring frequent therapy.
Clinical Evidence
Bench testing only. Verification testing performed per 21 CFR 820.30, including MR safety (ASTM F2503, F2052, F2119, F2182, F2213), catheter requirements (ISO 10555-1, ISO 10555-3), sterilization (ISO 11135-1), biocompatibility (ISO 10993-1, ISO 10993-7), and bacterial endotoxins (AAMI ST72). Risk management conducted per ISO 14971.
Technological Characteristics
Hard plastic port body; 8F polyurethane catheter; catheter lock compression fitting. Triangular shape with three palpation bumps. Sterilized via ethylene oxide. Complies with ISO 10555-1/3, ISO 10993-1/7, and ASTM standards for MR safety.
Indications for Use
Indicated for patients requiring repeated vascular access for infusion of medications, IV fluids, parenteral nutrition, blood products, and blood withdrawal. When used with PowerLoc® Safety Infusion Set, indicated for power injection of contrast media at max rate 5 mL/s.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
Titanium PowerPort® isp with 8F Polyurethane Catheter
Submission Summary (Full Text)
{0}------------------------------------------------
K122899
ﮯ
Section 5 – 510(k) Summary
Bard Access Systems, Inc.
PowerPort® ClearVUE® Slim Implantable Port
with 8F Polyurethane Catheter
Traditional 510(k) Premarket Notification
Image /page/0/Picture/3 description: The image shows the logo for BARD ACCESS SYSTEMS. The word "BARD" is in large, bold, sans-serif font. Below it, in a smaller font, are the words "ACCESS SYSTEMS."
NOV 15 2012
## 510(k) Summary 21 CFR 807.92(a)
| General<br>Provisions | Submitter Name: | Bard Access Systems, Inc. |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| | Address: | 605 North 5600 West<br>Salt Lake City, UT 84116 |
| | Contact Person: | Amy J. Honey<br>Regulatory Affairs Specialist |
| | Telephone Number: | (801) 522-5671 |
| | Fax Number: | (801) 522-5425 |
| | Date of Preparation: | September 19, 2012 |
| Subject<br>Device | Trade Name: | PowerPort® ClearVUE® Slim Implantable Port with 8F |
| | Common Name: | Polyurethane Catheter |
| | Classification Name: | Implanted Infusion Port & Catheter |
| | Product Code/<br>Regulation: | LIT/21 CFR §880.5965 |
| Predicate Device | Trade Name: | Titanium PowerPort® isp with 8F Polyurethane Catheter |
| | Common Name: | Implanted Infusion Port & Catheter |
| | Classification Name: | LIT - Subcutaneous, Implanted, Intravascular Infusion<br>Port & Catheter |
| | Product Code/<br>Regulation: | LIT/21 CFR §880.5965 |
| Device<br>Description | The PowerPort® ClearVUE® Slim with 8F Polyurethane Catheter is a member of<br>the PowerPort® series of power-injectable implanted ports. The subject device<br>consists of a hard plastic port and 8F polyurethane catheter that is attached to<br>the port with a catheter lock compression fitting. The subject device is<br>distinguishable as a member of BAS's power-injectable port series by the<br>triangular body shape and three palpation bumps on the septum. | |
| | PowerPort® implanted ports can be used for routine vascular access using a non-<br>coring access needle. However, for power injection procedures, PowerPort® ports<br>must be accessed with a Bard PowerLoc® Safety Infusion Set (SIS) to create a<br>power-injectable system. | |
| Intended Use | The PowerPort® ClearVUE® Slim Implantable Port with 8F Polyurethane Catheter<br>is intended to be an implanted vascular access device designed to provide long-<br>term, repeated access to the vascular system. | |
| Indications For<br>Use | The PowerPort® ClearVUE® Slim Implantable Port with 8F Polyurethane Catheter<br>is indicated for patient therapies requiring repeated access to the vascular<br>system. The port system can be used for infusion of medications, I.V. fluids,<br>parenteral nutrition solutions, blood products, and for the withdrawal of blood<br>samples. When used with a PowerLoc® Safety Infusion Set (SIS), the PowerPort®<br>device is indicated for power injection of contrast media. For power injection of<br>contrast media, the maximum recommended infusion rate is 5 mL/s. | |
| Technological<br>Characteristics | Technological characteristics of the PowerPort® ClearVUE® Slim Implantable Port<br>with 8F Polyurethane Catheter are substantially equivalent with respect to basic<br>design and function of the Titanium PowerPort® isp with 8F Polyurethane<br>Catheter. The differences between the two devices do not impact the intended<br>use, and do not raise any new questions regarding safety or efficacy. | |
| Safety &<br>Performance<br>Tests | Verification testing has been performed in accordance with Design Controls per<br>21 CFR §820.30. The following guidance documents and standards in conjunction<br>with in-house protocols were used to determine appropriate methods for<br>evaluating the performance of the device:<br>Guidance on 510(k) Submissions for Implanted Infusion Ports;<br>October, 1990 Guidance – Establishing Safety and Compatibility of Passive Implants<br>in the Magnetic Resonance (MR) Environment; August 21, 2008 ASTM F2503: 2008, Standard Practice for Marking Medical Devices<br>and Other Items for Safety in the Magnetic Resonance (MR)<br>Environment ASTM F 2052: 2006, Standard Test Methods for Measurement of<br>Magnetically Induced Displacement Force on Medical Devices in the<br>Magnetic Resonance Environment ASTM F2119: 2007, Standard Test Method for Evaluation of MR<br>Image Artifacts from Passive Implants ASTM F2182 Rev A: 2011, Standard Test Method for Measurement of<br>Radio Frequency Induced Heating On or Near Passive Implants During<br>Magnetic Resonance Imaging ASTM F2213: 2006 (R2011),Standard Test Method for Measurement<br>of Magnetically Induced Torque on Medical Devices in the Magnetic<br>Resonance Environment ISO 10555-1/Amd 1/ Amd 2: 1195/1999/2009, Sterile, Single-Use<br>Intravascular Catheters, Part 1: General Requirements BS/EN/ISO 10555-3: 1996 Cor 1:2002, Sterile, Single-Use Intravascular | |
{1}------------------------------------------------
{2}------------------------------------------------
Catheters, Part 3: Central Venous Catheters AAMI/ANSI/ISO 11135-1: 2007, Sterilization of Healthcare Products – . Ethylene Oxide – Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices AAMI/ANSI/ISO 10993-1/Cor 1: 2009/2010, Biological Evaluation of . Medical Devices Part 1: Evaluation and Testing, and the FDA Modified ISO 10993 Test Profile AAMI/ANSI/ISO 10993-7:2008, Biological Evaluation for Medical . Devices; Part 7 - Ethylene Oxide Sterilization Residuals USP<161>: Sept. 8, 2009, Transfusion and Infusion Assemblies and ● Similar Medical Devices AAMI ST72, Jan. 1, 2001 (R 2010): Bacterial Endotoxins—Test . Methodologies, Routine Monitoring, and Alternatives to Batch Testing The subject device met all predetermined acceptance criteria derived from the above listed references and demonstrated substantially equivalent performance as compared to the cited predicate device. Risk management, including a failure modes and effects analysis (FMEA) of the subject device, was conducted in accordance with ISO 14971: 2009, Medical Devices - Risk Management for Medical Devices. Based on the indications for use, technological characteristics, and safety and performance testing, the subject PowerPort ClearVUE® Slim Implantable Port Summary of with 8F Polyurethane Catheter meets the minimum requirements that are Substantial considered adequate for its intended use and is substantially equivalent in design, Equivalence materials, sterilization, principles of operation, and indications for use to the predicate device cited.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized, curved lines.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
## November 15, 2012 .
C.R. Bard, Incorporated Ms. Amy Honey Regulatory Affairs Specialist 605 North 5600 West Salt Lake City, Utah 84116
Re: K122899
Trade/Device Name: PowerPort® ClearVUE® Slim Implantable Port with 8F Polyurethane Catheter Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port and Catheter
Regulatory Class: II Product Code: LJT Dated: September 19, 2012 Received: September 21, 2012
Dear Ms. Honey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.
Sincerely vours.
DN: c=US, o=U.S. Government, ou=HHS, DA. ou=People. cn=Anthony D. Watson 342.19200300.100.1.1=1300092402 Date: 2012.11.15 14:01:44 -05'00
Anthony D. Watson, B.S., M.S., M.B.A Director
Division of Anesthesiology, General Hospital, Respiratory. Infection Control and
Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
Section 4 - Indications for Use Statement
Bard Access Systems, Inc.
PowerPort® ClearVUE® Slim Implantable Port with 8F Polyurethane Catheter Traditional 510(k) Premarket Notification
### Indications for Use
# K122899
510(k) Number (if known):
PowerPort® ClearVUE® Slim Implantable Port with 8F Polyurethane Catheter
Device Name:
Indications for Use:
The PowerPort® ClearVUE® Slim Implantable Port with 8F Polyurethane Catheter is indicated for The PowerPort" Clear VDE "Sim Insplantable Per The vascular system. The port system can be
patient therapies requiring repeated access to the vascular system. Incoducts a patient therapies requiring repeated access to the rate to nutions, blood products, and used for infusion of medications, I.V. halas; perchers a PowerLoce" Safety infusion Set (SS), the
for the withdrawal of blood samples. When used with a Power interion of for the withdrawal of blood samples. When ease in edia. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s.
Prescription Use __ (Part 21 CFR §801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR §801 Subpart C)
(please do not write below this line – continue on another page if needed
Concurrence of CDRH, Office of Device Evaluation (ODE) Digitally signed by Sajjad H. Syed
DN: c=US, o=U.S. Government, ou=HHS, Sajjad H. Syed ou=FDA, ou=People, cn=Sajjad H. Syed, 0.9 2342.19200300.100.1.1=2000601742 Date: 2012.11.15 16:07:57 -05'00' (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.