← Product Code [JEY](/productcode/JEY) · K122890

# INION CPS ORBITAL PLATES (K122890)

_Inion OY · JEY · Jan 29, 2013 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K122890

## Device Facts

- **Applicant:** Inion OY
- **Product Code:** [JEY](/productcode/JEY.md)
- **Decision Date:** Jan 29, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4760
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Inion CPS Orbital Plates are intended for use in trauma and reconstructive procedures of the orbital cavity as part of the Inion CPS 1.5/2.0/2.5 Bioabsorbable Fixation System.

## Device Story

Inion CPS Orbital Plates are bioabsorbable fixation implants used in orbital cavity trauma and reconstructive surgery. Fabricated from a resorbable polylactic acid/trimethylcarbonate copolymer, the plates degrade in vivo via hydrolysis into monomers metabolized by the body. Plates are 0.5 mm thick (25-40 mm length, 24-40 mm width) and secured using 1.5 mm screws. Designed for use with specialized instrumentation including templates, drill bits, bone taps, screw drivers, plate benders, and a heating device. The device provides fixation during the healing period; it is not intended to replace healthy bone or withstand full load-bearing stress. Bioresorption occurs over two to four years. The device is provided sterile for single use by surgeons in clinical settings.

## Clinical Evidence

Bench testing only. Evidence includes mechanical shear tests, biomechanical fixation tests, and in vitro degradation studies to verify mechanical stability and degradation profiles. Functional, handling, and simulated clinical use tests were performed to confirm device performance and compatibility with accessory instrumentation.

## Technological Characteristics

Material: Resorbable polylactic acid/trimethylcarbonate copolymer. Dimensions: 0.5 mm thickness, 25-40 mm length, 24-40 mm width. Fixation: 1.5 mm screws. Principle: Bioresorbable fixation via hydrolysis. Sterilization: Provided sterile, not to be resterilized. Connectivity: None (mechanical device).

## Regulatory Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

## Predicate Devices

- Inion CPS 1.5/2.0/2.5 Bioabsorbable Fixation System ([K010352](/device/K010352.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) SUMMARY Inion CPS® Orbital Plate

# JAN 2 9 2013

#### Manufacturer and submitter:

Date: Contact person

Establishment registration number Trade name of the device Device classification and product code

Predicate device

Conformance with performance standards

Inion Ov. Lääkärinkatu 2. FIN-33520 Tampere, FINLAND December 20, 2012 Kati Marttinen, Quality and Regulatory Director Phone: +358 10 830 6600 Fax: +358 10 830 6691 kati.marttinen@inion.com 9710629

Inion CPS® Orbital Plate Class II (Special controls) Classification Panel: Dental Product Code: JEY Common name: Bone plate Regulation number: 872.4760 Inion CPS 1.5/2.0/2.5 Bioabsorbable Fixation System K010352 No applicable mandatory performance standards exist for this device. Compliance to voluntary consensus standards is listed in the application.

#### Device description and principles of operation

Inion CPS® Orbital Plates are intended for use in trauma and reconstructive procedures of the orbital cavity as part of the Inion CPS 1.5/2.0/2.5 Bioabsorbable Fixation System.

The Inion CPS® Orbital Plate fixation plates made of resorbable polylactic acid/trimethylcarbonate copolymer: This copolymer degrades in vivo by hydrolysis into L-lactic, D-lactic and TMC monomers which are then metabolised by the body.

Inion CPS® Orbital Plates are of dimensions: length: 25-40 mm, width: 24-40 mm, thickness: 0.5 mm. The fixation of the plates with Inion CPS® ø1.5 mm screws. The implants retain sufficient strength to fulfil their intended function during the healing period of the fracture or osteotomy, and degrade gradually thereafter. Bioresorption takes place within two to four years. The implants are provided sterile to the user and are not to be resterilized.

The Inion CPS® Orbital Plates provide fixation and are not intended to replace healthy bone or withstand the stress of full load bearing.

The Inion CPS® Orbital Plates are designed to be used with customized instrumentation consisting of template, drill bits, bone taps, screw driver, plate benders and a heating device.

{1}------------------------------------------------

### 510(k) SUMMARY Inion CPS® Orbital Plate

#### Indications for use

Inion CPS Orbital Plates are intended for use in trauma and reconstructive procedures of the orbital cavity as part of the Inion CPS 1.5/2.0/2.5 Bioabsorbable Fixation System.

#### Performance testing for substantial equivalence determination

Mechanical shear test and biomechanical fixation test were performed to verify the strength and fixation properties of Inion CPS Orbital Plates and to compare them to the predicate devices. Testing was conducted initially and during in vitro degradation.

In vitro degradation testing was carried out to determine the degradation profile (i.e., change in material and mechanical properties) and verify the sufficiency of the mechanical stability over healing period as the polymer degrades during in vitro degradation and to ensure the degradation of the Inion CPS Orbital Plates.

Functional and handling test and simulated clinical use test were performed to verify that the implants, accessory instruments, packaging and instructions for use are functioning. together as intended, and conform to the defined user needs and intended uses.

The data demonstrates that the intended use, material composition and scientific technology, degradation profile and mechanical properties of Inion CPS Orbital Plates are substantially equivalent with the predicate device Inion CPS 1.5/2.0/2.5 Bioabsorbable Fixation System (K010352). The devices have passed the tests for functionality and handling in simulated clinical use settings.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

#### January 29, 2013

Ms. Kati Marttinen Quality and Regulatory Director Inion Oy Lääkärinkatu 2 Tampere, Finland 33520

Re: K122890

Trade/Device Name: Inion CPS® Orbital Plate Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY Dated: January 3, 2013 Received: January 18, 2013

Dear Ms. Marttinen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Kwame Q. Ulmer

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Statement of Indications for Use

510(k) Number:

K122890

Device Name:

Inion CPS® Orbital Plate

The INION CPS Orbital Plates are intended for use in trauma and reconstructive procedures The INTON OF OF Of the Inion CPS 1.5/2.0/2.5 Bioabsorbable Fixation System.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

2013.01.23 Susan Runner DDS, MA -1-2:38:45 -05'00'

(Division Sign-Off) (Division Sign-Off)
Division of Anestheslology, General Hospital
Division Control Dental Devices Division of Anesthes.com-gy.
Infection Control, Dental Devices

510(k) Number: KLZZ 890

---

**Source:** [https://fda-staging.innolitics.com/device/K122890](https://fda-staging.innolitics.com/device/K122890)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
