K122869 · Depuy Mitek, A Johnson & Johnson Company · MAI · Dec 13, 2012 · Orthopedic
Device Facts
Record ID
K122869
Device Name
MILAGRO INTERFERENCE SCREWS
Applicant
Depuy Mitek, A Johnson & Johnson Company
Product Code
MAI · Orthopedic
Decision Date
Dec 13, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DePuy Mitek MILAGRO BR Interference Screws are designed to attach soft tissues to bone in orthopedic surgical procedures for following indications: Shoulder: Proximal Biceps Tenodesis, Acromio-Clavicular Repair Elbow: Distal Biceps Tenodesis, Ulnar Collateral Ligament repair Knee: Collateral Ligament Repair
Device Story
Milagro Interference Screws are absorbable, tapered, cannulated, threaded fasteners used for interference fixation of soft tissue or bone-tendon grafts. Fabricated from a composite of absorbable Poly (lactide-co-glycolide) polymer and Tricalcium Phosphate (TCP). Used in orthopedic surgical procedures (shoulder, elbow, knee) to attach soft tissue to bone. Operated by surgeons using provided reusable accessories: modular drivers, ratchet handle, guidewires, tap, and notcher. Device provides mechanical fixation during the healing process. Benefit includes secure graft attachment using biocompatible, absorbable material.
Clinical Evidence
Bench testing only. Performance verification included insertion torque, anchor pullout strength at T=0, 3, 6, and 12 weeks of in-vitro physiological aging, and torque to failure testing.
Technological Characteristics
Absorbable, tapered, cannulated, threaded fastener. Material: Poly (lactide-co-glycolide) polymer and Tricalcium Phosphate (TCP) composite. Dimensions: 5mm and 6mm diameters, 12mm length. Accessories: Reusable modular drivers, ratchet handle, guidewires, tap, and notcher. Sterilization: Identical to previously cleared predicates.
Indications for Use
Indicated for patients requiring soft tissue-to-bone attachment in orthopedic procedures, specifically proximal biceps tenodesis, acromio-clavicular repair, distal biceps tenodesis, ulnar collateral ligament repair, and collateral ligament repair.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the DePuy Synthes logo. The logo consists of the word "DePuy" in bold black letters, with a black circle to the left of the word. Below the word "DePuy" is the text "Synthes" in a smaller font. To the right of the logo is the tagline "never stop moving" in a smaller font.
5, 6x12mm MILAGRO INTERFERENCE SCREWS
DEC =1-3-2012
#### ATTACHMENT 1
### 5 10(k) SUMMARY - DEPUY MITEK MILAGRO* INTERFERENCE SOREWS
#### SUBMITTER'S NAME AND ADDRESS
DePuy Mitek, Inc. a Johnson & Johnson company 325 Paramount Drive Raynham, MA 02767
#### CONTACT PERSON
Kristine Christo DePuy Mitek Regulatory Affairs
| TELEPHONE | 508-828-3359 |
|---------------|----------------------|
| FACSIMILE | 508-977-6911 |
| E-MAIL | KChristo@its.jnj.com |
| DATE PREPARED | September 18th, 2012 |
#### NAME OF MEDICAL DEVICE
CLASSIFICATION NAME Fastener, Fixation, Biodegradable, Soft Tissue
#### COMMON/USUAL NAME
Bone Anchor
#### PROPRIETARY NAME
DePuy Mitek Milagro® Interference Screws
:
#### SUBSTANTIAL EQUIVALENCE
The proposed 6, 7 and 8 x15mm DePuy Mitek Milagro® Interference Screws are substantially equivalent to the following devices.
- K120589, K103831, K060830, K032717:
Milagro Interference Screws
- · K051726, K041356, K020043:
Arthrex Tenodesis Screws
#### FDA PRODUCT CODE
MAI, HWC
#### DEVICE CLASSIFICATION
This type of fixation screw was originally classified as a Class II medical device by the Orthopedic Review Panel, regulated as 21 CFR 888.3040 Smooth or Threaded Metallic Bone Fixation Fastener.
{1}------------------------------------------------
Image /page/1/Figure/1 description: The image contains the logo for DePuy Synthes, a company specializing in medical devices. The logo features the company name in bold, with the tagline "never stop moving" in smaller font to the right. A black circle is positioned to the left of the company name. The overall image appears to be a cropped section of a larger document or advertisement.
5, 6x12mm MILAGRO INTERFERENCE SCREWS
#### ATTACHMENT 1
Continues ...
# 5 I 0(k) SUMMARY - DEPUY MITEK MILAGRO® INTERFERENCE SOREWS
#### DEVICE DESCRIPTION
The DePuy Mitek Milagro® Interference Screws are absorbable, tapered, cannulated, threaded fasteners for use in interference fixation of soft tissue grafts or bone-tendon grafts. The Interference Screw is made from a composite made of absorbable Poly (lactide-co-glycolide) polymer and Tricalcium Phosphate (TCP).
Modular Drivers, Ratchet Handle, Guidewires, a Tap and Notcher are provided separately as reusable accessories to assist in the placement of the MILAGRO BR Screw.
The proposed DePuy Mitek Milagro® Interference Screws are offered in diameters of 5, 6mm and in the length of 12mm.
#### INDICATIONS FOR USE
The DePuy Mitek MILAGRO BR Interference Screws are designed to attach soft tissues to bone in orthopedic surgical procedures for following indications: Shoulder: Proximal Biceps Tenodesis, Acromio-Clavicular Repair Elbow: Distal Biceps Tenodesis, Ulnar Collateral Ligament repair Knee: Collateral Ligament Repair
#### TECHNOLOGICAL CHARACTERSTICS
The design specifications of the proposed DePuy Mitek Milagro® Interference Screws are substantially equivalent to the existing DePuy Mitek Milagro® Interference Screws cleared under 510(k) K120589, K103831, K060830 and K032717 except that the proposed DePuy Mitek Milagro® Interference Screws are smaller in size. Technological characteristics including design construct, and indications are similar to the predicate devices. There have been no changes to the material used to manufacture the product, packaging, and or to the sterilization processes, sterilization contractors; these are identical to that used for the currently marketed DePuy Mitek Milagro® Interference Screw (previously cleared by the FDA under the premarket notification K032717, K060830, K103831 and K120589).
#### NONCLINICAL TESTING
Product Design Verification activities, such as, Insertion Torque, Anchor Pullout (at T=0, 3, 6 and 12 week in-vitro physiological aging), and Torque to Failure were performed on the implant.
#### SAFETY AND PERFORMANCE
Results of performance and safety testing have demonstrated that the nronosed device is substantially equivalent to the predicate devices.
Based on the indications for use, technological characteristics, and comparison to predicate devices, the proposed Small Size DePuy Mitek Milagro® Interference Screws have been shown to be substantially equivalent to predicate devices under the Federal Food, Drug and Cosmetic Act.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in a bold, sans-serif font. The text is black and the background is white. The image is a close-up of the text.
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all caps and appears to be in a sans-serif font.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-002
DePuv Mitek, A Johnson & Johnson Company % Ms. Kristine Christo Regulatory Affairs Manager 325 Paramount Drive Raynham, Massachusetts 02767
Letter dated: December 13, 2012
Re: K122869
Trade/Device Name: 5, 6x12mm Milagro Interference Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MAI, HWC Dated: September 18, 2012 Received: September 19, 2012
Dear Ms. Christo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
#### Page 2 - Ms. Kristine Christo
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the DePuy Synthes logo. The logo includes the company name "DePuy" in a stylized font, along with the tagline "never stop moving". The logo also includes a small graphic element, possibly a stylized representation of a joint or bone structure, which is a common theme for medical device companies specializing in orthopedics.
#### ATTACHMENT 2
# INDICATIONS FOR USE
510(k) Number (if known): K122869
Device Names: 5, 6x12mm Milagro® Interference Screws
Indications for Use:
The DePuy Mitek MILAGRO BR Interference Screws are designed to attach soft tissues to bone in orthopedic surgical procedures for following indications:
Shoulder: Proximal Biceps Tenodesis, Acromio-Clavicular Repair Elbow: Distal Biceps Tenodesis, Ulnar Collateral Ligament repair
Knee: Collateral Ligament Repair
Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of _ |
# Casey Hanley
For Division of Orthopaedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.