K122783 · Aesculap Implant System, Inc. · OQG · Oct 30, 2013 · Orthopedic
Device Facts
Record ID
K122783
Device Name
PLASMAFIT ACETABULAR CUP AND VITELENE INSERT
Applicant
Aesculap Implant System, Inc.
Product Code
OQG · Orthopedic
Decision Date
Oct 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Plasmafit Pro Acetabular Cup System and Vitelene Insert are intended for use with existing Aesculap femoral heads and femoral stems to replace the hip joint. The devices are intended for: Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral ephiphysis Patients suffering from disability due to previous fusion Patients with acute femoral neck fractures Plasmafit Pro Acetabular Cup System and Vitelene Insert are intended for cementless applications.
Device Story
Modular hip prosthesis system for total hip arthroplasty; consists of cementless titanium acetabular cup with plasma-sprayed outer surface and vitamin E-stabilized highly crosslinked UHMWPE (Vitelene) insert. Used with compatible Aesculap femoral heads (Biolox delta, ISODUR) and femoral stems. Optional titanium alloy screws provide primary stability. Implanted by orthopedic surgeons in clinical/hospital settings. Device replaces damaged hip joint surfaces; vitamin E insert provides oxidation resistance. Clinical benefit includes restoration of joint function and pain relief.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including wear testing per ISO 14242, oxidation index characterization per ASTM F2003/ISO 5834-4, and testing of metallic plasma-sprayed coatings and modular locking mechanisms per FDA guidance.
Technological Characteristics
Cementless titanium acetabular cup with titanium plasma spray coating; Vitelene (vitamin E-stabilized UHMWPE) insert. Modular design with optional titanium alloy anchoring screws. Compatible with 40mm Biolox delta and ISODUR (CoCrMo) femoral heads. Materials tested per ASTM F2003 and ISO 5834-4. No software or electronic components.
Indications for Use
Indicated for patients requiring hip joint replacement due to severe pain/disability from rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, nonunion of femoral fractures, congenital hip dysplasia, protrusion acetabuli, slipped capital femoral epiphysis, previous fusion, or acute femoral neck fractures.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
510(k) Premarket Notification
Plasmafit Pro Acetabular Cup and Vitelene Insert
K122783
Page 1 of 3
# 510(k) SUMMARY (as required by 21 CFR 807.92)
# Plasmafit Pro Acetabular Cup System and Vitelene Insert
October 30, 2013
#### COMPANY:
Aesculap Implant Systems, LLC 3773 Corporate Parkway Center Valley, PA 18034
Establishment Registration Number: 3005673311
OCT 3 0 2013
#### CONTACT:
Julie Tom Wing 610-984-9147 (phone) 610-791-6882 (fax) julie.tomwing@aesculap.com
### TRADE NAME:
Plasmafit Pro Acetabular Cup System and Vitelene Insert
#### COMMON NAME:
Acetabular Cup and Insert
# CLASSIFICATION NAME(s):
Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis;
Hip joint metal/polymer semi-constrained cemented prosthesis;
Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis;
Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis
REGULATION NUMBER: 888.3358; 888.3350; 888.3360; 888.3353
| CLASSIFICATION | |
|------------------|-----|
| PRODUCT CODE(S): | OQG |
| SUBSEQUENT | |
|------------------|------------------------------|
| PRODUCT CODE(S): | LPH; JDI; OQH; LWJ; LZO; OQI |
{1}------------------------------------------------
#### K122783
#### Page 2 of 3
# DEVICE DESCRIPTION
Aesculap Plasmafit Pro Acetabular Cup System and Vitelene Insert is a modular hip prostheses system intended to be used in conjunction with a prosthetic femoral head and hip stem in total hip arthroplasty.
Aesculap Plasmafit Pro consists of a cementless titanium acetabular cup, a highly crosslinked (B-75 kGv) ultra-high molecular weight polyethylene (UHMW-PE) vitamin E insert, 40mm Biolox® delta, 40mm ISODUR® heads and optional titanium screws.
The hemispherical titanium cup with a slightly flattened dome is coated with a titanium plasma spray on the outer surface which aims to provide a high level of primary stability with its rough finish. Plasmafit Pro cups are available in three (3) similar designs only varying in the number of holes (0, 3, 5/7) around the apex. The number of holes are adapted for optional anchoring 6.5mm diameter titanium alloy screws that may be used to ensure primary stability. Plasmafit Pro anchoring screws are an option available separately in sizes 16mm to 68mm lengths.
Vitelene is designed with a taper-fit to the interior surface of Plasmfit Pro cups. The vitamin E insert has been shown to be more resistant to oxidation than conventional UHMWPE. Characterization of the oxidation indices of aged (ASTM F2003) and un-aged Vitelene measured according to ISO 5834-4 did not decrease . Further, Vitelene remains resistant to oxidation after 5 million cycles of wear as demonstrated in testing performed in accordance to ISO 14242.
Plasmafit Pro cups are available in sizes ranging from 40mm - 70mm.
Plasmafit Pro Acetabular Cup System and Vitelene insert may be used with Aesculap Biolox® delta and ISODUR® (CoCrMo) 40mm ball heads along with the following cleared compatible components: Aesculap Hip Implant Systems Excia, Metha, Prevision, and Unisyn and cleared Aesculap femoral heads: Biolox® Option, Biolox® Forte, Biolox® Delta, and CoCrMo to offer several interoperative implant options taking into consideration different patient bone conditions.
#### PURPOSE FOR PREMARKET NOTIFICATION
The purpose for this submission is to gain marketing clearance for Aesculap Implant Systems Plasmafit Pro Acetabular Cup. Vitelene insert, optional titanium screws and 40mm Biolox® and 40mm ISODUR® femoral heads.
{2}------------------------------------------------
Plasmafit Pro Acotabular Cup and Vitelene Insert
#### K122783
# Page 3 of 3
# INDICATIONS FOR USE
The Plasmafit Pro Acetabular Cup System and Vitelene Insert are intended for use with existing Aesculap femoral heads and femoral stems to replace the hip joint.
The devices are intended for:
- Patients suffering from severe hip pain and disability due to rheumatoid arthritis. . osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur
- Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral . ephiphysis
- . Patients suffering from disability due to previous fusion
- Patients with acute femoral neck fractures .
Plasmafit Pro Acetabular Cup System and Vitelene Insert are intended for cementless applications.
# SUBSTANTIAL EQUIVALENCE
Aesculap Implant Systems, LLC believes that the Plasmafit Pro Acetabular Cup System and Vitelene insert is substantially equivalent to:
- . Excia Total Hip System (Aesculap - K042344)
- Excia Total Hip System (Aesculap K061699) .
- EXp Acetabular Shell Insert (StelKast, Inc. K094035) .
- Mectacer 40mm Biolox Delta Heads (K11215) .
- Mpact 40mm Ball Heads (K103721) t
# TECHNOLOGICAL CHARACTERISTICS (compared to Predicate(s)}
Aesculap Implant Systems Plasmafit Pro Acetabular Cup and Vitelene Insert is an acetabular cup system that is substantially equivalent to the predicate devices previously cleared by FDA. The subject device is shown to be substantially equivalent and has the same technological characteristics to predicate devices through comparison of indications for use, design, material composition, function and range of sizes.
# PERFORMANCE DATA
As recommended by the FDA Guidance for Industry and FDA Staff - Non-clinical testing was performed to demonstrate that Plasmafit Pro Acetabular Cup and Vitelene Insert System is substantially equivalent to other predicate devices. In addition, testing was performed where applicable per FDA Guidance documents:
- "Testing of Metallic Plasma Spraved Coating on Orthopedic Implant to Support . Reconsideration of Postmarket Surveillance";
- "Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone . Cement"
- "Testing Non-Articulation, "Mechanically Locked," Modular Implant Components ●
Testing demonstrated that the Plasmafit Pro Acetabular Cup System and Vitelene Insert is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the logo is a stylized image of an eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 30, 2013
Aesculap Implant Systems, Incorporated Ms. Julie Tom Wing Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K122783
Trade/Device Name: Plasmafit Pro Acetabular Cup System and Vitelene Insert Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: OQG, LPH, JDI, OQH, LWJ, LZO, OQI Date: October 24, 2013 Receiver: October 28, 2013
Dear Ms. Julie Tom Wing:
.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
### Page 2 - Ms. Julie Tom Wing
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Erin Keith
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Page 1 of 1
#### A. INDICATIONS FOR USE STATEMENT
K122783 510(k) Number:
# Device Name: Plasmafit Pro Acetabular Cup System and Vitelene Insert
#### Indications for Use:
The Plasmafit Pro Acetabular Cup System and Vitelene Insert are intended for use with existing Aesculap femoral heads and femoral stems to replace the hip joint.
The devices are intended for:
- Patients suffering from severe hip pain and disability due to rheumatoid arthritis, . osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur
- Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital . femoral ephiphysis
- Patients suffering from disability due to previous fusion .
- Patients with acute femoral neck fractures ♥
Plasmafit Pro Acetabular Cup System and Vitelene Insert are intended for cementless applications.
Prescription Use X and/or Over-the-Counter Use (per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth L. Frank -S
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.