SAKURA STEAM STERILIZAER MODEL ASSV-AH06; SAKURA STEAM STERILIZER MODEL ASSV-AH06E; SKYTRON INTEGRITY 175; SKYTRON INTEG
K122774 · Sakura Seiki Co., Ltd. · FLE · May 9, 2013 · General Hospital
Device Facts
Record ID
K122774
Device Name
SAKURA STEAM STERILIZAER MODEL ASSV-AH06; SAKURA STEAM STERILIZER MODEL ASSV-AH06E; SKYTRON INTEGRITY 175; SKYTRON INTEG
Applicant
Sakura Seiki Co., Ltd.
Product Code
FLE · General Hospital
Decision Date
May 9, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers are designed and optimized for use in health care facilities. Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers incorporate high-pressure steam to sterilize non-porous and porous, heat and moisture-stable items and materials used in the health care facilities. Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers are available in the following configuration. 17.5" x 17.5" x 26.5" chamber size, Single Door Prevacuum and Gravity. Note: Sakura Steam Sterilizer ASSV-AH06E / SKYTRON Integrity 175 SG Steam Sterilizer equip the integrated steam generator.
Device Story
Steam sterilizer for healthcare facilities; utilizes high-pressure steam to process non-porous and porous, heat/moisture-stable items. Input: user-selected sterilization cycle (e.g., Prevacuum, Gravity, Flash/IUSS). Operation: microprocessor-controlled cycles; integrated steam generator in 'E' models; chamber size 17.5" x 17.5" x 26.5". Output: sterilized instruments/materials; cycle status/alerts via controller. Used by healthcare staff to ensure instrument sterility; prevents cross-contamination. Benefits: provides validated sterilization assurance for surgical/medical instruments.
Clinical Evidence
Bench testing only. Validation performed per ANSI/AAMI ST8:2008 standards. Efficacy demonstrated by biological indicator kill and sterility assurance level (SAL) of at least 10^-6 reduction. Software validated per FDA moderate level of concern guidance.
Technological Characteristics
Steam sterilizer; 17.5" x 17.5" x 26.5" chamber; Prevacuum and Gravity cycles; microprocessor-controlled; integrated steam generator (select models). Complies with IEC 60601-1-2, IEC 61010-1, IEC 61010-2-040, and ASME Boiler and Pressure Vessel Code Section VIII Division 1.
Indications for Use
Indicated for sterilization of non-porous and porous, heat and moisture-stable items and materials in healthcare facilities. Includes immediate use steam sterilization (IUSS) cycles for unwrapped instruments.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
# Document 004
# MAY 0 9 2013
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K122774
- 1. Submitter's Identification: Sakura Seiki Co., Ltd. 46-1, Shinden Chikuma, Nagano 387-0018 Japan Phone: 81-26-273-3583 Fax: 81-26-273-2584
Date Summary Prepared: December 5, 2012
#### 2. Name of the Device:
Proprietary name: Sakura Steam Sterilizer ASSV-AH06 SKYTRON Integrity 175 Steam Sterilizer Sakura Steam Sterilizer ASSV-AH06E SKYTRON Integrity 175 SG Steam Sterilizer
Common Name: Steam Sterilizer Regulation: 21CFR Part 880:6880, Class II Product Code: FLE
#### 3. Common or Usual Name:
Common name: Steam Sterilizer Classification name: Sterilizer, Steam
#### 4. Predicate Device Information:
Skytron Integrity 215 Steam Sterilizer, Skytron Integrity 215 SG Steam Sterilizer, K120149
#### 5. Description:
Sakura Steam Sterilizer ASSV-AH06, SKYTRON Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E and SKYTRON Integrity 175 SG Steam
{1}------------------------------------------------
Sterilizers are Class II, Product Code FLE Medical Devices as defined by CFR§880.6880 and defined for use in healthcare facilities. Sakura Steam Sterilizer ASSV-AH06, SKYTRON Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E and SKYTRON Integrity 175 SG Steam Sterilizers provide efficient steam sterilization of non-porous and porous, heat and moisture stable materials. The sterilizers also dry items that have been sterilized with wet steam.
### છં Intended Use:
Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer. Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers are designed and optimized for use in health care facilities. Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers incorporate high-pressure steam to sterilize non-porous and porous, heat and moisture-stable items and materials used in the health care facilities. Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers are available in the following configuration.
17.5" x 17.5" x 26.5" chamber size, Single Door Prevacuum and Gravity.
# Note: Sakura Steam Sterilizer ASSV-AH06E / SKYTRON Integrity 175 SG Steam Sterilizer equip the integrated steam generator.
Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers are equipped with the following factory-programmed sterilization cycle values. The following factory-programmed cycles are validated according to ANSI/AAMI ST8: 2008.
| Cycle | Maximum Validated<br>Loads | Sterilize<br>Temp | Sterilize<br>Time | Dry Time |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|-------------------|------------|
| Flash 1<br>(IU 1) | One unwrapped,<br>nonporous instrument no<br>heavier than 0.22 lbs. (100<br>gram) | 270 F (132 C) | 3 minutes | 1 minute |
| Flash 2<br>(IU 2) | Maximum of two (2)<br>unwrapped instrument<br>trays, maximum weight 25<br>pound (11.3 kg) each.<br>The maximum of three (3)<br>2 mm inside diameter<br>(minimum) and 400 mm | 270 F (132 C) | 4 minutes | 1 minute |
| | long (maximum) lumen<br>instruments per cycle load. | | | |
| Prevacuum<br>270 | Maximum of two (2)<br>double-wrapped<br>instrument trays,<br>maximum weight 25<br>pound (11.3kg) each. The<br>maximum of three (3) 2<br>mm inside diameter<br>(minimum) and 400 mm<br>long (maximum) lumen<br>instruments per cycle load. | 270 F (132 C) | 4 minutes | 30 minutes |
| Prevacuum<br>275 | Maximum of two (2)<br>double-wrapped<br>instrument trays,<br>maximum weight 25<br>pound (11.3 kg) each.<br>The maximum of three (3)<br>2 mm inside diameter<br> | 275 F (135 C) | 3 minutes | 30 minutes |
| Express<br>Prevacuum<br>(IU 3) | Maximum of two (2) single<br>wrapped instrument trays,<br>maximum weight 25<br>pound (11.3 kg) each.<br>The maximum of three (3)<br>2 mm inside diameter<br>(minimum) and 400 mm<br>long (maximum) lumen<br>instruments per cycle load.<br>This cycle is intended for<br>immediate use and the<br>wrap used for this cycle is<br>not intended to be a sterile<br>barrier. | 270 F (132 C) | 4 minutes | 3 minutes |
| Gravity 1 | Maximum of four (4) -<br>fabric packs. | 250 F (121 C) | 30 minutes | 15 minutes |
| Gravity 2 | Maximum of two (2)<br>double-wrapped<br>instrument trays.<br>maximum weight 25<br>pound (11.3 kg) each. | 270 F (132 C) | 15 minutes | 30 minutes |
| Liquid | Maximum of three (3) | 250 F (121 C) | 45 minutes | 0 minutes |
| | with vented closures. This<br>cycle is intended for non-patient contacting devices<br>only. | | | |
| Leak Test | Empty Chamber, test time of 15 minutes. | N/A | N/A | N/A |
| Warm Up | Empty Chamber | 270 F (132C) | 4 minutes | 3 minutes |
| Bowie-Dick<br>Test | Bowie-Dick Test Pack | 273 F (134 C) | 3.5 minutes | 2 minutes |
{2}------------------------------------------------
# Document 004
.
:
: 上一篇:
.
-
.
:
-
·
.
・
.
{3}------------------------------------------------
Note:
> "IU" means "Immediate Use Steam Sterilization" cvcle.
> Liquid cycle is for non-patient contact use only.
### Technological Characteristics 7.
# Comparison to Predicate Device
The proposed sterilizers offered within this submission are equipped with the same options offered under the predicate device, in design and construction. The only difference between the predicate and the subject device is the size of the models being offered and less number of sterilization cycles. The subject device has different chamber dimensions and different load configulations and less sterilization cycles compared to the predicate device. The predicate devices are larger in size to fit more instruments in one sterilization cycle. This difference has no effect the safety and effectiveness of the device.
### EFFECTIVENESS:
Efficacy of sterilizer function and exposure time recommendations are ultimately shown by complete kill of biological indicators and verifying an appropriate safety factor or sterility assurance level (SAL) of at least 10° reduction. Sakura Seikl Co., Ltd. validates its sterilization cycles by recommended practices, standards and guidelines developed by various independent organizations such as the Association for Advancement of Medical Instrumentation (AAMI). Prior to release, the sterilizers were validated to meet the requirements of AAMIIANSI ST8: 2008.
The software validation for the cycle operation was performed according to FDA's moderate level of concern recommendations provided in the document "Guidance for the Content for Premarket Submissions for Software Contained in Medical Devices (May 2005).
### SAFETY:
Sakura Steam Sterilizer ASSV-AH06, SKYTRON Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E and SKYTRON Integrity 175 SG Steam
{4}------------------------------------------------
Sterilizers have been designed, constructed and tested to meet the safety and performance requirements of various national safety codes and standards. Sakura Steam Sterilizer ASSV-AH06, SKYTRON Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E and SKYTRON Integrity 175 SG Steam Sterilizers comply with the following safety standards:
IEC 60601-1-2:2007 IEC61010-1 (200 Edition) 2008 / CSA C22.2 No.61010-1 IEC61010-2-040: 2005 CISPR11: 2009 ED5.0: AMENDMENT1: 2010 IEC61000-3-2:2005 ED3.0: AMENDMENT1:2010 IEC61000-3-3:2008 ED2.0 (EMC) - PART 3-3 IEC61000-4-2: 2008 ED2.0 (EMC) - PART 4-2 IEC61000-4-3: 2006 ED3.0 +AMENDMENT1: 2007+AMENDMENT2:2010 (EMC) IEC61000-4-4: 2004 ED2.0 +AMENDMENT1:2010 IEC61000-4-5: 2005 ED2.0 IEC61000-4-6: 2003 ED2.0+AMENDMENT1:2004+AMENDMENT1:2006 IEC61000-4-8: 2009 ED2.0 IEC61000-4-11: 2004 ED2.0 American Society of Mechanical Engineers (ASME) Boiler and Pressure Vessel Code Section VIII Division 1 2010 Edition: Rules for the Construction of Pressure Vessels.
# HAZARDS-FAILURE OF PERFORMANCES
Failure of the sterilization process can lead to incidence of cross contamination, the transmission of potentially infectious organisms from one infected person to another who was not otherwise infected prior to the incident.
To avoid failure, the user must ensure the materials, instruments and devices to be sterilized are thoroughly cleaned, that the manufacturer's instructions for use are followed, that the cvcle to be used for each type of sterilizer load has been validated, that the sterilizer has been maintained in accordance with the sterilizer's manufacturer's maintenance schedule and is operating properly, and that each sterilizer load is monitored with available and validated biological and chemical sterilization process indicators.
Today, there are many steam sterilizers in daily use in hospitals throughout the United States. The incident of sterilizer malfunction or sterilization process failure is relatively rare considering the wide spread use of steam sterilizers. Further, there are no reports in the literature of patient infection that have resulted from steam sterilizer failure.
The technology designed in Sakura Steam Sterilizer ASSV-AH06, SKYTRON Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E and SKYTRON Integrity 175 SG Steam Sterilizers provide microprocessor controller safequards that abort the cycle and give appropriate signals, alerts and warnings when required conditions have not been met or when a malfunction occurs.
{5}------------------------------------------------
# USER INFORMATION
Sakura Seiki Co., Ltd. provides information to the user that is intended to insure safe and effective use of steam sterilization in its detailed User's Manual and other labeling.
Sakura Seiki Co., Ltd. also recommends the use and periodic review of the AAMI steam sterilization standards to ensure further assurance of the safe and effective use of steam sterilization equipment in health care facilities.
### 8. Performance Data:
Non-clinical: Verification, validation and testing activities were conducted to establish the performance, functionality and reliability characteristics of Sakura Steam Sterilizer ASSV-AH06, SKYTRON Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E and SKYTRON Integrity 175 SG Steam Sterilizer. The device passed all of the tests based on predetermined Pass/Fail criteria.
#### 9. Conclusions:
The data from consensus standard testing and comparison to the predicate device show that Sakura Steam Sterilizer ASSV-AH06, SKYTRON Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E and SKYTRON Integrity 175 SG Steam Sterilizers are as safe and effective as the legally marketed predicate device.
Therefore Sakura Seiki Co., Ltd. conclude that the each of the sterilizers included under this submission are substantially equivalent to the predicate device.
{6}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with three wing-like extensions.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 09, 2013
Sakura Seiki Company, Limited C/O Ms. Swapna Chirala Responsible Third Party Official TÜV SÜD America, Incorporated 1775 Old Hwy 8 NW, Suite 104 NEW BRIGHTON, MN 55112-1891
Re: K122774
Trade/Device Name: Sakura Sterilizer ASSV-AH06, Skytron Integrity 175 Steam Sterilizer, Sakura Sterilizer ASSV-AH06E Skytron Integrity 175 SG Steam Sterilizer
Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: April 4, 2013 Received: April 26, 2013
### Dear Ms. Chirala:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 – Ms. Chirala
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
if you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Ruth
urot.u.s.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use Statement
**Page** 1 of 3
510(k) Number (if known): 长122774
# Device Name:
Sakura Steam Sterilizer ASSV-AH06 Skytron Integrity 175 Steam Sterilizer Sakura Steam Sterilizer ASSV-AH06E Skytron Integrity 175 SG Steam Sterilizer
# Indications For Use:
Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers are designed and optimized for use in health care facilities. Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers incorporate high-pressure steam to sterilize non-porous and porous, heat and moisture-stable items and materials used in the health care facilities. Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers are available in the following configuration.
17.5" x 17.5" x 26.5" chamber size, Single Door Prevacuum and Gravity.
Note: Sakura Steam Sterilizer ASSV-AH06E / SKYTRON Integrity 175 SG Steam Sterilizers equip the integrated steam generator.
| Prescription Use | |
|----------------------------|---|
| (Per 21 CFR 801 Subpart D) | |
| OR | |
| Over-The Counter Use | X |
| (21 CFR 807 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------------------------------------------------------|------|
| ed by Elizabeth F. Claverie<br>Elizabeth= | |
| Claverie | |
| (Division Sign-Off)<br>ിvision of Anesthesiology, General Hospital<br>ifection Control, Dental Devices | |
| 510/k) Number: K \22774 | rage |
of 3
{9}------------------------------------------------
Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers are equipped with the following factory-programmed sterilization cycle values. The following factory-programmed cycles are validated according to ANSI/AAMI ST8: 2008.
| Cycle | Maximum Validated Loads | Sterilize Temp | Sterilize Time | Dry Time |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|----------------|------------|
| Flash 1<br>(IU 1) | One unwrapped, nonporous instrument no heavier than 0.22 lbs. (100 gram) | 270 F (132 C) | 3 minutes | 1 minute |
| Flash 2<br>(IU 2) | Maximum of two (2) unwrapped instrument trays, maximum weight 25 pound (11.3 kg) each. The maximum of three (3) 2 mm inside diameter (minimum) and 400 mm long (maximum) lumen instruments per cycle load. | 270 F (132 C) | 4 minutes | 1 minute |
| Prevacuum 270 | Maximum of two (2) double-wrapped instrument trays, maximum weight 25 pound (11.3kg) each. The maximum of three (3) 2 mm inside diameter (minimum) and 400 mm long (maximum) lumen instruments per cycle load. | 270 F (132 C) | 4 minutes | 30 minutes |
| Prevacuum 275 | Maximum of two (2) double-wrapped instrument trays, maximum weight 25 pound (11.3 kg) each. The maximum of three (3) 2 mm inside diameter (minimum) and 400 mm long (maximum) lumen instruments per cycle load. | 275 F (135 C) | 3 minutes | 30 minutes |
| Express Prevacuum<br>(IU 3) | Maximum of two (2) single wrapped instrument trays, maximum weight 25 pound (11.3 kg) each. The maximum of three (3) 2 mm inside diameter (minimum) and 400 mm long (maximum) lumen instruments per cycle load | 270 F (132 C) | 4 minutes | 3 minutes |
| | This cycle is intended for<br>immediate use and the<br>wrap used for this cycle is<br>not intended to be a sterile<br>barrier. | | | |
| Gravity 1 | Maximum of four (4) -<br>fabric packs. | 250 F (121 C) | 30 minutes | 15 minutes |
| Gravity 2 | Maximum of two (2)<br>double-wrapped<br>instrument trays,<br>maximum weight 25<br>pound (11.3 kg) each. | 270 F (132 C) | 15 minutes | 30 minutes |
| Liquid | Maximum of three (3)<br>1,000 mill-liter containers<br>with vented closures. This<br>cycle is intended for non-<br>patient contacting devices<br>only. | 250 F (121 C) | 45 minutes | 0 minutes |
| Leak Test | Empty Chamber, test time<br>of 15 minutes. | N/A | N/A | N/A |
| Warm Up | Empty Chamber | 270 F (132C) | 4 minutes | 3 minutes |
| Bowie-Dick<br>Test | Bowie-Dick Test Pack | 273 F (134 C) | 3.5 minutes | 2 minutes |
Page 2 of 3
{10}------------------------------------------------
Note:
、
・
. . . . . .
: ・
V V "IU" means "Immediate Use Steam Sterilization" cycle.
Liquid cycle is for non-patient contact use only.
.
. .
. . . . .
1 4. "
.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.