K122768 · Dornier Medtech America, Inc. · JAA · Jan 18, 2013 · Radiology
Device Facts
Record ID
K122768
Device Name
GENESIS
Applicant
Dornier Medtech America, Inc.
Product Code
JAA · Radiology
Decision Date
Jan 18, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Dornier Genesis is an image intensified, fluoroscopic x-ray system that is intended for use in a wide field of applications, including all general examinations in urology and gynecology, as well as endoscopic and contrast examinations, imaging with radiography and/or fluoroscopy on patients in either the horizontal or vertical position.
Device Story
Dornier GENESIS is an image-intensified fluoroscopic X-ray system; utilizes solid-state flat panel digital imager. Components include X-ray generator, tube housing, flat panel detector, monitors, and procedure table. System captures X-ray signals, processes them through internal electronics, and displays images for clinician review. Used in clinical settings (urology/gynecology departments) by trained medical professionals. Output assists clinicians in visualizing internal anatomy for diagnostic or procedural guidance. Benefits include improved diagnostic capability and procedural support for urological/gynecological interventions.
Clinical Evidence
Nonclinical and clinical testing completed per FDA guidance for solid-state X-ray imaging devices. Bench testing performed according to IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-7, 60601-2-28, and 60601-2-54 standards. Clinical concurrence study demonstrated performance at least equivalent to the predicate Urotract device.
Technological Characteristics
Image-intensified fluoroscopic X-ray system; solid-state flat panel digital imager. Includes X-ray generator, tube housing, and procedure table. Complies with IEC 60601 series standards for electrical safety, electromagnetic compatibility, radiation protection, and usability. Operates as a diagnostic imaging system.
Indications for Use
Indicated for patients requiring general urological and gynecological examinations, endoscopic procedures, and contrast studies, including radiography and fluoroscopy imaging in horizontal or vertical positions.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
# | | | | 22768
# JAN 1 8 2013
・・
### 510(k) SUMMARY
### Dornier MedTech America, Inc.'s GENESIS
| Submitter Name and<br>Address: | Dornier MedTech America, Inc.<br>1155 Roberts Blvd.<br>Kennesaw, GA 30144 |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | John Hoffer<br>Vice President, Quality/Regulatory/Clinical |
| Phone Number: | 770-514-6163 |
| Fax Number: | 770-514-6291 |
| Date Prepared: | December 18, 2012 |
| Device Trade Name(s): | GENESIS |
| Device Common Name /<br>Classification Name: | Image-intensified fluoroscopic x-ray system (Product Code: JAA) |
| Predicate Device(s): | Dornier Urotract (K955019)<br>Siemens Uroskop Omnia (K101491) |
| General Device<br>Description: | The Dornier GENESIS is an Image Intensified Fluoroscopic X-ray<br>System (i.e., a solid state x-ray imager (flat panel/digital imager)).<br>The GENESIS consists of the following components: an x-ray<br>generator and tube housing, flat panel detector, monitors and<br>procedure table. An X-ray cabinet contains system elements such<br>as the X-ray generator, power electronics and electronics for the<br>imaging chain. |
| Intended Use /<br>Indications for Use: | The Dornier Genesis is an image intensified, fluoroscopic x-ray<br>system that is intended for use in a wide field of applications,<br>including all general examinations in urology and gynecology, as<br>well as endoscopic and contrast examinations, imaging with<br>radiography and/or fluoroscopy on patients in either the horizontal or<br>vertical position. |
| Technological<br>Characteristics: | From a clinical perspective and comparing design specifications, the<br>Dornier GENESIS and the predicate devices are substantially<br>equivalent. Based on the technological characteristics and overall<br>performance of the devices, Dornier believes that no significant<br>differences exist between the proposed product and the predicate<br>devices. |
| Performance Data | The GENISIS was tested per the following recognized testing<br>standards:<br>IEC 60601-1: Medical electrical equipment. Part 1: General<br>requirements for safety (2005). IEC 60601-1-2: Medical electrical equipment. Part 1-2:<br>General requirements for safety; Electromagnetic<br>compatibility-requirements and tests (2007). IEC 60601-1-3: Medical electrical equipment. Part 1:<br>General requirements for safety; general requirements for<br>radiation protection in diagnostic X-ray equipment (2008). IEC 60601-1-6: Medical electrical equipment. Part 1-6:<br>General requirements for basic safety and essential<br>performance -Collateral standard: Usability (2008). IEC 60601-2-7: Medical electrical equipment. Part 2-7:<br>Particular requirements for the safety of high voltage<br>generators of diagnostic X-ray generators (1998). IEC 60601-2-28: Medical electrical equipment. Part 2:<br>Particular requirements for the safety of X-ray source<br>assemblies and X-ray tube assemblies for medical diagnosis<br>(2010). IEC 60601-2-54: Medical electrical equipment. Part 2-54:<br>Particular requirements for the basic safety and essential<br>performance of X-ray equipment for radiography and<br>radioscopy (2009). |
| | In addition, per FDA's Guidance Document entitled Guidance for the<br>Submission of 510(k)s for Solid State X-ray Imaging Devices<br>(August 6, 1999), the company completed all required nonclinical<br>and clinical testing for the subject device. In all instances, the<br>GENESIS functioned as intended and the results observed were as<br>expected. With regard to the completed concurrence study, the<br>results confirm that the GENESIS has at least equivalent<br>performance to the predicate Urotract. |
| Substantial Equivalence | Based on the technological characteristics and overall performance<br>of the devices, Dornier believes that the GENESIS and the predicate<br>devices selected are substantially equivalent and that the<br>differences between the devices are minor and do not raise new<br> |
| | In sum, the GENESIS is as safe and effective as the identified<br>predicate devices. The GENESIS has the same intended<br>uses/indications for use and similar technological characteristics,<br>and principles of operation as its predicate device. The minor<br>technological differences between the GENESIS and its predicate<br>devices raise no new issues of safety or effectiveness.<br>Performance data demonstrate that the GENESIS is as safe and<br>effective as the identified predicate devices. Thus, the GENESIS is<br>substantially equivalent. |
{1}------------------------------------------------
:
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is oriented towards the upper right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the left side of the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 18, 2013
Dornier MedTech America, Inc. % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 Thirteenth Street, NW WASHINGTON DC 20004
Re: K122768
Trade/Device Name: GENESIS Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Dated: January 2, 2013 Received: January 2, 2013
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Sean M. Boyd -S
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K122768
Device Name: Dornier GENESIS
Indications For Use:
The Dornier GENESIS is intended for use in a wide field of applications, including all general examinations in urology and gynecology, as well as endoscopic and contrast examinations, imaging with radiography and/or fluoroscopy on patients in either the horizontal or vertical position.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Sean M. Boyd -S
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) (Division Sigm Off)
Division of Radiologii
**Limits of In Vitro Diagnostics and Radiological Use**
51060 K122768
Page
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.