PHYSIOLOGICAL DATA PROCESSOR

K122763 · Zephyr Technology Corporation · DRG · Jun 20, 2013 · Cardiovascular

Device Facts

Record IDK122763
Device NamePHYSIOLOGICAL DATA PROCESSOR
ApplicantZephyr Technology Corporation
Product CodeDRG · Cardiovascular
Decision DateJun 20, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2910
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The Packet Decoder Process (PDP) is indicated for use as a prescription software device to receive MDDS digital data packets and decode, collate and deliver data to a Class I, MDDS, Enterprise Service Bus according to a proprietary scheme defined by Zephyr Technology.

Device Story

The Packet Decoder Process (PDP) is a software-based device that functions as a data interface. It receives proprietary digital data packets transmitted by the BioHarness 3.0 (K113045). The device operates within an MDDS (Medical Device Data System) computer environment. Using a proprietary decoding scheme (Zephyr's General Comms Link Specification), the PDP unpacks the complex data array, collates the information, and delivers the decoded data into registers on an Enterprise Service Bus for access by other MDDS devices. The device is intended for use in clinical or professional settings where MDDS infrastructure is present. By facilitating the translation of proprietary signals into accessible data registers, the PDP enables the integration of physiological data into broader clinical information systems, supporting healthcare providers in accessing and managing patient data.

Clinical Evidence

Bench testing only. The device underwent factory testing to verify signal accuracy, calibration maintenance, and functional performance of the decoding process. No clinical data was required or provided.

Technological Characteristics

Software-based device; operates on MDDS computer systems. Implements proprietary decoding scheme (Zephyr's General Comms Link Specification). Written in Java. Functions as a data interface between proprietary transmission systems and MDDS Enterprise Service Bus registers. Factory-calibrated.

Indications for Use

Indicated for prescription use to receive, decode, and collate MDDS digital data packets for delivery to a Class I MDDS Enterprise Service Bus.

Regulatory Classification

Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 5. 510(k) Summary # K122763 JUN 2 0 2013 (per 21 CFR 807.92) #### 19 June 2013 Consultant Sponsor . Mr. Richard Keen Mr. Sailor Mohler Compliance Consultants Quality Manager 1151 Hope Street Zephyr Technology Corporation Stamford, CT 06907-1659 1 Annapolis Street, Suite 200 203 329 2700 F 203 329 2345 Annapolis MD 21401 USA rkeen@fda-complianceconsultants.com Office: 443-569-3603 sailor.mohler@zephyranywhere.com | Proprietary Name: | Packet Decoder Process (PDP) | |-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | PDP | | Device Classification Name: | Transmitters and receivers, physiological signal, radiofrequency | | Classification Number: | 21 CFR 870.2190 | | Product Code | DRG | | Reviewing Group | Cardiovascular | | Device Classification | Class II | | Establishment registration<br>No. | # 233836 | | Predicate Device | PDP is substantially equivalent to three predicates determined<br>substantially Equivalent in K113045, BioHarness 3.<br>1) Firmware (9500.0085). The PDP is a mirror image of the<br>portion of the firmware (PEP) that implements the encoding of<br>digital data.<br>2) Test Application (9500.0091). The PDP performs the same<br>function as the Test Application to decode packet data and<br>places the decoded data in registers for MDDS access.<br>3) Config Tool (9500.0096). The PDP performs the same<br>features / functions as the Config Tool (9500.0096) to change<br>threshold alert settings in the hardware resisters in the | Trademark Notice: All Trademarks used other than those of Zephyr Technology Corporation are registered to their respective owners. #### Device Description The Packet Decoder Process (PDP) receives digital data that is transmitted by BioHarness 3.0, K 1 1 3045 in a proprietary structure. PDP receives this complex array of packet coded data. A decoding scheme is used to decode data and organize for presentation and access by Class I MDDS devices. The Packet Decoder Process is manufactured by Zephyr Technology Corporation. This software is loaded and executed in an MDDS operating systems in an MDDS computer. This complex process is defined by a proprietary scheme: Zephyr's General Comms Link Specification. ### INDICATIONS FOR USE The Packet Decoder Process (PDP) is indicated for use as a prescription software device to receive MDDS digital data packets and decode, collate and deliver data to a Class I, MDDS, Enterprise Service Bus according to a proprietary scheme defined by Zephyr Technology. | Company of Children<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>and the more of | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>And Concession of Children Company of<br>with and the country of the program of<br>00000<br>A 64 San Was and | And Children and Children Come of Children Come of<br>----------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>-------<br>No. of Cattle.<br>DAD<br>A And I | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Carol of Catalog<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br> | .<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>AND THE LEASE<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>And Call Property of Children Company of Children<br>A STATE A STATE A BEAT A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH A BEATH | {1}------------------------------------------------ 5. 510(k) Summary (per 21 CFR 807.92) ### INTENDED USE The BioHarness 3-M1 Packet Decoder Process (PDP) is intended for decoding digital data packets that were coded in the BioHarness 3.0 Packet Encoder Processor (PEP). The decoding process receives data from an MDDS data transmission system then decodes, collates and drives MDDS registers on the Enterprise Service Bus. This process is defined by the proprietary scheme: Zephyr's General Comms Link Specification, version 1.44 dated 11 June 2009 Part Number #9700.0028 (and subsequent revisions). The scientific concept on which this device is based is the principle that complexity and precise structure of packet data give rise to a repeatable mechanism to "pack up" digital data and transmit it using a qualified communication system. At the receive side, a scheme decodes or "un-pack" this data and delivers it to MDDS registers. This device functions by executing programmed commands written in the Java language. The calibration is established by the factory and yields accurate and calibrated signals that can maintain calibration over its useful life. #### Substantial Equivalence Zephyr Technology Corporation has determined that the Packet Decoder Process (PDP) is substantially equivalent to the performance to the predicate device. The differences between these systems are incidental and not significant. Both this device and the predicates use a similar technological characteristics and principles. #### Safety and Effectiveness This device is safe and effective for the application for which it is intended and has been tested to confirm safety and efficacy. A series of factory tests are conducted to verify the intended signals are accurate and can maintain calibration over its useful life. The Packet Decoder Process (PDP) has benefited from design, development, testing and production procedures that conform to Quality Systems. Zephyr Technology Corporation continues to search all appropriate sources for information relating to safety and effectiveness and maintains an in-house reporting device to identify adverse safety and effectiveness information and as such, applicable data is recorded for this product. # K122763 {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three distinct head-like shapes, suggesting multiple individuals or a sense of community. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the logo, with "DEPARTMENT OF HEALTH & HUMAN SERVICES" curving along the left side and "USA" positioned at the top right. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 20, 2013 Zephyr Technology Corp. c/o Richard Keen 1151 Hope St Stamford, CT 06907 Re: K122763 Trade/Device Name: Packet Decoder Process (PDP) Regulation Number: 21 CFR 870.2190 Regulation Name: Transmitters and receivers, physiological signal, radiofrequency Regulatory Class: Class II Product Code: DRG Dated: February 18, 2013 Received: April 25, 2013 Dear Mr. Keen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). -You-may, therefore, market-the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Richard Keen forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Owen P. Faris -S for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and. Radiological Health Enclosure {4}------------------------------------------------ # 4. Indications for Use Statement 510(K) Number assigned K 122763 Name: Packet Decoder Process (PDP) ### INDICATIONS FOR USE The Packet Decoder Process (PDP) is indicated for use as a prescription software device to receive MDDS digital data packets and decode, collate and deliver data to a Class I, MDDS, Enterprise Service Bus according to a proprietary scheme defined by Zephyr Technology. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | XXX | |--------------------------|--------------------------| | | or | | Over - The - Counter Use | (Per 21 CFR 801.109) | | | (Optional Format 1-2-96) | ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Owen P. Faris -S 2013.06.20 16:07:44 1 -04'00' Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...