FASTECT II PPX DRUG SCREEN DIPSTICK, FASTECT II DRUG SCREEN DIPSTICK, QUICKTOX DRUG SCREEN DIPCARD
K122752 · Branan Medical Corp. · JXN · Jan 30, 2013 · Clinical Toxicology
Device Facts
Record ID
K122752
Device Name
FASTECT II PPX DRUG SCREEN DIPSTICK, FASTECT II DRUG SCREEN DIPSTICK, QUICKTOX DRUG SCREEN DIPCARD
Applicant
Branan Medical Corp.
Product Code
JXN · Clinical Toxicology
Decision Date
Jan 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3700
Device Class
Class 2
Indications for Use
The Fastect® II PPX Drug Screen Dipstick is lateral flow immunoassay in vitro diagnostic screen test for qualitative detection of propoxyphene at or above 300ng/ml in human urine. The Fastect® II Drug Screen Dipstick and QuickTox® Drug Screen Dipcard are lateral flow immunoassay in vitro diagnostic screen tests for the qualitative detection of cocaine (at or above 300ng/ml), opiates (at or above 300ng/ml), methamphetamine (at or above 500ng/ml), THC (at or above 50ng/ml), amphetamine (at or above 1000ng/ml), phencyclidine (at or above 25ng/ml), benzodiazepines (at or above 300ng/ml), barbiturates (at or above 300ng/ml), methadone (at or above 300ng/ml), tricyclic antidepressants (at or above 1000ng/ml), MDMA (at or above 500ng/ml), oxycodone (at or above 100ng/ml), buprenorphine (at or above 10ng/ml), or propoxyphene (at or above 300ng/ml) in human urine. It is for prescription point-of-care use and not intended for over-the-counter sale to non-professionals.
Device Story
Lateral flow immunoassay devices (dipsticks/dipcards) for qualitative drug screening in human urine. Principle: competitive immunoassay; immobilized drug conjugate competes with drug in urine for limited antibody binding sites. User dips device into urine sample; visual qualitative results appear on membrane strip. Includes internal procedural control band to validate test completion. Used in point-of-care settings by healthcare professionals. Provides preliminary results; positive findings require confirmatory testing via GC/MS or LC/MS. Assists clinicians in rapid drug screening; supports clinical decision-making when combined with professional judgment.
Clinical Evidence
Bench testing only. Precision studies conducted at three point-of-care sites using spiked urine samples at various concentrations relative to cutoff. Method comparison study evaluated 84 urine specimens against GC/MS reference method. No clinical studies performed.
Technological Characteristics
Lateral flow immunochromatographic assay. Uses mouse monoclonal antibodies against propoxyphene. Qualitative visual readout. No instrumentation required. Shelf life 24 months at 15-30°C. Analyte cutoff 300 ng/mL.
Indications for Use
Indicated for rapid, qualitative, in vitro diagnostic screening of propoxyphene and other specified drugs/metabolites in human urine at defined cutoff concentrations. Intended for prescription point-of-care use by professionals. Preliminary results require confirmation by GC/MS or LC/MS.
Regulatory Classification
Identification
A propoxyphene test system is a device intended to measure propoxyphene, a pain-relieving drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of propoxyphene use or overdose or in monitoring levels of propoxyphene to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). A propoxyphene test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
ACON MULTI-CLIN Drug Screen Test Device (k041685)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k122752
B. Purpose for Submission:
New device and addition of new analyte
C. Measurand:
Propoxyphene (PPX)
D. Type of Test:
Qualitative, immunochromatographic
E. Applicant:
Branan Medical Corporation
F. Proprietary and Established Names:
QuickTox® Drug Screen Dipcard
Fastect® II PPX Drug Screen Dipstick
Fastect® II Drug Screen Dipstick
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JXN | II | 862.3700 – Propoxyphene test system | 91-Toxicology |
H. Intended Use:
1. Intended use(s):
See indications for use below
2. Indication(s) for use:
The Fastect® II PPX Drug Screen Dipstick Test is a lateral flow immunoassay for the rapid detection of propoxyphene in human urine at or above 300 ng/mL.
The Fastect® II Drug Screen Dipstick and QuickTox® Drug Screen Dipcard are
{1}
lateral flow immunoassay for the rapid detection of multiple drugs and drug metabolites in human urine at or above the following cutoff concentration:
| THC | 11-nor-Δ9-Tetrahydrocannabinol-9-carboxylic acid | 50 ng/ml |
| --- | --- | --- |
| COC | Benzoylecgonine | 300 ng/ml |
| OPI | Morphine | 300 ng/ml |
| MET | Methamphetamine | 500 ng/ml |
| AMP | Amphetamine | 1000 ng/ml |
| PCP | Phencyclidine | 25 ng/ml |
| BZO | Oxazepam | 300 ng/ml |
| BAR | Secobarbital | 300 ng/ml |
| MTD | Methadone | 300 ng/ml |
| TCA | Nortriptyline | 1000 ng/ml |
| MDMA | 3,4-methylenedioxymethamphetamine | 500 ng/ml |
| OXY | Oxycodone | 100 ng/ml |
| BUP | Buprenorphine | 10 ng/ml |
| PPX | Propoxyphene | 300 ng/ml |
These tests provide visual qualitative results and are intended for in vitro diagnostic use only. It is for prescription point-of-care use only and not intended for over-the-counter sale to non-professionals.
These tests provide only a preliminary test result. For a quantitative result or to confirm preliminary positive results obtained by the QuickTox® Drug Screen Dipcard, Fastect® II Drug Screen Dipstick or Fastect® II PPX Drug Screen Dipstick tests, a more specific alternative method such as Gas Chromatography/Mass Spectrometry (GC/MS) must be used. Clinical Consideration and professional judgment should be applied to any drug of abuse test results, particularly when a preliminary positive result is indicated.
3. Special conditions for use statement(s):
These tests are for prescription point-of-care use only and not intended for over-the-counter use.
4. Special instrument requirements:
Not applicable
I. Device Description:
The Fastect® II Drug Screen Dipstick with Propoxyphene and QuickTox® Drug Screen Dipcard with Propoxyphene contain multiple drugs and drug metabolites in addition to Propoxyphene. Propoxyphene is added as a new analyte. The Fastect® II PPX Drug Screen Dipstick only contains the propoxyphene analyte. All dipstick and dipcard devices are based on the principle of highly specific immunochemical reactions between antigens and antibodies and all devices utilize a competitive immunoassay procedure in which an immobilized drug conjugate competes with the
{2}
drug present in urine for limited antibody binding sites.
The Fastect® II PPX Drug Screen Dipstick, Fastect® II Drug Screen Dipstick and QuickTox® Drug Screen Dipcard devices are standardized to detect Propoxyphene in human urine at a cutoff concentration of 300 ng/ml. These tests can be performed without the use of any additional instruments.
A control band with a different antigen/antibody reaction is added to the immunochromatographic membrane strip and should always appear regardless of the presence of drug or metabolite. The appearance of the control band during testing indicates that the test has completed and the test is valid.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
ACON MULTI-CLIN Drug Screen Test Device
2. Predicate K number(s):
k041685
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Feature | Candidate Devices
(QuickTox® Drug Screen Dipcard, Fastect® II PPX Drug Screen Dipstick and Fastect® II Drug Screen Dipstick) | Predicate Device
(ACON® multi-CLIN™ Drug Screen Test Device k041685) |
| Intended Use | Screening Device | Same |
| Matrix | Human Urine | Same |
| Test Principle | Competitive immunoassay | Same |
| Analytes | Propoxyphene | Same |
| Cut-Off | 300 ng/ml | Same |
| Sample Volume | 10 ml | Same |
| Target User Population | For professional point of care | Same |
| Shelf Life | 24 Months | Same |
| Testing Method | Lateral Flow Immunoassay | Same |
| Antibody/Antigen | Mouse monoclonal antibody PPX antigen | Same |
{3}
| Differences | | |
| --- | --- | --- |
| Feature | Candidate Devices
(QuickTox® Drug Screen Dipcard and Fastect® II Drug Screen Dipstick) | Predicate Device
(ACON® multi-CLIN™ Drug Screen Test Device k041685) |
| Test Strip | Single Drug/Multi-Drug | Multi-Drug |
| Product Design | Dip method | Dip method and drop method |
| Storage | Sealed pouch at 15-30°C | Sealed pouch at 2-30°C |
| Reading Time | 5-30 minutes | 5-8 hours |
| Internal Procedural Controls | Negative control line | Positive and negative control lines |
K. Standard/Guidance Document Referenced (if applicable):
None were referenced
L. Test Principle:
The Fastect® II PPX Drug Screen Dipstick, Fastect® II Drug Screen Dipstick with Propoxyphene and QuickTox® Drug Screen Dipcard with Propoxyphene are based on the principle of highly specific immunochemical reactions between antigens and antibodies. All dipstick and dipcard devices utilize a competitive immunoassay procedure in which an immobilized drug conjugate competes with the drug present in urine for limited antibody binding sites.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision studies were performed over ten days at three point-of-care sites using drug-free normal urine spiked with drug standards to the following concentrations: cutoff, $\pm 25\%$, $\pm 50\%$, $\pm 75\%$ and $\pm 100\%$ of the cutoff. A total of 30 samples were tested at cutoff, $\pm 25\%$, $\pm 50\%$, $\pm 75\%$ and $\div 100\%$ cutoff levels, respectively. A total of 60 samples were tested at negative (zero) level. Sample concentrations were confirmed by GC/MS. The intended users performed the testing by following the instructions for use.
{4}
PPX measured by QuickTox® Drug Screen Dipcard
| Conc. | Site 1 | | Site 2 | | Site 3 | | Combined | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Neg | Pos | Neg | Pos | Neg | Pos | Neg | Pos |
| Negative | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 |
| -75% | 10 | 0 | 10 | 0 | 10 | 0 | 30 | 0 |
| -50% | 10 | 0 | 10 | 0 | 9 | 1 | 29 | 1 |
| -25% | 9 | 1 | 8 | 2 | 9 | 1 | 26 | 4 |
| Cutoff | 4 | 6 | 3 | 7 | 7 | 3 | 14 | 16 |
| +25% | 2 | 8 | 0 | 10 | 3 | 7 | 5 | 25 |
| +50% | 0 | 10 | 0 | 10 | 0 | 10 | 0 | 30 |
| +75% | 0 | 10 | 0 | 10 | 0 | 10 | 0 | 30 |
| +100% | 0 | 10 | 0 | 10 | 0 | 10 | 0 | 30 |
PPX measured by Fastect® II Drug Screen Dipsticks
| Conc. | Site 1 | | Site 2 | | Site 3 | | Combined | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Neg | Pos | Neg | Pos | Neg | Pos | Neg | Pos |
| Negative | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 |
| -75% | 10 | 0 | 10 | 0 | 10 | 0 | 30 | 0 |
| -50% | 10 | 0 | 10 | 0 | 10 | 0 | 30 | 0 |
| -25% | 9 | 1 | 8 | 2 | 8 | 2 | 25 | 5 |
| Cutoff | 6 | 4 | 1 | 9 | 7 | 3 | 14 | 16 |
| +25% | 2 | 8 | 1 | 9 | 2 | 8 | 5 | 25 |
| +50% | 0 | 10 | 0 | 10 | 0 | 10 | 0 | 30 |
| +75% | 0 | 10 | 0 | 10 | 0 | 10 | 0 | 30 |
| +100% | 0 | 10 | 0 | 10 | 0 | 10 | 0 | 30 |
b. Linearity/assay reportable range:
Not applicable, the devices are intended for qualitative use.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
A control band with a different antigen/antibody reaction is added to the immunochromatographic membrane strip and should always appear regardless of the presence of the drug or its metabolites. The appearance of the control band during testing indicates that the test has completed and the test is valid.
Device Stability:
Accelerated studies have been conducted for one lot of Fastect® II Drug Screen Dipsticks and two lots of QuickTox® Drug Screen Dipcard kits. The manufacturer claims that when stored at room temperature (15 - 30 °C), the product is stable for 24 months. Real time studies have been conducted to support a shelf life of 24 months at room temperature. Stability protocols and acceptance criteria were reviewed and considered acceptable.
{5}
Sample Storage and Stability:
Urine specimens may be refrigerated at 2-8°C for up to two days.
d. Detection limit:
Not applicable, this is a qualitative assay.
e. Analytical specificity:
Analytical specificity was established by spiking structurally related compounds into negative human urine. The lowest amount of compounds required to produce a positive response in Fastect® II Drug Screen Dipsticks and QuickTox® Drug Screen Dipcard and their corresponding % cross reactivity are displayed below.
| Compound | Concentration | % cross reactivity |
| --- | --- | --- |
| Propoxyphene | 300 ng/mL | 100% |
| Norpropoxyphene | 500 ng/mL | 60% |
Potential interference from structurally unrelated and endogenous compounds was tested by spiking the potentially interfering compound into ±50% of cutoff urine controls at a concentration of 100,000 ng/mL. No negative or positive interference was observed for the following compounds except furosemide in this study.
Acetaminophen (4-Acetamidophenol; APAP; N-Acetyl-p-aminophenol), Acetone, Acetylsalicylic acid (Aspirin), Albumin, Amoxapine, Amoxicillin, Ampicillin, l-Ascorbic Acid (Vitamin C), Aspartame, Atropine, Benzocaine (Ethyl p-Aminobenzoate), Bilirubin, d-Brompheniramine, Caffeine, Chloroquine, d,l-Chlorpheniramine, Chlorpromazine, l-Cotinine, Creatine, Creatinine, Cyclobenzaprine, Desipramine, Dextromethorphan, 4-Dimethylaminoantipyrine, Diphenhydramine, Dopamine (3-Hydroxytyramine), Doxylamine, l-Ephedrine, d,l-Ephedrine, l-Epinephrine, Erythromycin, Ethanol, 2-Ethylidene-1,5-Dimethyl-1-3,3-Diphenylpyrrolidone, Furosemide*, Gentisic acid, Glucose, Guaiacol, Glyceryl Ether, Hemoglobin, Hippuric acid, Hydrochlorothizide, Hydrocodone, Hydromorphone, Ibuprofen, Lidocaine, Methaqualone, Methoxyphenamine, Methylphenidate, Methadone, Nalidixic acid, Naloxone, d-Naproxen, Niacinamide, Norethindrone, Oxalic acid, Oxolinic acid, Papaverine, Penicillin-G (Benzylpenicillin), Pentazocine, Perphenazine, Pheniramine, Phenothiazine (Thiodiphenylamine), Phenylephrine, β-Phenylethylamine, Procaine, Promethazine, Quinidine, Ranitidine, Riboflavin, Salicylic acid, Serotonin, Sodium Chloride, Sulfamethazine, Sulindac, Thiamine, Trimethobenzamide, Thioridazine, Trifluoperazine, Trimipramine Maleate, Tryptamine, d,l-Tryptophan, Tyramine, d,l-Tyrosine, Uric Acid, Verapamil, Zomepirac.
6
{6}
Possible false negative result for PPX may be observed for Furosemide at $\geq 60,000\ \mathrm{ng/mL}$
The effects of pH and specific gravity of the specimen on the performance of QuickTox® Drug Screen Dipcard and Fastect® II Drug Screen Dipstick test kits at $\pm 50\%$ of the cutoff levels were also tested. The results demonstrated no interference to results with pH range of 4.5-8.5 and specific gravity range of 1.005 to 1.030.
f. Assay cut-off:
Characterization of how the device performs analytically around the claimed cutoff concentration appears in the precision section, M.1.a, above.
2. Comparison studies:
a. Method comparison with predicate device:
A total of 84 urine specimens were evaluated with the QuickTox® Drug Screen Dipcard and Fastect® II Drug Screen Dipsticks devices. Of these 84 urine specimens, 42 were negative urine samples with 5 of them between cut-off and $-50\%$ cut-off levels. 42 were positive urine samples with 7 of them between cut-off and $+50\%$ cut-off levels. Both the QuickTox® Drug Screen Dipcard and Fastect® II Drug Screen Dipstick devices were compared against the values obtained through GC/MS analysis. The results are presented in the table below:
| Candidate Device | Candidate Device Results | Clinical Urine Cut-off (CO) Concentrations | | | |
| --- | --- | --- | --- | --- | --- |
| | | -100% to <-50% | ≥ -50% to < CO | ≥CO to <+50% | ≥ +50% |
| QuickTox® Drug Screen Dipcard | Positive | 0 | 2 | 5 | 35 |
| | Negative | 37 | 3 | 2 | 0 |
| Fastect® II Drug Screen Dipsticks | Positive | 0 | 2 | 7 | 35 |
| | Negative | 37 | 3 | 0 | 0 |
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
{7}
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
The sponsor performed an optimal read time evaluation by testing 5 levels of urine controls in 20 replicates per level using both dip stick and dip card devices. The test results supported a reading time of 5 to 60 minutes. The sponsor chose a more conservative reading time of 5 to 30 minutes in their product inserts.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and does satisfy the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and does support a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.