Remesense for Sensitive Teeth is intended for use as a tooth desensitizer
Device Story
Sylphar Remesense for Sensitive Teeth is a tooth desensitizer consisting of a pre-formed dental tray and foam strips impregnated with a chemical formulation (glycerin, aqua, dipotassium oxalate, aroma, EDTA, methylparaben, citric acid, saccharin). The device functions by applying dipotassium oxalate crystals to the teeth to physically block dentinal tubules; this prevents thermal stimuli from reaching the dental nerve, thereby reducing hypersensitivity. It is intended for over-the-counter (OTC) use by patients at home. The device is identical to the previously cleared Remedent Remesense (K082594), with this submission seeking to transition the product from prescription-only to OTC status. Clinical benefit is derived from the rapid relief of tooth sensitivity symptoms.
Clinical Evidence
No clinical trials were performed. Evidence consists of a questionnaire sent to US dentists confirming that desensitizing products are routinely used at home and that additional information is merely reinforcement, and a usability study demonstrating that patients can correctly identify the product and recognize contraindications.
Technological Characteristics
The device consists of a pre-formed dental tray and foam strips impregnated with a chemical solution containing dipotassium oxalate, glycerin, aqua, aroma, EDTA, methylparaben, citric acid, and saccharin. It is a non-electronic, passive device. No specific ASTM standards are cited.
Indications for Use
Indicated for patients suffering from acute tooth hypersensitivity to block dental tubules and prevent pain signals from hot or cold stimuli.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "Sylphar" in a stylized font, with a line above it. To the left of the word is a circular logo with the letters "SP" inside. The logo and the word "Sylphar" are in a dark color, possibly black or dark blue. The image appears to be a logo or branding element for a company named Sylphar.
Type of 510(k) submission:
FDA Registration Number:
FDA Registration Number: Owner/Operator Number:
510{k} Submitter and Contact:
Date of Submission:
## Appendix E
JAN 1 4 2013
# 510(k) Summary in accordance with 21 CFR 807.92(c)
Device Name:
Manufacturer:
510(k) Owner:
Phone:
Fax:
Sylphar Remesense for Sensitive Teeth
Traditional
30 August 2012
Sylphar N.V. Xavier de Cocklaan 42 B-9831 Deurle Belgium
#### 3004847139
Sylphar N.V. Xavier de Cocklaan 42 B-9831 Deurle Belgium
+32 (0)9 321 70 80 +32 (0)9 321 70 90
3006673777
10034642
ાં કાર
872.2360
Dental
Dental
Class II
Varnish, Cavity
Tooth Desensitizer
Mr Roger Gray VP Quality and Regulatory Donawa Lifescience Consulting Piazza Albania 10 00153 Rome Italy
Phone: +39 06 578 2665 Fax: +30 06 574 3786 Email: rgray@donawa.com
FDA Product Code: FDA Regulation Number: FDA Classification Name: Classification Panel: Common Name: FDA Panel: FDA Classification:
FDA Identification:
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
510(k) for Sylphar Remesense for Sensitive Teeth
Appendix E
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Sylphar. The logo consists of a circular design with the letters "SF" inside, followed by the word "Sylphar" in a bold, sans-serif font. The word "Sylphar" has a line above it.
Indications for Use: Remesense for Sensitive Teeth is intended for use as a tooth desensitizer.
#### Device Description:
'Sylphar Remesense for Sensitive Teeth' is an identical device to 'Remedent Remesense', cleared for marketing in the US by FDA under K082594.
'Remedent Remesense' was cleared as a 'prescription only' (Rx) device. The purpose of this 510(k) submission is to present additional data so that 'Sylphar Remesense for Sensitive Teeth' may be cleared by FDA for marketing as an 'over-the-counter' (OTC) device in addition to a prescription device under a new name.
The labeling, both on the device box and in the instructions sheet provided in the box, has however been revised to include the new product name and provide additional information to better accommodate sale as an OTC device.
The component parts of the device are:
- . Outer box
.
- Pre-formed dental tray in a plastic bag .
- 3 x Two foam strips impregnated with glycerin, aqua, dipotassium oxalate, aroma, EDTA, methylparaben, citric acid and saccharin
- Plastic packaging for the foam strips, with foil covering .
- Instruction sheet .
#### Performance Data:
Sylphar Remesense for Sensitive Teeth uses dipotassium oxalate crystals to block the tubules in the teeth of patients suffering from acute tooth hypersensitivity. By blocking the tubules, it stops the signals from hot or cold food reaching the dental nerve, providing a rapid remedy for hypersensitivity.
in order to confirm the safety and effectiveness of Sylphar Remesense for Sensitive Teeth as an OTC product, two studies were performed:
- · The first was a questionnaire sent to US dentists
- · The second was a usability study of the subject device
The conclusion from the results of the questionnaire was that dentists are routinely prescribing Rx tooth desensitizing products for patient home use, and that any additional information provided is merely a reinforcement of product instructions, rather than anything new related to safety or effectiveness. The additional instructions provided by the dentists answering the questionnaire have been reviewed and, where relevant to the subject device, it has been ensured that these are included within the instructions drafted for the subject device.
The conclusion from the usability study on the subject device was that patients suffering from tooth hypersensitivity are able to select the product from a display of other dental products, and also that patients who have conditions that are contraindicated for use of the subject device are unlikely to select the product. Several recommendations for improvement of the product labeling resulted from the usability studv.
#### Comparison with predicate devices:
The predicate devices selected for comparison with the Sylphar Remesense for Sensitive Teeth are:
| Predicate Device 1: | Remedent Remesense |
|---------------------|--------------------|
| 510(k) Sponsor: | Remedent N.V. |
| 510(k) Number: | K082594 |
510(k) for Sylphar Remesense for Sensitive Teeth
Appendix E
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "Sulphar" in a stylized font, with a line above it. To the left of the word is a circular logo with the letters "SP" inside. The logo is made up of thick lines, and the letters are stylized to fit within the circle. The image is in black and white.
| Clearance Date: | 19 March 2009 |
|----------------------|---------------------|
| FDA Product Code: | LBH |
| Classification Name: | Varnish, Cavity |
| Regulation No: | 872.3260 |
| Predicate Device 2: | Centrix Senzzzz Awa |
| 510(k) Sponsor: | Centrix Inc |
| 510(k) Number: | K120176 |
| Clearance Date: | 25 July 2012 |
| FDA Product Code: | LBH |
| Classification Name: | Varnish, Cavity |
| Regulation No: | 872.3260 |
#### Conclusion:
Based on the information contained within this submission, it is concluded that 'Sylphar Remesense for Sensitive Teeth' is substantially equivalent to the identified predicate devices already in interstate commerce within the USA.
:
: . .
.
.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol. The text is in all capital letters and is in a sans-serif font.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## January 14, 2013
Sylphar N.V. C/O Mr. Roger Gray Vice President, Quality and Regulatory Donawa Lifescience Consulting Piazza Albania 10 Rome, Italy 00153
Re: K122708
Trade/Device Name: Sylphar Remesense for Sensitive Teeth Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: II Product Code: LBH Dated: January 3, 2013 Received: January 11, 2013
Dear Mr. Gray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kwame O. Ulmer
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
## Indications for Use Statement
1122708 510(k) Number (if known): Notknown
Device Name: Sylphar Remesense for Sensitive Teeth
Indications for Use:
Remesense for Sensitive Teeth is intended for use as a tooth desensitizer
Prescription Use (Part 21 CFR 801 Subpart D)
ਸੀ
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
ಸ
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
| 20<br>1 | C<br>را، لا، ا | | T | 1<br>1 | |
|---------|----------------|--|---|--------|--|
| | | | | | |
Susan Runner DDS, MA 14:35:39
-05'00'
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
**510(k) Number:** kla 1708
Page 1 of 1 Rev.2 Apr 4, 2008
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.