K122702 · Filligent Limited · ONT · Mar 18, 2013 · General, Plastic Surgery
Device Facts
Record ID
K122702
Device Name
BIOFRIEND BIOMASK N95 SURGICAL RESPIRATOR
Applicant
Filligent Limited
Product Code
ONT · General, Plastic Surgery
Decision Date
Mar 18, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The BioFriend™ BioMask™ N95 surgical respirator is a single use NIOSH-approved, disposable N95 surgical respirator with a hydrophilic plastic coating on the outer layer (active ingredient: citric acid 2% w/w, a pH-lowering agent), and a second inner layer treated with metal ions (active ingredients: copper 1.6% w/w and zinc 1.6% w/w which form ionic bonds with negatively-charged side-groups on influenza viruses). The BioFriend™ BioMask™ N95 surgical respirator inactivates 99.99% of tested influenza viruses on five minutes contact with the surface of the respirator in laboratory (in vitro) tests against the following seasonal, pandemic, avian, swine and equine influenza viruses: Influenza A subtypes and strains H1N1 (the 2009 pandemic flu subtype A/Californial04/2009, A/Brisbane/59/2007, A/Wisconsin/10/1998, A/New Jersey/8/1976, A/PuertoRico/8/1934), H3N2 (A/Brisbane/10/2007, A/Wisconsin/67/2005), H2N2 (A/2/Japan/305/1957); the bird flu subtypes: H5N1 (NIBRG-14), H9N2 (A/TurkeyWisconsin/1966), H5N2 (A/Duck/PA/10218/84); the swine fly sublype: H1N1 (A/Swine/1976/1931); the equine flu subtype: H3N8 (A/Equine/2/Miami/1963); and Influenza B strains: (B/Florida/4/2006, B/Lee/1940), under tested contact conditions. Correlation between in vitro testing results and any clinical event has not been tested. The BioFriend™ BioMask™ N95 surgical respirator, Model: Professional BF-200-3013AN is flatfolded and expands into a convex-shaped mask with polyamide/spandex elastic head-loops to secure the mask to the user's face, and a malleable aluminum strip positioned above the nose for a tighter seal around the nose and face. The device is intended to be worn during seasonal Influenza A or Influenza B, and an Influenza A or Influenza B pandemic. It is intended for occupational use, to help reduce wearer exposure to pathogenic biological airborne particulates, and to protect against the transfer of micro-organisms, body fluids, and particulate material.
Device Story
Flat-folded, convex-shaped N95 surgical respirator; secures via polyamide/spandex head-loops; includes malleable aluminum nose strip for seal. Four-layer construction: outer spun-bond polypropylene (citric acid coating), cellulose/polyester layer (copper/zinc ion treatment), melt-blown polypropylene filter, inner spun-bond polypropylene. Active layers inactivate influenza viruses via pH-lowering and ionic bonding. Used in occupational settings to reduce wearer exposure to airborne particulates and protect against fluid/microorganism transfer. Healthcare providers/occupational users wear device to mitigate infection risk during influenza outbreaks. Benefits include mechanical filtration combined with surface-level viral inactivation.
Clinical Evidence
No clinical data. Evidence consists of bench testing (Bacterial Filtration Efficiency, Sub-micron Particulate Filtration, Fluid Penetration Resistance, Breathing Resistance, Flammability) and in vitro virucidal efficacy testing against 15 influenza strains showing 99.99% inactivation after 5 minutes contact.
Indicated for occupational use by healthcare workers or others to reduce exposure to pathogenic biological airborne particulates and protect against transfer of microorganisms, body fluids, and particulate material during seasonal or pandemic Influenza A or B.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
1870 3M N95 Health Care Particulate Respirator and Surgical Mask (K063023)
Submission Summary (Full Text)
{0}------------------------------------------------
K122702
Image /page/0/Picture/1 description: The image shows the logo for "filligent" with the tagline "treating global health choices" underneath. To the left of the word "filligent" is a circular graphic with three curved lines inside. The font is sans-serif and the text is black.
# MAR 1 8 2013
### 5. 510(k) Summary
### 5.1 Applicant and Correspondent
Filligent (HK) Limited Name:
7th Floor, 69 Jervois Street Address: Sheung Wan Hong Kong
Kai Deusch, M.D., Ph.D. Contact Person: Chief Executive Officer
Phone: (852) 2542 2400
Date of Preparation: August 28, 2012
### 5.2 Manufacturer
Filligent (HK) Limited 7th Floor, 69 Jervois Street Sheung Wan Hong Kong
#### 5.3 Name of Device
Trade/Proprietary/Model Name: BioFriend™ BioMask™ N95 Surgical Respirator
Model: Professional BF-200-3013AN
N95 Respirator, Filtering Facepiece Respirator Common Name: N95 Respirator With Antimicrobial/Antiviral Agent Classification Name: Filtering Facepiece Respirator for Use by Healthcare Workers Classification Regulation: 878.4040 Panel: General Hospital Device Class: Class II Product Code: ONT Recognized Performance Std: ASTM F2100-11, NIOSH 42 CFR 84 (refer to submission)
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for "filligent". To the left of the word is a circle with three curved lines inside. Below the word "filligent" are the words "creating global health choices" in a smaller font.
#### Devices to Which New Device is Substantially Equivalent 5.4
| Device Name: | BioFriend™ BioMask™ "Premium" BF-200-3013A (Convex) |
|---------------|------------------------------------------------------------------|
| Manufacturer: | Filligent (HK) Limited |
| Reference: | K101128 |
| Device Name: | BioFriend™ BioMask™ "Universal" BF-200-2001A (Flat) |
| Manufacturer: | Filligent (HK) Limited |
| Reference: | K101128 |
| Device Name: | 1870 3M N95 Health Care Particulate Respirator and Surgical Mask |
| Manufacturer: | 3M Company |
| Reference: | K063023 |
#### ર્સ Device Description
The BioFriend™ BioMask™ N95 surgical respirator, Model: Professional (BF-200-3013AN) is flatfolded and expands into a convex-shaped mask with polyamide/spandex elastic head-loops to secure the mask to the user's face, and a malleable aluminum strip positioned above the nose for a tighter seal around the nose and face. The respirator is comprised of four layers of material: an outer layer of spun-bond polypropylene, a second layer of cellulose/polyester, a third layer of melt-blown polypropylene filter material and an inner (fourth) layer of spun-bound polypropylene. All of the construction materials used in this device are typical construction materials commonly used in surgical facemasks and N95 surgical respirators and being used in current legally marketed devices. The outer active layer of the respirator is coated with a hydrophilic plastic. The second inner layer is treated with copper and zinc ions. Both layers inactivate influenza viruses using different mechanisms of action.
#### 5.6 Statement of Intended Use
The BioFriend™ BioMask™ N95 surgical respirator is a single use NIOSH-approved, disposable N95 surgical respirator with a hydrophilic plastic coating on the outer layer (active ingredient: citric acid 2% w/w, a pH-lowering agent), and a second inner layer treated with metal ions (active ingredients: copper 1.6% w/w and zinc 1.6% w/w which form ionic bonds with negatively-charged side-groups on influenza viruses).
The BioFriend™ BioMask™ N95 surgical respirator inactivates 99.99% of tested influenza viruses on five minutes contact with the surface of the respirator in laboratory (in vitro) tests against the following seasonal, pandemic, avian, swine and equine influenza viruses: Influenza A subtypes and strains H1N1 (the 2009 pandemic flu subtype A/Californial04/2009,
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Filligent. The logo consists of a circle with three curved lines inside, followed by the word "filligent" in a sans-serif font. Below the word "filligent" are the words "creating global health choices" in a smaller font.
A/Brisbane/59/2007, A/Wisconsin/10/1998, A/New Jersey/8/1976, A/PuertoRico/8/1934), H3N2 (A/Brisbane/10/2007, A/Wisconsin/67/2005), H2N2 (A/2/Japan/305/1957); the bird flu subtypes: H5N1 (NIBRG-14), H9N2 (A/Turkey/Wisconsin/1966), H5N2 (A/Duck/PA/10218/84); the swine flu subtype: H1N1 (A/Swine/1976/1931); the equine flu subtype: H3N8 (A/Equine/2/Miami/1963); and Influenza B strains: (B/Florida/4/2006, B/Lee/1940), under tested contact conditions. Correlation between in vitro testing results and any clinical event has not been tested.
The BioFriend™ BioMask™ N95 surgical respirator, Model: Professional (BF-200-3013AN) is flatfolded and expands into a convex-shaped mask with polyamide/spandex elastic head-loops to secure the mask to the user's face, and a malleable aluminum strip positioned above the nose for a tighter seal around the nose and face. The device is intended to be worn during seasonal Influenza A or Influenza B, and an Influenza B pandemic. It is intended for occupational use, to help reduce wearer exposure to pathogenic biological airborne particulates, and to protect against the transfer of micro-organisms, body fluids, and particulate material.
### 5.7 Summary of Technological Characteristics
The BioFriend™ BioMask™ N95 surgical respirator is a single use NIOSH-approved, disposable N95 surgical respirator with a hydrophilic plastic coating on the outer layer (active ingredient: citric acid 2% w/w, a pH-lowering agent), and a second inner layer treated with metal ions (active ingredients: copper 1.6% w/w and zinc 1.6% w/w which form ionic bonds with negatively-charged side-groups on influenza viruses). Both the outer and second inner layers are treated with different compounds that independently inactivate viruses through different mechanisms of action.
The manufacturing methods and construction materials used in the BioFriend™ BioMask™ N95 surgical respirator Professional BF-200-3013AN, are identical to those used in the following FDAapproved predicate devices: the BioFriend™ BioMask™ Universal BF-200-2001A, and the BioFriend™ BioMask™ Premium BF-200-3013A (K101128). Filligent (HK) Limited is the manufacturer of these predicate devices and has complete documentation with respect to the manufacturing methods and materials employed. The above BioFriend™ BioMask™ facemasks are identically comprised of four layers of the same material: an outer layer of spun-bond polypropylene, a second layer of cellulose/polyester, a third layer of melt-blown polypropylene filter material and an inner (fourth) layer of spun-bound polypropylene. The BioFriend™ BioMask™ N95 surgical respirator is also identical in structural shape and design to the following FDA-approved predicate: the BioFriend™ BioMask™ Premium model (BF-200-3013A, convex), except that the meitblown material used in the third layer of the N95 respirator model has
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "filligent" with the tagline "creating global health choices". To the left of the word "filligent" is a circular icon with three curved lines inside. The text is in a simple, sans-serif font and is black.
increased filtration specifications (18g/m² meltblown polypropylene increased to a 50g/m² mettblown polypropylene) to meet the standards required for NIOSH Respirator Certification.
The anti-viral/anti-influenza materials used in the first two active layers of the BioFriend™ BioMask™ N95 surgical respirator are identical in all respects to those of the following FDAapproved predicate devices: the BioFriend™ BioMask™ Universal BF-200-2001A, and the BioFriend™ BioMask™ Premium BF-200-3013A (K101128). Both the outer and second inner layers are treated with different compounds that independently inactivate viruses through different mechanisms of action. Laboratory (in vitro) tests conducted demonstrate that, as the active layers are identical in all respects and processed by identical manufacturing methods to those in the above predicates, the BioFriend™ BioMask™ N95 surgical respirator inactivates 99.99% (≥4-logs) of 15 different strains of Influenza A and Influenza B viruses, including the circulating 2009 pandemic H1N1, recent vaccine isolates, major reassortments and avian, swine and equine isolates after 5 minutes contact with the mask surface.
The construction materials used to provide the mechanical filtration in the BioFriend™ BioMask™ N95 surgical respirator - the polypropylene and meltblown materials - are also identical or similar to all of the following predicate devices: the NIOSH-certified Type N95 3M 1870 N95 Health Care Particulate Respirator and Surgical Mask (K063023), and the BioFriend™ BioMask™ surgical facemasks - Universal BF-200-2001A and Premium BF-200-3013A (K101128). The device as a whole is substantially equivalent to all of the predicate devices.
The device as a whole is comprised of identical materials and is processed by identical manufacturing methods to the predicates: the BioFriend™ BioMask™ surgical facemasks Universal BF-200-2001A and Premium BF-200-3013A (K101128), and have been previously approved as biocompatible by irritation and sensitization testing, and toxicological assessment of the products ingredients potentially released with inhalation or salivary contact, and indicates that the facemask and respirator products are safe for use in the intended application. Filligent (HK) Limited is the manufacturer of the predicate devices: the BioFriend™ BioMask™ surgical facemasks - Universal BF-200-2001A and Premium BF-200-3013A (K101128), and has complete documentation with respect to the manufacturing methods and materials employed. Therefore the BioFriend™ BioMask™ N95 surgical respirator raises no new safety issues. The device as a whole is substantially equivalent to these predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "filligent" with the tagline "creating global health choices". To the left of the word "filligent" is a circle with three curved lines inside. The font is sans-serif and the text is black.
### 5.8 Brief description of the nonclinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence
- Bacterial Filtration Efficiency ASTM F2101 .
- Sub-micron Particulate Filtration Efficiency ASTM F2299 .
- Fluid Penetration Resistance ASTM F1862 .
- Breathing Resistance MIL-M-3654C .
- . Flammability Testing -- 16 CFR 1610
- . Biocompatibility - Irritation - ISO 10993-10
- Biocompatibility Sensitization ISO 10993-10 .
- . Biocompatibility - Chemical Characterization - ISO 10993-18
- Breathing Resistance NIOSH 42 CFR 84.180 .
- Particulate Filtration Efficiency NIOSH 42 CFR 84.181 .
### 5.9 Brief discussion of the clinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence
Not applicable.
#### 5.10 Conclusions drawn from the nonclinical and clinical tests
The performance testing demonstrates that the BioFriend™ BioMask™ N95 surgical respirator is substantially equivalent to the FDA-approved predicate devices, and that it conforms to the recognized FDA consensus standard ASTM F2100-11 Standard Specification for Performance of Materials used in Medical Face Masks and the NIOSH Respirator Certification - 42 CFR 84 -Approval of Respiratory Protective Devices.
The BioFriend™ BioMask™ N95 surgical respirator has been tested for, and appropriately conforms to the requirements set out in ASTM F2100-11 Standard Specification for Performance of Materials Used in Medical Face Masks and NIOSH Respirator Certification 42 CFR 84 -Approval of Respiratory Protective Devices. Performance testing on the BioFriend™ BioMask™ N95 surgical respirator Professional Model BF-200-3013AN has assessed bacterial filtration efficiency, differential pressure, sub-micron particulate filtration efficiency, resistance to penetration by synthetic blood, and flammability. These tests are defined in the FDA Guidance for Industry and FDA Staff, Surgical Masks - Premarket Notification 510(k) Submissions and in the
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Filligent. The logo consists of a circle with three curved lines inside, resembling a fingerprint or sound waves. To the right of the circle is the word "filligent" in a bold, sans-serif font. Below the word "filligent" are the words "creating global health choices" in a smaller font.
FDA-recognized consensus standard, ASTM F2100-11 Standard Specification for Performance of Materials used in Medical Face Masks and the NIOSH Respirator Certification - 42 CFR 84 -Approval of Respiratory Protective Devices. Tests were conducted at Nelson Laboratories Inc., Salt Lake City, UT. Virucidal Efficacy testing was carried out at Microbiotest Inc., Sterling, VA. All testing was carried out in compliance with the Good Laboratory Practice (GLP) regulations.
The construction materials used to provide the mechanical filtration in the BioFriend™ BioMask™ N95 surgical respirator - the polypropylene and meltblown materials - are also identical or similar to all of the following predicate devices: the NIOSH-certified Type N95 3M 1870 N95 Health Care Particulate Respirator and Surgical Mask (K063023), and the BioFriend™ BioMask™ surgical facemasks - Universal BF-200-2001A and Premium BF-200-3013A (K101128) The BioFriend™ BioMask™ N95 surgical respirator Professional Model BF-200-3013AN is substantially equivalent as a whole to the predicate devices and is comprised of identical materials and processed by identical manufacturing methods to the following FDA-approved predicate devices: the BioFriend™ BioMask™ Universal BF-200-2001A, and the BioFriend™ BioMask™ Premium BF-200-3013A. Filligent (HK) Limited is the manufacturer of these predicate devices and has complete documentation with respect to the manufacturing methods and materials employed. Therefore the BioFriend™ BioMask™ N95 surgical respirator raises no new safety issues. The BioFriend™ BioMask™ N95 surgical respirator Professional Model BF-200-3013AN is substantially equivalent as a whole to the above predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Filligent. The logo consists of a circular icon with concentric arcs inside, followed by the word "filligent" in a sans-serif font. Below "filligent" are the words "creating global health choices" in a smaller font size.
Table 5.1: Performance to ASTM F2100 Level 3 Barrier Medical Facemask - Comparison of the BioFriend™ BioMask™ N95 Surgical Respirator Model: Professional BF-200-3013AN to the Predicate Device, the BioFriend™ BioMask™ Surgical
Facemask Model: Premium BF-200-3013A (K101128) and the 1870 3M N95 Health Care Particulate Respirator and Surgical Mask (K063023)
| Description | Test Method | BioFriend™<br>BioMask™<br>N95 Surgical<br>Respirator<br>Professional<br>BF-200-3013AN | BioFriend ™<br>BioMask™ Surgical<br>Facemasks<br>Universal<br>BF-200-2001A and<br>Premium<br>BF-200-3013A | 1870 3M N95<br>Health Care<br>Particulate<br>Respirator and<br>Surgical Mask |
|-------------------------------------------------------------|---------------------|---------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| 510(k) Number | | K122702 | K101128 | K063023 |
| Product Code | | ONT | OUK | MSH |
| NIOSH Certificate<br>Number | | TC-84A-6252 | | TC-84A-3844 |
| Bacterial Filtration<br>Efficiency | ASTM F2101 | > 99.9 % | 99.9 % | > 99 % |
| Sub-micron<br>(0.1 µm) Particulate<br>Filtration Efficiency | ASTM F2299 | > 99.9 % | 99.7 % | Information Not<br>Declared |
| Differential Pressure<br>(Delta-P) | MIL-M-36954C | 4.9 mm H2O/cm² | 2.6 mm H2O/cm² | 4.9 mm H2O/cm² |
| Resistance to<br>Penetration by<br>Synthetic Blood | ASTM F1862 | 160 mm Hg | 160 mm Hg | 160 mm Hg |
| Flame Spread | 16 CFR Part<br>1610 | Class 1 | Class 1 | Class 1 |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Filligent. The logo consists of a circle with three curved lines inside, and the word "filligent" in bold letters. Below the word "filligent" are the words "creating global health choices" in a smaller font. The logo is simple and modern, and it conveys the company's focus on health and innovation.
Table 5.2: Performance to NIOSH 42 CFR Part 84 - N95 Respirator - Comparison of the BioFriend™ BioMask™ N95 Surgical Respirator Model: Professional BF-200-3013AN to the Predicate Device, the BioFriend™ BioMask™ Surgical Facemask Model: Premium BF-200-3013A (K101128) and the 1870 3M N95 Health Care Particulate Respirator and Surgical Mask (K063023)
| Description | Test Method | BioFriend™<br>BioMask™<br>N95 Surgical<br>Respirator<br>Professional<br>BF-200-3013AN | BioFriend™<br>BioMask™ Surgical<br>Facemasks<br>Universal<br>BF-200-2001A and<br>Premium<br>BF-200-3013A | 1870 3M N95<br>Health Care<br>Particulate<br>Respirator and<br>Surgical Mask |
|------------------------------------------------|---------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| 510(k) Number | | K122702 | K101128 | K063023 |
| Product Code | | ONT | OUK | MSH |
| NIOSH Certificate<br>Number | | TC-84A-6252 | | TC-84A-3844 |
| Particulate<br>Filtration Efficiency<br>(NaCl) | 42 CFR 84.181 | 98 % | N/A | Pass |
| Breathing<br>Resistance | 42 CFR 84.180 | 8.0 mm H2O<br>(Inhalation)<br>8.9 mm H2O<br>(Exhalation) | N/A | Pass. |
{8}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### March 18, 2013
Kai Deusch, M.D., Ph.D. Chief Executive Officer Filligent (HK) Limited 700 Floor 69 Jervois Street Sheung Wan, Hong Kong
Re: K122702
Trade/Device Name: BioFriend 101 BioMask 114 N95 Surgical Respirator Model: Professional BF-200-3013AN Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: ONT Dated: January 14, 2013 Received: February 28, 2013
## Dear Dr. Deusch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
Page 2 - Dr. Deusch
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Image /page/9/Picture/6 description: The image shows a logo or emblem with stylized text and geometric shapes. The text appears to be handwritten or calligraphic, with some words or letters overlapping. The geometric shapes consist of a series of nested triangles, creating a pyramid-like structure on the right side of the image. The overall design is intricate and abstract, with a combination of organic and geometric elements.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure -
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for "filligent" with the tagline "creating global health choices" underneath. To the left of the text is a circular graphic with curved lines inside, resembling a stylized sound wave or signal. The logo is simple and modern, using a clean font and a minimalist design.
# Indications for Use Statement
K122702
510(k) Number (if known):
Device Name:
प
BioFriend™ BioMask™ N95 Surgical Respirator Model: Professional BF-200-3013AN
### Indications for Use:
The BioFriend™ BioMask™ N95 surgical respirator is a single use NIOSH-approved, disposable N95 surgical respirator with a hydrophilic plastic coating on the outer layer (active ingredient: citric acid 2% w/w, a pH-lowering agent), and a second inner layer treated with metal ions (active ingredients: copper 1.6% w/w and zinc 1.6% w/w which form ionic bonds with negatively-charged side-groups on influenza viruses).
The BioFriend™ BioMask™ N95 surgical respirator inactivates 99.99% of tested influenza viruses on five minutes contact with the surface of the respirator in laboratory (in vitro) tests against the following seasonal, pandemic, avian, swine and equine influenza viruses: Influenza A subtypes and strains H1N1 (the 2009 pandemic flu subtype A/Californial04/2009, A/Brisbane/59/2007, A/Wisconsin/10/1998, A/New Jersey/8/1976, A/PuertoRico/8/1934), H3N2 (A/Brisbane/10/2007, A/Wisconsin/67/2005), H2N2 (A/2/Japan/305/1957); the bird flu subtypes: H5N1 (NIBRG-14), H9N2 (A/TurkeyWisconsin/1966), H5N2 (A/Duck/PA/10218/84); the swine fly sublype: H1N1 (A/Swine/1976/1931); the equine flu subtype: H3N8 (A/Equine/2/Miami/1963); and Influenza B strains: (B/Florida/4/2006, B/Lee/1940), under tested contact conditions. Correlation between in vitro testing results and any clinical event has not been tested.
The BioFriend™ BioMask™ N95 surgical respirator, Model: Professional BF-200-3013AN is flatfolded and expands into a convex-shaped mask with polyamide/spandex elastic head-loops to secure the mask to the user's face, and a malleable aluminum strip positioned above the nose for a tighter seal around the nose and face. The device is intended to be worn during seasonal Influenza A or Influenza B, and an Influenza A or Influenza B pandemic. It is intended for occupational use, to help reduce wearer exposure to pathogenic biological airborne particulates, and to protect against the transfer of micro-organisms, body fluids, and particulate material.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for "filligent" with a circle on the left that has three curved lines inside of it. To the right of the circle is the word "filligent" in a bold, sans-serif font. Below the word "filligent" are the words "creating global health choices" in a smaller font. The logo is black and white.
Prescription Use _ (21 CFR 801 Subpart D) AND/OR
Over-The Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
1
Page 1 of
# Elizabeth គ 2013.03.18 4'00'
(Division Sign-Off) Division of Anestheslology, General Hospital Infection Control, Dental Devices
510(k) Number;_ 1122702
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.