PROSBEN CONTACT LENS CASES MODEL CA66-02, CA66-03WB, CA66-04T SERIES
K122700 · Prosben, Inc. · LRX · May 31, 2013 · Ophthalmic
Device Facts
Record ID
K122700
Device Name
PROSBEN CONTACT LENS CASES MODEL CA66-02, CA66-03WB, CA66-04T SERIES
Applicant
Prosben, Inc.
Product Code
LRX · Ophthalmic
Decision Date
May 31, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
The PROSBEN Contact Lens Cases are intended for storage during chemical disinfection of soft (hydrophilic), rigid gas permeable and hard contact lenses. For use during chemical disinfection only; not to be used with heat.
Device Story
PROSBEN Contact Lens Cases are containers for storing contact lenses during chemical disinfection. Device consists of two wells with either hinged caps (CA66-04T series) or screw-top caps (CA66-02 and CA66-03WB series). Used by contact lens wearers or practitioners to hold lenses while submerged in chemical disinfectant solution. Device does not perform disinfection itself; it provides a storage vessel. Benefits include safe, clean storage of lenses during the chemical disinfection process. Not for use with heat-based disinfection systems.
Clinical Evidence
Bench testing only. Device underwent biological evaluation per ISO 10993 standards, including in vitro cytotoxicity (ISO 10993-5), irritation and skin sensitization (ISO 10993-10), and systemic toxicity (ISO 10993-11). Results confirmed safety for intended use.
Technological Characteristics
Materials: Polypropylene (PP, CAS# 9003-07-0) for CA66-02/CA66-03WB series; Polyethylene (PE, CAS# 9002-88-4) for CA66-04T series. Includes colorants (white, blue, green). Design: Two-well container with hinged or screw-top caps. Non-powered, passive storage device.
Indications for Use
Indicated for storage of soft (hydrophilic), rigid gas permeable, and hard contact lenses during chemical disinfection. For use by contact lens wearers or practitioners. Contraindicated for use with heat disinfection.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: This image contains the company information for PROSBEN INC. Their phone number is 886-2-2500 0586, and their fax number is 886-2-2500 0945. The company address is 11F, No.58, Ming Chuan East Road, Sec.3, Taipei 104, Taiwan. The image also contains the date MAY 31 2013.
# II. 510(K) Summary of Safety and Effectiveness (Per 21 CFR 807.92)
## 2.1. General Information Establishment
- . Manufacturer: PROSBEN INC.
- Address: 11F., No.58, Sec. 3, Mingchaun East Road, Taipei, 10477, Taiwan, R.O.C. .
- B Owner Number: 8022656
| Contact Person: | Dr. Jen, Ke-Min E-mail: ceirs.jen@msa.hinet.net |
|-----------------|-------------------------------------------------|
| | (Tel) +886-3-5208829 ; (Fax) +886-3-5209783 |
. Date Prepared: August 23, 2012
## Device
- Proprietary Name: PROSBEN Contact Lens Cases;
## CA66-02, CA66-03WB, CA66-04T series
- Common Name: . Contact Lens Case
- . Case, Contact Lens Classification Name:
- LRX, Class II, 886.5928 ● Product Code:
## 2.2. Safety and Effectiveness Information
#### . Predicate Device:
Claim of Substantial Equivalence (SE) is made to i-Promotions Contact Lens Case (K042578)
#### ● Device Description:
The PROSBEN Contact Lens Case is the subject of this 510(k) premarket notification. The Contact Lens Case is a lens care product to be used the contact lens wearer or practitioner for storing contact lenses while not being worn. The PROSBEN Contact Lens Cases are designed for use in chemical disinfection only; not to be used with heat. The PROSBEN Contact Lens Cases are not including the disinfection, only for shipping the cases in a dry state.
#### Summary of Safety Data: ●
PROSBEN Contact Lens Cases completed the relevant test including:
- 1) ISO 10993-5 Biological evaluation of medical devices, part 5: Test for in vitro cytotoxicity;
- 2) ISO10993-10 Biological evaluation of medical devices, part 10: Test for irritation and skin sensitization;
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the letterhead for PROSBEN INC. The letterhead includes the company's name, address, telephone number (886-2-2500 0586), and fax number (886-2-2500 0945). The address is 11F, No.58, Ming Chuan East Road, Sec. 3, Taipei 104, Taiwan. The letterhead also includes the phrase "NEW VISION BEGIN WITH".
- 3) ISO10993-11 Biological evaluation of medical devices, part 11: Test for system toxicity:
- 4) The materials are complied with the biological and toxicology evaluation and the results of these studies show that the Contact Lens Cases are safety for its intended use.
#### Intended Use: ●
The PROSBEN Contact Lens Cases are intended for storage during chemical disinfection of soft (hydrophilic), rigid gas permeable and hard contact lenses. For use during chemical disinfection only; not to be used with heat.
#### . Substantial Equivalence (SE)
A claim of substantial equivalence is made to i-Promotions Contact Lens Case (K042578). Both of them have the same indications for use, working principle and technologies. The major differences of the two devices are due to the different designed for adjoining wells with integral hinged caps or screw-top caps, and used the different PP or PE material with different color additives; please refer to the below "Comparison Table". These are not relating to the safety or effectiveness aspects. Thus they are substantially equivalent.
| Comparison<br>Feature | i-Promotions<br>Contact Lens Case<br>(K042578) | PROSBEN<br>Contact Lens Cases series<br>(K122700) |
|-----------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Design | Two adjoining wells with<br>integral hinged caps into<br>which respective lenses are<br>immersed. | hinged caps: CA66-04T series<br>screw-top caps: CA66-02 and<br>CA66-03WB series |
| Materials | Low density Polyethylene<br>#9931 with five colorants. | PP (CAS# 9003-07-0): CA66-02 and<br>CA66-03WB series<br>PE (CAS# 9002-88-4): CA66-04T<br>series with colorants:<br>white, blue, green |
## Comparison Table
Ke. Minte
Dr. Jen, Ke-Min 510Kcorrespondent person for PROSBEN INC.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.
## May 31, 2013
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Dr. Jen, Ke-Min Prosben, Inc. 11F, No. 58, Sec. 3 Mingchaun East Road Taipei China (Taiwan) 104
Re: K122700
Trade/Device Name: Prosben Contact Lens Cases, CA66-02, CA66-03WB, and CA66-04T Series Regulation Number: 21 CFR 886.5928 Regulation Name: Case, Contact Lens Regulatory Class: Class II Product Code: LRX Dated: May 11, 2013 Received: May 22, 2013
Dear Dr. Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstatecommerce.prior to.May. 28, 1976, the enactment.date.of the Medical.Device. Amendments, or todevices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Dr. Jen. Ke-Min
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia YAlexander -S
Center for Devices and Radiological Health
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation
Enclosure
{4}------------------------------------------------
# Indications for Use
510 (K) Number ( If Known ): K122700 Device Name: _ Contact Lens Cases; CA66-02, CA66-03WB, CA66-04T series
Indications for Use:
The PROSBEN Contact Lens Cases are intended for storage during chemical disinfection of soft (hydrophilic), rigid gas permeable and hard contact lenses.
For use during chemical disinfection only; not to be used with heat.
Prescription Use (21 CFR Part 801 Subpart D) AND/OR
V Over the Counter Use __ (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Claudine H. Krayçzy 2013.05.30 13:27:3540
Page_ 1 _of _1
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.