← Product Code [BTY](/productcode/BTY) · K122699

# SENTRYSUITE PRODUCT LINE (K122699)

_Carefusion Germany 234 GmbH · BTY · Nov 19, 2012 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K122699

## Device Facts

- **Applicant:** Carefusion Germany 234 GmbH
- **Product Code:** [BTY](/productcode/BTY.md)
- **Decision Date:** Nov 19, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.1890
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Pediatric

## Indications for Use

The SentrySuite Product line is intended to be used for measurements, data collection and analysis of lung function (PFT) parameters, aiding in the diagnosis of related conditions. All the measurements are performed via a mouthpiece, a mask or nasal adapters. The results of the test can be viewed on-line with the help of a computer screen and can be printed after the test. The test results can be saved for further referral or report generation purposes. Use of the Option Bronchial Challenge requires the supervision of a physician familiar with emergency medicine. The products can be utilized for patients from 4 years on and older as long as they can cooperate in the performance - no special limit to patient's sex or height. Measurements will be performed under the direction of a physician in a hospital environment, physician's office or similar settings.

## Device Story

SentrySuite Product Line is a software-based system for pulmonary function testing (PFT). It interfaces with hardware modules (MasterScreen Pneumo, IOS, APS Pro, PFT, Body/Diff) to collect respiratory data via mouthpiece, mask, or nasal adapter. The software processes inputs to perform measurements including spirometry, flow-volume, MVV, R-occlusion, impulse oscillometry, bronchial challenge, FRC helium rebreathing, diffusion (single breath/intra-breath), bodyplethysmography, respiratory drive (P0.1), and MIP/MEP. Operated by clinicians in hospitals or physician offices, the system displays results on a computer screen for real-time monitoring, report generation, and data storage. It aids physicians in diagnosing respiratory conditions. The software replaces legacy JLAB software, providing a new graphical interface while maintaining equivalent measurement functionality. It supports workstation and server-based deployment.

## Clinical Evidence

Bench testing only. Comparative study of human subjects using predicate JLAB software and new SentrySuite 2.7 software. Statistical analysis (bivariate correlation) performed on key parameters: Diffusion (DLCO, VA, Vin), Bodyplethysmography (TLC, FRCpl, RV, sReff, Reff), Respiratory Drive (P0.1), and PI/PE Max (MIP/MEP). All parameters showed statistically significant correlation (p-values <0.0004) between software versions, meeting pass criteria.

## Technological Characteristics

Software-based PFT system operating on high-performance computers/servers. Patient-contact parts: Bormed RG835 MO (single-use mouthpiece), silicone mouthpiece, polyacetal nose clip, EVA nose clip pad. Energy: 100-240V / 50-60Hz. Connectivity: Workstation/Server, online/Vlink/GDT/JINET. Sterilization: Not supplied sterile, not intended for sterilization. Software: SentrySuite 2.7x.

## Regulatory Identification

A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.

## Predicate Devices

- SentrySuite Product Line ([K113813](/device/K113813.md))
- MasterScreen PFT Body ([K072061](/device/K072061.md))
- MasterScreen Body / Diff ([K936108](/device/K936108.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary

# NOV 1 9 2012

## GENERAL INFORMATION

### Type of Submission 5.1

Traditional 510(k) Submission

Submission date: 08/16/2012

### 5.2 Submitter

Name:

CareFusion Germany 234 GmbH

Address: .

Leibnizstrasse 7 D-97204 Hoechberg Germany

Contact person in Germany: (Regulatory Affairs Specialist) Address:

Elmar Niedermeyer CareFusion Germany 234 GmbH Leibnizstrasse 7, 97204 Hoechberg Germany +49 931 49 72 - 361 +49 931 49 72 - 62361 elmar.niedermeyer@carefusion.com

Phone: FAX: E-mail

Contact person in the U.S .: (Official Correspondent) Address

### Carol Emerson

CareFusion 22745 Savi Ranch Parkway Yorba Linda, CA 92887 714-919-3342 714-283-8420 carol.emerson@carefusion.com

Phone: Fax: E-mail:

{1}------------------------------------------------

### Establishment Registration Number 5.3 9615102

### Common Name or Classification Name 5.4

Calculator, Predicted Values, Pulmonary Function (CFR 868.1890, Product Code BTY) Pulmonary function data calculator (CFR 868.1880, Product Code BZC) Diagnostic Spirometer (CFR 868.1840, Product Code BZG) Volume plethysmograph (CFR 868.1760, Product Code JEH)

#### રે. ર Trade Name

SentrySuite Product Line

#### Device Classification 5.6

This is a Class II device

#### 5.7 Classification Panel

73 Anesthesiology Part 868 Code BTY, BZG, BZC, JEH

### 5.8 Reason for Premarket Notification

--- Modification of legally marketed devices ---Change from previously JLAB software to SentrySuite software for some measurement programs of the device MasterScreen PFT Body K072061

### Legally predicate marketed device 5.9

SentrySuite Product Line MasterScreen PFT Body MasterScreen Body / Diff

K111053 Code BTY, BZC, BZG K072061 Code JEH K936108 Code JEH

#### 5.10 Predicate Device Company

CareFusion Germany 234 GmbH

#### 5.11 Device Description

The SentrySuite Product line when operating on the existing hardware for MasterScreen Pneumo, MasterScreen IOS, APS Pro and MasterScreen PFT will be as functional as the existing version of JLAB software for all the available measuring programs and options for these devices.

- The SentrySuite software replaces the JLAB software and got a brand-new . graphical surface.
- Measurement can be accomplished under SentrySuite software equivalent . as it was possible under the previously powerful JLAB software

-

CareFusion

{2}------------------------------------------------

- The results of the tests can be viewed on-line on the computer screen . during the test and can be saved on the computer hard disk for further referral or report generation purposes.
- SentrySuite provides the functionality currently available on the . MasterScreen devices using the JLAB software.
- SentrySuite can be operated on workstations and on servers. .

Measurements:

- Spiormetry .
- Flow Volume .
- Maximal Voluntary Ventilation (MVV) .
- Incentive Spirometry .
- R-Occlusion .
- . Impulse oscillometry
- Bronchial test .
- FRC Helium Rebreathing .
- . Real Time Single Breath Diffusion
- Intra Breath Diffusion .
- Bodyplethysmography .
- . Respiratory Drive P0.1
- PI/PE Max (MIP/MEP) .
- Single Breath Diffusion CO/He .

#### 5.12 Intended Use Statement

The SentrySuite Product line is intended to be used for measurements, data collection and analysis of lung function (PFT) parameters, aiding in the diagnosis of related conditions. All the measurements are performed via a mouthpiece, a mask or nasal adapters. The results of the test can be viewed on-line with the help of a computer screen and can be printed after the test. The test results can be saved for further referral or report generation purposes.

Use of the Option Bronchial Challenge requires the supervision of a physician familiar with emergency medicine.

The products can be utilized for patients from 4 years on and older as long as they can cooperate in the performance - no special limit to patient's sex or height.

Measurements will be performed under the direction of a physician in a hospital environment, physician's office or similar settings.

### 5.13 Required Components

- Measuring device MS-Pneumo or MS-IOS or APS Pro or MS-PFT or MS-. PFT Body or MS Body Diff
- Or Workstation / Server ●
- Trolley / Stand .
- High performance computer .

{3}------------------------------------------------

## . 5 510(k) Summary

- Printer .
- . Accessories
- SentrySuite software 2.7x .

・

.

- Instruction for Use .
# 5.14 Summary Table of Comparison

|                                | SentrySuite Product Line<br>(K113813)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | SentrySuite Product Line            |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| Indications<br>for Use         | The SentrySuite Product line is intended to<br>be used for measurements, data collection<br>and analysis of lung function (PFT)<br>parameters, aiding in the diagnosis of<br>related conditions. All the measurements<br>are performed via a mouthpiece, a mask or<br>nasal adapters. The results of the test can<br>be viewed on-line with the help of a<br>computer screen and can be printed after<br>the test. The test results can be saved for<br>further referral or report generation<br>purposes. Use of the Option Bronchial<br>Challenge requires the supervision of a<br>physician familiar with emergency<br>medicine.<br>The products can be utilized for patients<br>from 4 years on and older as long as they<br>can cooperate in the performance - no<br>special limit to patient's sex or height.<br>Measurements will be performed under the<br>direction of a physician in a hospital<br>environment, physician's office or similar<br>settings. | identical                           |
| Patient<br>population          | The products can be utilized for patients<br>from 4 years on and older as long as they<br>can cooperate in the performance - no<br>special limit to patient's sex or height.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | identical                           |
| Hardware                       | APS Pro<br>• Nebulizer head<br>• Compressor<br>• Trolley or Stand with power supply<br>• Desktop / Notebook<br>• Accessories<br>MasterScreen Pneumo<br>• Trolley or Stand with power supply<br>• Pneumotach handle<br>• Desktop / Notebook<br>• Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | identical                           |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | MasterScreen IOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                     |
|                                | · Trolley or Stand with power supply                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | · Pneumotach handle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                     |
|                                | · Desktop / Notebook                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | · Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | MasterScreen PFT / PFT Body                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                     |
|                                | · Trolley with power supply                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | · Analyzer box (gas)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                     |
|                                | • Pneumotach handle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                     |
|                                | · Body Box                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | · Desktop / Notebook                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                     |
|                                | · Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
| Software                       | SentrySuite Software (version 2.5)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | SentrySuite Software (version 2.7x) |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | APS Pro                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                     |
| Perfor-                        | · Bronchial test                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                     |
| mance                          | o Bronchospasmolysis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
| specifica-                     | Bronchoprovocation<br>0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                     |
| tion                           | Pulsed Nebulization<br>0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
| (measure-                      | Continuous Nebulization<br>o                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                     |
| ment                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
| programs)                      | MasterScreen Pneumo                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | · Spirometry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                     |
|                                | · Incentive Spirometry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | · Flow / Volume                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | • MVV                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                     |
|                                | · R-Occlusion (Airway Resistance)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | MasterScreen IOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | identical                           |
|                                | · Spirometry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                     |
|                                | · Incentive Spirometry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | · Flow / Volume                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | • MVV                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                     |
|                                | · Impulse oscillometry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | MasterScreen PFT / PFT Body                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                     |
|                                | · Spirometry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | · Incentive Spirometry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                     |
|                                | · Flow / Volume                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | • MVV                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | · R-Occlusion (Airway Resistance)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                     |
|                                | · FRC Helium Rebreathing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
|                                | · Real Time Single Breath Diffusion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                     |
|                                | · Intra Breath Diffusion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
| Energy                         | 100 - 240V / 50 - 60Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                     |
| type                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | identical                           |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
| Patient                        | · Single Use mouthpiece (material:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                     |
|                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                     |
| contacting                     | Bormed RG835 MO)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | identical                           |
| parts                          | · Silicone mouthpiece                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                     |
|                                |…

---

**Source:** [https://fda-staging.innolitics.com/device/K122699](https://fda-staging.innolitics.com/device/K122699)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).