The Femasys FemCerv is a sterile, disposable endocervical sampler indicated for single patient use in obtaining tissue samples from the endocervical canal for histological analysis. Clinical indications include: further evaluation of an abnormal Pap smear; cervical lesions extending into the endocervical canal; undiagnosed uterine bleeding.
Device Story
FemCerv is a sterile, disposable, non-metal endocervical sampler; used by clinicians to obtain tissue samples from the endocervical canal for histological analysis. Device features a sample collection chamber; operation involves insertion into the endocervical canal, rotational scraping, and sheath opening/closing to collect and release tissue into a specimen container. Used in clinical settings for patients with abnormal Pap smears, cervical lesions, or undiagnosed uterine bleeding. Benefits include single-pass sampling capability.
Clinical Evidence
Clinical study conducted on subjects prior to scheduled hysterectomy to demonstrate insertion, device opening/closing, and sampling capability. Pathologist evaluation of samples from 7 devices showed an 85.7% sample adequacy rate.
Technological Characteristics
Sterile, disposable, non-metal endocervical sampler. Features a sample collection chamber with rotational scraping mechanism. Biocompatibility verified per ISO 10993 (cytotoxicity, irritation, sensitization).
Indications for Use
Indicated for patients requiring endocervical tissue sampling for histological analysis, specifically for evaluation of abnormal Pap smears, cervical lesions extending into the endocervical canal, or undiagnosed uterine bleeding.
Regulatory Classification
Identification
An endocervical aspirator is a device designed to remove tissue from the endocervix (mucous membrane lining the canal of the cervix of the uterus) by suction with a syringe, bulb and pipette, or catheter. This device is used to evaluate endocervical tissue to detect malignant and premalignant lesions.
{0}------------------------------------------------
K122658
## 510(k) Summary for the FemCerv™ Endocervical Sampler
(Models REF FCV-009, FCV-011, and FCV-013)
| Date: | December 18, 2012 DEC 2 0 2012 |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter and Primary Contact: | Lisa Peacock<br>Vice President, Regulatory and Clinical Affairs<br>Femasys Inc.<br>5000 Research Court<br>Suite 100<br>Suwanee, GA 30024<br>Tel: 770-500-3910<br>Fax: 770-500-3980<br>LPEACOCK@FEMASYS.COM |
| Device Common Name: | endocervical sampler |
| FDA Device Classification Name: | endocervical aspirator |
| Product Code: | PCF |
| Classification Regulation: | 21 CFR 884.1050 |
| Device Class: | 2 |
| Panel: | Obstetrics/Gynecology |
| Indication for Use: | The Femasys FemCerv is a sterile, disposable endocervical sampler indicated for single patient use in obtaining tissue samples from the endocervical canal for histological analysis. Clinical indications include:<br><ul><li>further evaluation of an abnormal Pap smear;</li><li>cervical lesions extending into the endocervical canal;</li><li>undiagnosed uterine bleeding.</li></ul> |
| Device Description: | The FemCerv Endocervical Sampler (FemCerv) is a sterile, disposable device that collects an endocervical tissue sample for histological evaluation. |
| Predicate Device: | K060320 FemECC™ Endocervical Curette<br>The FemCerv is substantially equivalent to the predicate in intended use, sample collection method using a single-pass |
{1}------------------------------------------------
Summary of Testing:
rotational scraping mode within the endocervical canal, sample transfer method, and in its sterile, disposable, non-metal design.
Final, finished, sterilized FemCerv devices were tested by the following methods for verification of functionality, safety, and effectiveness.
- · Bench testing was conducted to evaluate device functionality and integrity over cycle testing of opening and closing the device's sample collection chamber. In addition, the bench testing included simulated use testing in a model of the endocervical canal evaluating insertion capability, sheath opening functionality, rotational functionality, sheath closure, and sample release into a specimen container. A sample set of 29 devices was tested and passed all criteria.
- · Shelf life integrity up to 1 year accelerated aging was evaluated using the same test methodology and acceptance criteria in the point above. A sample set of 29 devices was tested and passed all criteria.
- · Biocompatibility of components according to ISO 10993 standards: cytotoxicity, irritation, and sensitization was conducted and passed all criteria.
- · A clinical study was conducted involving endocervical sampling on subjects prior to scheduled hysterectomy to demonstrate insertion, device opening and closing, and adequate sampling capability based upon a pathologist's evaluation. An 85.7% sample adequacy rate was based on sampling with 7 devices.
Conclusions of Substantial Equivalence Demonstrations:
- · FemCerv is substantially equivalent to its proposed predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## December 20, 2012
Femasys, Inc. % Ms. Lisa Peacock Vice President, Regulatory and Clinical Affairs 5000 Research Court, Suite 100 SUWANEE GA 30024
Re: K122658
Trade/Device Name: FemCerv™ Endocervical Sampler Regulation Number: 21 CFR§ 884.1050 Regulation Name: Endocervical aspirator Regulatory Class: II Product Code: PCF Dated: December 11, 2012 Received: December 12, 2012
Dear Ms. Peacock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Herbert R. Lerner
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Femasys Inc.
Page 1 of 1
## Indications for Use
510(k) Number (if known): K122658
Device Name: FemCerv™ Endocervical Sampler (Models REF FCV-009, FCV-011, and FCV-013)
Indications for Use:
The Femasys FemCerv is a sterile, disposable endocervical sampler indicated for single patient use in obtaining tissue samples from the endocervical canal for histological analysis. Clinical indications include:
- further evaluation of an abnormal Pap smear; ●
- cervical lesions extending into the endocervical canal; 미
- undiagnosed uterine bleeding. 8
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Herbert** R. Lerner
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K122658 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.