← Product Code [HRS](/productcode/HRS) · K122623

# SHORT LATERAL SUPERIOR CLAVICLE PLATE, LEFT AND RIGHT MODEL 3017-000 (LEFT) AND 3018-000 (RIGHT); 6 HOLE MEDIUM MIDSHAFT (K122623)

_Advanced Orthopaedic Solutions, Inc. · HRS · Sep 26, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K122623

## Device Facts

- **Applicant:** Advanced Orthopaedic Solutions, Inc.
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Sep 26, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The AOS Clavicle Plate System provides fixation for fractures, fusions, or osteotomies for the Clavicle.

## Device Story

AOS Short Clavicle Plates are metallic bone fixation implants designed for clavicle fractures, fusions, and osteotomies. These plates are shorter versions of existing lateral and midshaft plates within the AOS Clavicle Plate System. They feature pre-shaped curvatures to match clavicle anatomy, accept locking and non-locking screws, and include bending notches for fine-tuning fit to the bone surface. The short lateral plates incorporate an elongated hole to facilitate provisional placement and positioning adjustments. Used by orthopedic surgeons in clinical settings, these plates provide internal fixation to stabilize bone segments, promoting healing and structural integrity. The device is manufactured from Ti-6Al-4V ELI.

## Clinical Evidence

No clinical data. Substantial equivalence was determined based on design similarity and material composition to the predicate device. Bench testing was deemed unnecessary due to the similarity in strength and geometry to the existing cleared system.

## Technological Characteristics

Material: Ti-6Al-4V ELI (ASTM F136). Design: Metallic bone plates with pre-shaped curvatures, bending notches for geometry adjustment, and elongated holes for positioning. Fixation: Accepts locking and non-locking screws. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- AOS Clavicle Plate System ([K103513](/device/K103513.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K12-623

Page 1/2

Image /page/0/Picture/2 description: The image shows the logo for Advanced Orthopaedic Solutions. The logo consists of the letters "AOS" in a bold, sans-serif font. Below the letters, the words "ADVANCED ORTHOPAEDIC SOLUTIONS" are written in a smaller, sans-serif font. The logo is simple and professional, and it conveys the company's focus on orthopaedics.

SEP 2 6 2012

#### 8. SPECIAL 510(K) SUMMARY

SUMMARY PREPARED ON:

August 27, 2012

SUBMITTED BY:

Advanced Orthopaedic Solutions, Inc. 386 Beech Avenue, Unit B6 Torrance, CA 90501 Phone: (310) 533-9966

CONTACT PERSON:

Allyson Parks Advanced Orthopaedic Solutions, Inc. 386 Beech Avenue, Unit B6 Torrance, CA 90501 Phone: (310) 533-9966

Class II, 21 CFR 888.3030 Single/multiple

component metallic bone fixation appliances and

AOS Short Clavicle Plates

Internal Fixation

accessories

HRS

DEVICE NAME: COMMON NAME: CLASSIFICATION:

DEVICE CODE:

SUBSTANTIALLY EQUIVALENT DEVICE:

DEVICE DESCRIPTION:

AOS Clavicle Plate System (K103513, October 24, 2011)

The AOS Clavicle Plate System consists of bone plates and screws for fractures, fusions, and osteotomies of the Clavicle bone. The bone plates are pre-shaped to fit the curves and angles of the Clavicle and are provided in a range of curvatures and sizes. The plates accept both locking and nonlocking screws.

The proposed new plates are shorter versions of the lateral and midshaft plates currently in the system. The 6 Hole Medium Midshaft Superior Clavicle Plate and 6 Hole Straight Midshaft Superior Clavicle Plate both come in right and left versions and follow the same basic curvature as that of the existing 8 Hole Medium and Straight Midshaft Clavicle Plates, respectively. The Short Lateral Superior Clavicle Plates come in right and left versions and follow the same basic geometry as the current lateral superior plates, with the addition of an elongated hole. The

Page 8-2

{1}------------------------------------------------

122623

Page. 2/2

elongated hole will be used for provisional placement with the ability to adjust the plate positioning.

All plates except for the short lateral plates have at least one bending notch. This notch feature is used for fine adjustment of the plate geometry in order to fit the plate more intimately with the bone surface.

The AOS Clavicle Plate System provides fixation for fractures, fusions, or osteotomies for the Clavicle.

### INDICATIONS FOR USE:

## SUBSTANTIAL EQUIVALENCE:

Information presented supports substantial equivalence of the AOS Short Clavicle Plates to the predicate device. The proposed plates have the same indications for use, are similar in geometry and design, have the same fundamental scientific technology, and are made of the same material (Ti-6Al-4V ELI, per ASTM F136) as the predicate plates. As detailed in the submission, the proposed plates do not present a worst-case scenario with respect to strength characteristics, and because of their similarity to the current plates, physical testing was deemed unnecessary, and substantial equivalence was determined in strength and geometry between the proposed plates and the predicate plates.

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the circumference of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SEP 2 6 2012

Advanced Orthopaedic Solutions, Incorporated % Ms. Allyson Parks Regulatory Associate 386 Beech Avenue, Unit B6 Torrance, California 90501

Re: K122623

Trade/Device Name: AOS Short Clavicle Plates Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II

Product Code: HRS Dated: August 27, 2012 Received: August 28, 2012

### Dear Ms. Parks:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

# Page 2 -Ms. Allyson Parks

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

Image /page/4/Picture/1 description: The image shows the logo for Advanced Orthopaedic Solutions. The logo consists of the letters "AOS" in large, bold, black font. Below the letters, the words "ADVANCED ORTHOPAEDIC SOLUTIONS" are written in a smaller, sans-serif font.

# 7. INDICATIONS FOR USE STATEMENT

Special 510(k) Premarket Notification Indication for Use Statement AOS Short Clavicle Plates

510(k) Number (if known): K122623

Device Name: AOS Short Clavicle Plates

### Indications for Use:

The AOS Clavicle Plate System provides fixation for fractures, fusions, or osteotomies for the Clavicle.

AND/OR Prescription Use: X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use: (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Asht

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number k122623

---

**Source:** [https://fda-staging.innolitics.com/device/K122623](https://fda-staging.innolitics.com/device/K122623)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).