Real-world clinical samples were used to perform method comparison studies against a predicate device and to establish normal reference intervals for the assay.
Method comparison; Reference range establishment; Clinical field study
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
US field studies; Method comparison study
Patient samples (Normal & Abnormal); Sample Size: Total n > 600 across 3 sites; Number of Sites: 3
HemosIL RecombiPlasTin 2G
PT (sec), PT (INR), Fibrinogen correlation
Normal range study; Observational reference range study
Healthy subjects who were not coagulated; Sample Size: 199
Not applicable for this study
PT (sec) and Fibrinogen (mg/dL) reference intervals
Indications for Use
HemosIL ReadiPlasTin is an in vitro diagnostic thromboplastin reagent, based on recombinant human tissue factor, for the quantitative determination, in human citrated plasma, of Prothrombin Time (PT) and Fibrinogen, on the ACL TOP Family of analyzers. The product is intended to be used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Vitamin K Antagonist Therapy.
Device Story
HemosIL ReadiPlasTin is a liquid liposomal reagent containing recombinant human tissue factor (RTF), synthetic phospholipids, and calcium chloride. Used on ACL TOP Family analyzers, it initiates the extrinsic coagulation pathway in citrated plasma samples. The analyzer detects clot formation via absorbance or light scatter. The device quantifies PT and fibrinogen by relating clotting signals to a calibrator. Used in clinical laboratory settings by trained personnel. Output assists clinicians in evaluating coagulation status and monitoring patients on Vitamin K antagonist therapy.
Clinical Evidence
Bench testing only. Precision/reproducibility assessed per CLSI EP05-A2 (n=80/instrument/lot). Linearity evaluated per CLSI EP6-A (r≥0.95). Analytical specificity tested for interferences (Heparin, Hemoglobin, Triglycerides, Bilirubin). Method comparison (n=100+ samples) against predicate showed high correlation (r>0.99). Reference range established per CLSI C28-A3 (n=199).
Technological Characteristics
Liquid liposomal reagent; recombinant human tissue factor (RTF); synthetic phospholipid blend; calcium chloride; buffer; preservative. Operates on ACL TOP Family analyzers using absorbance/light scatter clot detection. No FDA performance standards established.
Indications for Use
Indicated for quantitative determination of Prothrombin Time (PT) and Fibrinogen in human citrated plasma for evaluation of extrinsic coagulation pathway and monitoring of Oral Vitamin K Antagonist Therapy. For prescription use only.
Regulatory Classification
Identification
A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).
{0}------------------------------------------------
K122584
MAR 2 4 2013
# 510K SUMMARY
# ASSAY ONLY TEMPLATE
#### A. 510(k) Number
# TBD
New Assay (ReadiPlasTin only)
Instrumentation Laboratory Co.
Absorbance clot detection
HemosIL ReadiPlasTin
Prothrombin Time (PT) and Fibrinogen
- Purpose for Submission B.
- C. Measurand
- D. Type of Test
- E. Applicant
- F. Proprietary Name
- G. Regulatory Information
| Product Name | HemosIL ReadiPlasTin | |
|----------------------|-------------------------|------------------|
| Regulation section: | 21 CFR 864.7750 | 21CFR 864.7340 |
| Classification: | Class II | Class II |
| Product code: | GJS | GIS |
| Common Name | Prothrombin Time Test | Fibrinogen Test |
| Classification name: | Test, Time, Prothrombin | Test, Fibrinogen |
| Panel: | Hematology | Hematology |
### H. Intended Use
- Intended use(s): 1.
HemosIL ReadiPlasTin is an in vitro diagnostic thromboplastin reagent, based on recombinant human tissue factor, for the guantitative determination, in human citrated plasma, of Prothrombin Time (PT) and Fibrinogen, on the ACL TOP Family of analyzers. The product is intended to be used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Vitamin K Antagonist Therapy.
- 2. Indication(s) for use:
Same as intended use(s).
- Special conditions for use statement(s): 3.
For in-vitro diagnostic use only. For prescription use.
- Special instrument requirements: 4.
ACL TOP Family analyzers
### I. Device Description
The thromboplastin reagent included in the ReadiPlasTin kit is a liquid liposomal preparation that contains recombinant human tissue factor (RTF) in a synthetic phospholipid blend combined with calcium chloride, buffer and a preservative. In the PT test, the addition of tissue thromboplastin to the patient plasma, in the presence of calcium ions, initiates the activation of the extrinsic pathway. This results in the conversion of fibrinogen to fibrin, with the formation of a solid gel. The fibrinogen is then quantitated by relating the absorbance to a calibrator.
ReadiPlasTin can be used with II's Normal control, Routine controls 1-3, Low abnormal control, high abnormal control, ISI calibrate and INR validate.
{1}------------------------------------------------
### J. Substantial Equivalence Information:
- Predicate device name(s): 1.
HemosIL RecombiPlasTin 2G K070005
.
- 2. Predicate 510(k) #: Comparison with predicate: 3.
Table of similarities:
| Characteristic | Predicate | Applicant |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | Instrumentation Laboratory (self) | Same |
| Device Name | HemosIL RecombiPlasTin 2G | HemosIL ReadiPlasTin |
| K# | K070005 | TBD |
| Intended Use | HemosIL RecombiPlasTin 2G is a<br>high sensitivity thromboplastin<br>reagent, which is based on<br>recombinant human tissue factor<br>(RTF), for the quantitative<br>determination, in human citrated<br>plasma, of Prothrombin Time<br>(PT) on IL Coagulation and<br>ELECTRA Systems and Fibrinogen<br>on IL Coagulation Systems only.<br>The product is used for the<br>evaluation of the extrinsic<br>coagulation pathway, and the<br>monitoring of Oral Anticoagulant<br>Therapy (OAT). | HemosIL ReadiPlasTin is an<br>in vitro diagnostic<br>thromboplastin reagent,<br>based on recombinant<br>human tissue factor (RTF),<br>for the quantitative<br>determination, in human<br>citrated plasma, of<br>Prothrombin Time (PT) and<br>Fibrinogen, on the ACL TOP<br>Family of analyzers. The<br>product is intended to be<br>used for the evaluation of<br>the extrinsic coagulation<br>pathway and the<br>monitoring of Oral Vitamin<br>K Antagonist Therapy. |
| Mode of | Prothrombin Measurement<br>In the PT test, the addition of the<br>tissue thromboplastin to the<br>patient sample, in the presence<br>of calcium, initiates the<br>activation of the extrinsic<br>pathway. This results in the<br>conversion of fibrinogen to fibrin,<br>with the formation of a solid gel.<br>PT derived Fibrinogen<br>Fibrinogen is quantitated (PT<br>based method) by relating the<br>absorbance or light scatter<br>during clotting to a calibrator. | Same |
| Operation | | |
| Sample Type | Citrated Plasma | Same |
# Table of differences
| Characteristic | Predicate | Applicant |
|----------------|---------------------------|----------------------|
| Device Name | HemosIL RecombiPlasTin 2G | HemosIL ReadiPlasTin |
| Delivery | Lyophilized | Liquid |
{2}------------------------------------------------
### K. Standard/Guidance Document Referenced (if applicable):
No FDA performance standard or FDA guidance has been established.
### L. Test Principle
The thromboplastin reagent included in the ReadiPlasTin kit, after mixing with the ReadiPlasTin Diluent, is a liposomal preparation that contains recombinant human tissue factor (RTF), re-lipidated in a synthetic phospholipid blend.
In the PT test, the addition of the tissue thromboplastin (ReadiPlasTin reagent) to the patient plasma in the presence of calcium ions initiates the activation of the extrinsic pathway. This results ultimately in the conversion of fibrin, with formation of a solid gel.
The Fibrinogen is quantitated (PT-based method) by relating the absorbance or lightscatter during clotting to a calibrator.
#### M. Performance Characteristics
# 1. Analytical performance:
- Precision/Reproducibility a.
Precision and Reproducibility was assessed utilizing 3 lots of reagent on 3 representative members of the ACL TOP Family (2 ACL TOP 700 and an ACL TOP 500 CTS) by 3 independent operators. Precision was evaluated in accordance with CLSI EP05-A2, for 20 days, with 2 runs per day and 2 replicates per run for each sample level (n=80/ instrument/ lot).
## Prothrombin Time (seconds)
| System | | HemosIL<br>ReadiPlasTin | | | Mean PT | | | | | |
|-------------------|--|-------------------------|-----|-----|-------------------|------|------|-------------|-----|-----|
| ACL TOP<br>Family | | Normal Control | | | 11.4 | 11.7 | 12.1 | | | |
| | | Low Abnormal | | | 21.6 | 21.6 | 22.1 | | | |
| | | High Abnormal | | | 36.2 | 35.5 | 34.8 | | | |
| | | | | | | | | | | |
| Controls | | Within Run (%CV) | | | Between Run (%CV) | | | Total (%CV) | | |
| Normal | | 0.6 | 0.6 | 0.6 | 0.1 | 0.1 | 0.1 | 1.2 | 1.4 | 1.1 |
| Low Abnormal | | 0.7 | 0.6 | 1.1 | 0.6 | 0.3 | 0.0 | 2.9 | 1.5 | 1.2 |
| High Abnormal | | 0.5 | 0.7 | 0.9 | 0.5 | 0.6 | 0.3 | 1.7 | 1.9 | 1.4 |
{3}------------------------------------------------
# Fibrinogen (mq/dL)
| | | Mean Fibrinogen | | |
|----------------|-------------------|-------------------|-------------------|--|
| System | Controls | Lot 1 Lot2 Lot3 | | |
| ACL TOP Family | Normal · | 297.5 284.6 287.9 | | |
| | Low Abnormal. | 121.8 114.9 117.2 | | |
| | Low Fibrinogen | 135.4 135.7 135.8 | | |
| | | | | |
| | Within Run (%CV) | Between Run (%CV) | Total (%CV) | |
| Controls | Lot1 Lot2 Lot3 | Lot1 Lot2 Lot3 | Lot1 Lot2 Lot3 | |
| Normal | 1.2<br>1.1<br>1.2 | 1.4<br>1.4<br>1.8 | 2.0<br>2.2<br>2.0 | |
| Low Abnormal | 1.5<br>1.4<br>1.8 | 1.6<br>1.0<br>1.3 | 2.2<br>2.4<br>1.8 | |
| Low Fibrinogen | 2.9<br>3.2<br>3.4 | 0.0<br>0.0<br>0.0 | 3.5<br>3.1<br>3.5 | |
| | | | | |
b. Linearity/assay reportable range:
### Fibrinogen:
A fibrinogen linearity study was performed in accordance with CLSI EP6-A. Results met the acceptance criteria of r≥0.95, supporting the device's insert claim of a Fibrinogen Analytical Range of 60-700mg/dL.
- Traceability, Stability, Expected values (controls, calibrators, or methods): C.
### Traceability
Each batch of ReadiPlasTin reagent is calibrated against an ISI certified House Standard which in turn is calibrated against the corresponding international reference preparation according to the WHO recommendations.
The reported INR Reference values were determined over multiple runs on IL coagulation systems running a specific lots of PT reagent against a House Standard ISI Calibrate, which is traceable to the current WHO International Standard identified in the INR Reference table of the product insert and consensus values from > 200 laboratories.
### Stability
Unopened reagent and diluent are stable until the expiration date shown on the vial, when stored at 2-8°C. Once prepared for use, the reagent is stable for 10 days at 2-8°C in closed original vial, or for 10 days at 15°C on the ACL TOP Family in the original vial with no stirring. For optimal stability remove the reagent from the system and store it closed at 2-8°C in the original vial. Do not freeze.
- d. Detection limit:
### Extrinsic Factor Linearity:
Extrinsic factor linearity studies were performed on factors II, V, VII and X. Results met the acceptance criteria at r≥0.95.
### Analytical Specificity: e.
Testing for interfering substances was performed according to CLSI EP7-A. The assay was insensitive to Heparin, Hemoglobin, Triglycerides, and Bilirubin up to the following levels:
| | Heparin | | | | |
|------------|-----------|-----------|------------|---------------|-----------|
| Parameter | UFH | LMW | Hemoglobin | Triglycerides | Bilirubin |
| PT (sec) | 1.0 IU/mL | 1.4 IU/mL | 500 mg/dL | 1000 mg/dL | 50 mg/dL |
| Fibrinogen | 1.5 IU/mL | 1.7 IU/mL | 500 mg/dL | 600 mg/dL | 50 mg/dL |
{4}------------------------------------------------
### f. Assay cut-off: Not Applicable
### 2. Comparison studies:
#### Method Comparison with predicate device: a.
An in-house method comparison was performed in accordance with EP09-A2 on 100+ samples (Normal & Abnormal), comparing HemosIL ReadiPlasTin on representative members of the ACL TOP Family (ACL TOP 700 & ACL TOP 500 CTS), against HemosIL RecombiPlasTin 2G, with the following result(s)10.
| Assay | Instrument | n | Slope (95% CI) | Intercept (95% CI) | r |
|------------|------------|-----|---------------------|----------------------|--------|
| 'PT(sec) | ACL TOP | 230 | 0.913 (0.907-0.919) | 0.948 (0.835-1.062) | 0.9989 |
| PT (INR) | 700 | 65 | 0.979 (0.961-0.996) | 0.026 (-0.020-0.072) | 0.9975 |
| Fibrinogen | | 241 | 0.946 (0.935-0.958) | -4.3 (-9.1-0.5) | 0.9955 |
| PT(sec) | ACL | 238 | 0.903 (0.896-0.910) | 1.091(0.945-1.236) | 0.9981 |
| PT (INR) | TOP 500 | 65 | 0.937 (0.914-0.959) | 0.016 (0.100-0.232) | 0.9955 |
| Fibrinogen | CTS | 242 | 0.975 (0.962-0.988) | -7.6 (-12.9--2.2) | 0.9945 |
In US field studies, the following data were obtained using a specific lot of ReadiPlasTin on the ACL TOP:
| Parameter | Site | n | Slope (95% CI) | Intercept (95% CI) | r |
|------------|--------|-----|---------------------|---------------------|--------|
| PT(sec) | Site 1 | 254 | 0.909 (0.900-0.918) | 0.45 (0.24-0.66) | 0.9966 |
| | Site 2 | 312 | 0.970 (0.962-0.978) | 0.51 (0.34-0.68) | 0.9972 |
| | Site 3 | 135 | 0.889 (0.874-0.903) | 1.76 (1.40-2.12) | 0.9955 |
| PT (INR) | Site 1 | 76 | 0.938 (0.916-0.961) | 0.138 (0.079-0.197) | 0.9947 |
| | Site 2 | 89 | 0.928 (0.892-0.963) | 0.119 (0.033-0.205) | 0.9838 |
| | Site 3 | 71 | 0.914 (0.891-0.937) | 0.120 (0.060-0.181) | 0.9945 |
| Fibrinogen | Site 1 | 237 | 0.947 (0.936-0.957) | 4.5 (-0.3-9.3) | 0.9964 |
| | Site 2 | 284 | 0.971(0.961-0.982) | 0.56 (-4.4-5.5) | 0.9956 |
| | Site 3 | 129 | 0.998 (0.978-1.019) | -4.2 (-13.6 to 5.2) | 0.9931 |
- b. Matrix comparison:
A fresh versus frozen sample study was performed (n=245) on both normal (n=140) and abnormal (n=105) patient samples. The results demonstrated comparable performance on fresh and once-thawed samples.
### 3. Clinical studies:
- Clinical Sensitivity: NA a.
- Clinical Specificity: NA b.
- Other clinical supportive data (when a. and b. are not applicable): NA C.
- Clinical cut-off: NA 4.
{5}------------------------------------------------
#### 5. Expected values/Reference range *:
A normal range study (n=199), was performed in accordance with CLSI C28-A3, testing healthy subjects who were not coagulated, with ReadiPlasTin reagent, on a representative member of the ACL TOP Family (ACL TOP 700). The reference range was established as:
| Assay | NR Reference Interval* | Lower Limit (95% CI) | Upper Limit (95% CI) |
|--------------------|------------------------|----------------------|----------------------|
| PT (sec) | | 10.2 (9.7-10.4) | 12.9 (12.7-14.1) |
| Fibrinogen (mg/dL) | | 282 (274-332) | 553 (531-562) |
*These results were obtained using a specific lot of reagent. Due to many variables which may affect clotting times, each laboratory should verify its own normal range.
### N. Proposed Labeling
The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.
#### 0. Conclusion
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
### P. Administrative Information
Applicant Contact Information
| Name of applicant: | Instrumentation Laboratory Co. |
|--------------------|------------------------------------------------------|
| Mailing address: | 180 Hartwell Road<br>Bedford, MA 01730, USA |
| Phone #: | 781-861-4350 |
| Fax #: | 781-861-4207 |
| E-mail address: | jemery@ilww.com |
| Contact: | Jacqueline Emery, BSEE<br>Regulatory Affairs Manager |
Date Prepared
August 21, 2012
{6}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, which is a traditional symbol of medicine.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 24, 2013
Instrumentation Laboratory Co. c/o Ms. Jacqueline Emery Regulatory Affairs Manager 180 Hartwell Road Bedford, MA 01730
Re: k122584
Trade/Device Name: HemosIL® ReadiPlasTin Regulation Number: 21 CFR § 864.7750 Regulation Name: Prothrombin time test Regulatory Class: Class II Product Code: GJS, GIS Dated: March 15, 2013 Received: March 18, 2013
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{7}------------------------------------------------
Page 2 - Ms. Jacqueline Emery
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Leonthena R. Carrington -S
for
Maria Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): _K122584
HemosIL ReadiPlasTin Device Name:
Indications for Use:
HemosIL ReadiPlasTin is an in vitro diagnostic thromboplastin reagent, based on recombinant human tissue factor, for the quantitative determination, in human citrated plasma, of Prothrombin Time (PT) and Fibrinogen, on the ACL TOP® Family of analyzers. The product is intended to be used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Vitamin K Antagonist Therapy.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Leonthena R. Carrington -S
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k)_K122584
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.