DIGITAL ULTRASONIC DIAGNOSTIC IMAGING SYSTEM, MODEL D3 AND D6
K122574 · Edan Instruments, Inc. · IYO · Oct 12, 2012 · Radiology
Device Facts
Record ID
K122574
Device Name
DIGITAL ULTRASONIC DIAGNOSTIC IMAGING SYSTEM, MODEL D3 AND D6
Applicant
Edan Instruments, Inc.
Product Code
IYO · Radiology
Decision Date
Oct 12, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The D3 and D6 Digital Ultrasonic Diagnostic Imaging System are applicable for adult or children ultrasound evaluation in hospitals, clinics, gynecology rooms, obstetrics rooms, examination rooms, intensive care units, and emergency rooms The D3 and D6 are intended for use by or on the order of a physician or similarly qualified health care professional for ultrasound evaluation of Fetus; Abdomen; Pediatrics; Small Organ; Neonatal head; Cardiology; Peripheral Vessel; Musculo-skeleton (both Conventional and Superficial); Urology (including prostate); Transrecta and Transvagina.
Device Story
Portable digital ultrasonic diagnostic imaging system (Models D3, D6) for clinical diagnostic examination. Device emits ultrasound waves into body tissue via array transducers (linear, convex, microconvex, transrectal, transvaginal; 2.0-10.0 MHz) and processes returned echoes. Displays images in B-Mode, 2B-Mode, 4B-Mode, M-Mode, and B+M Mode. Operated by physicians or qualified healthcare professionals in hospitals, clinics, ICUs, and emergency rooms. Provides real-time visualization of internal structures to assist in clinical diagnosis and evaluation. Supports needle guide bracket kits for interventional procedures. No contrast agents used.
Clinical Evidence
No clinical data. Bench testing performed including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), acoustic output (per 2008 FDA guidance), and biocompatibility (ISO 10993-1, 5, 10).
Technological Characteristics
Portable ultrasound system with main unit and array transducers (linear, convex, microconvex). Frequency range 2.0-10.0 MHz. Display modes: B, 2B, 4B, M, B+M. Software-controlled. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, and ISO 10993 standards.
Indications for Use
Indicated for adult and pediatric ultrasound evaluation of fetus, abdomen, small organs (thyroid, galactophore, prostate), neonatal head, cardiac, peripheral vessel, musculoskeletal (conventional/superficial), urology, transrectal, and transvaginal applications. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Digital Ultrasonic Diagnostic Imaging System (K091680)
Submission Summary (Full Text)
{0}------------------------------------------------
Digital Ultrasonic Diagnostic Imaging System
510K Submission
OCT 12 2012
# Section 5- 510(k) Summary of Safety and Effectiveness
Prepared in accordance with the requirements of 21 CFR Part 807.92
1. Submitter:
Edan Instruments, Inc.
3/F-B, Nanshan Medical Equipments Park, Nanhai Rd 1019#, Shekou, Nanshan Shenzhen, 518067 P.R. China
Telephone: 0086-755-6856469
Fax: 0086-755-26882223
Contact Person: Randy Jiang Prepare date: Feb 21, 2012
2. Device name and Device name: Digital Ultrasonic Diagnostic Imaging System classification:
### Models D3 and D6
Classification: 892.1560 System, Imaging, Pulsed echo, Ultrasonic
#### Product code: IYO
892.1570 Transducer, Ultrasonic, Diagnostic
#### Product code: ITX
Regulatory Class: Class II
3. Predicate Device:
Digital Ultrasonic Diagnostic Imaging System. K091680 Manufacturer: Edan Instruments, Inc.
4. Device Description: D3 and D6 Digital Ultrasound Diagnostic Imaging System is a portable digital ultrasonic diagnostic system applied in ultrasound diagnostic examination of abdominal, obstetrical, small parts, gynecological, orthopedic, cardiac, and urological applications.
It is designed to produce ultrasound waves into body tissue and to present the returned echo information on the monitor, the resulting information is displayed in five display modes: B-Mode, 2B-Mode, 4B-Mode, M-Mode, B+M Mode. This system controlled by software is a Track 1 device that employs an array of probes that include linear array, convex linear array, microconvex linear array, transrectal and transvaginal with a frequency range of approximately 2.0MHz-10.0MHz.
The system consists of a main unit, transducers and other accessories.
{1}------------------------------------------------
#### 5. Intended Use:
The D3 and D6 Digital Ultrasonic Diagnostic Imaging System are applicable for adult or children ultrasound evaluation in hospitals, clinics, gynecology rooms, obstetrics rooms, examination rooms, intensive care units, and emergency rooms The D3 and D6 are intended for use by or on the order of a physician or similarly qualified health care professional for ultrasound evaluation of Fetus; Abdomen; Pediatrics; Small Organ; Neonatal head; Cardiology; Peripheral Vessel; Musculo-skeleton (both Conventional and Superficial); Urology (including prostate); Transrecta and Transvagina.
#### 6. Effectiveness and Safety Considerations:
#### Clinical test:
Clinical testing is not required.
#### Non-clinical test:
The following safety standards are conducted on the subject device:
- 1. IEC 60601-1 Electrical Safety
- 2. IEC 60601-1-2 Electromagnetic Compatibility
- 3. Acoustic output testing as per the guideline "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" dated September 9, 2008.
- 4. UD-2, IEC 60601-2-37
- 5. ISO 10993-1, ISO 10993-5 and ISO 10993-10
### 7.Comparison to the predicate device
Comparison to the predicate device, the subject device has the similar technology characteristics and has the same intended use, same design principle, same electrical classification and same accuracy. The different between the subject device and predicate device primarily includes physical specifications, display type and display mode, all the above differences do not affect the usage, safety and effectiveness, and no new question is raised regarding the safety and effectiveness.
## 8. Substantially Equivalent Determination
Verification and validation testing was conducted on the subject device. This premarket notification submission demonstrates that D3 and D6 Digital Ultrasonic Diagnostic Imaging System is substantially equivalent to the predicate device.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is surrounded by a circular inscription that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
OCT 12 2012
Edan Instruments, Inc. % Mr. Ned Devine Senior Staff Engineer/FDA Office Coordinator Underwriters Laboratories, Inc. 333 Pfingsten Road NORTHBROOK IL 60062
. Re: K122574
Trade/Device Name: Digital Ultrasonic Diagnostic Imaging System, Models D3 and D6 Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO and ITX Dated: September 21, 2012 Received: September 26, 2012
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Digital Ultrasonic Diagnostic Imaging System, Models D3 and D6, as described in your premarket notification:
Transducer Model Number
| D3 | D6 |
|-------------|--------|
| C361-1/C341 | C363-1 |
| C321-1 | C362 |
| L741 | C343-1 |
| E741 | C321 |
| E611-1 | L743 |
| | E743 |
| | E613 |
{3}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.
Sincerely Yours,
Zumhof D O'Hern Lier
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
## Section 6- Indications for Use
510(k) Number (if known):
Device Name: Digital Ultrasonic Diagnostic Imaging System Models D3 and D6
The D3 and D6 Digital Ultrasonic Diagnostic Imaging System are applicable for adult or children ultrasound evaluation in hospitals, clinics, gynecology rooms, obstetrics rooms, examination rooms, intensive care units, and emergency rooms The D3 and D6 are intended for use by or on the order of a physician or similarly qualified health care professional for ultrasound evaluation of Fetus; Abdomen; Pediatrics; Small Organ; Neonatal head; Cardiology; Peripheral Vessel; Musculo-skeleton (both Conventional and Superficial); Urology (including prostate); Transrecta and Transvagina.
Prescription Use (21 CFR Part 801 Subpart D)
Or Over the Counter Use
(21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
(Division Sign-Off)
Division of Radiotogical Devices Office of In Vitro Diagnostic Device Evaluation and Safety
Ki22574
510K.
{5}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
## D3 Digital Ultrasonic Diagnostic Imaging System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|---------------|--------------------|------------------|
| | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | N | N | | | | N | Note 1,Noter2 |
| Abdominal | N | N | | | | N | Note 1,Noter2 |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | N | N | | | | N | Note 1,Noter2 |
| Small Organ (Specify) | N | N | | | | N | Note 1,Noter2 |
| Neonatal Cephalic | N | N | | | | N | Note 1,Noter2 |
| Adult Cephalic | | | | | | | |
| Transrectal | N | N | | | | N | Note 1,Noter2 |
| Transvaginal | N | N | | | | N | Note 1,Noter2 |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | N | N | | | | N | Note 1,Noter2 |
| Musculo-skeletal (Superficial) | N | N | | | | N | Note 1,Noter2 |
| Intravascular | | | | | | | |
| Other (Gynecology) | N | N | | | | N | |
| Cardiac | N | N | | | | N | Note 1,Noter2 |
| Intravascular | | | | | | | |
| Peripheral vascular | N | N | | | | N | Note 1,Noter2 |
| Other (Urology) | N | N | | | | N | Note 1,Noter2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mull D'the
(Division Sign-Off)
Division of Radiological Devices
Office In Vitro Diagnostic Device Evaluation and Safety
510K K122574/
{6}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
### D3 with C361-1 / C341 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|---------------|--------------------|---------------------|
| Clinical Application | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other*<br>(Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | P | P | | | | P | Note 1,Noter2 |
| Abdominal | P | P | | | | P | Note 1,Noter2 |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | P | P | | | | P | |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Other (Gynecology) | P | P | | | | P | Note 1,Noter2 |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | P | P | | | | P | Note 1,Noter2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K122574
{7}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
## D3 with C321-1 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | P | P | | | | P | Note 1,Noter2 |
| Abdominal | P | P | | | | P | Note 1,Noter2 |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | P | P | | | | P | Note 1,Noter2 |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Other (Gynecology) | P | P | | | | P | Note 1,Noter2 |
| Cardiac | P | P | | | | P | Note 1,Noter2 |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | P | P | | | | P | Note 1,Noter2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
510K -
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Office of In Vitro Diagnostic Device Evaluation and Safety
Division of Radiological Devices
(Division Sign-Off)
6-4
{8}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
## D3 with L741 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | Other*<br>(Specify) |
|---------------------------------|--|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| | | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | |
| Ophthalmic | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | |
| Abdominal | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | |
| Intra-operative (Neurological) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | | | | | | | | |
| Small Organ (Specify) | | P | P | | | | P | Note 1,Noter2 |
| Neonatal Cephalic | | P | P | | | | P | Note 1,Noter2 |
| Adult Cephalic | | | | | | | | |
| Transrectal | | | | | | | | |
| Transvaginal | | | | | | | | |
| Transurethral | | | | | | | | |
| Musculo-skeletal (Conventional) | | P | P | | | | P | Note 1,Noter2 |
| Musculo-skeletal (Superficial) | | P | P | | | | P | Note 1,Noter2 |
| Intravascular | | | | | | | | |
| Other (Gynecology) | | | | | | | | |
| Cardiac | | | | | | | | |
| Intravascular | | | | | | | | |
| Peripheral vascular | | P | P | | | | P | Note 1,Noter2 |
| Other (Urology) | | P | P | | | | P | Note 1,Noter2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit
Small organ includes galactophore, thyroid gland, prostate
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Devices
Office of In Vitro Diagnostic Devices
Office of In Vitro Diagnostic Devices
510k K122574
{9}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
## D3 with E741 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | | | | | | | |
| Abdominal | | | | | | | |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | | | | | | | |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | P | P | | | | P | Note 1,Noter2 |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Other (Gynecology) | | | | | | | |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | P | P | | | | P | Note 1,Noter2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
*(Division Sign-Off)*
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K K12257X/
ર-ઉ
{10}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
## D3 with E611-1 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
|---------------------------------|---|---|-----|-----|---------------|--------------------|------------------|
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | P | P | | | | P | Note 1,Noter2 |
| Abdominal | | | | | | | |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | | | | | | | |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | P | P | | | | P | Note 1,Noter2 |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Other (Gynecology) | P | P | | | | P | Note 1,Noter2 |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | P | P | | | | P | Note 1,Noter2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
***(Division Sign Off)***
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K K122574/
{11}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
## D6 Digital Ultrasonic Diagnostic Imaging System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | N | N | | | | N | Note 1,Note2 |
| Abdominal | N | N | | | | N | Note 1,Note2 |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | N | N | | | | N | Note 1,Note2 |
| Small Organ (Specify) | N | N | | | | N | Note 1,Note2 |
| Neonatal Cephalic | N | N | | | | N | Note 1,Note2 |
| Adult Cephalic | | | | | | | |
| Transrectal | N | N | | | | N | Note 1,Note2 |
| Transvaginal | N | N | | | | N | Note 1,Note2 |
| Transurethral | | | | | | | |
| Musculo-skeletal<br>(Conventional) | N | N | | | | N | Note 1,Note2 |
| Musculo-skeletal (Superficial) | N | N | | | | N | Note 1,Note2 |
| Intravascular | | | | | | | |
| Other (Gynecology) | | | | | | | |
| Cardiac | N | N | | | | N | Note 1,Note2 |
| Intravascular | | | | | | | |
| Peripheral vascular | N | N | | | | N | Note 1,Note2 |
| Other (Urology) | N | N | | | | N | Note 1,Note2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
*(Division Sign-Off)*
Division of Radiological Devices Office of In Vitro Diagnostic Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K122574/
{12}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
### D6 with C363-1 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | Other* (Specify) |
|---------------------------------|-------------------|---|-----|-----|---------------|---------------------|------------------|
| | B | M | PWD | CWD | Color Doppler | Combined (Specify). | |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | P | P | | | | P | Note 1,Noter2 |
| Abdominal | P | P | | | | P | Note 1,Noter2 |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | P | P | | | | P | Note 1,Noter2 |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Other (Gynecology) | P | P | | | | P | Note 1,Noter2 |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | P | P | | | | P | Note 1,Noter2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
510K.
Prescription Use (Per 21 CFR 801.109)
Mmhnd DK
(Division Sign-Off)
Division of Radiological Devices
Vitro Diagnostic Devices Devices
Vitro Diagnostic Devices Custom Custom Office of In Vitro Diagnostic Devices
Office of In Vitro Diagnostic Devices
Constitution Diagnostic Devices Evelulation and Safety
Constitution of One Commend Safety
6-9
{13}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
## D6 with C362 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | Other*<br>(Specify) |
|---------------------------------|-------------------|---|-----|-----|---------------|-----------------------|---------------------|
| | B | M | PWD | CWD | Color Doppler | Combined<br>(Specify) | |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | P | P | | | | P | Note 1,Note2 |
| Abdominal | P | P | | | | P | Note 1,Note2 |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | P | P | | | | P | Note 1,Note2 |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Other (Gynecology) | P | P | | | | P | Note 1,Note2 |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | P | P | | | | P | Note 1,Note2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K122574
{14}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
### D6 withC343-1 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | P | P | | | | P | Note 1,Noter2 |
| Abdominal | P | P | | | | P | Note 1,Noter2 |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | P | P | | | | P | Note 1,Noter2 |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Other (Gynecology) | P | P | | | | P | Note 1,Noter2 |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | P | P | | | | P | Note 1,Noter2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
*(Division Sign-Off)*
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K. K.122574/
{15}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
## D6 with C321 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|---------------|--------------------|------------------|
| | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | P | P | | | | P | Note 1,Noter2 |
| Abdominal | P | P | | | | P | Note 1,Noter2 |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | P | P | | | | P | Note 1,Noter2 |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Other (Gynecology) | P | P | | | | P | Note 1,Noter2 |
| Cardiac | P | P | | | | P | Note 1,Noter2 |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | P | P | | | | P | Note 1,Noter2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
510K
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
*(Division Sign-Off)*
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
{16}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
## D6 with L743 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | | | | | | | |
| Abdominal | | | | | | | |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | | | | | | | |
| Small Organ (Specify) | P | P | | | | P | Note 1,Noter2 |
| Neonatal Cephalic | P | P | | | | P | Note 1,Noter2 |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | P | P | | | | P | Note 1,Noter2 |
| Musculo-skeletal (Superficial) | P | P | | | | P | Note 1,Noter2 |
| Intravascular | | | | | | | |
| Other (Gynecology) | | | | | | | |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | P | P | | | | P | Note 1,Noter2 |
| Other (Urology) | P | P | | | | P | Note 1,Noter2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit
Small Organ includes galactophore, thyroid gland, prostate
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
510K.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
MundlADR
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
K122574
{17}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
## D6 with E743 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | | | | | | | |
| Abdominal | | | | | | | |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | | | | | | | |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | P | P | | | | P | Note 1,Noter2 |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Other (Gynecology) | | | | | | | |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | P | P | | | | P | Note 1,Noter2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent.
Note 2: Needle guide bracket kit
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
---
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K K12252/
{18}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
### D6 with E613 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|---------------|--------------------|------------------|
| | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | P | P | | | | P | Note 1,Noter2 |
| Abdominal | | | | | | | |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | | | | | | | |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | P | P | | | | P | Note 1,Noter2 |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Other (Gynecology) | P | P | | | | P | Note 1,Noter2 |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | P | P | | | | P | Note 1,Noter2 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note 1: This feature does not use contrast agent ..
Note 2: Needle guide bracket kit.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Michael D. O'Brien
(Division Sign-Off)
Division of Radiological Devices Office of In Vitre Diagnostic Devices
Office of In Vitce Diagnostic Device Evaluation and Safety
510K K122574
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.