DENTO TEMP

K122549 · Itena Clinical · EMA · Apr 26, 2013 · Dental

Device Facts

Record IDK122549
Device NameDENTO TEMP
ApplicantItena Clinical
Product CodeEMA · Dental
Decision DateApr 26, 2013
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.3275
Device ClassClass 2
AttributesTherapeutic

Indications for Use

It is intended for: 1) quick temporary cementation of temporary crowns and bridges; 2) rebasing of crowns; and 3) permanent cementation of implant-retained crowns and bridges while maintaining retrievability, when desired. DentoTemp is fully compatible with both acrylic and composite (bis-acryl) temporary materials and as a result, is indicated on both natural and metal implants abutments.

Device Story

DentoTemp is a long-term acrylic-urethane polymer-based dental cement; non-eugenol and zinc-oxide free. Features a 2-stage polymerization process; designed to protect temporary crowns and avoid interference with permanent cements. Used by dental professionals in clinical settings for luting temporary restorations and implant-retained crowns. Provides retrievability for implant-retained restorations. Benefits include compatibility with various temporary materials (acrylic/bis-acryl) and suitability for both natural and metal implant abutments.

Clinical Evidence

Bench testing only. Evidence includes physical property testing (working time, setting time, solubility, film thickness, linear shrinkage) and biocompatibility assessment. Compliance with ISO 4049 (Polymer-based filling, restorative and luting materials) and FDA guidance for Dental Cements demonstrated.

Technological Characteristics

Acrylic-urethane polymer-based dental cement. Non-eugenol, zinc-oxide free. 2-stage polymerization. Physical properties: 90-120s working time, 2.5-3 min setting time at 37°C, 15 µg/mm³ film thickness, 2.5% linear shrinkage. Complies with ISO 4049.

Indications for Use

Indicated for temporary cementation of crowns/bridges, rebasing of crowns, and permanent cementation of implant-retained crowns/bridges requiring retrievability. Compatible with acrylic and composite temporary materials; indicated for use on natural and metal implant abutments.

Regulatory Classification

Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K122549 # Abbreviated 510(k) Summary APR 2 6 2013 - 1. Name/Address of Submitter: Itena Clinical 83 avenue Foch 75116 Paris FRANCE 2. Contact Person: Louis-Paul Marin Co-President, BCF Certification inc. Phone: (514) 397-5546 (514) 397-8515 Fax: Email: lpmarin@bcf.ca - Date Summary Prepared: 3. July 30, 2012 - Device Names: DentoTemp 4. - Device Classification: II ਹੈ. - Common name: Dental Cement 6. - Classification Product Code: EMA 7. - 8. Predicate Device: | DentoTemp | Premier Implant Cement | |-----------|------------------------| | | K033309 | ### Device Description: 9. DentoTemp: DentoTemp is a dental long term acrylic-urethane polymer based temporary cement. It does not contain eugenol or zinc oxyde. It is a unique 2-stage polymerization composition which does not harm temporary crowns and does not interfere with permanent cements. The technological characteristics for this subject device are as follows: | Technological Characteristics | Subject Device | Premier Implant<br>Cement | |-------------------------------|----------------|---------------------------| | | | K033309 | {1}------------------------------------------------ | Type | Long term temporary Cement | Long term temporary Cement | |-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------| | Intended Use | It is intended for:<br>1) quick temporary cementation<br>of temporary crowns and<br>bridges;<br>2) rebasing of crowns; and<br>3) permanent cementation of<br>implant-retained crowns and<br>bridges while maintaining<br>retrievability, when desired. | is a non-eugenol temporary<br>cement for luting implant-<br>retained crowns. | | Working time (sec) | 90-120 | 90 | | Solubility (µg/mm³) | Low | Low | | Setting time at 37 °C (min) | 2.5-3 | 4-5 | | Film thickness (µg/mm³) | 15 | 15 | | Shelf life (Years) | 2 | 2 | | Linear Shrinkage (%) | 2.5 | 2.5 | #### Statement of Intended Use: 10. ## Device Name: DentoTemp The device is intended for: 1) quick temporary cementation of temporary crowns and bridges; 2) rebasing of crowns: and 3) permanent cementation of implant-retained crowns and bridges while maintaining retrievability, when desired. DentoTemp is fully compatible with both acrylic and composite (bis-acryl) temporary materials and as a result, is indicated on both natural and metal implants abutments. Brief Description of Clinical and Non-clinical Testing: This submission is an 11. Abbreviated 510(k) as described in FDA's guidance document entitled "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications". In support of this, Itena Clinical has provided information to demonstrate conformity with FDA's guidance document entitled Dental Cements - Premarket Notification, August 1998 and ISO 4049 -- Dentistry - Polymer-based filling, restorative and luting materials. {2}------------------------------------------------ Conclusion Drawn: Based on their indications for use, technological characteristics and comparison to predicate devices, the subject device has been shown to be safe and effective for their intended use. Further, combined with biocompatibility testing and based on a comparison of intended use and indications for use, physical properties and composition, Itena Clinical concludes that the subject device is substantially equivalent to the predicate device (Premier Implant Cement K033309). {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 26, 2013 Itena Clinical C/O Mr. Louis-Paul Marin Co-President BCF Certification, Incorporated 500 Boul Cartier West Laval, Canada H7V 5B7 Re: K122549 Trade/Device Name: DentoTemp Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: March 27, 2013 Received: March 29, 2013 ## Dear Mr. Marin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Susan Runner, DDSMA for Digitally signed by Mary S. Runneг -5 DN: C=US, OU.S. Government, ou=HHS, ou-FDA, ou People, cn=Mary S. Runner - 0092342.19200300.100.1.1-1300087950 Date: 2013.04.26 09:15:02-04'00' Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 122549 ## Indication for Use ## Device Name: DentoTemp Indication for Use: It is intended for: 1) quick temporary cementation of temporary crowns and bridges; 2) rebasing of crowns; and 3) permanent cementation of implant-retained crowns and bridges while maintaining retrievability, when desired. DentoTemp is fully compatible with both acrylic and composite (bis-acryl) temporary materials and as a result, is indicated on both natural and metal implants abutments. ## Concurrence of CDRH Office of Device Evaluation Prescription Use _ X_ OR Over-the-counter Use __ (per 21CFR 801.109) Susan Runner, DDS, MA Digitally signed by Mary S. Runner -S5 CN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Mary S. Runner 0.9.2342.19200300.100.1.1=1300087950 Date: 2013.04.26 09:14:09 -04'00' on Bign-Om) on of Anesthesiology, General Hospital ction Control, Dental Devices **510(k) Number:** K
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...