The Surefire® High Flow Angiographic Catheter is intended for use where angiographic diagnosis is indicated.
Device Story
Surefire High Flow Angiographic Catheter (SAC) is a single-lumen, fixed-length guide catheter used to facilitate advancement of a microcatheter to a target vessel. Device features a reinforced proximal section for insertion, barium sulfate-filled extrusion for fluoroscopic visibility, and a braided tungsten-filled shaped tip for visual feedback and atraumatic tracking. Compatible with 0.038" guidewires, Luer lock infusion syringes, and rotating hemostatic valves. Supplied sterile with a peel-away introducer for single-patient use. Operated manually by clinicians in a clinical setting under fluoroscopic guidance. Provides visual feedback of catheter location; assists in delivering microcatheters to target sites for diagnostic angiography.
Clinical Evidence
Bench testing only. Testing included kink radius, trackability, pull strength, high-pressure injection, infusion agent compatibility, package integrity, device corrosion, and shape retention. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, irritation, toxicity, pyrogenicity, hemolysis, coagulation, complement system, and thrombogenicity) and USP 788 (particulate).
Technological Characteristics
Single-lumen, fixed-length intravascular catheter. Materials include barium sulfate-filled extrusion and braided tungsten-filled tip. Compatible with 0.038" guidewires. Sterile, single-use. Complies with ISO 10555-1 and ISO 10555-2. Biocompatibility per ISO 10993. Manual operation.
Indications for Use
Indicated for use in patients requiring angiographic diagnosis. Prescription use only.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
K122506
Surefire® High Flow Angiographic Catheter Premarket Notification Traditional 510(k) Submission
ーー ー.
Section 5: 510(k) Summary 16 July 2012
| Owner/Manufacturer: | Owner<br>Surefire Medical, Inc.<br>8601 Turnpike Dr.<br>Suite 206<br>Westminster, CO 80031 | Manufacturer<br>Surefire Medical, Inc.<br>12415 SW136 Avenue<br>Unit 3<br>Miami, FL 33186 | SEP 2 0 2012 |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|--------------|
| Contact Person: | Mario Arbesu<br>Director, Quality Assurance and Regulatory Affairs<br>305.378.2651 | | |
| Date of Summary<br>Preparation: | 16 July 2012 | | |
| Trade Name: | Surefire® High Flow Angiographic Catheter | | |
| Common Name: | Intravascular Catheter | | |
| Classification Name: | Intravascular Diagnostic Catheter | | |
| Classification: | Class II | | |
| Classification Regulation: | 21 CRF Part 870.1200 - Diagnostic intravascular catheter. | | |
| Product Code: | DQO | | |
| Intended Use: | The Surefire® High Flow Angiographic Catheter is intended for<br>use where angiographic diagnosis is indicated. | | |
| Device Description: | The Surefire High Flow Angiographic Catheter (SAC) is<br>used to facilitate advancement of the Surefire High Flow<br>Microcatheter to the target vessel. The SAC is a single<br>lumen, fixed length guide catheter with a Luer Lock Hub. It<br>is compatible with standard 0.038" guide wires, Luer lock<br>infusion syringes, and rotating hemostatic valves (RHVs).<br>The angiographic catheter has an approximate length of 65<br>cm (usable length). A reinforced proximal section allows<br>for ease of insertion and the barium sulfate filled extrusion<br>provides clear fluoroscopic images of the shaft. A braided<br>tungsten filled shaped tip provides visual feedback for the<br>location of the guide catheter under fluoroscopy. The distal<br>tip is rounded for atraumatic tracking. A shaped tip will<br>assist to deliver the Surefire High Flow Angiographic<br>Catheter to the desired target site. A peel-away Introducer<br>is supplied with the SAC to insert the shaped tip into a<br>catheter sheath introducer. The system is provided sterile<br>(FO) for single patient use. The Angiographic Catheter is | | |
12
{1}------------------------------------------------
Surefire® High Flow Angiographic Catheter Premarket Notification Traditional 510(k) Submission
Section 5: 510(k) Summary 16 July 2012
packaged in sealed sterile protective pouches and product boxes.
## Principals of Operation/
Technology:
The Surefire® High Flow Angiographic Catheter is operated manually.
## Performance Testing & Verification Testing
- · Kink Radius Testing
- · Trackability Testing
- · Pull Strength Testing
- · High Pressure Injection Testing
- · Infusion Agent Compatibility Testing
- · Package Integrity (Pouch Bubble) Testing
- · Device Corrosion Testing
# Biocompatibility Testing
- · Cytotoxicity Tested in accordance with ISO 10993-5
- · Sensitization Tested in accordance with ISO 10993-10
- · Intra-cutancous irritation Tested in accordance with ISO 10993-10
- · Toxicity Tested in accordance with ISO 10993-11
- · Pyrogenicity Tested in accordance with USP General Chapter < 151> Pyrogen Test recommended in ISO 10993-11
- · Hemolysis Tested in accordance with ASTM F756 and ISO 10993-4
- · Coagulation ~ Tested in accordance with ASTM F2382
- · Particulate Tested in accordance with USP 788
- · Complement System Testing was performed in accordance with ISO 10993-4
- · Thrombogenicity Testing Testing was performed
Performance/Safety: A risk/hazard analysis was conducted according to EN ISO 14971
(Medical Devices-Application of Risk management to medical devices). Performance characteristics for this indication for use were identified which included a review of both ISO 10555-1 (Intravascular catheters -Sterile and single-use catheters - Part 1: General requirements) and ISO 10555-2 (Sterile, single-use intravascular catheters - Part 2: Angiographic catheters). It was then determined that the performance of the Surefire High Flow Angiographic Catheter is substantially equivalent to the performance and safety of the Angiodynamics Soft-Vu Angiographic Catheter. A battery of tests was performed according to protocols based on the requirements of recognized standards and was shown to meet the acceptance criteria that were determined to be applicable to the safety and efficacy of the device.
#### Additional Safety Information:
Manufacturing controls include visual, functional, dimensional and sterility tests. Blood contacting materials were tested in accordance with
- · Visual and Dimensional Inspections
- · Shape Retention Testing
- · Particulate Testing
- · Tensile Testing
- · Torque Testing
- · Shelf Life Testing
{2}------------------------------------------------
| | Surefire® High Flow Angiographic Catheter<br>Premarket Notification Traditional 510(k) Submission | Section 5: 510(k) Summary<br>16 July 2012 |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | the tests recommended in the FDA General program Memorandum.<br>Biocompatibility testing was conducted in accordance with International<br>Standard ISO 10993, "Biological Evaluation of Medical Devices Part -1<br>Evaluation and testing". |
| Substantial<br>Equivalence: | The Surefire® High Flow Angiographic Catheter is substantially<br>equivalent in intended use, design, and technology/principles of operation<br>to the predicate. Both devices share the same Indications for Use. Both<br>devices have a 5F outer diameter. The tip of both the predicate and the<br>proposed device have an equivalent geometry. Both products are designed<br>to be compatible with 0.038" guidewires. The Microcatheter is<br>substantially equivalent to the Angiodynamics Soft-Vu Angiographic<br>Catheter, cleared under K112452. Differences between the devices do not<br>raise any issues of safety or effectiveness. | |
| | | Test data provided in bench tests demonstrate that the device is as safe, as<br>effective, and performs at least as safely and effectively as the predicate<br>device. |
| Submitter<br>Information: | Prepared by: | Mario Arbesu<br>Director, Quality Assurance and Regulatory Affairs |
| | Prepared for: | Surefire Medical, Inc.<br>12415 SW 136 Avenue<br>Unit 3<br>Miami, FL 33186 |
| | Date: | July 16, 2012 |
:
।
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 11, 2013
Surefire Medical, Inc. c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313
Re: K122506
Trade/Device Name: Surefire® High Flow Angiographic Catheter Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II Product Code: DQO Dated: June 8, 2012 Received: August 16, 2012
Dear Mr. Job:
This letter corrects our substantially equivalent letter of September 20, 2012.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Matthew G. Hillebrenner
for
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
-9-
### Indication for Use
510(k) Number (if known):
Device Name:
Indication for Use:
Surefire® High Flow Angiographic Catheter
The SUREFIRE HIGH FLOW ANGIOGRAPHIC CATHETER is intended for use where angiographic diagnosis is indicated.
X Prescription Use (part 21 CFR 801 Subpart D) Over-The-counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Le
(Dhisicri 9:7):1. Off) Division in Cardiovascular Devices 122506 510:-., Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.