K122449 · Acacia, Inc. · LGZ · Jun 13, 2013 · General Hospital
Device Facts
Record ID
K122449
Device Name
ENTERAL NUTRITION WARMER
Applicant
Acacia, Inc.
Product Code
LGZ · General Hospital
Decision Date
Jun 13, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Enteral Nutrition Warmer is an electronically powered dry warmer, which supplies external heat to the plastic tubing incorporated administration sets for intermittent use and low flow rate applications at 0 to 7mL/min.
Device Story
The Enteral Nutrition Warmer is an electronically powered dry warmer designed for use in clinical settings, such as the NICU, to warm enteral nutrition fluids. It operates by applying controlled external heat to the outer wall of an administration set tubing. The device features a single heating plate, a thermocouple for temperature sensing, and a temperature fuse as a safety mechanism to shut off power if the sensor fails. It is operated by hospital staff via an on/off switch and includes an LED indicator for operational status. By maintaining enteral fluids at a controlled temperature (32°C to 41°C) during delivery, the device aims to improve patient comfort and tolerance during feeding. It is a reusable, non-sterile device powered by a standard 120V electrical outlet.
Clinical Evidence
A human factors study was conducted to assess clinical risks related to the use of the device for warming enteral fluids in neonatal patients by NICU staff. All identified risks were mitigated. Non-clinical bench testing was performed to verify specifications, including maximum temperatures of the housing, heat sink, and enteral fluid, as well as time to reach nominal temperature.
Technological Characteristics
Electronically powered dry warmer; utilizes a thermocouple for temperature sensing and a temperature fuse for over-temperature protection. Heating plate maximum temperature is 48.8°C. Operates on standard 120V AC. Dimensions: 4.67in x 3.42in x 0.74in. Weight: 22.8 ounces. Materials are fire-rated. Provided non-sterile, reusable.
Indications for Use
Indicated for warming enteral fluids in plastic administration tubing for intermittent, low flow rate (0-7 mL/min) applications in neonatal patients.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
# JUN 1 3 2013
### 510(k) Summary of safety and effectiveness
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | Acacia, Inc. |
|------------------------|-------------------------------------------|
| APPLICANT ADDRESS | 785 Challenger Street<br>Brea, CA 92821 |
| APPLICANT PHONE | (714) 257-0470 |
| CONTACT PERSON | Fergie Ferguson |
| PREPARATION DATE | 05/15/13 |
| TRADE NAME: | Enteral Nutrition Warmer |
| COMMON NAME: | Warmer, Thermal, Infusion Fluid |
| CLASSIFICATION NAME: | Warmer, Thermal, Infusion Fluid |
| DEVICE CLASSIFICATION: | Unclassified |
| PRODUCT CODE | LGZ |
| PREDICATE DEVICES: | Elltec Enteral Nutrition Warmer (K024373) |
#### Purpose
The purpose of this submission is to introduce a new device for Enteral Nutrition Warming.
#### Substantially Equivalent To:
The Acacia, Inc. Enteral Nutrition Warmer is substantially equivalent in intended use, principle of operation and technological characteristics to the Elltec Enteral Nutrition Warmer (K024373) for the delivery of enteral fluids.
The Acacia, Inc. Enteral Nutrition Warmer and the Elltec Enteral Nutrition Warmer are both electronically powered dry warmers, which supplies external heat to the outside of the administration set tubing. The flow rates, output temperature, along with dimensions and weight are equivalent.
The table below describes the characteristics reviewed to determine substantial equivalence.
| Item Reviewed | Proposed Device | Predicate Device | Equivalence |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Enteral Nutrition<br>Warmer is an<br>electronically powered<br>dry warmer, which<br>supplies external heat<br>to the plastic tubing<br>incorporated<br>administration sets for<br>intermittent use and low | The Enteral Nutrition<br>Warmer LIFORT LT-1 is<br>an electronically<br>powered, dry warmer<br>which supplies external<br>heat to the plastic tubing<br>incorporated<br>administration sets for<br>intermittent use and low | Same |
| Principle of<br>Operation | flow rate applications at<br>0 to 7mL/min. | flow rate applications at<br>0 to 7ml/min. The LT-1<br>is available in two sizes:<br>4mm and 5mm channels<br>for administration sets. | Same |
| | Electrically powered<br>device that incorporates<br>controlled heat to the<br>outer wall of an<br>extension set that is<br>delivering enteral fluid<br>to the patient. | Electrically powered<br>device that incorporates<br>controlled heat to the<br>outer wall of an<br>extension set that is<br>delivering enteral fluid to<br>the patient. | |
| Materials | All materials are fire<br>rated for the proposed<br>device's intended use | Unknown | All materials are fire<br>rated for the<br>proposed device's<br>intended use |
| Flow rate | 0 to 7mL/min | 0 to 7mL/min | Same |
| Tubing OD size | 2.4mm | 4.1mm to 5.0mm | Acacia utilizes a<br>more commonly used<br>tubing outer diameter<br>for enteral feeding of<br>premature infants |
| Number of<br>heaters | One | Unknown | Only one sensor is<br>required for the<br>proposed device to<br>perform as intended |
| Maximum<br>heating plate<br>temperature | 48.8 C | 46 C | The proposed<br>device's maximum<br>heating plate<br>temperature does not<br>pose a burn hazard |
| Output<br>temperature | 32 C to 41 C | 32 C to 41 C | Same |
| Temperature<br>sensor | Thermocouple | Thermocouple | Same |
| Safety feature | Temperature fuse shuts<br>off power in case<br>sensor becomes<br>inoperable | Temperature fuse shuts<br>off power in case sensor<br>becomes inoperable | Same |
| Electrical<br>source | Standard 120V<br>electrical outlet | Standard 120V electrical<br>outlet | Same |
| Operating<br>condition | 0 C to 40 C | 0 C to 40 C | Same |
| Warmer weight | 22.8 ounces | 19.75 ounces | The proposed device<br>weighs 3.05 ounces<br>more than the<br>predicate device;<br>however, the unit that<br>warms the tubing |
| | | | near bedside only<br>weighs 3 ounces.<br>The predicate device<br>includes the<br>temperature<br>controller and the<br>warming unit all in<br>one unit, which<br>makes the device<br>more difficult to place<br>at the bedside. |
| Warmer<br>dimensions | 4.67in(length) x 3.42in<br>(width) x<br>0.74in(height) | 6.9in(length) x<br>2.6in(width) x<br>1.5in(height) | The warming unit<br>footprint of both the<br>proposed and<br>predicate devices is<br>similar. The<br>differences are not<br>significant. |
Acacia, Inc. Enteral Nutrition Warmer
Page 1 of 4
{1}------------------------------------------------
## 510(k) Summary of safety and effectiveness
Acacia, Inc. Enteral Nutrition Warmer
:
Page 2 of 4
:
{2}------------------------------------------------
#### 510(k) Summary of safety and effectiveness
#### Description of the Device Subiect to Premarket Notification:
The Enteral Nutrition Warmer is an electronically powered dry warmer, which supplies external heat to the outside of the administration set tubing. The warmer includes an LED light to indicate the warmer is operating and a switch that will turn the warmer on and off.
The Enteral Nutrition Warmer is reusable and provided non-sterile.
#### Indications for Use:
The Enteral Nutrition Warmer is an electronically powered dry warmer, which supplies external heat to the plastic tubing incorporated administration sets for intermittent use and low flow rate applications at 0 to 7mL/min.
## Technical Characteristics:
The Enteral Nutrition Warmer is the same in indications for use, principle of operation and overall technological characteristic compared to the predicate device.
#### Performance Data:
A human factors study was conducted to assess the clinical risks related to the use of the Enteral Nutrition Warmer for warming enteral fluids during the feeding of neonatal patients by the hospital NICU staff. All identified risks have been identified and adequately mitigated to ensure the system is safe for use.
The following non-clinical tests were performed to ensure the device is capable of meeting the predicate device's specifications when tested for the addition of enteral fluids to the intended use:
- ◆ Maximum temperature outside housing (top)
- Maximum temperature outside housing (bottom) .
- Maximum temperature heat sink (tubing track) .
Acacia, Inc. Enteral Nutrition Warmer
Page 3 of 4
{3}------------------------------------------------
## 510(k) Summary of safety and effectiveness
- Maximum temperature of enteral fluid (portion under heater) �
- Maximum temperature of enteral fluid (at extension set feeding tube portion) .
- Time to reach nominal temperature .
.
## Basis for Determination of Substantial Equivalence:
Since the proposed and predicate devices have no significant technological differences, it is concluded that the proposed device is as safe, as effective, and performs at least as safely and effectively as the legally marketed predicate device.
.
## Acacia, Inc. Enteral Nutrition Warmer
Page 4 of 4
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three overlapping wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The text is in all caps and is in a sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
· June 13, 2013
Ms. Fergie Ferguson Director of Operations 785 Challenger Street BREA California 92821
Re: K122449
Trade/Device Name: Enteral Nutrition Warner Regulation Number: Unclassified Regulation Name: Warmer, Thermal, Infusion Fluid Regulatory Class: Unclassified Product Code: LGZ Dated: May 15, 2013 Received: May 24, 2013
Dear Ms. Ferguson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 - Ms. Ferguson
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Image /page/5/Picture/8 description: The image contains the names "Susan Runner DDS, MA" and "Mary S. Runner -S". The text is in a variety of fonts and sizes. The image appears to be a business card or advertisement.
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K122449
Device Name: Enteral Nutrition Warmer
Indications for Use:
The Enteral Nutrition Warmer is an electronically powered dry warmer, which supplies external heat to the plastic tubing incorporated administration sets for intermittent use and low flow rate applications at 0 to 7mL/min.
Prescription Use ਮ (Part-21-CFR 801-Subpart-D)- AND/OR Over-The-Counter Use (21-CFR 801 Subpart.C).
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
FDA
Richard C. Chapman 2013.06.12 13:47:56 -04'00'
- Ivision Sign-Off)
ാടion of Anesthesiology, General Hospital ection Control, Dental Devices
Page 1 of 1
510(k) Number: K122449
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.