DEPUY CTA REVERSE SHOULDER SYSTEM
Device Facts
| Record ID | K122442 |
|---|---|
| Device Name | DEPUY CTA REVERSE SHOULDER SYSTEM |
| Applicant | Depuy France |
| Product Code | PHX · Orthopedic |
| Decision Date | Sep 6, 2012 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 888.3660 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
A Delta CTA™ Reverse Shoulder Prosthesis is indicated for use in - Grossly rotator cuff deficient joint with severe arthropathy or a previous failed joint replacement with a grossly rotator cuff deficient joint. - The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device. - The metaglene component is HA coated and is intended for cementless application with the addition of screws for fixation. For US Use Only: All other components are intended for cemented use only.
Device Story
Reverse shoulder prosthesis; inverted articulation design (ball on glenoid, cup on humerus) to stabilize shoulder in absence of functional rotator cuff. Used in orthopedic surgery; implanted by surgeons. Device components include humeral cups (UHWMPE) and HA-coated metaglene. Provides mechanical joint replacement; restores shoulder function via deltoid muscle. Line extension of existing Delta CTA system; maintains compatibility with existing mating components (Delta CTA epiphysis and Delta Xtend glenospheres).
Clinical Evidence
No clinical tests were conducted to demonstrate substantial equivalence; bench testing only.
Technological Characteristics
UHWMPE humeral cups; HA-coated metaglene for cementless fixation; other components cemented. Semi-constrained design. Dimensions: 38mm/42mm diameters; +3mm, +6mm, +9mm thicknesses. Compatible with Delta CTA epiphysis and Delta Xtend glenospheres.
Indications for Use
Indicated for patients with grossly rotator cuff deficient joints and severe arthropathy or failed previous joint replacement. Requires anatomically/structurally suitable joint and functional deltoid muscle.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,” (vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
- Delta CTA™ Reverse Shoulder System - Humeral Cups (K050315)