K122423 · Medical Components, Inc. · LJS · Dec 12, 2012 · General Hospital
Device Facts
Record ID
K122423
Device Name
VASU-PICC
Applicant
Medical Components, Inc.
Product Code
LJS · General Hospital
Decision Date
Dec 12, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Peripherally Inserted Central Vein Access Catheters are designed for long or shortterm peripheral access to the central venous system for intravenous therapy and blood sampling and allows for central venous pressure monitoring.
Device Story
Vascu-PICC is a taperless, radiopaque polyurethane catheter for peripheral vein catheterization. Available in single, double, and triple lumen configurations with varying French sizes. Features include in-line clamps for fluid control, female luer connectors for IV administration, and depth markings. Used by clinicians for central venous access to facilitate therapy, blood sampling, and pressure monitoring. Device is packaged sterile with insertion accessories. Operation involves percutaneous insertion into peripheral veins to reach the central venous system. Output is the established venous access path for clinical interventions. Benefits include reliable, long-term or short-term vascular access.
Clinical Evidence
Bench testing only. In-vitro testing performed in accordance with ISO standards included air leakage, liquid leakage, force at break, elongation, gravity flow rate, infusion simulation, priming volume, catheter collapse, aging, and chemical testing. Biocompatibility testing demonstrated that all materials meet ISO 10993 requirements.
Technological Characteristics
Radiopaque polyurethane catheter; taperless lumen design; single, double, and triple lumen configurations; includes in-line clamps and female luer connectors; depth markings every 5 cm; sterile packaging. Principles of operation are identical to predicate devices.
Indications for Use
Indicated for patients requiring long or short-term peripheral access to the central venous system for intravenous therapy, blood sampling, and central venous pressure monitoring.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
## DEC 1 2 2012
Image /page/0/Picture/1 description: The image shows a logo with the text "medCOMP." The logo consists of a geometric shape above the text. The geometric shape is made up of black and white stripes, forming an abstract design. The text "medCOMP" is in a lowercase, italicized font.
1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax: 215-256-1787
www.medcompnet.com
| Section 5 | 510(k) SUMMARY | K12242 |
|------------------------|----------------------------------------------------------------------------------------------------------------------|--------|
| Submitter Information: | | |
| Submitter: | MEDCOMP®<br>1499 Delp Drive<br>Harleysville, PA 19438<br>Telephone: (215) 256-4201, ext. 2271<br>Fax: (215) 256-9191 | |
| Contact: | Jessica Leo | |
| Date Prepared: | October 1, 2012 | |
| Device Name: | Vascu-PICC® | |
| Common Name: | Catheter, Intravascular, Therapeutic, Long-Term | |
| Classification Name: | Long Term Intravascular Catheter (LJS 78) | |
| C.F.R. Section: | 880.5970 | |
| Class: | II | |
| Predicate Devices: | K003682 Z-Cath PICC<br>K121094 Vascu-PICC and Midline Catheters<br>K091586 Vascu-PICC and Midline Catheters | |
#### Device Description:
The Vascu-PICC taperless catheters are designed for peripheral vein catheterization. The Vascu-i TOO taponoos cansion comprised of a soft radiopaque polyurethane The lument is an open oncediopacity. The lumen is connected to the extensions material with bandin suitate for facts passement. Clamps are provided on the extension tubes to prevent air/fluid communication. Female luer connectors provide the connection for intravenous administration.
The catheters are available in a range of French sizes in single, double and triple The Cathelers are available in a fange on a fange on throughout the length of the lument "The outside diameter of the lamon to sculature consistent luminal sizing. The Catherer, providing patiente with capseter and numerical marks every 5" centimeter and lumen has depth mants overy continues have in-line clamps to control fluid are available in a thin uble of am. I the gauge size and French size and French size.
The Vascu-PICC® catheter product line is packaged sterile with the necessary accessories to facilitate catheter insertion.
#### Intended Use:
The Vascu-PICC catheters are designed for short or long term peripheral access to the The Vasual TOO catherors are associations therapy and blood sampling. For central venous central venous system for intraverled that catheter lumen of 20 gauge or larger be used.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a geometric shape that is black and white. The word "medCOMP" is written in a lowercase, italicized font below the geometric shape.
- 1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax: 215-256-1787
www.medcompnet.com
### Indications for Use:
The Peripherally Inserted Central Vein Access Catheters are designed for long or r he r enpherally insertou ookly in exentral venous system for intravenous therapy and blood sampling and allows for central venous pressure monitoring.
#### Comparison to Predicate Devices:
The Vascu-PICC® catheters are substantially equivalent to the predicate devices in The vasou 1 1000 battlers area location, basic design, performance, labeling, manufacturing process, method of sterilization and materials.
The difference between the Vascu-PICC® and the predicate Vascu-PICC® device is the taperless lumen design which is based on the original Z-Cath.
### Bench / Performance Data:
The following in-vitro testing was performed on the Vascu-PICC® Catheter to assure reliable design and performance in accordance with ISO standards.
Air Leakage Liquid Leakage Force at Break Elongation Gravity Flow Rate Infusion Simulation Priming Volume Catheter Collapse Aging Chemical Testing
#### Biocompatibility:
All materials used for the Vascu-PICC are identical to the predicate Vascu-PICC All matenals used for the Vacca Price (K003682) and have been submitted in (K12 1094 and R00 1000) and = Satir 100 (All biocompatibility testing demonstrated thematerials used meet the requirements of ISO 10993.
### Technological characteristics:
The principles of operation are the same as the predicate devices. The lumen design change does not affect the safety or effectiveness of the device.
## Conclusion:
The proposed devices meet the performance criteria of design verification as specified The proposed as not test protocols. The proposed devices are substantially equivalent to the legally marketed predicate devices.
Page 2 of 2
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a bird or a stylized human figure with outstretched arms.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-002
#### December 12, 2012
Ms. Jessica Leo Regulatory Associate Medical Components Incorporated 1499 Delp Drive Harleysville, PA 19438
Re: K122423
Trade/Device Name: Vascu-PICC® Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: November 27, 2012 Received: November 27, 2012
#### Dear Ms. Leo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Leo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Cinthorn Lo m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): _k|22423
Vascu-PICC® Device Name:
Indications for Use:
The Peripherally Inserted Central Vein Access Catheters are designed for long or shortterm peripheral access to the central venous system for intravenous therapy and blood sampling and allows for central venous pressure monitoring.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by Richard C. Chapman Date: 2012.12.10 12:16:00 -05'00'
(Division Sign-Off) (Division Sign-off)
Division of Anesthesiology, General Hospital Division of Antsoulocion.org/ Divises
1122423
510(k) Number:
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.