SCOPEVALET DISPOSABLE BIOPSY VALVE

K122417 · The Ruhof Corp. · OCX · May 17, 2013 · Gastroenterology, Urology

Device Facts

Record IDK122417
Device NameSCOPEVALET DISPOSABLE BIOPSY VALVE
ApplicantThe Ruhof Corp.
Product CodeOCX · Gastroenterology, Urology
Decision DateMay 17, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The ScopeValet Disposable Biopsy Valves, Model Numbers 345BVO and 345BVP, are intended to provide access for endoscopic device passage and exchange, helps maintain sufflation, minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure, and provides access for irrigation.

Device Story

ScopeValet Disposable Biopsy Valves are single-use, non-sterile accessories for gastrointestinal endoscopes (Olympus, Fujinon, Pentax). Device consists of injection-molded thermoplastic elastomer (TPE) with a slit diaphragm opening. During endoscopic procedures, device attaches to biopsy/suction channel inlet; allows passage/exchange of endoscopic instruments; maintains insufflation pressure; minimizes bio-burden leakage; provides access for irrigation. Operated by clinicians in hospital or clinic settings. Device performance verified via leak testing, insufflation air flow rate testing, and biocompatibility assessment per ISO 10993-1. Benefits include reduced risk of cross-contamination through single-use design and effective maintenance of endoscopic working conditions.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including leak testing, insufflation air flow rate testing, and biocompatibility testing per ISO 10993-1.

Technological Characteristics

Material: Thermoplastic elastomer (TPE). Design: Injection-molded, single-use, non-sterile. Interface: Slit diaphragm for instrument passage. Compatibility: Olympus, Fujinon (G5 series and newer), and Pentax gastrointestinal endoscopes. Standards: ISO 10993-1 (biocompatibility), ISO 10993-5 (cytotoxicity).

Indications for Use

Indicated for patients undergoing gastrointestinal endoscopic procedures requiring access for device passage, irrigation, and maintenance of insufflation.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K122417 page 1 of 4 で # 510(k) Summary ### 1. Submission Sponsor Ruhof Corporation 393 Sagamore Avenue . Mineola New York 11501 USA Phone: (516) 294.5888 Fax: (516) 248.6456 Contact: Douglas J MacKay, VP of Sales and Marketing ## 2. Submission Correspondent Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: (720) 838.4113 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Carrie Hetrick, Senior Consultant, RA Email: project.management@emergogroup.com ### 3. Date Prepared April 26, 2013 ### 4. Device Identification Trade/Proprietary Name: ScopeValet Disposable Biopsy Valves Common/Usual Name: Biopsy Valves Classification Name: Endoscope and accessories Classification Regulation: 876.1500 Product Code: OCX Device Class: Class II Classification Panel: Gastroenterology/Urology ## 5. Predicate Devices EndoChoice - Seal™ Biopsy Valve (K111821) ### 6. Device Description The Ruhof Corporation ScopeValet Disposable Biopsy Valves can be manufactured with a choice of two types - regular disposable biopsy valve and irrigating biopsy valve. Each of these types can be ordered for Olympus, Fujinon, and Pentax gastrointestinal endoscopes. ## 7. Intended Use MAY 1 7 2013 {1}------------------------------------------------ The ScopeValet Disposable Biopsy Valves are intended to provide access for endoscopic device passage and exchange, helps maintain sufflation, minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure, and provides access for irrigation. #### 8. Substantial Equivalence Discussion The following table compares the ScopeValet Disposable Biopsy Valves to the predicate device with respect to the intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence. | Manufacturer | Ruhof Corporation | EndoChoice, Inc. | |-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name | ScopeValet Disposable Biopsy<br>Valves | Seal Biopsy Valves | | 510(k) Number | K122417 | K111821 | | Product Code | OCX | OCX | | Regulation Number | 876.1500 | 876.1500 | | Regulation Name | Endoscope and accessories | Endoscope and accessories | | Design | The ScopeValet disposable biopsy<br>valve – non-sterile is used to<br>cover the opening to the<br>biopsy/suction channel inlet of<br>Olympus® and Fujinon® (G5<br>series and newer),<br>gastrointestinal endoscopes or<br>for use with Pentax®<br>gastrointestinal endoscopes.<br>The ScopeValet disposable biopsy<br>valve - non-sterile provides<br>access for endoscopic device<br>passage and exchange, helps<br>maintain insufflation and<br>minimizes leakage of bio-burden<br>from the biopsy port throughout<br>the gastrointestinal endoscopic | The EndoChoice biopsy valve can<br>be manufactured with a choice of<br>two types - regular disposable<br>biopsy valve and irrigating biopsy<br>valve. Each of these types can be<br>ordered for Olympus, Fujinon, and<br>Pentex gastrointestinal<br>endoscopes. | | Indications for Use | procedure.<br>The ScopeValet Disposable<br>Biopsy Valves are intended to<br>provide access for endoscopic<br>device passage and exchange,<br>helps maintain sufflation,<br>minimizes leakage of<br>biomaterial from the biopsy<br>port throughout the<br>endoscopic procedure, and<br>provides access for irrigation. | Biopsy valves are intended to<br>provide access for endoscopic<br>device passage and exchange,<br>helps maintain sufflation,<br>minimizes leakage of biomaterial<br>from the biopsy port throughout<br>the endoscopic procedure, and<br>provides access for irrigation. | | Compatible Endoscope(s) | Olympus and Fujinon, (G5 series<br>and newer) gastrointestinal | Olympus and Fujinon, (G5 series<br>and newer) gastrointestinal | ## Table 5A – Comparison of Characteristics {2}------------------------------------------------ | Manufacturer | Ruhof Corporation | EndoChoice, Inc. | |---------------------------|-----------------------------------------------------|-----------------------------------------------------| | Trade Name | ScopeValet Disposable Biopsy<br>Valves | Seal Biopsy Valves | | | endoscopes or Pentax<br>gastrointestinal endoscopes | endoscopes or Pentax<br>gastrointestinal endoscopes | | Patient Population | Patients who are undergoing an<br>endoscopy | Patients who are undergoing an<br>endoscopy | | Environment | Hospital and/or Clinic | Hospital and/or Clinic | | Manufacturing Method | Injection molded | Injection molded | | Opening for Instruments | Slit in diaphragm | Slit in diaphragm | | Material | TPE | TPE | | Sterile | No | No | | Single-Use, Disposable | Yes | Yes | | Shelf Life | One (1) year | Not known | | Complies with ISO 10993-1 | Yes | Yes | ## 9. Non-Clinical Performance Data The following testing has been performed to support substantial equivalence: - Leak testing Leak testing was performed with the ScopeValet Disposable Biopsy Valves . to simulate the withdrawal of biopsy samples, bowel irrigation and bowel insufflation. The ScopeValet Disposable Biopsy Valves revealed no evidence of fluid leakage during the sample biopsy or bowel irrigation. Thus, the ScopeValet Disposable Biopsy Valves are an effective means of preventing fluid leakage in gastrointestinal endoscopic procedures. - Insufflation air flow rate testing - Tests were performed simulating bowel insufflations, whereby the ScopeValet Disposable Biopsy Valves tested maintained a pressure of at least 10 psi with no testing. The insufflation air flow rate is comparable to that of the predicate device, delivering enough air for insufflation in the time typically used for insufflation. Thus, the ScopeValet Disposable Biopsy Valves is an effective means of achieving insufflation in gastrointestinal endoscopic procedures. - . Biocompatibility – Biocompatibility testing of the ScopeValet Disposable Biopsy Valves was conducted in accordance with the ISO 10993-1 "Biological evaluation of medical devices" standards and FDA/CDRH/ODE Blue Book Memorandum G95-1, "Use of International Standard ISO 10993", 'Biological Evaluation of Medical Devices, Part 1: ' Evaluation and Testing' using a representative device. Based on ISO 10993-1, the biopsy valves are body contact (mucosal membranes) and limited contact duration (less than 24 hours). The ScopeValet Disposable Biopsy Valves are made of thermoplastic elastomer (TPE) materials that are Generally Recognized As Safe (GRAS) materials according to ISO 10993 for cytotoxicity, sensitization, and irritation, are USP Grade VI, and the colorants are cleared (or FDA listed). The device was tested for cytotoxicity as per ISO 10993-5, and passed. The results and analysis demonstrate the ScopeValet Disposable Biopsy Valves are biocompatible as per ISO 10993-1. - t Shelf-life – Stability of the subject device was established from the results of physical testing data using a protocol.. Based on the evaluation of the results of the physical testing data, the expiring date has been set at one (1) year. {3}------------------------------------------------ As part of demonstrating safety and effectiveness of ScopeValet Disposable Biopsy Valves and in showing substantial equivalence to the predicate device that is subject to this 510(k) submission, Rohuf Corporation completed a number of tests. The ScopeValet Disposable Biopsy Valves meets all the requirements for the overall design,and biocompatibility that the output meets the design inputs and specifications. The ScopeValet Disposable Biopsy Valves passed all testing stated above as shown by the acceptable results obtained. The ScopeValet Disposable Biopsy Valves complies with the applicable voluntary standards for biocompatibility. The device passed all the testing in accordance with national and international standards. ### 10. Clinical Testing There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate device, have been on the market for many years with proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device. ### 11. Statement of Substantial Equivalence By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device. It has been shown in this 510(k) submission that the difference between ScopeValet Disposable Biopsy Valves and the predicate device do not raise any questions regarding its safety and effectiveness. The ScopeValet Disposable Biopsy Valves, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device. {4}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 17, 2013 Ruhof Corporation % Ms. Carrie Hetrick Senior Consultant, RA Emergo Group, Inc. 816 Congress Avenue, Suite 1400 AUSTIN TX 78701 Re: K122417 Trade/Device Name: ScopeValet Disposable Biopsy Valves, Models 345BVO and 345BVP Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCX Dated: April 30, 2013 Received: May 1, 2013 Dear Ms. Hetrick: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {5}------------------------------------------------ Page 2 - Ms. Carrie Hetrick You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDeyices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): K122417 Device Name: ScopeValet Disposable Biopsy Valve, Model Numbers 345BVO and 345BVP Indications for Use: The ScopeValet Disposable Biopsy Valves, Model Numbers 345BVO and 345BVP, are intended to provide access for endoscopic device passage and exchange, helps maintain sufflation, minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure, and provides access for irrigation. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR . Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Benjamin R 2013.05.17 115:4 (Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K122417 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...