K122414 · Leto Enterprises Incorporation · NGX · Dec 21, 2012 · Physical Medicine
Device Facts
Record ID
K122414
Device Name
AB COMMAND DUAL CHANNEL FITNESS BELT
Applicant
Leto Enterprises Incorporation
Product Code
NGX · Physical Medicine
Decision Date
Dec 21, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5850
Device Class
Class 2
Indications for Use
The AB COMMAND Dual Channel Fitness Belt is intended for use by healthy persons to apply trans-coetancous electrical muscle stimulation (EMS) through skin contact electrodes for the following purposes: Improvement of muscle tone of the muscles in the abdomen.
Device Story
Battery-operated, two-channel muscle stimulation system; designed for abdominal muscle conditioning. Comprised of electronic stimulator module, fixation belt, and four integrated skin-contact electrodes. User operates device via buttons to adjust electrical current intensity. Stimulator delivers gentle electrical pulses to targeted abdominal muscles to induce contraction. Used by healthy individuals for fitness purposes; not for clinical therapy. Benefits include improved muscle tone and firmness. Device is self-operated by the user.
Clinical Evidence
Bench testing only. Compliance verified with IEC 60601-1 (electrical safety), IEC 60601-2-10 (particular requirements for nerve and muscle stimulators), IEC 60601-1-2 (electromagnetic compatibility), and IEC 60601-1-4 (software). Waveform parameters verified via bench testing.
Technological Characteristics
Powered muscle stimulator (Product Code NGX). Two-channel battery-operated system. Includes electronic stimulator module, fixation belt, and four integrated electrodes. Operates via transcutaneous electrical muscle stimulation (EMS). Complies with IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, and IEC 60601-1-4 standards.
Indications for Use
Indicated for improvement of muscle tone, strengthening, and development of firmer abdominal muscles in healthy individuals. Contraindicated for use on injured or impaired muscles. Not for therapy or treatment of medical conditions/diseases.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
# K122414
| Sponsor: | Leto Enterprises Incorporation |
|-----------------|--------------------------------------|
| Subject Device: | AB COMMAND Dual Channel Fitness Belt |
| File No.: | 510(k) submission report, Section 9 |
#### Section 9 510 (k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
### 1. Submitter's Information
510(k) Owner's Name: Leto Enterprises Incorporation Address: 19/F, Nan Dao Commercial Bldg, 359-361 Queen's Road Central, Hong Kong, China Phone: (852) 3694 0110 Fax: (852) 2515 3995 Contact Person: Choi Chui Ngan E-mail: dir@leto-ent.com
#### Contact Person of the Submission: A
Ms. Sabrina Wei
MEDLAB (Shenzhen) Information Service Co., LTD B102, Nanshan Medical Device Industrial Park, No. 1019 of Nanhai Avenue, Nanshan District,Shenzhen, Guangdong, P. R. China, 518000 TEL: +86-755- 8632 9133 FAX: +86-755- 8632 9134 E-mail: sabrinawei@hotmail.com
#### 2. Device Information
| Trade Name: | AB COMMAND Dual Channel Fitness Belt |
|----------------------|------------------------------------------------------|
| Common Name: | Powered muscle stimulator |
| Classification name: | Stimulator, Muscle, Powered, For muscle conditioning |
| Review Panel: | Physical Medicine |
| Product Code: | NGX |
| Regulation Class: | II |
| Regulation Number: | 890.5850 |
#### 3. Predicate Device Information
| Submitter: | Leto Enterprises Incorporation |
|----------------|---------------------------------|
| Device Name: | X2ABS Dual Channel Fitness Belt |
| 510(K) Number: | K102295 |
Report by MEDLAB (Shenzhen) Information Service Co., Ltd.
Page 9-1
{1}------------------------------------------------
Leto Enterprises Incorporation Sponsor: AB COMMAND Dual Channel Fitness Belt Subject Device: 510(k) submission report, Section 9 File No .:
### 4. Device description
AB COMMAND Dual Channel Fitness Belt is a two channel battery operated muscle stimulation system specifically designed to exercise the abdominal muscles. It is comprised of an electronic stimulator module for signal generation, a belt for fixation, and four electrodes for signal connection to skin. The built-in electrodes are located on the inner surface of the belts.
Power is derived from 2 batteries located in a compartment protected by a removable battery cover for the Fitness Belt. There is no current passed from side to side. The user cannot access the wiring or connectors within the belt.
The stimulator sends gentle electrical current to targeted muscle group through the electrodes placed on the skin. The parameters of the unit are controlled by the buttons. Its intensity level can be adjustable by user.
#### 5. Intended Use
The AB COMMAND Dual Channel Fitness Belt is intended for use by healthy persons to apply trans-coetancous electrical muscle stimulation (EMS) through skin contact electrodes for the following purposes:
- Improvement of muscle tone of the muscles in the abdomen.
#### 6. Indications for Use
The AB COMMAND Dual Channel Fitness Belt is indicated for the improvement of muscle tone, for strengthening of muscles and for the development of firmer muscles. It is indicated for use on the abdominal muscles.
Contraindicated use on injured or otherwise impaired muscles.
Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
#### 7. Performance Summary
Testing has been carried out to assure compliance with recognized electrical safety standards:
- IEC 60601-1 and -2-10 standards for electrical safety
- IEC 60601-1-2 standard for electromagnetic compatibility
- IEC 60601-1-4 standard for the software
Performance data has been verified according to the requirements of the FDA Guidance for Pre Market Submissions and for Software Contained in Medical Devices.
The waveform test report has also been provided to verify the parameters of the device.
#### 8. Comparison to predicate device and conclusion
The Intended Use, Indications for Use, technological characteristics, features,
Report by MEDLAB (Shenzhen) Information Service Co., Ltd.
Page 9-2
{2}------------------------------------------------
Leto Enterprises Incorporation Sponsor: AB COMMAND Dual Channel Fitness Belt Subject Device: 510(k) submission report, Section 9 File No.:
specifications, materials, operation, and software of AB COMMAND Dual Channel Fitness Belt are substantial cquivalent to the predicate device X2ABS Dual Channel Fitness Belt.
The main differences between the subject device and predicate device are the trade mark, dimension and art design of belt, package bag and box. These differences will not raise new issues of safety and effectiveness.
9. Date of the summary prepared: September 29, 2012
#### Report by MEDLAB (Shenzhen) Information Service Co., Ltd.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name arranged in a circular pattern around a stylized emblem. The emblem consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with three overlapping lines forming the staff and a serpent winding around it.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
21 December 2012
Leto Enterprises Incorporation % Ms. Sabrina Wei Medlab (Shenzhen) Information Service Co., Ltd. Suite A-2706, Zhongfang Jingyuan, Xinwen Road, Futian District Shenzhen, Guangdong 518100 China
Re: K122414
Trade/Device Name: AB Command Dual Channel Fitness Belt Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: Undated Received: November 26, 2012
Dear Ms. Wei:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your been wined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the energed of to the enactment date of the Medical Device Amendments, or to commerce pror to they 20, 2017, 12, 12, 2017, 11:22 provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adulton. Thease noter u, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be devisou alle 1 Dr mination that your device complies with other requirements of the Act that I Dri has made a associations administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 - Ms. Sabrina Wei
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Of K I art 007), acomig (21 CFR 803); good manufacturing practice requirements as set de rece related daverse overse) (2 - Entation (2) CFR Part 820); and if applicable, the electronic roral in also quino control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please the Concertified, "Misbranding by reference to premarket notification" (21CFR Part note the regulations regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Victor Krauthamer -S
Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Leto Enterprises Incorporation Sponsor: AB COMMAND Dual Channel Fitness Belt Subject Device: 510(k) submission report, Section 5 File No.:
### 3. Indications for Use
510(k) Number (if known): KIZZYIY Device Name: AB COMMAND Dual Channel Fitness Belt
Indications for Use:
The AB COMMAND Dual Channel Fitness Belt is indicated for the improvement of muscle tone, for strengthening of muscles and for the development of firmer muscles. It is indicated for use on the abdominal muscles.
Contraindicated use on injured or otherwise impaired muscles.
Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ X AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Neurological and Physical
Medicine Devices
510(k) Number K122414
Report by MEDLAB (Shenzhen) Information Service Co.; Ltd.
Page 5-10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.