K122393 · Miga Medical Co., Ltd. · NGX · Dec 10, 2013 · Physical Medicine
Device Facts
Record ID
K122393
Device Name
TG-BODY MI&BODY S-TRAINER
Applicant
Miga Medical Co., Ltd.
Product Code
NGX · Physical Medicine
Decision Date
Dec 10, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Indications for Use
TG-BODY, MI&Body, S-Trainer are intended to stimulate healthy muscles in order to improve or facilitate muscle performance and not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind.
Device Story
Portable, battery-operated neuromuscular electronic stimulation system; models TG-Body, S-Trainer, and Mi&Body. Device delivers electrical stimulation to healthy muscles via electrodes; output waveform is symmetrical biphasic square wave. User selects from 6 pre-programmed modes; output strength adjustable 0-100mA; session time 5-60 minutes. LCD displays operation status, mode, intensity, time, and battery level. Used by healthy adults for muscle performance improvement; not for medical treatment. Safety features include automatic overload trip, no-load trip, and automatic shut-off. Healthcare provider/user monitors output via LCD; device provides non-invasive muscle stimulation to facilitate performance.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through compliance with consensus standards EN 60601-1, EN 60601-2-10, and EN 60601-1-2.
Indicated for healthy adults to stimulate healthy muscles to improve or facilitate muscle performance. Not indicated for therapy or treatment of any medical disease or condition.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
# DEC 1 0 2013
### 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
Submitter Information: Miga Medical Co., Ltd 925, Pyongsan-Dong, Yangsan-Si Kyongsangnam-Do, Korea
Contact Person:
Yang Ho Dong Onbix Corporation #821 Samil Plaza, 837-26 Yeuksam-dong Gangnam-gu, Seoul, 135-768, Korea Tel: *82-2-5663360 / Fax: *82-2-62803360 Email: onbix@naver.com
Date Summary Prepared: Dec 09, 2013
### Device Name:
| Trade Name(s): | TG-BODY, MI&Body, S-Trainer |
|----------------------|-----------------------------|
| Classification Name: | Powered Muscle Stimulator |
| Panel: | Physical Medicine |
| Product Code: | NGX |
### Predicate Device Information:
K011880 Compex sport stimulator, Compex
#### Device Description:
The MIGA Co.,Ltd OTC EMS System . model TG-Body. S-Trainer and Mi&Body are intended for use by healthy adults for the stimulation of healthy muscles in order to improve or facilitate muscle performance; it is not intended as a therapy for any medical condition. Basically the model TG-body, powered by 4.8V, S-Trainer, powered by 3.6V and Mi&Body, powered by 3.7V are muscle stimulation devices and identical to the predicate device -K011880 (Compex Sport) with the following features:
It is a portable, battery operated neuromuscular electronic stimulation system. [1]
It contains 6 programs as similar to the predicate device; the output waveform is (2)
selectable pre-programming change among P1-P6 (in case of TG-body 10)
[3] The output strength is adjustable at 0-100mA via regulated voltage, with variable set time 5-60 minutes counting from starting session.
[4] The LCD display is provided for the indication of operation status including operation mode, output program mode, output intensity, time to cut-off, and battery low indication.
#### Intended Use:
TG-BODY, MI&Body, S-Trainer are intended to stimulate healthy muscles in order to improve or facilitate muscle performance and not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind.
#### Comparison to Predicate Device(s):
These devices are equivalent to the predicate devices in its intended use and technological characteristics, including:
> *indications for use *technological characteristics
{1}------------------------------------------------
・
# *performance properties
Comparison table is provided as below.
| Features | New Device | New Device | New Device | Predicate<br>Device |
|----------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------|
| 510(K) Number | - | - | - | K011880 |
| Device Name, Model | TG Body | Mi&Body | S-Trainer | Compex sport<br>stimulatior |
| Manufacturer | Miga | Miga | Miga | Compex |
| Power Source(s) | 4.8 | 3.7 | 3.6 | 7.2 |
| - Method of Line<br>Current Isolation: | Transformers<br>isolation | Transformers<br>isolation | Transformers<br>isolation | Transformers<br>isolation |
| - Patient Leakage<br>Current: | - | - | - | - |
| - Normal Condition | 2uA | 2uA | 2uA | 2uA |
| - Single Fault<br>Condition | 13uA | 13uA | 13uA | 15uA |
| Number of Output Mode | 1 | 1 | 1 | 1 |
| Number of Output<br>Channels | 4 | 2 | 1 | 4 |
| - Synchronous or<br>Alternating | Synchronous | Synchronous | Synchronous | Synchronous |
| - Method of Channel<br>Isolation | galvanic | galvanic | galvanic | galvanic |
| Regulated Current or<br>Regulated Voltage? | 500ohm RMS<br>26mA / RMS<br>13.0V<br>2Kohm RMS<br>8.4mA / RMS<br>16.8V | 500ohm RMS<br>15mA / RMS<br>7.5V<br>2Kohm RMS<br>4.0mA / RMS<br>8.1V | 500ohm RMS<br>15mA / RMS<br>7.5V<br>2Kohm RMS<br>4.0mA / RMS<br>8.1V | 500ohm RMS<br>21mA / RMS<br>11.8V<br>2Kohm RMS<br>7mA / RMS 16V |
| Software/Firmware/Micr<br>oprocessor Control | Yes | Yes | Yes | Yes |
| Automatic Overload<br>Trip | Yes | Yes | Yes | Yes |
| Automatic No-Load Trip | Yes | Yes | Yes | Yes |
| Automatic Shut Off | Yes | Yes | Yes | Yes |
| Patient Override Control | Yes | Yes | Yes | Yes |
| Indicator Display | | | | |
| - On/Off Status | Yes | Yes | Yes | Yes |
| - Low Battery<br>signal | Yes | Yes | Yes | Yes |
| - Voltage/Current Level | No | No | No | No |
| Timer Range(Minutes) | 1 ~ 60 | 20 ~ 33 | 20 ~ 33 | 1 ~ 60 |
2
{2}------------------------------------------------
K122393
.
:
·
,
.
・
| | 60601-1.60601-1-1<br>60601-1-<br>2.60601-2-10 | 60601-<br>1.60601-1-1<br>60601-1-<br>2.60601-2-10 | 60601-<br>1.60601-1-1<br>60601-1-<br>2.60601-2-10 | Page 3 of 4<br>60601-<br>1.60601-1-1<br>60601-1-<br>2.60601-2-10 |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|---------------------------------------------------|---------------------------------------------------|------------------------------------------------------------------|
| Compliance With<br>Voluntary Standards | | | | |
| Compliance with 21<br>CFR898?<br>(Becomes Mandatory<br>Beginning May 9,2000) | Yes | Yes | Yes | Yes |
| Weight | 360g | 120g | 70g | 350g |
| Dimensions(W*H*D) | 106*162*38.5(m<br>m) | 64.8*103.6*26(m<br>m) | 51*51*14(mm) | 142*99*36(mm) |
| Housing Materials and<br>Construction | ABS | ABS | ABS | Plastic case |
| the name of the output<br>Mode | Square wave | Square wave | Square wave | Square wave |
| cleary labeled<br>amplitude and times<br>axes | 0~300 | 0~300 | 0~300 | 0~300 |
| identification of the<br>amplitude baseline | - | - | - | - |
| a list of all output<br>parameter setting<br>the load resistance(in<br>ohms) | - | - | - | - |
| Waveform(e.g., pulse<br>monophasic,biphasic) | Symmetrical<br>biphasic square | Symmetrical<br>biphasic square | Symmetrical<br>biphasic square | Symmetrical<br>biphasic square |
| Shape(e.g.,<br>rectangular, spike, rectifi<br>ed sinusoldal) | Rectangular | Rectangular | Rectangular | Rectangular |
| Maximum output<br>Voltage(specify units) | 100V peak | 40V peak | 40V peak | 100V peak |
| Maximum output<br>Current(specify units) | 26mA | 14.4mA | 13.3mA | 21mA |
| Pulse Width(specify<br>units) | 50 ~300 | 150-300 | 150~300 | 200 ~ 400 |
| Frequency(Hz) | 4 ~100 | 4~100 | 4~100 | 1 ~ 120 |
| Beat Frequency(Hz) | - | - | - | - |
| For multiphasic<br>waveforms only | - | - | - | - |
| Symmetrical phases | Yes | Yes | Yes | Yes |
| Phase duration(include<br>units)(state range, if<br>applicable)(both<br>phases, if<br>asymmetrical) | 1 - 0.005 | 0.25 ~ 0.01 | 0.25 ~ 0.01 | 0.1 - 0.001 |
| net charge(m C per<br>pulse)(if zero, state<br>method of achieving<br>zero net charge) | 0 | 0 | 0 | 0 |
| Maximum Phase charge<br>Density(uC) | 36 | 24 | 24 | 18 |
| Maximum Current<br>Density(mA/cm2) | 1.0432 | 0.20571 | 0.26473 | 9.6 |
| Maximum Power<br>Density(W/cm2)(using<br>smallest electrode<br>conductive surface<br>area) | 0.06259 | 0.01059 | 0.00823 | 0.046 |
| Burst mode(i.e., Pulse<br>trains) | - | - | - | - |
| a. pulses per burst | - | - | - | - |
| b. bursts per second | - | - | - | - |
| c. burst<br>duration(seconds) | - | - | - | - |
| d. duty Cycle (line A *<br>line B) | - | - | - | - |
| ON time(seconds) | 10 | 10 | 10 | - |
| OFF time(seconds) | 10 | 10 | 10 | - |
| Additional Features(if<br>applicable) | - | - | - | - |
:
. . . . . .
·
:
:
:
.
3
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
{3}------------------------------------------------
### K122393
4
### Performance Test Data and Conclusions:
These devices meet consensus standards EN 60601-1, EN 60601-2-10, EN 60601-1-2. Bench testing was also conducted to demonstrate performance specifications. Based on the information provided in this premarket notification Miga Medical Co., Ltd concludes that the TG-BODY, MI&Body, S-Trainer are safe and effective and substantially equivalent to predicate device as described herein.
. '
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized eagle-like symbol with three curved lines representing the body and wings.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### December 10, 2013
Miga Medical Co., Ltd c/o Yang Ho Dong Onbix Corporation 821 Samil Plaza, 837-26 Yeuksam-dong Gangnam-gu, 135-768 Seoul South Korea
Re: K122393
Trade/Device Name: TG-Body, MI&Body, and S-Trainer Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: NGX Dated: November 20, 2013 Received: November 27, 2013
Dear Yang Ho Dong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. However, we remind you that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Yang Ho Dong
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Victor Krauthamer -A
Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
. In rant 800); fabening (2
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K122393
TG-BODY, Mi&Body, S-Trainer Device Name:
Indications For Use:
TG-BODY, MI&Body, S-Trainer are intended to stimulate healthy muscles in order to improve or facilitate muscle performance and not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind.
Prescription Use (Part 21 CFR 801 Subpart D)
13 4, 188 (21 CAN 1280 000), moonan as are
. . . . . . . . . . . .
AND/OR
Over-The-Counter Use _ x (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
# Victor K 2013.12
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
s rate on transmill (21 Cark fall
.
.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.