SECA MEDICAL BODY COMPOSTION ANALYZER 514, SECA MBCA 514, SECA MBCA, SECA 514
K122388 · Seca GmbH & Co. KG · MNW · May 2, 2013 · Cardiovascular
Device Facts
Record ID
K122388
Device Name
SECA MEDICAL BODY COMPOSTION ANALYZER 514, SECA MBCA 514, SECA MBCA, SECA 514
Applicant
Seca GmbH & Co. KG
Product Code
MNW · Cardiovascular
Decision Date
May 2, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The seca mBCA 514 measures body Weight and records body Height for calculation of Body Mass Index (BMI), Resting Energy Expenditure (REE) and Total Energy Expenditure (TEE) to monitor growth process and weight changes in individuals of all ages. In healthy individuals aged 18 years and older the device also measures body impedance (Reactance Xc and Resistance R) for estimation of body composition. Estimation of body composition includes Total Body Water (TBW), Intra Cellular Water (ICW), Extra Cellular Water (ECW), Lean Soft Tissue (LST), Fat-Free Mass (FFM), Fat Mass (FM), Fat Mass Indices (FMI and FMMI), Skeletal Muscle Mass (SMM), Bioelectrical Impedance Vector Analysis (BIVA), Energy stored in body (Ebody) and Phase Angle (Φ). The seca mBCA 514 can be used in conjunction with the optional PC software accessory seca analytics 115 for data management, calculations and display of information.
Device Story
Scale and impedance plethysmograph; measures body weight and height; performs multi-frequency segmental bioelectrical impedance analysis (BIA) (right arm, trunk, right leg, left leg, right/left body sides). Inputs: weight, height, bioelectrical impedance (resistance, reactance). Outputs: BMI, REE, TEE, and body composition metrics (TBW, ICW, ECW, LST, FFM, FM, FMI, FMMI, SMM, BIVA, Ebody, Phase Angle). Used in clinical settings; operated by healthcare professionals or patients. Optional seca analytics 115 PC software provides data management, calculations, and display. Output assists clinicians in monitoring growth, weight changes, and body composition status.
Clinical Evidence
Bench and clinical testing performed. Impedance and weight measurements compared to InBody 720. TBW compared to deuterium oxide dilution; ECW compared to sodium bromide dilution; LST compared to GE Lunar iDXA. FFM estimated via four-compartment model incorporating DXA, BOD POD, and deuterium oxide dilution. No specific sensitivity/specificity metrics provided.
Technological Characteristics
Multi-frequency segmental bioelectrical impedance analyzer. Measures resistance, reactance, and phase angle. Integrates weight scale and height recording. Optional connectivity to PC software (seca analytics 115).
Indications for Use
Indicated for individuals of all ages to monitor growth and weight changes via BMI, REE, and TEE. For healthy individuals aged 18+, indicated for estimation of body composition parameters including TBW, ICW, ECW, LST, FFM, FM, FMI, FMMI, SMM, BIVA, Ebody, and Phase Angle.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
BOD POD Sonamet Body Composition Analyzer (K060848)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/20 description: The image shows the word "seca" in a bold, sans-serif font. The letters are connected to each other, creating a continuous, flowing design. The color of the text is black, and the background is white.
| seca medical Body Composition Analyzer 514 | 510(k) Summary - section 05 |
|--------------------------------------------|-----------------------------|
| 510(k) Premarket Notification - K122388 | page 1 of 2 |
2 2013 MAY
#### Submitter Information 1
Owner: Address:
Internet:
Contact:
E-mail:
seca gmbh & co. kg. Hammer Steindamm 9 - 25 22089 Hamburg Germany www.seca.com
Djavaneh Blalaie - Quality Management +49 40 20 00 00 - 259 +40 40 20 00 00 - 3259 djavaneh.blalaie@seca.com
Date prepared:
Telephone Number: FAX Number:
April 22, 2013
# 2 Device Name
Common Device Name: Trade names:
Device Class: Classification Code: Regulation number:
Body Composition Analyzer seca medical Body Composition Analyzer 514, seca mBCA 514, seca mBCA, seca 514 Class II MNW - Analyzer, Body Composition 21 CFR 870.2770
#### Predicate Device 0
### BioSpace InBody 720
Common Device Name: Trade names: Device Class: Classification Code: Regulation number: 510(k) number:
Body fat meter Biospace Body Composition Analyzer, Model Inbody 720 Class II MNW - Analyzer, Body Composition 21 CFR 870.2770 K052646
#### Device Description র্ব
The seca mBCA 514 is a scale and impedance plethysmograph body composition analyzer for estimation of the body composition of individuals based on the Bioelectrical Impedance Analysis (BIA). The device measures bioelectrical impedance values by means of multi frequency,
{1}------------------------------------------------
| seca medical Body Composition Analyzer 514 | 510(k) Summary - section 05 |
|--------------------------------------------|-----------------------------|
| 510(k) Premarket Notification - K122388 | page 2 of 2 |
segmental measurements (right arm, trunk, right leg, left leg, right body side and left body side).
# 5 Intended Use
The seca mBCA 514 measures body Weight and records body Height for calculation of Body Mass Index (BMI), Resting Energy Expenditure (REE) and Total Energy Expenditure (TEE) to monitor growth process and weight changes in individuals of all ages.
In healthy individuals aged 18 years and older the device also measures body impedance (Reactance Xc and Resistance R) for estimation of body composition. Estimation of body composition includes Total Body Water (TBW), Intra Cellular Water (ICW), Extra Cellular Water (ECW), Lean Soft Tissue (LST), Fat-Free Mass (FFM), Fat Mass (FM), Fat Mass Indices (FMI and FMMI), Skeletal Muscle Mass (SMM), Bioelectrical Impedance Vector Analysis (BIVA), Energy stored in body (Ebody) and Phase Angle (Φ).
The seca mBCA 514 can be used in conjunction with the optional PC software accessory seca analytics 115 for data management, calculations and display of information.
# 6 Safety and Effectiveness Comparison
The results of bench and clinical testing demonstrate that the seca mBCA 514 is as safe and effective as the predicate device. Impedance raw data measurements (Impedance. Resistance, Reactance, and Phase Angle) and measurements of Weight were compared against measurements made by the Biospace InBody 720 (K052646). Estimation of Total Body Water (TBW) was compared against the deuterium oxide dilution method. Estimation of Extra Cellular Water (ECW) was compared against the sodium bromide dilution method. Estimation of Lean Soft Tissue (LST) was compared against measurements obtained from the GE Lunar iDXA Bone Densitometer (K052581). Fat-Free Mass (FFM) was estimated by applying the four compartment model, which incorporates measurements from the GE Lunar iDXA Bone Densitometer (K052581), the BOD POD Sonamet Body Composition Analyzer (K060848), and the deuterium oxide dilution method.
# 7 Conclusion
The differences with the predicate devices and methods are sufficiently minor to establish substantial equivalence, and do not impact the safety or the effectiveness of the device when used as labeled.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 2, 2013
seca gmbh & co. kg % Mr. Seth A. Mailhot Special Counsel Sheppard Mullin Richter & Hampton LLP 13001 Street, NW. 11th Floor East WASHINGTON DC 20005-3314
Re: K122388
Trade/Device Name: seca mBCA 514 Regulation Number: 21 CFR§ 870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: April 24, 2013 Received: April 25, 2013
Dear Mr. Mailhot:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Seth A. Mailhot
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
## 510(k) Number (if known): K122388
Device Name: seca mBCA 514
### Intended Use:
The seca mBCA 514 measures body Weight and records body Height for calculation of Body Mass Index (BMI), Resting Energy Expenditure (REE) and Total Energy Expenditure (TEE) to monitor growth process and weight changes in individuals of all ages.
In healthy individuals aged 18 years and older the device also measures body impedance (Reactance Xc and Resistance R) for estimation of body composition. Estimation of body composition includes Total Body Water (TBW), Intra Cellular Water (ICW), Extra Cellular Water (ECW), Lean Soft Tissue (LST), Fat-Free Mass (FFM), Fat Mass (FM), Fat Mass Indices (FMI and FMMI), Skeletal Muscle Mass (SMM), Bioelectrical Impedance Vector Analysis (BIVA), Energy stored in body (Ebody) and Phase Angle (Φ).
The seca mBCA 514 can be used in conjunction with the optional PC software accessory seca analytics 115 for data management, calculations and display of information.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
# Herbert - S
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K122388 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.