K122361 · Tandem Diabetes Care, Inc. · OUG · Feb 22, 2013 · General Hospital
Device Facts
Record ID
K122361
Device Name
T:CONNECT DATA MANAGEMENT SYSTEM
Applicant
Tandem Diabetes Care, Inc.
Product Code
OUG · General Hospital
Decision Date
Feb 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6310
Device Class
Class 1
Attributes
Software as a Medical Device
Indications for Use
The t:connect Diabetes Management Application is intended for use by individuals with diabetes mellitus who use Tandem Diabetes Care insulin pumps, their care givers, and their healthcare providers in home and clinical settings. The t connect Application supports diabetes management through the display and analysis of information downloaded from Tandem Diabetes Care insulin pumps and specified blood glucose meters.
Device Story
t:connect is a web-based software system for diabetes management; retrieves history logs, diagnostic info, and pump settings from Tandem insulin pumps and blood glucose meters via USB connection. System components include a client-side Data Uploader, Web Application, and secure database server. Users (patients, caregivers, clinicians) view tabular and graphical reports (BG, insulin delivery, logbooks) via web browser. System does not modify therapy or pump parameters; provides read-only data visualization to assist clinical decision-making and diabetes management. Used in home and clinical settings; requires user authentication for secure access. Benefits include centralized data visualization and reporting to support therapy adjustments.
Clinical Evidence
No clinical data. Performance verified through software verification and validation testing, GUI verification, and human factors evaluation. Summative human factors validation test conducted with 30 participants (15 healthcare professionals, 15 type 1 diabetics) to assess usability and task performance; no safety-related issues identified.
Technological Characteristics
Web-based software system; client-side Data Uploader (PC-based) and server-side Web Application. Connectivity via USB for data retrieval and encrypted internet connection for server communication. Supports Windows (XP, Vista, 7) and Mac OS (10.5-10.7). Compatible with standard web browsers (IE, Safari, Firefox, Chrome).
Indications for Use
Indicated for individuals with diabetes mellitus using Tandem Diabetes Care insulin pumps, their caregivers, and healthcare providers in home and clinical settings to display and analyze data from insulin pumps and specified blood glucose meters.
Regulatory Classification
Identification
(1) A medical device data system (MDDS) is a hardware device that is intended to provide one or more of the following uses, without controlling or altering the functions or parameters of any connected medical devices:(i) The electronic transfer of medical device data; (ii) The electronic storage of medical device data; (iii) The electronic conversion of medical device data from one format to another format in accordance with a preset specification; or (iv) The electronic display of medical device data. (2) An MDDS may include electronic or electrical hardware such as a physical communications medium (including wireless hardware), modems, and interfaces. This identification does not include hardware devices intended to be used in connection with active patient monitoring. Hardware devices for active patient monitoring are classified under other regulations and are not included in this regulation.
Predicate Devices
Medtronic MiniMed Diabetes Data Management System (K032164)
{0}------------------------------------------------
K122361
# FEB 2 2 2013
## 510(k) SUMMARY
# Tandem Diabetes Care, Inc.'s t:connect Data Management System
Tandem Diabetes Care, Inc. Submitter's Name: 11045 Roselle Street, Suite 200 Address: San Diego, California 92121 Phone: (858) 366-6963 Facsimile: (858) 202-6707
Contact Person: Nora C.R. York Contact Email Address: nyork@tandemdiabetes.com
July 31, 2012 Date Prepared:
# Common or Usual Name / Classification Regulation
The t:connect Data Management System is an accessory to the Tandem t:slim Insulin Pump system which is classified under 21 C.F.R. 880.5725, Pump, Infusion, Insulin, Class II, Product Code, LZG.
#### Predicate Devices
Medtronic MiniMed Diabetes Data Management System (K032164)
Aidera Diasend System (K101806)
#### Intended Use / Indications for Use
The t:connect Diabetes Management Application is intended for use by individuals with diabetes mellitus who use Tandem Diabetes Care insulin pumps, their care givers, and their healthcare providers in home and clinical settings. The t connect Application supports diabetes management through the display and analysis of information downloaded from Tandem Diabetes Care insulin pumps and specified blood glucose meters.
#### Device Description
The Tandem t:connect is a web based software system which allows pump users, care rhe Tunders troomfoot rofessionals to view data retrieved from a Tandem Ambulatory Insulin Pump and Blood Glucose meters. t:connect is comprised of a Data Uploader, a Web Application, a Secure Web and Database Server, and an Admin Web Application.
The t:connect Data Uploader is a client based software application installed on the pump user's personal computer. Using a USB cable connecting the pump to a personal computer, the application uploads pump data to a secure online database server through a secure, encrypted, internet connection. The t.connect Data Uploader also uploads data from a limited set of Blood Glucose meters.
The t:connect Web Application is a reporting application that provides users with the ability to view and print (to PDF) both tabular and graphical reports representing the data uploaded to the online database. The f:connect Web Application is the "t:connect Website" shown in Figure 1.
The tconnect Web and Database Server is an application that runs on the secure server. Data displayed on the t:connect Web Application is retrieved from the secure Database
| <br><br>Title: | Page Number: |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Tandem Diabetes Care - t:connect 510(k) Notification<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 18 of 66 |
{1}------------------------------------------------
"intermediary" providing an extra layer of security and restricted access to the Database Server.
For both the Data Uploader and the Web Application, users are authenticated against a "or both the extir oploads to insure maximum security and secure access to only their data.
The Admin Web Application is an application used to manage accounts. It also allows holders of certain types of accounts to Logon As User to the t.connect Web Application as other users.
t:connect does not have the ability to modify any therapy provided by or parameters internal to the devices it communicates with. t:connect only reads data from these devices.
The major functions of the device include:
- Retrieve History Log data from the pump and transmit to the database server .
- Retrieve History Log data from various BG meters and transmit to the database server .
- Retrieve Diagnostic information from the pump and transmit to the database server .
- Provide a means to view the pump's complete history log �
- Retrieve Pump Settings from the pump for easy viewing by the user o
- Display Therapy Data textually and graphically .
- Manage User Accounts for retrieving and viewing pump data .
- Generate Notifications for the user の
- Print reports to PDF .
- Export data for use in generating customer reports using other software applications o
- Provision of user help .
- Administrator functionality related to creation and managing user accounts
## Performance Data
Tandem has completed the necessary performance testing to verify that the design input requirements have been met. Software verification and validation testing was completed to demontrate the t:connect Data Management System meets the established design inputs and its intended use.
- 1. Software Validation Report
- 2. Software Verification Report
- 3. GUI Verification Test Report
- 4. Human Factors Research Validation Report
- Software V&V Report 5.
- Labeling Verification Test report ర్.
- 7. Final Traceability Report
- 8. Risk Management Summary
#### Clinical and Human Factors Evaluation
Tandem completed three formative human factors studies and a summative human factors validation test using 30 participants. Data from these formative studies were used to refine the graphical user interface of the t:connect Data Management System. A summative human factors validation test was completed with 30 participants.
A total of 30 representative adult users stratified into two population segments: (15) healthcare professionals and (15) type 1 diabetics. The participants completed a set of representative task scenarios presented to them in as efficient and timely a manner as
| Title: | r auch Number. |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 19 of 66 |
| Tandem Diabetes Care - t:connect 510(k) Notification | |
| The first to the first the first and the are a minute a firm a fine a firm a fine a firm a fine a fire<br>Property of Children Compression a consideration of CLE LE STANLINE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE L | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
{2}------------------------------------------------
possible. They were asked to provide feedback regarding the usability and arrists the possible. They were actived to provicipants received 30 minutes of training prior to the session from a Certified Diabetes Educator.
The participant's interaction with the system was observed by the moderator seated in an The participant 3 interdollor than the System of the Morae usability software suite, observation room and were addior nous room so and well as system interaction and which oapterou partionmative study did not reveal any safety related issues.
#### Assurance Case Reports
Tandem has demonstrated the t:connect Data Management System is safe and effective as intended, through the use of Assurance Case Reports.
## Substantial Equivalence
The t:connect Data Management System is as safe and effective is the Medtronic MiniMed The Coomeor Data Management System (K032164) and the Aidera Diasend System (K101806), Diabetes Data Managomont Oyken (Nota For) and equivalence analysis. It has the same intended as oefnonstrated by performance outd and ogentions and principles of operation,
use/indications for use, and similar technological characteristics and principles of created disemindications for use, and similar toonhological one it Data Management System and the minor technological annoversons of safety or effectiveness. Thus, the t.connect and its prodicate actions faller no the necessaritially equivalent to the predicate devices.
| Element of Comparison | Tandem<br>Diabetes<br>t:connect | Medtronic<br>DMS<br>(K032164) | Aiedera<br>Diasend<br>(K101806) |
|-----------------------------------------|---------------------------------|-------------------------------|---------------------------------|
| Application Type | Web App | Web App | Web App |
| Pie Charts of BG & Insulin | Yes | Yes | Yes |
| Modal Day BG Charts | Yes | Yes | Yes |
| Charts of Basal & Bolus Delivery | Yes | Yes | Yes |
| User Selectable Date Ranges | Yes | Yes | Yes |
| Logbook | Yes | Yes | Yes |
| Displays Device Settings | Yes | Yes | Yes |
| "Print Report" Option | Yes | Yes | Yes |
| Allows User to Annotate Therapy<br>Data | Yes | Yes | Yes |
| Dashboard Summary | Yes | No | No |
| Notifications | Yes | No | No |
| Exports to Excel Compatible Files | Yes | Yes | Yes |
| Supported Insulin Pumps | Tandem | Medtronic | Multiple |
# Comparison of Technological Characteristics
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Company of the commend on a frequent of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the contribution of th<br>For a manager of the commended to the comments of the comments of the comments of the comments of the country of<br>THE REE RESERVED ENTERES AND | Page Number: |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Title: | |
| | 20 of 66 |
| Care - I connect 510(k) Notification<br>soloholos<br>andam | |
| | and the county of the production of the count of the complete of the comments of course of the comments of coursely |
| <br>A . L . L . B . B | |
{3}------------------------------------------------
| Element of Comparison | Tandem<br>Diabetes<br>t:connect | Medtronic<br>DMS<br>(K032164) | Aiedera<br>Diasend<br>(K101806) |
|---------------------------------|----------------------------------|-----------------------------------|----------------------------------|
| Supports J&J/LifeScan BG Meters | Yes | Yes | Yes |
| Supports Abbott BG Meters | Yes | Yes | Yes |
| Supports Roche BG Meters | Yes | Yes | Yes |
| Supports Additional BG Meters | No | Yes | Yes |
| Supports Microsoft OS | Win XP | Win XP | Win XP |
| | Win Vista | Win Vista | Win Vista |
| | Win 7 | Win 7 | Win 7 |
| Supports Apple Mac OS | OSX 10.5<br>OSX 10.6<br>OSX 10.7 | OSX 10.5<br>OSX 10.6<br>OSX 10.7 | OSX 10.5<br>OSX 10.6<br>OSX 10.7 |
| | Supported Browsers | IE<br>Safari<br>Firefox<br>Chrome | IE<br>Safari<br>Firefox |
. . . . .
| Title: | Tandem Diabetes Care - t:connect 510(k) Notification | Page Number: | 21 of 66 |
|--------|------------------------------------------------------|--------------|----------|
|--------|------------------------------------------------------|--------------|----------|
.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 22, 2013
Ms. Nora C.R. York Tandem Diabetes Care, Incorporated 11045 Roselle Street Suite 200 SAN DIEGO CA 92121
Re: K122361
Trade/Device Name: Tandem t: connect Data Management System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: February 12, 2013 Received: February 13, 2013
Dear Ms. Nora C.R. York:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
{5}------------------------------------------------
# Page 2 - Ms. Nora C.R. York
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21) CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Image /page/5/Picture/8 description: The image shows a stylized signature or logo. The text "For" is at the top left, and the text "with" is in the middle. The text "wrouts" is at the bottom. The signature is enclosed in an oval shape.
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indications for Use Statement
510(k) Number (if known): K122361
#### Device Name: Tandem t:connect Data Management System
Indications for Use:
The t.connect Diabetes Management Application is intended for use by individuals with diabetes mellitus who use Tandem Diabetes Care insulin pumps, their caregivers, and their healthcare providers in home and clinical settings. The t:connect Application supports diabetes management through the display and analysis of information downloaded from Tandem Diabetes Care insulin pumps and specified blood glucose meters.
| Prescription Use |
|------------------|
|------------------|
X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Richard C. Chapman 2013.02.21 11:58:51 -05'00'
#### (Division Sign-Off) Division of Anestheslology, General Hospital Infection Control, Dental Devices
510(k) Number:
Draket : 000 1 million in 19, 1
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
1: 4 .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.