MERITS MODEL P322FWD POWERED WHEELCHAIR

K122357 · Merits Health Products Co., Ltd. · ITI · Jan 9, 2013 · Physical Medicine

Device Facts

Record IDK122357
Device NameMERITS MODEL P322FWD POWERED WHEELCHAIR
ApplicantMerits Health Products Co., Ltd.
Product CodeITI · Physical Medicine
Decision DateJan 9, 2013
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 890.3860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Merits Model P322 Powered Wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair.

Device Story

The Merits Model P322 is a front-wheel-drive powered wheelchair designed for indoor and outdoor mobility. It utilizes a PG VR-2 50amp controller and a joystick interface for user operation. Input from the joystick signals the controller to release solenoid electromechanical brakes, allowing movement; releasing the joystick automatically re-engages brakes to slow and stop the chair. The device is powered by two 12 VDC 34ah batteries with an approximate 28km range. Construction features a welded steel frame with front drive units and rear pivoting casters. The device is intended for use by patients capable of operating a powered wheelchair on dry, level surfaces such as concrete or asphalt. It provides mobility assistance to users with limited physical capacity, enhancing independence in daily navigation.

Clinical Evidence

Bench testing only. The device underwent extensive performance testing per ISO 7176 standards, including static and dynamic stability, brake effectiveness, energy consumption, obstacle climbing, and climatic testing. Electrical safety and electromagnetic compatibility were verified per IEC 60601-1 and IEC 60601-1-2. Upholstery ignitability was tested per EN 1021-1/-2.

Technological Characteristics

Welded steel frame; front-wheel drive; PG VR-2 50amp controller; two 12 VDC 34ah batteries; solenoid electromechanical brakes; joystick interface. Complies with ISO 7176 series (stability, dimensions, strength, power control), IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and EN 1021-1/-2 (upholstery ignitability).

Indications for Use

Indicated for individuals with mobility limitations who are restricted to a seated position and possess the physical and cognitive capability to operate a powered wheelchair for indoor and outdoor mobility.

Regulatory Classification

Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K122357 JAN 9 2013 #### 510(k) SUMMARY Merits Health Products Co., LTD. 510(k) Premarket Notification ### Submitter: Merits Health Products Co., LTD. No.18, Jingke Rd., Nantun Dist. Taichung City 40852, Taiwan (R.O.C.) #### Contact Person: Vincent Chen Merits Health Products Co., LTD. No.18, Jingke Rd., Nantun Dist., Taichung City 40852, Taiwan (R.O.C.) Phone: +886-4-23594985 ext.200 Fax: +886-4-23503962 #### Date Prepared: July 30, 2012 #### Proprietary Name: Merits Model P322 Powered Wheelchair . #### Common name: Powered Wheelchair #### Classification name: 1 Powered Wheelchair #### Comparison to Predicate Devices: This submission indicates the Substantial Equivalence of Merits Model P322 FWD Powered Wheelchair with the predicate Hoveround Teknique FWD mk4 Power Chair (K090108). P322 has the same intended uses and similar indications, technological characteristics and principles of operation with predicate device. {1}------------------------------------------------ #### Device Description Model P322 Powered Wheelchair is battery powered, front wheel motor driven and is controlled by the PG power wheelchair VR-2 50amp controller. The user interface is a joystick. P322 is powered by two 12 VDC 34ah (U1) batteries. The batteries are charged by 4A off-board charger connect with 3-pin Microphone Connector to charging socket on joystick. The approximate driving range on fully charged batteries is up to 28km (18mi). The chair frame is a rived nut and welded steel construction and includes two front drive wheels with drive units (including motor, gear, brake), batteries and rear pivoting casters. Depending on users needs, the joystick motor control is mounted to the left or right armrest. When the user activates the joystick, the controller receives a signal to release the brakes. With the brakes released, the wheelchair is allowed to move in the direction the joystick is actuated. When the user releases the joystick, the chair slows to a stop and the brakes are automatically re-engaged. The solenoid electromechanical brakes allow the user stop by letting go of the joystick. The upholstery of the device complies with EN 1021-1/-2:2006: Furniture: Assessment of the ignitability of upholstered furniture: Ignition source: Smouldering cigarette/ Match flame equivalent The device can be operated on dry, level surfaces composed of concrete, blacktop, or asphalt under normal driving conditions. The Merits Model P322 is substantially equivalent to Hoveround Teknique FWD mk4 Power Chair (K090108). Both products are battery power, motorized designed for use with powered wheelchairs. Performance characteristics and drive mechanisms are similar and all have the same intended function and use which is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair. Additional, they are all constructed from the same basic materials, have the same basic operational principles and all use DC batteries as their source of power. Although there are some minor differences between P322 and its predicate device. But they raise no new issues of safety or effectiveness. Performance data demonstrate that P322 is safe. The non-clinical testing and the predicate comparisons demonstrate that any differences in their technological characteristics do no raise any new questions of safety or effectiveness. #### Intended Use The Merits Model P322 Powered Wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair. {2}------------------------------------------------ Discussion of Non-clinical Tests Performed for Determinations of Substantial equivalence are as follows: ISO 7176-1:1999 Determination of Static Stability ISO 7176-2:2001 Determination of Dynamic Stability of electric wheelchairs ISO 7176-3:2003 Determination of effectiveness of brakes Energy consumption of electric wheelchairs and scooters for ISO 7176-4:2008 determination of theoretical distance range - Determination of overall dimensions, mass and maneuvering ISO 7176-5:2008 space of speed. Determination maximum acceleration and ISO 7176-6:2001 deceleration of electric wheelchairs - ISO 7176-7:1998 Method of Measurement of Seating and Wheel Dimensions Requirements and test methods for static, impact and fatigue ISO 7176-8:1998 strengths ISO 7176-9:2009 Climatic tests for wheelchairs Determination of obstacle-climbing ability of electrically power ISO 7176-10:2008 wheelchairs ISO 7176-11:1992 Test dummies ISO 7176-13:1989 Determination of coefficient of friction of test surfaces ISO 7176-14:2008 Power control systems for electrically powered and wheelchairs and scooter- Requirements and test methods. - Requirements for Information Disclosure, Documentation and ISO 7176-15:1996 Labeling Medical electrical equipment Part 1: General requirements for IEC 60601-1:1988 safety - IEC 60601-1-2:2007 Electromagnetic Compatibility Requirements and Tests - Medical devices -- Application of risk management to medical ISO 14971:2007 devices - Furniture Assessment of the ignitability of upholstered EN 1021-1:2006 furniture : Ignition source: Smouldering cigarette Furniture Assessment of the ignitability of upholstered EN 1021-2:2006 furniture : Ignition source: Match flame equivalent The results of the testing confirm that the device meets specifications and is substantially equivalent to the predicate device. #### Conclusions According to comparison table, the differences on weight capacity, function of elevate, recline and dimension of width do not deleteriously affect the safety and effectiveness of the device. So based on the design, performance specifications and testing and intended use, the Merits Model P322 Powered Wheelchair is substantially equivalent to Hoveround Teknique FWD mk4 Power Chair (K090108). {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Merits Health Products CO., LTD Mr. Vincent Chen No. 18 Jingke Road Nantun Dist Taichung China (Taiwan) 40852 JAN 9 2013 Re: 510(k) Number: K122357 Trade/Device Name: Merits Model P322 Powered Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: December 21, 2012 Received: December 26, 2012 Dear Mr.Chen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ Page 2 - Mr. Vincent Chen device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Joyce M. Whang for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT 510(k) File Number: Device Name: Merits Model P322 Powered Wheelchair Indications for Use: The Merits Model P322 Powered Wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair. Prescription Use_ V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Brian D.Pullin -S Division of Neurological and Physical Medicine Devices 510(k) Number: 1122357
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...