K122350 · Airway Management, Inc. · BZD · Nov 1, 2012 · Anesthesiology
Device Facts
Record ID
K122350
Device Name
TAP PAP NASAL PILLOW MASK
Applicant
Airway Management, Inc.
Product Code
BZD · Anesthesiology
Decision Date
Nov 1, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TAP PAP Nasal Pillow Mask is intended to be used by individuals who have been diagnosed by a Physician as requiring CPAP or Bi-Level ventilator treatment. The TAP PAP Nasal Pillow Mask is intended for single patient adult (>66lb/30kg) use in the home, hospital or other clinical setting,
Device Story
TAP PAP Nasal Pillow Mask is a non-invasive patient interface for CPAP/Bi-Level therapy. Device consists of a boil-and-bite dental appliance with an attached bracket, mask frame, and soft silicone nasal pillows. User customizes dental tray for fit; bracket holds nasal pillows in nares. Device connects to CPAP/Bi-Level equipment via 22 mm corrugated tubing. Integrated exhalation vents (bias holes) in mask frame allow CO2 washout. Used in home, hospital, or clinical settings by adult patients. Healthcare providers prescribe the device; patients self-administer. Output is delivery of positive airway pressure to patient's nostrils. Benefits include effective ventilation and CO2 clearance for OSA treatment.
Clinical Evidence
Bench testing only. Evaluated pressure drop, bias flow, dead space, CO2 washout (ISO 17510-2), mask weight, cleaning durability, and environmental stress (hot/cold/drop). Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation, genotoxicity, implantation). CO2 washout testing confirmed performance within ISO 17510-2 criteria across pressures (4-10 cm H2O).
Technological Characteristics
Boil-and-bite dental tray with heat-sensitive impression material; bracket-mounted silicone nasal pillows; 22 mm corrugated tubing connector. Materials tested per ISO 10993-1. Operating pressure range 4–20 cm H2O. Dead space 95 ml. Dimensions 280 mm x 67 mm x 220 mm. Weight 59 gm. Mechanical interface; no software or electronics.
Indications for Use
Indicated for adult patients (>66 lbs/30 kg) diagnosed by a physician as requiring CPAP or Bi-Level ventilator treatment.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
K122350 510(k) Summary Page 1 of 6
| Date Prepared | 31-Oct-12 |
|----------------------------|-------------------------------------------------------|
| Airway Management, Inc. | 3418 Midcourt Rd. |
| | Suite 114 |
| | Carrollton, TX 75006 |
| | Tel - 214-369-0978 |
| | Fax - 214-691-3151 |
| Official Contact: | Dale Siebenmorgen - Manager of Quality and Regulatory |
| Proprietary or Trade Name: | TAP PAP Nasal Pillow Mask |
| Common/Usual Name: | CPAP Patient interface |
| Classification Code/Name: | BZD - non-continuous ventilator (IPPB) |
| | 21 CFR 868.5905 |
| | Class 2 |
| Predicate Devices: | K992384 - Stevenson CPAP/Pro |
### Device Description:
The TAP PAP Nasal Pillow Mask consists of a boil & bite type dental appliance. A small bracket that extends beyond the lips to attach the mask frame (Mask - Upper Mouth Piece Adjustment Joint). Two soft silicone nasal pillows are located on the mask frame to seal against the nostrils. The nasal pillows come in three sizes and can be individually rotated to align with the user's nares. A corrugated flexible tube extends upwards from the mask frame to a 22 mm connector which connects to a breathing tube of a CPAP or bi-level device.
The mask frame has exhalation vents (bias holes) that allow exhaled gases to be continually flushed.
#### Indications for Use:
The TAP PAP Nasal Pillow Mask is intended to be used by individuals who have been diagnosed by a Physician as requiring CPAP or Bi-Level ventilator treatment. The TAP PAP Nasal Pillow Mask is intended for single patient adult (>661b/30kg) use in the home, hospital or other clinical setting,
Patient Population: Adults (>66 lbs / 30 kg)
Environment of Use: Home, hospital or other clinical settings
{1}------------------------------------------------
# 510(k) Summary Page 2 of 6 31-Oct-12
#### Predicate Device Comparison:
The TAP PAP Nasal Pillow Mask has been compared to the predicate and is viewed as substantially equivalent because:
#### Indications -
- The TAP PAP Nasal Pillow Mask is intended to be used by individuals who have been ● diagnosed by a physician as requiring CPAP or Bi-Level ventilator treatment. The TAP PAP Nasal Pillow Mask is intended for single patient adult (>661b/30kg) use in the home, hospital or other clinical setting.
- The predicate CPAP/Pro (K992384) is intended to treat adult OSA, but it is a patient . interface which must connect to CPAP and bi-level devices, therefore the indications for use are considered substantially equivalent.
- . Substantially equivalent to Stevenson CPAP/Pro (K992384).
Discussion - The indications for use are equivalent to the predicate.
#### Patient Population -
- The TAP PAP Nasal Pillow Mask and the predicate CPAP/Pro are both intended for . adults (>66 lbs. / 30 kg.).
Discussion - The intended patient population is identical to the predicate.
#### Technology -
- Technology is the use of an upper dental tray with boils & bite impression material . which has a bracket to hold nasal pillows and tubing. The nasal pillows are offered in multiple sizes and fit in the nares of the patient. They each on a means for relieving excess flow, vent holes, and connect to the positive pressure equipment with a standard 22 mm corrugated tubing.
Discussion - The technology of tubing, nasal pillow2s and a manifold is identical to the predicate. The difference in placing the manifold closer to the patient in the proposed device is a matter of comfort for the user vs. the manifold remote from the patient and bracket holding 2 separate tubes with the nasal pillows as the predicate has. The difference in this design configuration increases the dead space, but the performance of the proposed device in Pressure vs. Flow and CO2 washout testing demonstrates that it is equivalent and the differences are not clinically significant and safety and effectiveness is not affected.
#### Materials -
The materials are standard medical grade materials and have been tested per ISO 10993. . Discussion - The materials have been tested per ISO 10993-1 and passed the requirements of each test.
#### Environment of Use -
The masks are intended for use in the home, hospital, or sleep laboratories. . Discussion - The environments of sue are identical to the predicate.
{2}------------------------------------------------
# 510(k) Summary Page 3 of 6 31-Oct-12
.
# Differences –
There are no differences between the predicate and the proposed device which would raise any new safety or risks and thus can be found to be substantially equivalent.
| Attributes | TAP PAP Nasal Pillow Mask | Stevenson<br>CPAP/Pro K992384 | | |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|------------------------------------|------------------------------|
| Intended Use | Non-invasive ventilator interface for CPAP or bi-level treatment | Non-invasive ventilator interface for CPAP or bi-level treatment | | |
| Indications for Use | The TAP PAP Nasal Pillow Mask is intended to be used by individuals who have been diagnosed by a Physician as requiring CPAP or Bi-Level ventilator treatment. The TAP PAP Nasal Pillow Mask is intended for single patient adult (>66lb/30kg) use in the home, hospital or other clinical setting. | Intended to treat adult OSA | | |
| Patient Population | Adults (>66 lbs. / 30 kg.) | Adults (>66 lbs. / 30 kg.) | | |
| Environment of Use | Home<br>Hospital<br>Other clinical settings | Not specified | | |
| Duration of Use | Single patient, multi-use | Single patient, multi-use | | |
| Prescriptive | Yes | Yes | | |
| Method of fixation | Boil & bite dental tray with bracket attachment | Boil & bite dental tray with bracket attachment | | |
| Sealing method | Soft nasal pillows | Soft nasal pillows | | |
| Positive pressure source | CPAP or bi-level equipment | CPAP or bi-level equipment | | |
| Means to relieve excess flow | Vent holes | Diffuser near pillows | | |
| Tray configurations | Boil & Bite with heat sensitive impression material | Boil & Bite with heat sensitive impression material | | |
| Bracket for adjusting tubing and nasal pillows | Yes | Yes | | |
| Tubing from pillows to connector to gas source | Yes | Yes | | |
| Different Pillow sizes | 3 | 3 | | |
| Technical Specifications / Performance Testing | | | | |
| Operating Pressure range | 4 – 20 cm H₂O | 4 – 18 cm H₂O | | |
| Dead Space | 95 ml | 45 ml | | |
| Pressure Drop | Flow<br>50 lpm<br>100 lpm | Drop<br>0.2 cm H₂O<br>0.6 cm H₂O | Not available or reported | |
| Exhaust flow / Pressure | Pressure (cm H₂O)<br>6<br>10<br>20 | Flow (lpm)<br>23<br>30<br>42 | Pressure (cm H₂O)<br>6<br>10<br>20 | Flow (lpm)<br>15<br>21<br>33 |
| Weight | 59 gm | 55 gm | | |
| Gross Dimensions (no head gear) | 280 mm (H) x 67 mm (W) x 220 mm (D) | 260 mm (H) x 100 mm (W) x 100 mm (D) | | |
{3}------------------------------------------------
# . 510(k) Summary ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------31-Oct-12
### Non-clinical Testing Summary -
We have performed a number of tests typical of CPAP patient interfaces. These tests include:
### Pressure Drop
- Standard test method of using a standard flow and measuring pressure drop .
- No pass / fail criteria, reportable values only .
- Discussion The proposed device has a low pressure drop at 50 and 100 lpm and is only a . reported value. CPAP machines are able to supply the required flow at these low resistance values. As the patient exhaled through their mouth there is no resistance to breathing.
### Bias (vent) Flow (Pressure / Exhaust Flow
- Standard test method using CPAP unit at various flow rates and measuring the leak at the . patient interface
- . No pass / fail criteria, reportable values only. It should be comparable to the predicate.
- Discussion The proposed device has a slightly higher "Leak or Flow" rate at all pressure . vs. the predicate which means greater CO2 washout. The differences do not have any clinical significance.
#### Dead space
- Standard test method of measuring the volume of the parts near the patient.
- . No pass / fail criteria, reportable values only. It should be comparable to the predicate.
- Discussion The proposed device has a "manifold" near the patient with a single hose . coming from the CPAP unit and the nasal pillows attached to the manifold. The predicate has the manifold remote from the patient and then 2 separate tubes with individual nasal pillows coming from the manifold to the patient. The higher reported dead space is based upon measuring the proposed device with the manifold, tubing, and nasal pillows vs. the predicate was the tubing and nasal pillows only but not the remote manifold. While the reported value for the proposed is higher 95 ml vs. 45 ml, the Pressure vs. Flow and the CO2 washout data demonstrated that the proposed device effectively flushed the patient interface as it had higher leak rates than the predicate at the same pressures and the proposed device passed the CO2 washout per ISO 17510-2 criteria. Therefore the higher dead space is not clinically significant nor does it affect safety and effectiveness.
#### CO2 washout per ISO 17510-2
- Tested per ISO 17510-2 .
- Pass / fail criteria per ISO 17510-2 allow for changes in CO2 at various pressures to be < . 20% and < 60% from baseline).
| Pressure | Mask ETCO₂ | % Increase |
|-----------|------------|------------|
| 4 cm H₂O | 5.4 % | 4% |
| 5 cm H₂O | 5.4 % | 4% |
| 10 cm H₂O | 5.3 % | 2% |
| Occluded | 5.6 % | 7% |
{4}------------------------------------------------
# 510(k) Summary Page 5 of 6 31-Oct-12
- . Discussion - The proposed mask demonstrated that it meet the requirements of CO2 washout per ISO 17510-2.
## Mask Weight
- Simply weighed the devices ●
- . No pass / fail criteria, reportable values only. It should be comparable to the predicate.
- . Discussion - The proposed device vs. the predicate were equivalent.
### Cleaning durability
- Repeated cleaning was performed per the recommended cleaning instructions .
- Visual and performance testing was performed and compared pre and post- cleaning and ● found to be similar
- . Discussion - The proposed mask can be cleaned as intended and meet the performance specifications
### Environmental (Hot / cold exposure and drop test)
- Test method was subjected the samples to 50℃ for 72 hours and 20℃ for 24 hours ●
- . Pass / fail criteria was that they would meet the performance specifications which was performed after the cleaning durability
- Discussion The proposed device met the performance specifications after conditioning . and cleaning.
#### Biocompatibility of Materials -
Materials were evaluated per ISO 10993-1. There are parts and thus their respective materials which have the following patient contact and duration.
Parts which are Externally communicating, Tissue - gas pathway contact, permanent duration require the following tests. They are the mask shell, assembly, pillow seals and mouth tray and tubing.
- Cytotoxicity (tested per ISO 10993-5) ಂ
- Sensitization (tested per ISO 10093-10) O
- Irritation (tested per ISO 10993-10) ಂ
- Genotoxicity (Tested per ISO 10993-3) O
- Implantation (tested per ISO 10993-6) O
- Pass / fail criteria was for the each respective ISO 10993 test o
Parts which are Surface communicating, Skin, permanent duration require the following tests. They are the headgear and chin strap.
- Surface communicating, Skin, permanent duration .
- o Cytotoxicity (tested per ISO 10993-5)
- Sensitization (tested per ISO 10093-10) O
- Irritation (tested per ISO 10993-10) O
- Pass / fail criteria was for the each respective ISO 10993 test O
{5}------------------------------------------------
# 510(k) Summary Page 6 of 6 3 1-Oct-12
Parts which are Surface communicating, Skin, limited duration require the following tests. They are the nut and bolt.
- Cytotoxicity (tested per ISO 10993-5) o
- Sensitization (tested per ISO 10093-10) O
- Irritation (tested per ISO 10993-10) O
- Pass / fail criteria was met for each respective ISO 10993 test o
### Discussion -
All materials were tested according to ISO 10993-1 for the appropriate level of patient contact and duration and found pass the applicable ISO 10993-1 test requirements.
## Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to substantially equivalent.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
**NOV** 1 **2012**
Airway Management, Incorporated C/O Paul Dryden President Promedic. Incorporated Bonita Springs, Florida 34134
Re: K122350
Trade/Device Name: Tap PAP Nasal Pillow Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: August 2, 2012 Received: August 3, 2012
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
## Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh. for
Anthony D. Watson, B.S., M.S., M.B.A.
Director
Division of Anesthesiology, General Hospital,
Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use Statement
Page 1 of 1
510(k) Number:
K122350
TAP PAP Nasal Pillow Mask Device Name:
Indications for Use:
The TAP PAP Nasal Pillow Mask is intended to be used by individuals who have been diagnosed by a physician as requiring CPAP or Bi-Level ventilator treatment. The TAP PAP Nasal Pillow Mask is intended for single patient adult (>66lb/30kg) use in the home, hospital or other clinical setting.
Over-the-counter use _ Prescription Use XX or (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Schultheis
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _ K 122350
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.