The splinting system is designed for use in oral splintings or to support laboratory-processes composite bridge/denture work - QUARTZ SPLINT™ UD (Unidirectional): Composite laboratory bridge strengthening/ reinforcement - QUARTZ SPLINT™ WOVEN: Direct Splinting/Bridge and Implant positioning reinforcement. Splinting of mobile teeth/temporary stabilization in Clinical use. Retention of teeth subsequent to orthodontic therapy. - OUARTZ SPLINT™ ROPE: Laboratory splint fabrication for bridge reinforcement or denture repair .- Periodontal splinting of mobile teeth in Clinical use. - QUARTZ SPLINT™ MESH: Denture reinforcement and denture repair. - QUARTZ SPLINT™ RESIN: for additional wetting during the application of the finished splints - QS FLOWM: a nano filled, light cured flowable composite designed to be used with the QUARTZ SPLINT system for reinforcement and stabilizing of mobile teeth.
Device Story
Quartz Splint system consists of continuous quartz fibers impregnated with light-cured resin matrix; available in various structures (UD, Woven, Rope, Mesh) plus flowable nano-hybrid composite (QS Flow) and resin for wetting. Used by dental professionals in clinical or laboratory settings for splinting mobile teeth, reinforcing bridges/dentures, and post-orthodontic retention. Fibers are placed and polymerized using standard halogen or LED dental light systems. System provides mechanical reinforcement for dental structures; improves stability of mobile teeth; facilitates laboratory fabrication of dental prosthetics.
Clinical Evidence
No clinical studies were utilized. Evidence based on bench testing (ISO 4049:2009 for mechanical properties) and biological evaluation (ISO 10993-5 for cytotoxicity).
Technological Characteristics
Continuous quartz fibers impregnated in light-cured resin matrix. Includes nano-hybrid light-cured flowable composite. Polymerization via halogen or LED dental light systems. Complies with ISO 4049:2009 for mechanical properties and ISO 10993-5 for biological safety.
Indications for Use
Indicated for oral splinting, laboratory-processed composite bridge/denture reinforcement, stabilization of mobile teeth, temporary stabilization, retention of teeth post-orthodontic therapy, and denture repair.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
K|22348
Image /page/0/Picture/1 description: The image shows a logo with the letters "RTD" in a stylized font on the left side. To the right of the logo are the words "RECHERCHES TECHNIQUES" stacked on top of each other. Below these words is a left-pointing triangle followed by the word "DENTAIRES". The logo and text appear to be for a dental research or technology company.
OCT 1 5 2012
# 510(k) Summary [As described in 21CFR 807.92]
1. Submitter by: RECHERCHES TECHNIQUES DENTAIRES (R.T.D.) 3 RUE LOUIS NEEL TECHNOPARC, ESPACE GAVANIERE 38120 SAINT EGREVE FRANCE 38120 Phone: 33-4-76565663
## 2. Contact Person:
Manh Quynh Chu R.T.D. 3 Rue Louis Neel St. Egreve, FRANCE 38120 Phone: 33-4-76565663
3. Date Prepared: 26-Jul-2012
4. Trade Name: Quartz Splint
#### 5. Common Name:
- Dental splinting .
- . Tooth shade resin material21
#### Classification Name: 6.
- Dental/Resin, denture, relining, repairing, rebasing (21 CFR 872.3760, Product Code EBI) ●
- Material, tooth shade, resin (21 CFR 872.3690, Product Code EBF) .
# 7. Predicate Device(s): Device: EVERSTICK Manufacturer: STICK TECH LTD 510(k) Number: K011788
Device: SPLINT-IT Manufacturer: PENTRON INC 510(k) Number: K972985
Device: RIBBOND Manufacturer: RIBBOND Inc
Section 5- Page 1 of 3
RTD . 3, rue Louis Neel . Technoparc Espace Gavanière . 38120 SAINT-EGREVE . FRANCE Tél .: 33(0)4 76 56 56 66 . Fax: 33(0)4 76 56 11 77 . E-mail: infostiond fr . Internet: http://www.rdd.fr SARL-RCS GRENOBLE B 065 501 055/ APE 331 B / TVA Inca Nº FR 44 068 501 055
ISO 13485 (G-MED)
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the letters "RTD" in bold, black font. To the right of the logo, there is text that reads "RECHERCHES TECHNIQUES DENTAIRES" in a smaller, sans-serif font. The text is stacked vertically, with "RECHERCHES" at the top, "TECHNIQUES" in the middle, and "DENTAIRES" at the bottom. There is a small triangle pointing to the left before the word "DENTAIRES".
#### 13. Technological characteristics:
The predicate devices are made of continuous glass fibers/polyethylene fibers. The predicate devices are offered in a dry, non-impregnated fibers or in a light-cured resin impregnated fibers. QUARTZ SPLINT have the same technological characteristics as these predicate devices as they are made of continuous quartz fibers impregnated in a light-cured resin. Polymerization or curing is the same for all these devices with help of halogen or LED dental light system. A flowable nano-hybrid, light-cured composite especially designed for use with the fiber products for the stabilization/splinting of mobile teeth is available. All predicate devices have the same intended use as QUARTZ SPLINT system.
#### 14. Performance
Physical and mechanical properties testing of the subject and predicate devices have been evaluated according to ISO 4049:2009. The testing results indicate that the subject and predicate device comply with the requirements of ISO 4049:2009. Internal and external tests demonstrate that QUARTZ SPLINT system compared to the predicate devices perform as well as equal or better in mechanical features which are the main function for a splint.
Biological evaluation such as cytotoxicity tests (according to ISO 10993-5) has been conducted. and results comply with the requirements.
#### 15. Conclusion
In vitro tests and clinical evaluation demonstrate that the QUARTZ SPLINT system (subject device) is substantially equivalent in safety and effectiveness to the predicate device.
#### Section 5- Page 3 of 3
RTD . 3; rue Louis Neel . Technoparc Espace Gavanière . 38120 SAINT-EGREVE . FRANCE Tel.: 33(0)4 76 56 56 66 - Fax: 33(0)4 76 56 11 77 - E-mail: infost@rid.fr - Internet: http://www.rdd.fr SARL-RCS GRENOBLE B 068 501 055: APE 331 B/ TVA Intra Nº FR 44 068 501 055
ISO 13485 (G-MED)
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the letters RTD in a bold, stylized font on the left side. To the right of the letters, there is text that reads "RECHERCHES TECHNIQUES DENTAIRES". The text is stacked vertically, with "RECHERCHES" on top, followed by "TECHNIQUES", and then "DENTAIRES" at the bottom. A small arrow is to the left of the word "DENTAIRES".
510(k) Number: K913040 .
Device: EVERSTICK Manufacturer: STICK TECH LTD 510(k) Number: K031341
### 8. Description of the Device:
Quartz Splint is made of quartz fibers impregnated a light-cured resin matrix. Quartz Splint is a complete system that contains different structures and it is available in different structures and sizes:
Quartz Splint UD Quartz Splint WOVEN Ouartz Splint ROPE Quartz Splint MESH Ouartz Splint Resin/Flow
#### 9. Intended Use:
The splinting system is designed for use in oral splintings or to support laboratory-processes composite bridge/denture work
- QUARTZ SPLINT™ UD (Unidirectional): Composite laboratory bridge strengthening/ t reinforcement
- QUARTZ SPLINTTM WOVEN: Direct Splinting/Bridge and Implant positioning . reinforcement. Splinting of mobile teeth/temporary stabilization in Clinical use. Retention of teeth subsequent to orthodontic therapy.
- OUARTZ SPLINT™ ROPE: Laboratory splint fabrication for bridge reinforcement or . denture repair .- Periodontal splinting of mobile teeth in Clinical use.
- QUARTZ SPLINTTM MESH: Denture reinforcement and denture repair. .
- QUARTZ SPLINT™ RESIN: for additional wetting during the application of the finished . splints
- QS FLOWM: a nano filled, light cured flowable composite designed to be used with the . QUARTZ SPLINT system for reinforcement and stabilizing of mobile teeth.
#### 10. Technological Characteristics:
The subject device has the same technological characteristics and indications for use as the predicate device.
#### 11. Non-Clinical Tests:
・・
Verification and validation were conducted to ensure the expected performance of the Quartz Splint system.
#### 12. Clinical Performance Data:
No clinical studies were utilized for the purpose of obtaining safety or effectiveness data.
Section 5- Page 2 of 3
RTD . 3, rue Louis Neel . Technopare Espace Gavanière . 38120 SAINT-EGREVE . FRANCE Tél.: 33(0)4 76 56 56 66 · Fax: 33(0)4 76 56 11 77 · E-mail: infos(@gd.fr · Internet: http://www.rdd.fr SARL-RCS GRENOBLE B 068 501 0554 APE 335 B / TVA Intra No FR 44 068 501 055
ISO 13485 (G-MED)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Manh-Quynh Chu Technical Director Recherches Techniques Dentaires 3 Rue Louis Neel St. Egreve, 38120 France
Re: K122348
Trade/Device Name: Quartz Splint System (Quartz Splint UD, Quartz Splint Woven, Quartz Splint Rope, Quartz Splint Mesh, Quartz Splint Resin, and OS Flow)
Regulation Number: 21 CFR 872.3760
Regulation Name: Denture Relining, Repairing, or Rebasing Resin Regulatory Class: II
Product Codes: EBI, EBF, EBG
Dated: July 26, 2012
Received: August 3, 2012
Dear Mr. Chu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
OCT 1 5 2012
{4}------------------------------------------------
# Page 2 - Mr. Chu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. ma
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use
¥122348
510(k) Number (if known): To Be Assigned
# Device Name: Quartz Splint System
Indications For Use:
The splinting system is designed for use in oral splintings or to support laboratory-processes composite bridge/denture work
- QUARTZ SPLINT™ UD (Unidirectional): Composite laboratory bridge strengthening/ . reinforcement
- QUARTZ SPLINT™ WOVEN: Direct Splinting/Bridge and Implant positioning . reinforcement. Splinting of mobile teeth/temporary stabilization in Clinical use. Retention of teeth subsequent to orthodontic therapy.
- QUARTZ SPLINT™ ROPE: Laboratory splint fabrication for bridge reinforcement or . denture repair. Periodontal splinting of mobile teeth in Clinical use.
- QUARTZ SPLINT™ MESH: Denture reinforcement and denture repair: #
- QUARTZ SPLINT™ RESIN: for additional wetting during the application of the finished � splints
- QS FLOWTM: a nano filled, light cured flowable composite designed to be used with the . QUARTZ SPLINT system for reinforcement and stabilizing of mobile teeth.
Prescription Use X X AND/OR Over-The-Counter Use
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE | |
|-------------------------------------------------------|--|
|-------------------------------------------------------|--|
| (Division Sign-Off) | |
|----------------------------------------------|--|
| Division of Anesthesiology, General Hospital | |
| Infection Control, Dental Devices | |
| 510(k) Number: | K122348 |
|----------------|---------|
|----------------|---------|
Section 4- Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.