K122294 · Mui Scientific · FFX · Nov 1, 2012 · Gastroenterology, Urology
Device Facts
Record ID
K122294
Device Name
PRESSURIZED INFUSION PUMP
Applicant
Mui Scientific
Product Code
FFX · Gastroenterology, Urology
Decision Date
Nov 1, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1725
Device Class
Class 2
Indications for Use
The Pressurized Infusion Pump (Electrically Powered Models #PIP-4-4SS, PIP-4-6SS, PIP-4-8SS, PIP-4-12SS, and Portable Models #PIP-6-4SS, PIP-6-6SS, PIP-6-8SS, PIP-6-12SS, PIP-6-16SS, PIP-6-22SS, PIP-6-24SS, PIP-6-36SS) is intended to provide a regulated and channeled perfusate of water to a motility catheter to perform manometric motility studies along the gastrointestinal tract. The Pressurized Infusion Pump is intended to be used with a separate legally marketed compatible motility catheter and separate computerized gastrointestinal monitoring system.
Device Story
Pressurized Infusion Pump provides regulated water perfusion to motility catheters for GI manometry. System components: electrical compressor motor, drying cylinder, pressure switches/gauges, flow restrictors, water reservoir. Operation: motor compresses air into drying cylinder; pressure switch maintains 18-42psi range; regulator reduces air to 17psi; regulated air drives water from reservoir through flow restrictors (0.60mL/min) into motility catheter. When GI muscles contract, water flow through catheter ports is constricted; pressure change transmits back through water column to pump-integrated pressure transducer; signal sent to external computerized monitoring system for display. Used in clinical settings by healthcare professionals. Benefits: enables standardized, constant-rate perfusion for accurate manometric pressure measurement during GI motility assessment.
Clinical Evidence
Bench testing only. Performance compared to predicate using pinch tests over 8 pressure ports on a motility catheter to simulate muscle contractions. Results demonstrated that pressure exerted and measured by the electrical/portable pumps was equal to the predicate pump.
Technological Characteristics
System includes electrical compressor motor, drying cylinder, pressure switches, gauges, and flow restrictors. Delivers water at constant 0.60mL/min rate. Regulates air pressure to 17psi. Connects to external motility catheter and computerized monitoring system. Non-sterile pump unit; uses sterilized irrigation water.
Indications for Use
Indicated for patients requiring manometric motility studies of the gastrointestinal tract. Used with compatible motility catheters and monitoring systems.
Regulatory Classification
Identification
A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.
Predicate Devices
Mui Scientific's Nitrogen gas model of the pressurized infusion pump
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a black and white graphic. The graphic appears to be an abstract shape or design. The shape is irregular and has some jagged edges. The image is simple and minimalist.
Ki22294
NOV 1 2012 PAGE 1 OF 2
Mivision of Mui Scientific Enterprises Inc 145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3 Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523 Website: www.muiscientific.com Email: mail@muiscientific.com
October 31, 2012
## 510(k) Summary
#### RE: Electrical and Portable Pressurized Infusion Pump
Summary prepared by:
Contact Person: Tammy Mui Title: Operations Manager Manufaxturer: Mui Scientific, a Division of H&A Mui Enterprises Address: 145 Traders Blvd. East, Unit #33-34, Mississauga, Ontario, Canada L4Z 3L3 Phone: (905) 890-5525 Fax: (905) 890-3523 Email: tammy.mui@muiscientific.com
Trade name: Pressurized Infusion Pump Common name: Water-perfused motility pump Classification name: Gastrointestinal motility system
This 510(k) Summary is for the Pressurized Infusion Pump.
Indications for Use: The Pressurized Infusion Pump (Electrically Powered Models #PIP-4-4SS, PIP-4-6SS, PIP-4-8SS, PIP-4-12SS, and Portable Models #PIP-6-4SS, PIP-6-6SS, PIP-6-8SS, PIP-6-12SS, PIP-6-16SS, PIP-6-22SS, PIP-6-24SS, PIP-6-36SS) is intended to provide a regulated and channeled perfusate of water to a motility catheter to perform manometric motility studies along the gastrointestinal tract. The Pressurized Infusion Pump is intended to be used with a separate legally marketed compatible motility catheter and separate computerized gastrointestinal monitoring system.
These pressurized infusion pumps consist of an electrical compartment, a water chamber, a regulator and a set of gauges, and flow restrictors. The electrical compartment unit consists of a compressor motor, a pressure switch, a pressure gauge, and a drying cylinder. When turned on, the motor will compress air into the drying cylinder, which will remove the water byproduct from the compression. The pressure switch will turn the motor off at 42psi (as displayed on the pressure gauge), and will restart it when the pressure drops to 18psi. For the Portable Pressurized Infusion Pump, an air hose connects the electrical unit to the main pump housing to deliver the pressurized air to the pump. The pressurized air will flow through a high pressure gauge (to display the pressure of the supplied compressed air), to a regulator (which will regulate the pressurized air down to 17psi), then a low pressure gauge
{1}------------------------------------------------
# 122294 AGE 2 OF 2
### Division of Mui Scientific Enterprises Inc.
145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3 Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523 Website: www.muiscientific.com Email: mail@muiscientific.com
(to display the pressure of the regulated air), to a water reservoir filled with sterilized irrigation water. The requlated air will push the water out of the water reservoir at a constant rate through resistors that will further reduce the water flow rate to 0.60mL/min before it travels through a pressure transducer to the motility catheter. At the proximal end of the motility catheter are female luers that connect to the top of the pressure transducers on the infusion The water will flow through the entire lenath of the motility catheter to the distal end pump. where the pressure ports are located. The motility catheter is then inserted into the patient and placed along the specific section of the gastrointestinal system being measured. With the requlated pressurized water flowing through the catheter at a constant rate, when the qut muscles contract, it will constrict the flow of water through the pressure ports, and the pressure change will be transmitted back through the water flow to the pressure transducer on the pressurized infusion pump, where the signal will be displayed onto the computer.
We are claiming this system to be substantially equivalent to the following predicates:
- Mui Scientific's Nitrogen gas model of the pressurized infusion pump -
The Electrical and Portable pump versions are similar to this predicate in that they use pressure to push water from a water reservoir through resistors that further reduce the flow rate before passing through a motility catheter. The difference is the source of the pressure: compressed nitrogen gas in the predicate, and an air compressor motor in the Electrical and Portable pump models.
Bench tests were conducted with the Electrical and Portable Pressurized Infusion Pumps, comparing performance data with that of its predicates. With the systems on and perfusing, a pinch test was performed over each of the 8 pressure ports along the motility catheter, to mimic a muscle contraction. The pressure exerted was shown to equal the pressure measured and transmitted on the predicate pump.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of human figures.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Letter date: November 1, 2012
Mui Scientific Division of H&A Mui Enterprises Inc. % Ms. Tammy Mui Operations Manager 145 Traders Blvd. East, Unit #34 MISSISSAUGA ON CANADA L4Z 3L3
Re: K122294
Trade/Device Name: Pressurized Infusion Pump Regulation Number: 21 CFR§ 876.1725 Regulation Name: Gastrointestinal motility monitoring system Regulatory Class: II Product Code: FFX Dated: October 15, 2012 Received: October 19, 2012
Dear Ms. Mui:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Tammy Mui
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K122294
Device Name: Pressurized Infusion Pum
Indications For Use:
The Pressurized Infusion Pump (Electrically Powered Models #PIP-4-4SS, PIP-4-8SS PIP-4-12SS, and Portable Models #PIP-6-4SS, PIP-6-6SS, PIP-6-8SS, PIP-6-12SS, PIP-6-16SS, PIP-6-22SS, PIP-6-24SS; PIP-6-36SS) is intended to provide a regulated and channeled perfusate of water to a motility catheter to perform manometric motility studies along the gastrointestinal tract. The Pressunzed Infusion Pump is intended to be used with a separate legally marketed compatible motility catheter and separate computerized gastrointestinal monitoring system.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
oncurrence of CDRH, Office of Device Evaluation (ODE)
Benjamin R. Fisher 1 16:46:29 -04'00
(Division Sign-Off) wision of Reproductive, Gastro-Renal, and Proiogical Devices SIU(K) NUr
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.